← Product Code [DQO](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQO) · K170406

# Plato MICROCATH 27B Microcatheter (K170406)

_Scientia Vascular, LLC · DQO · Apr 18, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQO/K170406

## Device Facts

- **Applicant:** Scientia Vascular, LLC
- **Product Code:** [DQO](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQO.md)
- **Decision Date:** Apr 18, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Plato MICROCATH 27B Microcatheter is intended for the introduction of interventional devices and infusion of diagnostic agents to the peripheral and neurovascular systems. The catheter is also intended for the introduction of therapeutic agents to the peripheral system. The catheter is not intended for use in the coronary vasculature.

## Device Story

Plato MICROCATH 27B is a single-lumen, variable-stiffness microcatheter; 150 cm length; 3.4F proximal to 3.2F distal diameter. Used with guide catheter and steerable guidewire (up to 0.025") for navigation of peripheral and neurovasculature. Features hydrophilic coating on distal shaft for lubricity; radiopaque tip marker for fluoroscopic visualization; steam-shapeable distal tip. Operated by physicians in clinical settings. Provides mechanical support to prevent collapse in tortuous vessels; facilitates delivery/withdrawal of interventional devices and infusion of diagnostic/therapeutic agents. Benefits patient by enabling access to complex vascular anatomy for minimally invasive procedures.

## Clinical Evidence

Bench testing only. Performance evaluated via ISO 10555-1:2013 standards, including tensile strength, coating lubricity/integrity, torque to failure, burst pressure, kink resistance, and simulated use in tortuous neurovascular models. Cadaver testing performed to compare access time and device deployment/retrieval against predicate. Results confirm comparable performance.

## Technological Characteristics

Single-lumen, variable-stiffness microcatheter. Materials: polymeric laminate with embedded reinforcing braid; polycarbonate female Luer adapter. Distal end features polymeric skeletal support. Dimensions: 150 cm length, 3.4F-3.2F OD, accommodates 0.025" guidewires. Hydrophilic coating on distal shaft. Sterilization: not specified. Standards: ISO 10555-1:2013, ASTM D747-10 (flexural fatigue/stiffness), ASTM F88-00/F2096-04 (packaging). MRI Unsafe.

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- ev3 Marksman Catheter ([K111490](/device/K111490.md), [K091559](/device/K091559.md))

## Reference Devices

- Scientia Vascular, LLC Plato MICROCATH 27 Microcatheter ([K143398](/device/K143398.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a family.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 18, 2017

Scientia Vascular LLC % Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street. NW Buffalo, Minnesota 55313

Re: K170406

Trade/Device Name: Plato MICROCATH® 27B Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO, KRA, DQY Dated: April 13, 2017 Received: April 14, 2017

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K170406

Device Name Plato MICROCATH® 27B Microcatheter

#### Indications for Use (Describe)

The Plato MICROCATH 27B Microcatheter is intended for the introduction of interventional devices and infusion of diagnostic agents to the peripheral and neurovascular systems. The catheter is also intended for the introduction of therapeutic agents to the peripheral system. The catheter is not intended for use in the coronary vasculature.

| Type of Use (Select one or both, as applicable)                                                                        |                                                                                                                       |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Scientia. The logo features a stylized, blue, overlapping "S" shape, resembling a chain link. Below the symbol, the word "SCIENTIA" is written in a modern, sans-serif font, with the letters appearing in a light gray color.

## 510(K) SUMMARY (Per 21 CFR 807.92)

## SCIENTIA VASCULAR LLC PLATO MICROCATH® 27B MICROCATHETER

| 510(k) Sponsor:                          | Scientia Vascular LLC<br>3487 West 2100 South Suite 100<br>West Valley City, UT 84119<br>Tel: (775) 657-6330<br>Fax: (888) 503-6529                                |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                          | David Sabodski, Director of Quality Assurance<br>Tel: 801.573.5897<br>Fax: 888.503.6529<br>E-mail: dsabodski@scientiavascular.com                                  |
| Date Prepared:                           | April 10, 2017                                                                                                                                                     |
| Prepared by:                             | Phil Triolo PhD, RAC<br>President, Phil Triolo and Associates LC<br>Tel: 801.699.9846<br>Fax: 801.328.2399<br>E-mail: philt@philt.com                              |
| Trade Name:                              | Plato MICROCATH® 27B Microcatheter                                                                                                                                 |
| Common Name:                             | Microcatheter for neurovascular use                                                                                                                                |
| Classification Name<br>and Product Code: | Diagnostic Intravascular Catheter per 21 CFR 870.1200, DQO<br>Continuous Flush Catheter per 21 CFR 870.1210, KRA<br>Percutaneous Catheter per 21 CFR 870.1250, DQY |
| Predicate Device:                        | ev3 Marksman™ Catheter (K111490, K091559)                                                                                                                          |
| Reference Device:                        | Scientia Vascular, LLC Plato MICROCATH 27 Microcatheter<br>(K143398)                                                                                               |

{4}------------------------------------------------

### Device Description:

The Plato MICROCATH® 27B Microcatheter is a single lumen microcatheter with a flexible polymer shaft of varying stiffness to aid in navigation of the neurovasculature. The catheter is designed to be used with a guide catheter and a steerable guidewire. The proximal end of the catheter has a diameter of 3.4F tapering to a distal OD of 3.2F. The catheter can be inserted into a 5F guide catheter. The inner diameter is constant over the entire length of the catheter shaft and accommodates guidewires up to 0.025" in diameter. The catheter is 150 cm in length with a straight tip configuration which can be shaped, using steam heat and the provided mandrels, to the users' preferred shapes. A steam shaping mandrel is included in the packaging. The microcatheter has a hydrophilic coating on the outer distal shaft to reduce friction during catheter manipulation in the vasculature and has one radiopaque tip marker at its distal tip to facilitate fluoroscopic visualization.

The subject microcatheter is identical with respect to its technological characteristics, design and materials to Scientia's currently marketed Plato MICROCATH 27B Microcatheters cleared under K143398. This 510k was prepared to expand the indications for use of the device to include its use in the neurovasculature.

### Indications for Use:

The Plato MICROCATH 27B Microcatheter is intended for the introduction of interventional devices and infusion of diagnostic agents to the peripheral and neurovascular systems. The catheter is also intended for the introduction of therapeutic agents to the peripheral system. The catheter is not intended for use in the coronary vasculature.

NOTE: The indications for use are the same as those of the predicate ev3 Marksman Catheter, with the exception that the Marksman Catheter is also indicated for coronary use. The identical reference MICROCATH 27B is currently cleared for the intended use of the introduction of interventional devices and infusion of diagnostic or therapeutic agents to the peripheral system; not the neurovascular system.

### Technological Characteristics:

The Plato MICROCATH 27B Microcatheter is a single lumen, variable stiffness microcatheter designed to provide increased flexibility for navigating the peripheral and neurovasculature. The proximal end of the catheter incorporates a polycarbonate female Luer adapter to facilitate the attachment of accessories and devices terminating in a male Luer. The distal end has a polymeric skeletal support for improved distal navigation. The device is compatible with 5 F or larger guiding catheters and can be advanced over guidewires up to 0.025" in diameter. The steam-shapeable distal shaft has a hydrophilic coating applied to impart lubricity.

The technological characteristics of the new microcatheter are comparable to those of the predicate device, the ev3 Marksman Catheter. Both the Marksman Catheter and the Plato 27B have a single PTFE lumen within a polymeric laminate. Both catheters have variable stiffness along the length designed for a flexible distal end and a stable proximal end enabling the catheter to be introduced over a steerable guidewire into the peripheral

{5}------------------------------------------------

Scientia Vascular LLC Traditional 510(k) Plato MICROCATH® 27B Microcatheter

and neurovasculature systems. The proximal end of both catheters incorporates a standard female luer adapter to facilitate the attachment of devices and accessories that terminate in male luer adapters. The proximal ends of both catheters also have a reinforcing braid embedded in the polymeric coatings to impart strength and stiffness. Both catheters are compatible with 5F or larger guiding catheters and can accommodate guidewires up to 0.021" in diameter. The distal ends of both catheters have a steam shapeable tip. Further, the outer surface of the distal ends of both catheters has a hydrophilic coating to aid catheter movement in the vasculature.

The Plato MICROCATH 27B differs from the predicate Marksman Catheter in that the predicate has a coiled support structure embedded in the polymeric laminate at its distal end while the Plato 27B has a polymeric skeletal support embedded in the polymeric laminate at its distal end. Both designs provide mechanical support and minimize the potential for catheter collapse in tortuous vessels. Additionally, the predicate and MICROCATH 27B differ in material composition.

Results of tests performed on the new Plato MICROCATH 27B Microcatheter demonstrate that the new catheters perform as well as the predicate devices and/or meet requirements of relevant standards. Further, the differences in technological characteristics, including differences in materials used in the manufacture of the predicate and new catheters, do not raise different questions of safety and effectiveness.

### Non-Clinical Performance Data:

### Biocompatibility

The materials used in the manufacture of the subject Microcatheters are identical to those used in the manufacture of Plato MICROCATHETER 27B Microcatheters also manufactured by Scientia Vascular LLC, cleared 26 June, 2015 after review of K143398; in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents).

There are no additional biologic interactions that need to be considered in order to use the cleared Microcatheters in the neurovasculature. Consequently, no additional evaluations are needed to determine that the subject Microcatheters present a low and acceptable biological and toxicological risk when used in accordance with their intended and indicated uses, and no additional biological safety information is provided in this 510(k).

### Functional Testing

The results of functional testing completed to comply with ISO 10555-1:2013 requirements, the relevant portions of the FDA document, Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters (Part VIII. section 13), issued September 8, 2010, and the relevant portions of the FDA document, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, issued March 16, 1995 are reported herein. The following tables summarize the functional tests performed and test results obtained to demonstrate substantial equivalence.

{6}------------------------------------------------

| Test                                                                                                                                                                                                                                                                         | Test Method Summary                                                                                                                                                                   | Results                                                                                                                                                                                                                             |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tensile Strength                                                                                                                                                                                                                                                             | Tensile testing performed per<br>ISO 10555-1 on distal,<br>medial and proximal catheter<br>sections after simulated use                                                               | All catheters meet minimum force<br>breakage requirements specified<br>in ISO 10555-1.                                                                                                                                              |
| Coating Lubricity and<br>Durability                                                                                                                                                                                                                                          | Frictional force of uncoated<br>and coated catheters was<br>determined                                                                                                                | All catheters met specified<br>frictional force requirements.                                                                                                                                                                       |
| Coating Integrity                                                                                                                                                                                                                                                            | Coating uniformity and<br>integrity visually examined<br>on dyed samples after<br>simulated use in a tortuous<br>path                                                                 | All samples showed acceptable<br>coating coverage and no defects<br>were identified post simulated use.                                                                                                                             |
| Particulates                                                                                                                                                                                                                                                                 | Particulates of various size<br>ranges counted after inserting<br>a guidewire and advancing<br>the Microcatheter through a<br>guide catheter multiple times.                          | The test and predicate catheters<br>had comparable numbers of<br>particles in each size range.                                                                                                                                      |
| Torque to Failure                                                                                                                                                                                                                                                            | Torque turns to failure in an<br>anatomical model to provide<br>a 4 fold safety factor                                                                                                | All catheters showed no signs of<br>breakage, twists or collapsed<br>lumens after specified number of<br>torque turns.                                                                                                              |
| Plato 27B Neurovascular<br>Model Evaluation                                                                                                                                                                                                                                  | Evaluate ability to deliver<br>and withdraw interventional<br>devices through catheter in a<br>model designed to simulate<br>the tortuous anatomy of the<br>neurovasculature.         | Interventional devices were<br>successfully delivered into and<br>withdrawn from the 27B<br>Microcatheter while it was locatec<br>in the model simulating the<br>tortuous anatomy of the<br>neurovasculature.                       |
| Table 2. Summaries of Tests that were Repeated to Evaluate the Performance of<br>13-month Real Time or Accelerated Aged Microcatheters to Establish a<br>Shelf Life Claim<br>NOTE:  Summaries of tests conducted on unaged catheters were<br>previously reported in k143398. |                                                                                                                                                                                       |                                                                                                                                                                                                                                     |
| Test                                                                                                                                                                                                                                                                         | Test Method Summary                                                                                                                                                                   | Results                                                                                                                                                                                                                             |
| Liquid Leakage under<br>Pressure/Leakage at Hub                                                                                                                                                                                                                              | Test for leakage at 300- 320<br>kPa per ISO 10555-1                                                                                                                                   | No leakage from hub or catheter<br>body                                                                                                                                                                                             |
| Static Burst Pressure                                                                                                                                                                                                                                                        | Burst pressure tested per ISO<br>10555-1                                                                                                                                              | Maximum peak pressures all<br>exceeded 300 psi.                                                                                                                                                                                     |
| Air Ingress/Negative<br>Collapse                                                                                                                                                                                                                                             | Tests per ISO 10555                                                                                                                                                                   | Hub fittings do not allow air<br>ingress and no evidence of lumen<br>collapse                                                                                                                                                       |
| Kink Resistance                                                                                                                                                                                                                                                              | Tests for kinks after distal tip<br>of catheter is wrapped around<br>cylindrical forms                                                                                                | Device was resistant to kinking<br>around small diameter turns per<br>specification                                                                                                                                                 |
| Tip Stiffness                                                                                                                                                                                                                                                                | Test for stiffness per ASTM<br>D747-10                                                                                                                                                | Tip stiffness was comparable that<br>of the predicate devices.                                                                                                                                                                      |
| Flexural Fatigue and<br>Profile                                                                                                                                                                                                                                              | Worst case bend of 90° with<br>an 8-fold safety factor for<br>repetitions per ASTM D747-<br>10                                                                                        | All catheters showed no signs of<br>cracks or breakage after worst<br>case simulated use.                                                                                                                                           |
| Tip Shape Retention                                                                                                                                                                                                                                                          | Catheters must retain angle<br>shaped by steam and meet<br>tensile strength specifications<br>after conditioning and<br>simulated use.                                                | All tips formed by steam met<br>shape and tensile requirements<br>after conditioning and simulated<br>use.                                                                                                                          |
| Simulated Use                                                                                                                                                                                                                                                                | Anatomical model designed<br>to simulate the tortuous<br>anatomy of the<br>neurovasculature used for<br>simulated use testing.                                                        | Catheters and predicate devices<br>were found to perform acceptably<br>in evaluations of guidewire<br>tracking and guide catheter<br>movement. Interventional devices<br>were successfully deployed and<br>withdrawn.               |
| Table 2. Summaries of Tests that were Repeated to Evaluate the Performance of<br>13-month Real Time or Accelerated Aged Microcatheters to Establish a<br>Shelf Life Claim<br>NOTE: Summaries of tests conducted on unaged catheters were<br>previously reported in k143398.  |                                                                                                                                                                                       |                                                                                                                                                                                                                                     |
| Test                                                                                                                                                                                                                                                                         | Test Method Summary                                                                                                                                                                   | Results                                                                                                                                                                                                                             |
| Visual/Dimensional<br>Inspection                                                                                                                                                                                                                                             | Tests per ISO 10555-1<br><br>Visual inspection for extraneous matter, process and surface defects or defects that may cause trauma to vessels.<br>Dimensional inspection per drawings | No surface defects or visible droplets of coating on catheters.<br>All catheters met dimensional specifications. Plato MICROCATH 27B Microcatheters do not exhibit clinically significant hydration as defined in ISO 10555-1:2013. |
| Pressure vs. Flow<br>Characterization                                                                                                                                                                                                                                        | Flow rates measured at two typical injection pressures per ISO 10555-1                                                                                                                | Flow rates reported in Instructions for Use at 100 and 300 psi                                                                                                                                                                      |
| Dynamic Flow                                                                                                                                                                                                                                                                 | Product used with power injector at pressure-limited setting to 700 psi                                                                                                               | No leaks, breaks or occluded lumens at 700 psi                                                                                                                                                                                      |
| Packaging Testing                                                                                                                                                                                                                                                            | Pouch evaluated for seal strength per ASTM F 88-00 and leak tests (bubble test) per ASTM F 2096-04                                                                                    | All sterile barrier pouches met minimum seal strength and released no bubbles under leak test conditions.                                                                                                                           |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

| Test                   | Test Method Summary                                                                                                                                         | Results                                                                                                                      |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Cadaver Testing        | Human Cadaver used to<br>evaluate catheter and<br>predicate devices by<br>physicians for performance,<br>access time, and ability to<br>deploy and retrieve | Test and predicate devices both<br>exhibited comparable performance<br>with similar access time to the<br>designated target. |
| Radiopacity            | Catheters and predicate<br>devices evaluated by<br>physicians in human cadaver<br>model                                                                     | Both test and predicate catheters<br>exhibited acceptable radiopacity.                                                       |
| MRI Compatibility      | Catheters contain metallic<br>materials and should not be<br>exposed to MRI procedures.                                                                     | Catheters are labeled MRI Unsafe<br>in IFU.                                                                                  |
| Chemical compatibility | Catheters were exposed to<br>saline and contrast<br>agent/saline solutions and<br>examined for degradation.                                                 | All catheters showed no signs of<br>degradation, corrosion or physical<br>decomposition.                                     |
| Latex Content          | Tested for trace latex proteins<br>per ASTM D6499-07                                                                                                        | No detectable traces of latex were<br>found.                                                                                 |
| Corrosion Resistance   | Test for corrosion resistance<br>per ISO 10555-1                                                                                                            | No signs of corrosion on metallic<br>components of catheters                                                                 |

#### Table 3. Summaries of Tests Whose Results Were Already Reported in K143398 nd are Being Leveraged for this Premarket Notifi

### Conclusion:

Scientia Vascular, LLC has presented information in this premarket notification supporting its contention that the Plato MICROCATH 27B Microcatheter is substantially equivalent with respect to technological characteristics and indications for use to the ev3 predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQO/K170406](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQO/K170406)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
