Zio XT ECG Monitoring System, Zio AT ECG Monitoring System

K190593 · iRhythm Technologies, Inc. · DQK · Aug 23, 2019 · Cardiovascular

Device Facts

Record IDK190593
Device NameZio XT ECG Monitoring System, Zio AT ECG Monitoring System
ApplicantiRhythm Technologies, Inc.
Product CodeDQK · Cardiovascular
Decision DateAug 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Zio ECG Utilization Service (ZEUS) System is intended to analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring by Zio XT or Zio AT Patch, a final report is generated based on the beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients.

Device Story

ZEUS System processes/analyzes continuous single-lead ECG data recorded by Zio XT or Zio AT patches. System downloads, stores, and aggregates ECG data; automated analysis identifies cardiac events. Certified Cardiographic Technician (CCT) reviews automated findings to generate final clinical report. Used in clinical settings by trained technicians to assist physicians in cardiac diagnosis. Output provides comprehensive ECG analysis to support clinical decision-making; enables long-term rhythm monitoring for transient symptom evaluation.

Clinical Evidence

No clinical testing was performed. Evidence consists of bench testing, system performance testing, and software verification testing to ensure compliance with design specifications and consensus standards.

Technological Characteristics

Programmable diagnostic computer for ECG processing. Complies with AAMI/ANSI ISO 14971:2007(R)2010, AAMI/ANSI IEC 60601-2-47:2012, and AAMI/ANSI EC57:2012. System performs automated ECG analysis and aggregation of single-lead data. Cybersecurity managed per FDA guidance.

Indications for Use

Indicated for patients 18 years or older, symptomatic (palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, anxiety) or asymptomatic, requiring long-term ECG monitoring. Not for critical care patients.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. iRhythm Technologies, Inc. Gabrielle Logan Senior Regulatory Affairs Specialist 650 Townsend Street, Ste 500 San Francisco, California 94103 Re: K190593 Trade/Device Name: Zio® ECG Utilization Service (ZEUS) System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DQK Dated: July 19, 2019 Received: July 23, 2019 ## Dear Gabrielle Logan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Nicole Goodsell Assistant Director (Acting) External Heart Rhythm and Rate Devices Team Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190593 Device Name Zio® ECG Utilization Service (ZEUS) System #### Indications for Use (Describe) The Zio ECG Utilization Service (ZEUS) System is intended to analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring by Zio XT or Zio AT Patch, a final report is generated based on the beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Notification K190593 #### l. General Information #### Applicant: iRhythm Technologies, Inc. 650 Townsend Street, Suite 500 San Francisco, CA 94103, USA Phone: 415-632-5700 Fax: 415-632-5701 ### Contact Person: Gabrielle Logan Sr. Regulatory Affairs Specialist Phone: 415-214-7092 Email: glogan@irhythmtech.com Date Prepared: August 22, 2019 #### ll. Device Information Trade Name: Zio® ECG Utilization Service (ZEUS) System ### Generic/Common Name: Programmable diagnostic computer ### Classification Names: Programmable diagnostic computer [21CFR§870.1425] ### Regulatory Class: Class II Product Codes: DQK #### III. Predicate Devices The following predicate device has been selected: - iRhythm Technologies, Inc. Zio® AT ECG Monitoring System [K181502] | No reference devices were used in this submission. {4}------------------------------------------------ ### Section 1b 510(k) Summary #### IV. Indications for Use The Zio ECG Utilization Service (ZEUS) System is intended to analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring. After patient monitoring by Zio XT or Zio AT Patch, a final report is generated based on the beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients. #### V. Device Description The ZEUS System is an electrocardiogram (ECG) processing and analysis system, designed to handle continuously recorded, single-lead ECG data. It downloads, stores, analyzes and aggregates the ECG data for a Certified Cardiographic Technician (CCT) to review and generate a report of the findings contained within the data; thereby enabling the provision of a complete ECG processing and analysis service. Automated ECG analysis performance was quantified for any claimed analysis metrics. The resulting statistics demonstrate sensitivity and positive predictivity levels which satisfy requirements and minimize safety or efficacy concerns. {5}------------------------------------------------ ### Section 1b 510(k) Summary #### VI. Substantial Equivalence The proposed indications for use statement for the Zio Utilization Service (ZEUS) System is substantially equivalent to the intended use in the cleared Indications for Use statement for the predicate device. The performance testing results demonstrate that the differences in the technological characteristics between the devices do not raise any new issues of safety or efficacy. Therefore, the ZEUS System is determined to be substantially equivalent to the predicate device. #### VII. Nonclinical Testing in Support of Substantial Equivalence Determination There are no required FDA performance standards for the ZEUS System. Overall system performance testing was conducted as part of the verification activities on incremental changes to the ZEUS System to ensure performance as intended per specifications and to support a determination of substantial equivalence to the predicate device. Nonclinical testing included: - . System performance testing - . Software verification testing The results confirm by examination and provision of objective evidence that the design output met the design input requirements in conformance with the following list of recognized consensus standards: | FDA# | Body | Number / Version | Title | |-------|------------------|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 5-70 | AAMI<br>ANSI ISO | 14971:2007(R)2010<br>(Corrected 4 October 2017) | Medical Devices – Applications Of Risk<br>Management To Medical Devices | | 3-127 | AAMI<br>ANSI IEC | 60601-2-47:2012 | Medical Electrical Equipment -- Part 2-<br>47: Particular Requirements For The<br>Basic Safety And Essential Performance<br>Of Ambulatory Electrocardiographic<br>Systems | | 3-118 | AAMI<br>ANSI | EC57:2012 | Testing And Reporting Performance<br>Results Of Cardiac Rhythm And ST-<br>Segment Measurement Algorithms | | N/A | U.S. FDA | October 2, 2014 | Guidance for Industry and FDA Staff –<br>Content of Premarket Submissions for<br>Management of Cybersecurity in<br>Medical Devices | No clinical testing was performed in support of this premarket notification. {6}------------------------------------------------ Section 1b 510(k) Summary #### VIII. Conclusion The ZEUS System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...