← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K023300

# HI-TORQUE WHISPER LS; HI-TORQUE WHISPER MS; HI-TORQUE WHISPER MS CS-J; HI-TORQUE WHISPER ES; HI-TORQUE WHISPER ES CS-J (K023300)

_Guidant Corp. · DQX · Oct 28, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K023300

## Device Facts

- **Applicant:** Guidant Corp.
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** Oct 28, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The HI-TORQUE WHISPER Guide Wire is intended to aid in the placement of a Guidant implantable coronary venous lead in the coronary venous vasculature.

## Device Story

Guide wire designed to assist placement of implantable coronary venous leads; used in coronary venous vasculature. Features stainless steel tapered core; flexible tip; polyurethane distal jacket; PTFE proximal jacket; hydrophilic coating for lubricity. Available in 0.014" diameter, 190 cm length; straight or pre-formed 'J' (CS-J) distal tips. Straight tips manually shapeable; CS-J tips provide pre-formed access. Operated by physicians in clinical settings. Device provides trackability and support for lead delivery; facilitates navigation through venous anatomy.

## Clinical Evidence

Bench testing only. Verification testing demonstrated that the guide wires met established acceptance criteria and performed equivalently to the predicate devices.

## Technological Characteristics

Stainless steel tapered core; polyurethane distal jacket; PTFE proximal jacket; hydrophilic coating. 0.014" diameter; 190 cm length. Straight or pre-formed 'J' distal tips. Sterilization via identical processes as predicate devices.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- HI-TORQUE WHISPER Guide Wires ([K020340](/device/K020340.md))
- HI-TORQUE WHISPER Guide Wires ([K021285](/device/K021285.md))

## Reference Devices

- Guidant VI Whisper LS and MS (Enhanced Radiopacity) guidewires

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# GUIDANT

and the state of the research and

and the may

..

..............................................................................................................................................................................

OCT 2 8 2002

K023300

| 7. 510(K) SUMMARY           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name:           | Guidant Corporation<br>CRM Division                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Submitter's Address:        | 4100 Hamline Avenue<br>Mail Stop F330<br>St. Paul, Minnesota 55112                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Telephone:                  | (651) 582-6691                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Fax:                        | (651) 582-5134                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Contact Person:             | Janell A. Colley                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Date Prepared:              | October 1, 2002                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Trade Name:          | HI-TORQUE® WHISPER™ LS (Light Support) Guide<br>Wire with Hydrocoat Hydrophilic Coating,<br>Model 6726<br>HI-TORQUE® WHISPER™ MS (Medium Support) Guide<br>Wire with Hydrocoat Hydrophilic Coating,<br>Model 6737<br>HI-TORQUE® WHISPER™ MS CS-J (Medium Support<br>Coronary Sinus J) Guide Wire with Hydrocoat Hydrophilic<br>Coating,<br>Model 6738<br>HI-TORQUE® WHISPER™ ES (Extra Support) Guide Wire<br>with Hydrocoat Hydrophilic Coating,<br>Model 4482<br>HI-TORQUE® WHISPER™ ES CS-J (Extra Support<br>Coronary Sinus J) Guide Wire with Hydrocoat Hydrophilic<br>Coating,<br>Model 4483 |
| Device Common Name:         | Guide Wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Classification Name: | Catheter Guide Wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Classification:      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{1}------------------------------------------------

#### Summary of Substantial Equivalence:

The design, materials, method of delivery and intended use features of the HI-TORQUE WHISPER LS, MS, ES, MS CS-J, and ES CS-J Guide Wires are substantially equivalent with regard to these features in their predicate device, the HI-TORQUE WHISPER Guide Wires (K020340/March 1, 2002, and K021285/May 2, 2002).

## Device Description:

The general guide wire design consists of a stainless steel tapered core with a flexible tip, jacketed by polyurethane on the distal portion of the wire and polytetrafluoroethylene (PTFE) on the proximal portion of the wire. The polyurethane is subsequently covered with a hydrophilic coating to increase lubricity. The guide wires are available with either a straight distal tip that is shapeable, or as a pre-formed 'J' distal tip shaped specifically for accessing the coronary sinus (designated 'CS-J'). The straight tip can be manually shaped to accommodate specific patient anatomy or physician preference; the CS-J shape provides the convenience of an existing 'J' shape without manual shaping.

The HI-TORQUE WHISPER LS, MS/MS CS-J and ES/ES CS-J guide wires with Hydrocoat hydrophilic coating are guide wires with a 0.014" maximum diameter and a length of 190 cm. These wires are constructed with a high tensile (Hyten) stainless steel core wire having a nominal diameter of 0.0128" (LS/MS versions) or 0.0132" (ES versions). The distal segment of each guide wire consists of a series of unique tapers and grinds which reduce the diameter and support at the distal core, thus yielding the desired flexibility and performance characteristics.

## Intended Use:

The HI-TORQUE WHISPER Guide Wire is intended to aid in the placement of a Guidant implantable coronary venous lead in the coronary venous vasculature.

## Technological Characteristics:

Comparisons of the proposed and predicate device show that the technological characteristics such as materials, performance characteristics, sterilization and packaging are identical or substantially equivalent to the currently marketed device.

## Performance Data:

The results of the verification testing demonstrate that the HI-TORQUE WHISPER LS, MS, ES, MS CS-J, and ES CS-J Guide Wires meet the established acceptance criteria and perform in a manner equivalent to the predicate device. No new safety or effectiveness issues were raised during the testing program.

## Conclusions:

The Guidant CRM HI-TORQUE WHISPER LS, MS, ES, MS CS-J, and ES CS-J Guide Wires have the same intended use, materials, technological characteristics, performance properties, and use identical sterilization processes as the Guidant VI Whisper LS and MS (Enhanced Radiopacity) guidewires and the Guidant CRM HI-TORQUE guidewires; therefore, there are no new safety or effectiveness issues. The modified HI-TORQUE WHISPER Guide Wires are substantially equivalent to the predicate devices.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 -

OCT 28 2002

Guidant Corporation c/o Ms. Janell A. Colley Regulatory Affairs Associate Cardiac Rhythm Management 4100 Hamline North St. Paul. MN 55112-5798

Re: K023300

Trade Name: Hi-Torque Whisper LS, MS, MS CS-J, ES, and ES CS-J Guide Wires with Hydrocoat Hydrophilic Coating Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (two) Product Code: DQX

Dated: October 1, 2002 Received: October 3, 2002

Dear Ms. Colley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA mav publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Ms. Janell A. Colley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Odell Tun

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

2. Indications for Use Statement

| 510(k) Number (if known) |                                                                                                                                                        |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name              | HI-TORQUE WHISPER LS, MS, MS CS-J, ES, and ES CS-J Guide Wires with Hydrocoat Hydrophilic Coating                                                      |
| Indications for Use      | The HI-TORQUE WHISPER Guide Wire is intended to aid in the placement of a Guidant implantable coronary venous lead in the coronary venous vasculature. |

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

1

: 

11. 11. 11. 11.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number K023300

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ Prescription Use OR (Per 21 CFR 801.109) Guidant CRM HI-TORQUE WHISPER Guide Wire

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K023300](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K023300)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
