ACUNAV DIAGNOSTIC ULTRASOUND CATHETER

K992631 · Acuson Corp. · DQO · Dec 15, 1999 · Cardiovascular

Device Facts

Record IDK992631
Device NameACUNAV DIAGNOSTIC ULTRASOUND CATHETER
ApplicantAcuson Corp.
Product CodeDQO · Cardiovascular
Decision DateDec 15, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1200
Device ClassClass 2

Indications for Use

For use directly within the vasculature and/or the right heart for intravascular or intracardiac ultrasound imaging. The device is specifically indicated for use in visualization of vascular anatomy, cardiac and great vessel anatomy and physiology, or other devices in the heart, as well as measurement of blood flow. The company anticipates that the device will be used for a variety of intravascular and intracardiac imaging applications, consistent with its intended use, including vascular stent placement, monitoring left ventricular function post-surgery; identifying congenital abnormalities before therapeutic procedures; visualizing the relative orientation of diagnostic and therapeutic catheters; and visualizing procedures such as transseptal insertions of other catheters, valvuloplasties, balloon septostomies, septal defect closures, and pacemaker or defibrillator lead insertion or extraction.

Device Story

AcuNav is an ultrasound-tipped catheter for intravascular and intracardiac imaging. It utilizes a single-use, disposable phased-array ultrasonic transducer at the distal tip. The catheter is 10 French in diameter and 90 cm long, with a four-way deflection mechanism (left-right, anterior-posterior) for precise positioning. It connects to an Acuson ultrasound imaging platform (e.g., Aspen or Sequoia) which generates acoustic waves and processes signals for display. The transducer operates at frequencies between 4.0 and 10.0 MHz, providing a 2D vector image in a plane parallel to the catheter axis. Used by physicians in clinical settings to visualize cardiac structures, monitor post-surgical function, identify congenital abnormalities, and guide interventional procedures like stent placement, valvuloplasties, and lead insertions. The real-time visual feedback assists clinicians in navigating the heart and vasculature, potentially improving procedural accuracy and patient outcomes.

Clinical Evidence

Substantial equivalence supported by performance testing, animal testing, and human clinical testing. No specific quantitative clinical metrics (e.g., sensitivity, specificity) are provided in the summary document.

Technological Characteristics

10 French, 90 cm long ultrasound-tipped catheter; phased-array imaging transducer; 4.0-10.0 MHz frequency range; 4-way distal deflection mechanism (orthogonal planes); disposable; connects to external Acuson ultrasound imaging platform; biocompatible materials.

Indications for Use

Indicated for intra-cardiac and intra-vascular ultrasound imaging in the right heart and vasculature for visualization of anatomy, physiology, and other devices, and measurement of blood flow in patients requiring such diagnostic monitoring.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 199263 # DEC 1 2 1999 ## 510(k) Summary #### Sponsor: Acuson Corporation 1220 Charleston Road PO Box 7393 Mountain View, California 94039-7303 Telephone: (650) 969-9112 Facsimile: (650) 962-8018 #### Contact Person: Gladys May-Cooper Acuson Corporation 1220 Charleston Road PO Box 7393 Mountain View, California 94039-7303 Telephone: (650) 969-9112 Facsimile: (650) 962-8018 ## or Howard M. Holstein, Esq. Hogan & Hartson, LLP 555 Thirteenth Street, N.W. Washington, DC 20004 Tel: (202) 637-5813 Fax: (202) 637-5910 ## Submission Date: August 5, 1999 ## Device Name: AcuNav™ Diagnostic Ultrasound Catheter ## Classification: Diagnostic intravascular catheters and ultrasonic transducers-class II (21 C.F.R. §§ 870.1200 and 892.1570) {1}------------------------------------------------ ## Predicate Devices: - 1. Boston Scientific Corporation's Sonicath Ultra™ Imaging Catheter (9 Fr/ 9 MHz) (K970049), (used in conjunction with EP Technologies' Galaxy Intravascular Ultrasound Imaging System (K980851)) - 2. Cardiovascular Imaging Systems' (CVIS) Insight Catheter for Intracardiac Use (10 Fr) (K921148). ## Device Description: The AcuNav is an ultrasound-tipped catheter device which is used directly within the vasculature and/or the right heart for intravascular or intracardiac ultrasound imaging. The AcuNav incorporates a single-use, disposable ultrasonic phased-array imaging transducer, which must be used in conjunction with an Acuson ultrasound imaging platform to generate the acoustic waves and process the information for display to the physician. The catheter is 10 French in diameter and 90 cm in insertable length. The distal portion of the catheter can be deflected in four directions in two orthogonal planes: left-right (in a plane perpendicular to the image plane) and anterior-posterior (in a plane coincident with the image plane). The distal end of the catheter contains the transducer, which is oriented to provide a two-dimensional (90 degree vector) image in a plane parallel to the axis of the catheter. The transducer offers all imaging modes at frequencies between 4.0 and 10.0 MHz. {2}------------------------------------------------ ## Intended Use: For use directly within the vasculature and/or the right heart for intravascular or intracardiac ultrasound imaging. The device is specifically indicated for use in visualization of vascular anatomy, cardiac and great vessel anatomy and physiology, or other devices in the heart, as well as measurement of blood flow. The company anticipates that the device will be used for a variety of intravascular and intracardiac imaging applications, consistent with its intended use, including vascular stent placement, monitoring left ventricular function post-surgery; identifying congenital abnormalities before therapeutic procedures; visualizing the relative orientation of diagnostic and therapeutic catheters; and visualizing procedures such as transseptal insertions of other catheters, valvuloplasties, balloon septostomies, septal defect closures, and pacemaker or defibrillator lead insertion or extraction. ## Substantial Equivalence: Acuson's AcuNav Catheter and the predicate devices have the same intended use and very similar principles of operation and technological characteristics. All of the devices are intended for use in intravascular and/or intracardiac imaging. The minor technological differences between the AcuNav Catheter and the predicate devices, i.e., in acoustic parameters, transducer configuration, imaging format, steering mechanism, materials, or 3 {3}------------------------------------------------ sterilization methods, do not raise any new questions of safety or effectiveness, as confirmed by performance testing, animal testing, and human clinical testing. The AcuNav Catheter also has been demonstrated to be biocompatible and non-pyrogenic. Therefore, Acuson's AcuNav Catheter is substantially equivalent to the predicate devices. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 8 2001 8 2001 Acuson Corporation c/o Mr. Howard M. Holstein Hogan and Hartson, L.L.P. Columbia Square 555 Thirteenth Street, N.W. Washington, D.C. 20004 Re: K992631 > Trade Name: AcuNav Diagnostic Ultrasound Catheter Regulation Number: 21 CFR 870.1200 and 892.1570 Regulation Name: Intravascular Diagnostic Catheter Diagnostic Ultrasound Transducer Regulatory Class: II (two) Product Code: DQO and ITW Dated: November 12, 1999 Received: November 12, 1999 Dear Mr. Holstein: This letter corrects our substantially equivalent letter of December 15, 1999, regarding the indications for use. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Howard M. Holstein Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ | AcuNav™ Diagnostic Ultrasound System | | | | | | | | | | |--------------------------------------------------------------------------------------------------|---|---|---|-----|-----|---------------|-----------------------|--------------------|------------------------| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | Clinical Applications | A | B | M | PWD | CWD | Color Doppler | Power (Ampl.) Doppler | Combined (Specify) | Other Harmonic Imaging | | Ophthalmic | | | | | | | | | | | Fetal | | P | P | P | P | P | P | *P | P | | Abdominal | | P | P | P | P | P | P | *P | P | | Intra-operative<br>(vascular) | | P | P | P | P | P | P | *P | P | | Intra-operative<br>Neurological | | | | | | | | | | | Pediatric | | P | P | P | P | P | P | *P | P | | Small Organ<br>- Thyroid -<br>- Breast<br>- Testicle | | P | P | P | P | P | P | *P | P | | Neonatal Cephalic | | P | P | P | P | P | P | *P | P | | Adult Cephalic | | P | P | P | P | P | P | *P | P | | Cardiac | | P | P | P | P | P | P | *P | P | | Trans-esophageal | | P | P | P | P | P | P | *P | P | | Trans-Rectal | | | | | | | | | | | Trans-Vaginal | | P | P | P | P | P | P | *P | P | | Trans-Urethral | | | | | | | | | | | Intra-Luminal | | N | N | N | N | N | N | *N | N | | Peripheral Vascular | | P | P | P | P | P | P | *P | P | | Laparoscopic | | | | | | | | | | | Musculo-Skeletal<br>Conventional | | P | P | P | P | P | P | *P | P | | Musculo-Skeletal<br>Superficial | | P | P | P | P | P | P | *P | P | | Other (Intra-Cardiac) | | N | N | N | N | N | N | *N | N | ## Diagnostic Ultrasound Indications for Use Form For Suration; P = previously cleared by FDA; E = added under Appendix E Additional Comments: Adultional Colinnents: * Combinations: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+Color Doppler, B+PWD+Color Contonations: D+M, B+P+ver Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler WEI DOPPIET (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Device Use (Per 21 CFR 801.109) Division of Cardiovascular & Respiratory Devices 510(k) Number K992631 Acuson Corporation IIIARDC - 59177/1 - #981901 v1 May 6, 1999 L {7}------------------------------------------------ #### Diagnostic Ultrasound Indications for Use Form For AcuNav™ Diagnostic Ultrasound Catheter Intended Use: The AcuNav™ Diagnostic Ultrasound Catheter is intended for intra-cardiac and intra-Intended Ose. The Acarder - Diagnosat vessel anatomy and physiology as well as visualization of other devices in the heart. The AcuNav™ is intended for use in right heart only. Ultrasound System: Aspen™ Transducer: AcuNav™ | Clinical<br>Applications | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Ampl.)<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) | | |----------------------------------------------------|---|---|---|-----|-----|------------------|-----------------------------|-----------------------|--------------------|--| | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intra-operative<br>- vascular | | | | | | | | | | | | Intra-operative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>- Thyroid<br>- Breast<br>- Testicle | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | N | N | N | N | N | N | *N | N | | | Trans-esophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Luminal | | N | N | N | N | N | N | *N | N | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-Skeletal<br>Conventional | | | | | | | | | | | | Musculo-Skeletal<br>Superficial | | | | | | | | | | | | Other (Intra-<br>Cardiac) | | N | N | N | N | N | N | *N | N | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments: * Combinations: B+M, B+PWD, B+Color Doppler, B+M+Color Doppler, B+PWD+Color Contonations: D+M)/ D Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Device Use (Per 21 CFR 801.109) Acuson Corporation \\\ARDC - 59177/1 - #981901 v1 May 6, 1999 {8}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form For | Sequoia™ Diagnostic Ultrasound System | | | | | | | | | | |--------------------------------------------------------------------------------------------------|---|---|-----------|------------------|-----------------------|-----------------------|------------------------------|--|--| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | Clinical<br>Applications | A | B | M PWD CWD | Color<br>Doppler | Power Doppler (Ampl.) | Combined<br>(Specify) | Other<br>Harmonic<br>Imaging | | | | Ophthalmic | | | | | | | | | | | Fetal | | P | P | P | P | *P | P | | | | Abdominal | | P | P | P | P | *P | P | | | | Intra-operative<br>(vascular) | | P | P | P | P | *P | P | | | | Intra-operative<br>Neurological | | | | | | | | | | | Pediatric | | P | P | P | P | *P | P | | | | Small Organ<br>- Thyroid-<br>- Breast<br>- Testicle | | P | P | P | P | *P | P | | | | Neonatal Cephalic | | P | P | P | P | *P | P | | | | Adult Cephalic | | P | P | P | P | *P | P | | | | Cardiac | | P | P | P | P | *P | P | | | | Trans-esophageal | | P | P | P | P | *P | P | | | | Trans-Rectal | | | | | | | | | | | Trans-Vaginal | | P | P | P | P | *P | P | | | | Trans-Urethral | | | | | | | | | | | Intra-Luminal | | N | N | N | N | *N | N | | | | Peripheral Vascular | | P | P | P | P | *P | P | | | | Laparoscopic | | | | | | | | | | | Musculo-Skeletal<br>Conventional | | P | P | P | P | *P | P | | | | Musculo-Skeletal<br>Superficial | | P | P | P | P | *P | P | | | | Other (Intra-<br>Cardiac) | | N | N | N | N | *N | N | | | Caranae) N = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments: Adultional Colinnents: * Combinations: B+M, B+PWD, B+Color Doppler, B+M+Color Doppler, B+PWD+Color " Commilations: DTM, DTTWD, B+CWD, B+CHD, B+Color Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler WEIT DOPPIE! (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Device Use_ (Per 21 CFR 801.109) Acuson Corporation \\\ARDC - 59177/1 - #981904 v1 May 6, 1999 V {9}------------------------------------------------ #### Diagnostic Ultrasound Indications for Use Form For AcuNav™ Diagnostic Ultrasound Catheter Intended Use: The AcuNav™ Diagnostic Ultrasound Catheter is intended for intra-cardiac and intra-Intelided Ose. The Acurav - Blaglioste Oxase anatomy and physiology as well as visualization of other luminal viSualization of carants ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ultrasound System: SequoiaTM Transducer: AcuNav™ | Clinical<br>Applications | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Ampl.)<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) | |----------------------------------------------------|---|---|---|-----|-----|------------------|-----------------------------|-----------------------|--------------------| | Ophthalmic | | | | | | | | | | | Fetal | | | | | | | | | | | Abdominal | | | | | | | | | | | Intra-operative<br>- vascular | | | | | | | | | | | Intra-operative<br>Neurological | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ<br>- Thyroid<br>- Breast<br>- Testicle | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Cardiac | | N | N | N | N | N | N | *N | N | | Trans-esophageal | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | Intra-Luminal | | N | N | N | N | N | N | *N | N | | Peripheral Vascular | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Musculo-Skeletal<br>Conventional | | | | | | | | | | | Musculo-Skeletal<br>Superficial | | | | | | | | | | | Other (Intra-<br>Cardiac) | | N | N | N | N | N | N | *N | N | N = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments: Additional Colinnents. * Combinations: B+M, B+PWD, B+Color Doppler, B+M+Color Doppler, B+PWD+Color Collibulations: D+WD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler WEIT DOPPIEI (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Device Use Acuson Corporation \\\ARDC - 59177/1 - #981904 v1 May 6, 1999 (Per 21 CFR 801.109) L
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...