RADIFOCUS OPTITORQUE

K992051 · Terumo Medical Corp. · DQO · Aug 10, 1999 · Cardiovascular

Device Facts

Record IDK992051
Device NameRADIFOCUS OPTITORQUE
ApplicantTerumo Medical Corp.
Product CodeDQO · Cardiovascular
Decision DateAug 10, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1200
Device ClassClass 2

Indications for Use

The RADIFOCUS® Optitorque Angiographic Catheter is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the vascular system.

Device Story

The RADIFOCUS® Optitorque Angiographic Catheter is a manual vascular access device used by clinicians during angiographic procedures. It consists of a two-layer construction featuring a stainless steel mesh sandwiched between polyurethane polyamide elastomer layers, which contain barium sulfate for fluoroscopic visibility. The device includes a flexible, supple polyurethane 'soft-tip' permanently welded to the shaft to minimize vessel wall trauma. The catheter is used to deliver radiopaque contrast media and therapeutic agents to specific vascular sites. It is operated manually by a physician. The device is designed for limited contact (≤24 hours) with circulating blood.

Clinical Evidence

Bench testing only. Verification tests included tensile strength (shaft, hub, soft tip), rigidity, maximum pressure, torque transmission, flow rate, tip flexibility, delivery of therapeutic agents, and biocompatibility testing per ISO-10993. No clinical data was required or provided.

Technological Characteristics

Two-layer construction: stainless steel mesh between polyurethane polyamide elastomer layers. Contains barium sulfate for radiopacity. Flexible polyurethane soft-tip (0.5-2.5mm). Sizes: 4-6 Fr. Lengths: 650-1200 mm. Max injection pressure: 750-1000 psi. Manual operation. Biocompatible materials tested per ISO-10993.

Indications for Use

Indicated for use in angiographic procedures to deliver radiopaque media and therapeutic agents to selected sites in the vascular system.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 10 1999 K992051 ### SUMMARY AND CERTIFICATION II. # SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE #### A. Device Name RADIFOCUS® Optitorque Proprietary Name: Classification Name: Angiographic Catheter Common Name: Angiographic Catheter #### B. Intended Use The RADIFOCUS® Optitorque Angiographic Catheter is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the vascular system. Note: This is the same intended use as the RADIFOCUS® Angiographic Catheter, K863137/A. #### C. Device Description The RADIFOCUS® Optitorque Angiographic Catheter is a two-layer construction comprised of a stainless steel mesh sandwiched between layers of polyurethane polyamide elastomer. The polyurethane polyamide elastomer contains barium sulfate for visibility and contrast under fluoroscopy. The Catheter has a "soft-tip" whose purpose is to minimize trauma to the vessel wall. The soft-tip is a flexible, supple polyurethane tip that is permanently welded to the catheter shaft. The soft tip is approximately 0.5~1.5 mm long on the 4Fr. Optitorque Catheter. It is approximately 1.0~2.5mm long on the 5 and 6Fr. Optitorque Catheters. #### D. Substantial Equivalence The RADIFOCUS® Optitorque Angiographic Catheter submitted in this 510(k) is substantially equivalent in intended use, design, technology/principles of operation, materials, and performance to the cleared RADIFOCUS® Angiographic Catheter, (K863137/A), the Terumo Angiographic Catheter (K915414) and the Cordis Infiniti Angiographic Catheter (K960975). #### E. Principle Of Operation / Technology The RADIFOCUS® Optitorque Angiographic Catheter is operated manually or by a manual process. #### F. Design / Materials Differences in materials between the RADIFOCUS® Optitorque Angiographic Catheter and the RADIFOCUS® Angiographic Catheter, K863137/A, raise no new issues of safety and effectiveness. {1}------------------------------------------------ #### G. Specifications | | Optitorque Angiographic<br>Catheter | Angiographic Catheter,<br>K863137/A | |-------------------------------|----------------------------------------------------|--------------------------------------------------------------------------| | Available Sizes | 4 - 6 Fr. | 5 – 8 Fr. | | Catheter Length | 650 - 1200 mm | 500 – 1300 mm | | Maximum Injection<br>Pressure | 4 Fr.: 750 psi<br><br>5, 6 Fr.: 1000 psi<br><br>-- | --<br><br>5, 6 Fr.: 1000 psi<br><br>7 Fr.: 800 psi<br><br>8 Fr.: 700 psi | #### H. Performance The following verification tests were performed to demonstrate the substantial equivalence of the Optitorque Angiographic Catheter to the Angiographic Catheter, K863137/A. - Tensile strength on the shaft, hub and soft tip . - Rigidity . - Maximum Pressure Test ● - Torque Transmission Test . - Flow Rate Test . - . Tip Flexibility Test - Delivery of Therapeutic Agents . - Biocompatability Tests . None of the data raises any new issues of safety and effectiveness. Additionally, a risk analysis was conducted and there were no new issues of safety and effectiveness. Therefore the performance of the Optitorque Angiographic Catheter is substantially equivalent to the performance of the Angiographic Catheter, cleared under K863137/A. #### l. Additional Safety Information ・.. Manufacturing controls include visual, functional (which including dimensional), chemical, biological and sterility tests. The Optitorque Angiographic Catheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hrs). The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". Results of the testing demonstrate that the blood contacting materials are biocompatible. {2}------------------------------------------------ #### J. Conclusion The RADIFOCUS® Optitorque Angiographic Catheter is substantially equivalent in intended use, design, technology / principles of operation, materials and performance to the RADIFOCUS® Angiographic Catheter (K863137/A), the Terumo Angiographic Catheter (K915414) and the Cordis Infiniti Angiographic Catheter (K960975). Differences between the devices do not raise any significant issues of safety or effectiveness. . . - - | Date Prepared: | June 17, 1999 | |----------------|----------------------------------------------------------------------------------------------------------------------------------------| | Prepared By: | Yuk-Ting Lewis<br>Senior Regulatory Specialist | | Prepared For: | Terumo Medical Corporation<br>125 Blue Ball Road<br>Elkton, MD 21921<br>Phone: (410) 392-7213 or (410) 392-7231<br>Fax: (410) 398-6079 | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 10 1999 Ms. Yuk-Ting Lewis Senior Requlatory Specialist Terumo Medical Corporation 125 Blue Ball Road Elkton, MD 21921 Re : K992051 > RADIFOCUS® Optitorque™ Angiographic Catheter Trade Name: Requlatory Class: II Product Code: DQO Dated: July 12, 1999 Received: July 13, 1999 Dear Ms. Lewis: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation {4}------------------------------------------------ Page 2 - Ms. Yuk-Ting Lewis you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Note: This is the same intended use as the unmodified device, K863137/A 510(k) Number (if known): RADIFOCUS® Optitorque™ Angiographic Catheter Device Name: ## Indications For Use: The RADIFOCUS® Optitorque Angiographic Catheter is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the vascular system. Ohtyr M. for Callahan. ring Off Fascilla: Respiratory, 510(k) Number (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...