The ILUMIEN OPTIS with C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The ILUMIEN OPTIS will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
Device Story
ILUMIEN OPTIS is a cart-mounted OCT imaging system and FFR measurement tool. System inputs include reflected near-infrared light from an internal rotating fiber optic core within a Dragonfly OPTIS Imaging Catheter and radio frequency pressure signals from distal intracoronary and proximal aortic transducers. The system processes optical signals to construct high-resolution OCT images and calculates FFR values. Used in clinical settings by physicians to visualize coronary artery anatomy and assess physiological significance of lesions to guide therapeutic intervention. The system includes a Drive-motor and Optical Coupler (DOC) for catheter connection, dual monitors for operator/physician, and wireless FFR signal receivers. Benefits include improved visualization and physiological assessment for interventional decision-making.
Clinical Evidence
Bench testing and pre-clinical animal testing performed. Compliance with electrical safety (UL 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and DICOM standards verified. Software verification and validation conducted per FDA guidance.
Technological Characteristics
Cart-mounted OCT system with optical engine. Catheter: sterile, single-use, stainless steel torque wire, fiber optic core, TiO2 doped window, RFID. Connectivity: DICOM, USB-based FFR receivers. Energy: Near-infrared light, electrical power. Software: Revision E.1, GUI-based control. Sterilization: Not specified.
Indications for Use
Indicated for patients who are candidates for transluminal interventional procedures requiring coronary artery imaging. Catheter use restricted to vessels 2.0-3.5 mm diameter. Contraindicated for left main coronary artery or target vessels with previous bypass procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2014
Lightlab Imaging, Inc. Jeffrey Roberts Principal Regulatory Affairs Specialist 4 Robbins Road Westford, Massachusetts 01886
Re: K141453
Trade/Device Name: Ilumien Optis, Dragonfly Optis Imaging Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DOO. NOO Dated: August 7, 2014 Received: August 8, 2014
Dear Jeffrey Roberts.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely
Image /page/1/Picture/8 description: The image shows a signature that appears to be "KSL" in cursive. Below the signature, the text "Ken Skodacek for" is printed in a smaller, sans-serif font. The signature is bold and slightly stylized, with each letter distinct but connected. The background is plain, allowing the signature and text to stand out.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 4. INDICATIONS FOR USE
510(k) Number (if known):
Device Name: ILUMIEN OPTIS
Indications for Use:
The ILUMIEN OPTIS with C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The C7 Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The ILUMIEN OPTIS will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 5. 510(K) SUMMARY
# for the LightLab Imaging, Inc. ILUMIEN OPTIS and Dragonfly OPTIS Imaging Catheter (per 21CFR 807.92)
#### 1. SUBMITTER/510(K) HOLDER
LightLab Imaging, Inc. 4 Robbins Road Westford, MA 01886
| Contact Person: | Jeffrey Roberts |
|-----------------|-----------------|
| Telephone: | 978-577-3451 |
Date Prepared: 5/30/14
#### 2. DEVICE NAME
| Proprietary Name: | ILUMIEN OPTIS |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Ultrasonic pulsed echo imaging system |
| Classification Name: | Ultrasonic pulsed echo imaging system |
| The Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call Call | mi a mana mara ya mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mana mara mara marka |
| Proprietary Name: | The Dragonfly OPTIS Imaging Catheter |
|----------------------|--------------------------------------|
| Common/Usual Name: | Diagnostic Intravascular Catheter |
| Classification Name: | Diagnostic Intravascular Catheter |
#### 3. DEVICE CLASSIFICATION
| The ILUMIEN OPTIS medical device comprises the following: | |
|-----------------------------------------------------------|---------------------------------------|
| Classification Name | Ultrasonic Pulsed Echo Imaging System |
| Classification Regulation | 21 CFR 892.1560 |
| Product Code | NQQ |
The Dragonfly OPTIS Imaging Catheter device comprises the following: Classification Name: Diagnostic Intravascular Catheter Classification Regulation: 21 CFR 870.1200 Product Code: DQO
{4}------------------------------------------------
### 4. PREDICATE DEVICE
- ILUMIEN OPTIS with Dragonfly I Catheter manufactured by LightLab Imaging, Inc. ● K123369
# 5. DEVICE DESCRIPTION
The ILUMIEN OPTIS is a cart-mounted computer and Imaging Engine (or optical engine) placed inside an ergonomically designed mobile cart. It also includes the Drive-motor and Optical Coupler (DOC), which provides the interconnection between the ILUMIEN OPTIS System and the Dragonfly Catheter. The cart is equipped with two display monitors (one for the console operator, and the other for the physician), as well as a keyboard and mouse. The cart also contains an isolation transformer for electrical safety. The cart includes two USB mounted FFR receivers which provide wireless reception of distal intracoronary and aortic pressure signals originating at the PressureWire® Aeris and the aortic pressure transducer (AIU). These signals are used to calculate and display the patient's Fractional Flow Reserve (FFR) on the system monitor.
The Dragonfly OPTIS Imaging Catheter is a sterile, single-use intravascular catheter consisting of a catheter body external sheath and an internal rotating fiber optic imaging core. The external sheath serves two primary functions: 1) to facilitate placement of the device into the coronary artery and 2) to cover and protect the inner rotating fiber optic imaging core.
The inner rotating fiber optic imaging core emits near infrared light to the tissue and receives reflected light. It is driven by a stainless steel torque wire visible under fluoroscopy and pulled back through the window tube of the external sheath by the DOC. The emitted and returned reflected light are combined and processed by the Ilumien Optis System software to construct an OCT image. The patient is never exposed to moving parts as the external sheath completely covers the rotating imaging core.
{5}------------------------------------------------
#### 6. INTENDED USE
The ILUMIEN OPTIS with C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The ILUMIEN OPTIS will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
### 7. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
# ILUMIEN OPTIS
The ILUMIEN OPTIS is equivalent to the predicate device in terms of hardware and firmware components. They both contain a DOC which provides the interconnection between the ILUMIEN OPTIS and the optical imaging catheters that emit near-infrared light to produce high-resolution real-time images. This process is accomplished for both the ILUMIEN OPTIS and the predicate device through a graphical user interface (GUI) and software control to obtain Optical Coherence Tomography (OCT) imaging modality and fractional flow reserve (FFR) measurements.
The ILUMIEN OPTIS represents an upgrade to the predicate device in terms of performance through the same hardware and firmware design and technological characteristics. The software has been upgraded to revision E.1 for the following features:
- A curtain GUI interface.
- Dragonfly OPTIS Imaging Catheter Support
- Continuous calibration for TiO2 doped catheter window
- DICOM modality integration
- Manual pull back triggering ●
- Catheter user connection interface ●
- Improved acquisition workflow display ●
{6}------------------------------------------------
# Dragonfly OPTIS Imaging Catheter
The Dragonfly OPTIS Imaging Catheter is equivalent to the predicate device in terms of hardware components and operational use. They both are comprised of a catheter body external sheath and internal rotating fiber optic imaging core which emits near infrared light to the tissue and receives reflected light. They both are driven by a stainless steel torque wire by the DOC which is connected to the ILUMIEN OPTIS OCT Imaging System. They are both purged through the central catheter with 100% contrast media prior to use. In both the Dragonfly OPTIS Imaging Catheter and the predate device emitted and returned reflected light are combined and processed by the ILUMIEN OPTIS software to construct an OCT image.
The Dragonfly OPTIS Imaging Catheter represents an upgrade to the predicate device in terms of performance through the same operational characteristics, and fundamental technological characteristics to include the following:
- TiO2 Doped window
- Flexible proximal end ●
- o Improved break away joint
- Improved purge tube
- Dual lumen tip
- High speed proximal end ●
- 155um fiber
- o RFID.
### 8. PERFORMANCE TESTING
The ILLUMIEN OPTIS has been tested and is in compliance with UL Standard No 60601-1, Medical Electrical Equipment Part I: General Requirements for Safety, IEC 60601-1-2 Ed. 2.1, Electromagnetic emissions and immunity requirements for medical electrical equipment - Group 1 Equipment, Class B for non-life supporting equipment, EN 60601-1-2:2007, Electromagnetic emissions and immunity requirements for medical electrical equipment -Group 1 Equipment, Class B for non-life supporting equipment, IEC 60825-1, 2nd, Ed., 2007, SAFETY OF LASER PRODUCTS - Part 1: Equipment classification and requirements, DICOM Standard (PS 3.2-2008), 21 CFR 1040.10, Performance Standards for Light-Emitting Products, Laser Products, and CFR 47 FCC Part 15 Subpart B Class B emissions requirements (USA)
In addition to the electrical safety testing performed, software verification and validation was conducted to FDA regulations, standards and guidance document requirements. The results of this testing conclude the software has met these requirements. Design verification and
{7}------------------------------------------------
validation was also performed on the ILUMIEN OPTIS and Dragonfly OPTIS Imaging Catheter in compliance with internal design control procedures which included bench testing and pre-clinical animal testing. The results of this testing concludes the ILUMIEN OPTIS and Dragonfly OPTIS Imaging Catheter is determined to be safe and effective and is substantially equivalent to the predicate ILUMIEN OPTIS device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.