← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K011905

# SNIPER ELITE MODEL-OR-E3872XX SERIES (K011905)

_Maxxim Medical · DQX · Aug 28, 2001 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K011905

## Device Facts

- **Applicant:** Maxxim Medical
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** Aug 28, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

Sniper Elite™ Hydrophilic Guidewires are designed for use in angiographic procedures to introduce/position catheters and interventional devices within the vasculature.

## Device Story

The Sniper Elite™ Hydrophilic Guidewire is a vascular access tool used by clinicians during angiographic procedures. It features a super-elastic Nitinol core for kink resistance, covered by a polymer cladding impregnated with a radiopaque agent. A platinum/8% tungsten coil at the distal tip provides enhanced fluoroscopic visibility. The device surface includes a durable hydrophilic coating, which is more lubricous distally to facilitate navigation through the vasculature. The guidewire is used to introduce and position catheters and other interventional devices. It is supplied sterile for single use and includes a torque handle to assist the clinician in maneuvering the wire. By providing a low-friction, flexible, and radiopaque path, the device enables precise placement of interventional equipment, potentially improving procedural efficiency and patient outcomes during vascular interventions.

## Clinical Evidence

Bench testing only. Testing conducted in accordance with FDA Coronary and Cerebrovasculature Guidewire Guidance (#964) and ISO 11070. Evaluated parameters included lubricity, coating adherence, torque strength, torqueability, tip softness, tip flexibility, tensile strength, fracture resistance, kink resistance, radiopacity, and catheter compatibility.

## Technological Characteristics

Core: Nitinol. Cladding: Polymer impregnated with radiopaque agent. Distal tip: Platinum/8% tungsten coil. Coating: Hydrophilic. Sterilization: Ethylene oxide (EtO) per ANSI/AAMI/ISO 11135-1994. Form factor: Guidewire with torque handle.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Maxxim Medical Heparin Coated Stainless Steel/Teflon Coated Spring Guidewire ([K832030](/device/K832030.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# AUG 2 8 2001

SECTION D 510(k) SUMMARY

| 510(k) Number:       | K011905                                                                                               |
|----------------------|-------------------------------------------------------------------------------------------------------|
| Trade Name:          | Sniper Elite™ Hydrophilic Guide Wire                                                                  |
| Common Name:         | Guidewire                                                                                             |
| Classification Name: | Catheter Guidewire (per CFR 21 Part 870.1330, 74 DQX)                                                 |
| Product Code:        | DQX                                                                                                   |
| Classification:      | Class II                                                                                              |
| Submitted by:        | Maxxim Medical<br>1445 Flat Creek Rd.<br>Athens, TX 75751<br>Phone: 903-675-9321<br>Fax: 903-677-9397 |
| Contact person:      | Gail Doherty, Compliance Manager                                                                      |
| Date prepared:       | 6/18/01                                                                                               |

# Legally marketed devices to which equivalence is claimed:

Maxxim Medical Heparin Coated Stainless Steel/Teflon Coated Spring Guidewire, K832030

# Description of Device:

The Sniper Elite™ Hydrophilic Guidewire is constructed from a super elastic, Nitinol core wire. A plastic cladding is applied over the Nitinol core, with a hydrophilic coating covering the plastic cladding. The polymer cladding is impregnated with a radiopacive agent and a platinum/8% tungsten coil is located in the distal tip for enhanced contrast under fluoroscopy.

{1}------------------------------------------------

## Scientific concepts that form the basis for the device:

This guide wire has a core made from Nitinol, a metal that has a high resistance to kinking. The Nitinol core is cladded with a polymer of low enough durometer not to inhibit the super-elastic properties of Nitinol. The plastic cladding is coated with a durable hydrophilic coating, to provide a low friction surface when wet. The radiopacive agent that is impregnated in the polymer cladding and a platinum/8% tungsten coil in the distal tip provide enhanced radiopacity, when imaged under fluoroscopy. The hydrophilic coating is more lubricous distally than proximally to promote increased handling characteristics. The guidewire tip is color coded to indicate the size of the wire.

#### Intended Use of Device:

Sniper Elite™ Hydrophilic Guidewires are designed for use in angiographic procedures to introduce/position catheters and interventional devices within the vasculature.

# Comparison of Fundamental Scientific Technology to legally marketed device:

The Sniper Elite™ Hydrophilic Guidewire is being compared to the Maxxim Medical Heparinized, Teflon Coated Spring Guidewire.

The predicate device has:

- 1) An internal core that is tapered distally for increased tip softness, as does the Shiper Elite™ Hydrophilic Guidewires.
- 2) A stainless steel coil covering the core that extends the length of the wire, while the Sniper Elite™ possesses a polymer cladding that also extends the length of the wire, which incorporates the distal radiopaque coil. The function of the stainless steel coil and the cladding are to retain a consistent outer dimension for the wire and provide radiopacity, while retaining the flexibility imparted to the wire by the core wire.
- 3) A Heparinized-Teflon coating and the Sniper Elite™ has a hydrophilic coating for increased lubricity through the vasculature.

#### Performance Data:

Testing was conducted in accordance with the FDA's Coronary and Cerebrovasculature Guidewire Guidance (FDA document #964, 1-1995) and the ISO Standard 11070, Sterile single-use intravascular catheter introducers (1998-5-1, guide wire section). The following tests demonstrated substantial equivalence:

- Lubricity ו
- Coating Adherence -
- Torque Strength
- Torqueablility
- Tip Softness
- Tip Flexibility

{2}------------------------------------------------

- Tensile Strength
- Fracture Test
- Kink Resistance
- Radiopacity
- Catheter Compatibility

Test results demonstrate that the Sniper Elite™ Hydrophilic Guidewire is substantially equivalent to the Maxxim Medical Heparinized, Teflon Coated Spring Guidewire.

## Packaging and Sterilization Information

Maxxim Medical packages the Maxxim Medical Heparinized Teflon Coated Spring Guidewire (enclosed in dispenser) in a single put-up pouch and sterilized. The Spring Guidewire is packaged in a Tyvek pouch, 10 pouches to a shelf carton as the typical packaging configuration. Sterilization for Maxxim Medical is by a validated ethylene oxide sterilization (EtO) method that is referenced in ANSI/AAMI/ISO 11135-1994 "MEDICAL DEVICES- Revalidation and Routine Control of Ethylene Oxide Sterilization." The packaging and sterilization of the Sniper Elite™ will be identical to that for the predicate with a couple exceptions (torque handle included and 5 pouches/ shelf carton). The Sniper Elite™ is packaged with a torque handle (made by Maxxim Medical) in a Tyvek pouch, 5 pouches to a shelf carton and 10 shelf cartons in a shipper carton.

### CONCLUSIONS:

Maxxim Medical concludes that the Sniper Elite™ Hydrophilic Guidewires are substantially equivalent to the Maxxim Medical Heparinized, Teflon Coated Spring Guidewire:

- 1) functionally and
- 2) with regards to safety and efficacy

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three swooping lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 8 2001

Ms. Gail Doherty Manager Maxxim Medical 1445 Flat Creek Rd. Athens. TX 75751

K011905 Re: Sniper Elite Model-Or-E3872XX Series Regulation Number: 870.1330 Regulatory Class: II Product Code: DQX Dated: July 24, 2001 Received: August 2, 2001

Dear Ms. Doherty:

We have reviewed your Section 510(k) notification of intent to market the devidentians for t we have leviewed your becally in the device is substantially equivalent (for the indications for use above and we nave determined the actested predicate devices marketed in interstate commerce stated in the encrosule) to logally manote of the Medical Device Amendments, or to devices that provision in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosment Act (Act). Tourmaly, mercess), increases, and include requirements for annual provisions of the Act. "The goneral consiness of the practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remancer rippe raty in the Code of Federal Regulations, Title 21, Parts 800 to 895. arrouning your as novelant determination assumes compliance with the Current Good A substantially oqurraters as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspections, and Programs Drag in may result in regulatory action. In addition, FDA may publish comply with the GMT regaration into your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Gail Doherty

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

Kollaos 510(k) Number (if known):_

Device Name: Sniper Elite Hydrophilic Guidewire

Indications For Use:

Sniper Elite Hydrophilic Guidewires are designed for use in angiographic procedures to introduce/position catheters and interventional devices within the vasculature.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number K611906

(Optional Format 3-10-98)

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K011905](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K011905)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
