K153125 · Texasmedical Technologies, Inc. · DQO · Jan 13, 2016 · Cardiovascular
Device Facts
Record ID
K153125
Device Name
IntraNovo Microcatheter
Applicant
Texasmedical Technologies, Inc.
Product Code
DQO · Cardiovascular
Decision Date
Jan 13, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IntraNovo Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremittes, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.
Device Story
IntraNovo Microcatheter is a single-lumen infusion catheter for intravascular use. Device features PTFE inner liner, stainless steel coil reinforcement, polyether block amide outer shaft, and hydrophilic distal coating. Includes radiopaque marker band and polycarbonate hub. Available in 110 cm and 130 cm lengths. Operated by physicians in clinical settings for peripheral and coronary procedures. Delivered over-the-wire to target vessels. Used for contrast media delivery, drug infusion, and hemostasis via embolic material delivery. Manual tip shaping enabled by included stainless steel mandrel. Device facilitates access to smaller vessels due to reduced distal diameter (2.4 Fr) compared to predicate. Benefits include precise therapeutic agent delivery and vessel access.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated via 24 bench tests including guidewire compatibility, track force, coating durability/lubricity, burst pressure (static/dynamic), tensile strength, flow rate, kink resistance, radiopacity, torque, and particulate evaluation. Testing performed on T=0 and T=2 year accelerated aged samples. All tests passed.
Technological Characteristics
Single-lumen catheter; PTFE inner liner; stainless steel coil; polyether block amide outer shaft; platinum/iridium marker band; polycarbonate hub. Hydrophilic coating. Dimensions: 2.4-2.9 Fr OD, 0.021" ID. Sterilization: EtO gas (SAL 10^-6). Standards: ISO 10555 (mechanical), ISO 594 (luer), ASTM F640-12 (radiopacity), ASTM F88-09/1929-98 (packaging), USP <788> (particulates), ISO 10993 (biocompatibility).
Indications for Use
Indicated for infusion of contrast media, drugs, and embolic materials into peripheral, cervical, and coronary vessels. Contraindicated for use in cerebral vessels.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2016
Texas Medical Technologies, Inc. % E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114
Re: K153125
Trade/Device Name: IntraNovo Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO Dated: December 10, 2015 Received: December 14, 2015
Dear E.J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name IntraNovo Microcatheter
#### Indications for Use (Describe)
The IntraNovo Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremittes, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
- 1. Company Information & Contact Person
| Company Name:<br>Company Address | Texas Medical Technologies, Inc<br>9005 Montana Ave. Ste. A<br>El Paso, Texas 79925 |
|----------------------------------|-------------------------------------------------------------------------------------|
| Telephone:<br>Fax: | (915) 774-4321<br>(915) 774-4323 |
| Date Prepared: | 10/20/2015 |
- 2. Device Name & Classification
| Proprietary Name: | IntraNovo Microcatheter |
|----------------------|-----------------------------------|
| Common Name: | Microcatheter |
| Classification Name: | Diagnostic Intravascular Catheter |
| Regulation Number: | 21 CFR 870.1200 |
| Product Code: | DQO |
| Device Class: | Class II |
- 3. Predicate Device
| Proprietary Name: | IntraNovo 25 Microcatheter |
|----------------------|-----------------------------------|
| Company Name: | Texas Medical Technologies, Inc. |
| Common Name: | Microcatheter |
| Classification Name: | Diagnostic Intravascular Catheter |
| Regulation Number: | 21 CFR 870.1200 |
| Product Code: | DQO |
| Device Class: | Class II |
| 510(k) Number: | K142817 |
#### 4. Device Description
The IntraNovo Microcatheter (the "IntraNovo 21" or the "Microcatheter") is an infusion catheter intended for intravascular use. The Microcatheter consists of a single lumen. The shaft consists of a lubricous inner liner made from polytetrafluoroethylene (PTFE) with a stainless steel coil over the inner liner. The outer liner consists of different lengths of colored polyether block amide with varying durometers hardness. A lubricious hydrophilic coating covers the distal end of the IntraNovo Microcatheter. One radiopaque marker band is placed at the distal end of the Microcatheter and above the stainless steel coil. A polycarbonate hub is attached to the proximal end of the Microcatheter with a strain relief placed over the distal end of the hub.
{4}------------------------------------------------
The device shaft has an outer diameter of 2.9 French size on the proximal end and 2.4 French size on the distal end with an inner diameter of 0.021" throughout the shaft. The device is available in lengths of 110 cm and 130 cm.
The following accessories are included with the Microcatheter: a stainless steel steam shaping mandrel to allow for manual shaping of the distal tip and a 3.0 mL syringe.
| Design &<br>Development<br>Model<br>Number | Commercial<br>Model<br>Number | Shaft<br>Length<br>(cm) | Outer Diameter | | Inner<br>Diameter | Marker Band<br>Configuration | Tip Shape |
|--------------------------------------------|-------------------------------|-------------------------|----------------|---------|-------------------|------------------------------|-------------------------|
| SMC-21110-1S | MC-2411-1SN | 110 | 2.7 Fr. | 2.9 Fr. | 0.025" | 1 Marker Band | Straight<br>(shapeable) |
| SMC-21130-1S | MC-2413-1SN | 130 | 2.7 Fr. | 2.9 Fr. | 0.025" | 1 Marker Band | Straight<br>(shapeable) |
Table 4.1 IntraNovo Microcatheter Models and Sizes
- 5. Indications for Use
The IntraNovo Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.
- Predicate Product Comparison Table 6.
| Technical<br>Characteristics /<br>Principle of<br>Operation | IntraNovo<br>Microcatheter | IntraNovo 25<br>Microcatheter<br>(Predicate) | Comments |
|-------------------------------------------------------------|----------------------------------------|----------------------------------------------|--------------------------|
| 510(k) Number | | K142817 | |
| Length | 100cm - 130cm | 100cm -130 cm-<br>150cm | Same |
| Outer Diameter | Distal: 2.4 Fr.<br>Proximal 2.9 Fr. | Distal: 2.7 Fr.<br>Proximal 2.9 Fr. | Different 1<br>Identical |
| Inner Diameter | 0.021" | 0.025" | Different 2 |
| Maximum Pressure | 5,171 kPa (750 psi) | 5,171 kPa (750 psi) | Identical |
| Distal Curve | Straight (shapeable) | Straight<br>(shapeable) | Identical |
| Inner Liner<br>Material | Polytetraflouroethyl<br>ene<br>(PTFE)* | Polytetraflouroethy<br>lene<br>(PTFE)* | Identical |
| Coil Reinforcement<br>Material | Stainless Steel* | Stainless Steel* | Identical |
{5}------------------------------------------------
| Radiopaque Marker | Platinum/Iridium* | Platinum/Iridium* | Identical |
|----------------------------|---------------------------------------|---------------------------------------|-------------|
| Outer Shaft<br>Material | Polyether Block<br>Amide* | Polyether Block<br>Amide* | Identical |
| Luer Material | Polycarbonate* | Polycarbonate* | Identical |
| Luer Connector | Female Luer<br>Connector* | Female Luer<br>Connector* | Identical |
| Hydrophilic Coated | Yes* | Yes* | Identical |
| Anatomical Site<br>Use | Peripheral,<br>Coronary | Peripheral,<br>Coronary | Identical |
| Supplied<br>Accessories | Shaping Mandrel,<br>Injection Syringe | Shaping Mandrel,<br>Injection Syringe | Identical |
| Delivery to Site | Over-the-wire | Over-the-wire | Identical |
| Guidewire<br>Compatibility | Maximum 0.018" | Maximum 0.021" | Different 3 |
| Packaging | Polyethylene Hoop<br>and Tyvek Pouch | Polyethylene Hoop<br>and Tyvek Pouch | Identical |
| Sterilization | EtO Gas | EtO Gas | Identical |
## 7. Discussion of Technological Characteristics
Both devices are equivalent with respect to their indications for use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging.
The IntraNovo Microcatheter does differ from the predicate in outer diameter of 2.4 fr versus predicate's 2.7 fr, inner diameter of 0.021" versus the predicate's 0.025" and guidewire compatibility of a maximum 0.018" versus the predicate's 0.021". The following is a discussion of why no new issues of safety and effectiveness have been raised by these differences:
- 1. Difference on the distal outer diameter does not impact the product performance or modifies the intended use. The smaller size is intended to reach smaller target vessels in the anatomy. The Microcatheter's instructions for use indicate the minimum sizes of devices compatible with the Microcatheter outer diameter. This difference does not raise any new issues of safety and effectiveness.
- 2. Difference on the inner diameter does not impact on the product performance or modifies the intended use. The smaller size is intended to reach smaller target vessel in the anatomy. The Microcatheter's instructions for use indicate the maximum sizes of therapeutic agents compatible with the Microcatheter inner diameter. This difference does not raise any new issues of safety and effectiveness.
- 3. Difference in the guidewire compatibility does not impact on the product performance or modifies the intended use. Device lumen is smaller and therefore the maximum guidewire compatible is smaller. This difference does not raise any new issues of safety and effectiveness.
## 8. Testing Summary
{6}------------------------------------------------
The following bench tests were performed to evaluate the design elements and performance characteristics of the IntraNovo Microcatheter and to demonstrate substantial equivalence to the predicate device. The IntraNovo Microcatheter met the predetermined acceptance criteria. Testing was performed on non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2). Tests results show that the IntraNovo Microcatheter is substantially equivalent to the predicate device.
| Test<br># | Test Name | Applicable<br>Standard or<br>Internal Test<br>Method | Test<br>Results<br>(T=0)<br>and<br>(T=2) |
|-----------|---------------------------------------------|------------------------------------------------------|------------------------------------------|
| 1 | Guidewire & Guide<br>Catheter Compatibility | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 2 | In-Vitro Track Force | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 3 | Durability of Hydrophilic<br>Coating | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 4 | Lubricity of Hydrophilic<br>Coating | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 5 | Tip Shape Retention | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 6 | Static Burst Pressure | ISO 10555 | T=0 Pass<br>T=2 Pass |
| 7 | Dimensional & Physical<br>Attributes | ISO 10555 | T=0 Pass<br>T=2 Pass |
| 8 | Corrosion Resistance | ISO 10555 | T=0 Pass<br>T=2 Pass |
| 9 | Dynamic Burst Pressure | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 10 | Air Leak | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 11 | Liquid Leakage | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 12 | Tensile Strength | ISO 10555 | T=0 Pass<br>T=2 Pass |
| 13 | Flow Rate | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 14 | Kink Resistance | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 15 | Radiopacity | ASTM-F640-12 | T=0 Pass<br>T=2 Pass |
| 16 | Torque to Failure | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 17 | Catheter Stiffness | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 18 | Packaging Integrity | ASTM F-88-09<br>ASTM-1929-98 | T=0 Pass<br>T=2 Pass |
| 19 | Therapeutic Agents | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
# 9. Bench Testing Table
| Table 7.1. Bench Testing Performed on the IntraNovo Microcatheter. | |
|--------------------------------------------------------------------|--|
|--------------------------------------------------------------------|--|
{7}------------------------------------------------
| 20 | Female Luer Hub<br>Verification | ISO 594 | T=0 Pass<br>T=2 Pass |
|----|-------------------------------------------|-------------------------|----------------------|
| 21 | Shipping and Transportation<br>Simulation | ISTA 3PA | T=0 Pass<br>T=2 Pass |
| 22 | Coating Integrity | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 23 | Torque Response | Internal Test<br>Method | T=0 Pass<br>T=2 Pass |
| 24 | Particulate Evaluation | USP <788> | T=0 Pass<br>T=2 Pass |
# 10. Biocompatibility
The IntraNovo Microcatheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (<24) hours.
The final sterilized device is identical to IntraNovo 25 Microcatheter (K142817) in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents).
Biocompatibility testing was not performed since the same materials of the IntraNovo 25 Microcatheter were used and biocompatibility testing results shown that these materials are biocompatible according to ISO 10993. The IntraNovo 25 Microcatheter was cleared under 510K number K142817 on 4/23/2015.
## 11. Sterilization Testing Summary
| Adoption Cycle Sterilization<br>Process | Sterility Assurance Level<br>(SAL) | Result |
|-----------------------------------------|------------------------------------|--------|
| Ethylene Oxide Gas | $10^{-6}$ | Pass |
| LAL - AAMI ST 72:2010 | | Pass |
# 12. Conclusion
The IntraNovo Microcatheter is substantially equivalent in intended use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.