K181577 · Acutus Medical, Inc. · DQK · Feb 7, 2019 · Cardiovascular
Device Facts
Record ID
K181577
Device Name
AcQMap High Resolution Imaging and Mapping System
Applicant
Acutus Medical, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Feb 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The AcQMap® High Resolution Imaging and Mapping System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either dipole density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
Device Story
Diagnostic recording system for EP labs; images, navigates, and maps atrial heart chambers. Inputs: ultrasound data from AcQMap catheters, ECG/EGM signals, and impedance-based localization. System reconstructs 3D cardiac chamber anatomy; displays electrical activity as waveform traces; generates dipole density-based or voltage-based maps overlaid on 3D reconstructions. Operated by physicians to identify arrhythmia sources. Updated features include Anatomic Reference for respiratory motion subtraction, Marker Enhancement for snapping markers to anatomy, and GUI refresh. System consists of console, workstation, patient interface unit, and ablation interface unit. Benefits include improved visualization of complex arrhythmias difficult to identify with conventional systems.
Clinical Evidence
Prospective, non-randomized, open-label study (DDRAMATIC-SVT) conducted at eight sites outside the U.S. with 84 patients. Evaluated safety and performance of dipole density mapping in supraventricular tachycardia. Results demonstrated the system is safe and effective for its intended use.
Technological Characteristics
System components: console, workstation, patient interface unit, ablation interface unit. Sensing: ultrasound transducers and electrical electrodes. Connectivity: wired interfaces for catheters and ECG. Standards: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2, IEC 60601-1-6. Software: includes anatomic reference for respiratory motion subtraction and marker enhancement.
Indications for Use
Indicated for patients undergoing electrophysiology procedures for the visualization of cardiac anatomy and mapping of complex arrhythmias.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
AcQMap High Resolution Imaging and Mapping System (K170948)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.
# February 7, 2019
Acutus Medical, Inc. Greg Geissinger Director, Regulatory Affairs & Ouality Assurance 2210 Faraday Ave. Suite 100 Carlsbad, California 92008
Re: K181577
Trade/Device Name: AcQMap High Resolution Imaging and Mapping System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, IYO, ITX Dated: January 2, 2019 Received: January 3, 2019
## Dear Greg Geissinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K181577
Device Name
AcQMap® High Resolution Imaging and Mapping System
#### Indications for Use (Describe)
The AcQMap® High Resolution Imaging and Mapping System is intended for use in patients for whom electrophysiology procedures have been prescribed.
When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either dipole density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
AND
When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Notification K181577
## GENERAL INFORMATION [807.92(a)(1)]
#### Applicant:
Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6080 Fax: 1-442-232-6081
#### Contact Person:
Karla Schaffner Principal Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6161 Fax: 1-442-232-6081 Email: karla.schaffner(@acutus.com
### Date Prepared: 02 January 2018
#### DEVICE INFORMATION [807.92(a)(2)]
#### Trade Name:
AcQMap® High Resolution Imaging and Mapping System
### Generic/Common Name:
Programable diagnostic computer and Ultrasonic pulsed echo imaging system
#### Classification:
Class II/21 CFR § 870.1425 and Class II/21 CFR § 892.1560
#### Product Code:
DQK, IYO, ITX
{4}------------------------------------------------
# PREDICATE DEVICE [807.92(a)(3)]
AcQMap® High Resolution Imaging and Mapping System (K170948)
# DEVICE DESCRIPTION [807.92(a)(4)]
The AcQMap System operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation, the AcQMap Patient Interface Unit, and the AcQMap Ablation Interface Unit.
The AcQMap® High Resolution Imaging and Mapping System ("AcQMap System") is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping the atrial chambers of the heart.
The AcQMap System hardware consists of three functional subsystems:
- Ultrasound imaging.
- ECG and EGM recording; and ●
- Impedance based electrode Localization. 0
The AcQMap System is used in conjunction with the associated AcQMap 3D Imaging and Mapping Catheter. The AcOMap System provides:
- 3-D cardiac chamber reconstruction imaging, ●
- 0 Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters,
- Cardiac electrical activity as waveform traces, ●
- Remapping of the chamber at any time during the procedure; and
- Three-dimensional, dipole density-based maps overlaid on the cardiac chamber ● reconstruction to show chamber-wide electrical activation.
The AcQMap System is a diagnostic recording system consisting of ultrasound, electrical mapping components, a console, and a workstation. The AcQMap System is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.
# INDICATIONS FOR USE [807.92(a)(5)]
The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.
When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either dipole density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
# AND
When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
{5}------------------------------------------------
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [807.92(A)(6)]
This submission is seeking the clearance of the updated AcQMap System, which has the included software features, Anatomic Reference, Markers Enhancement, and Graphical User Interface Refresh.
Anatomic Reference allows the Anatomic Reference Channels to be used to subtract respiratory motion from all devices navigated by the system. With Marker Enhancement, markers can now be optionally "snapped" to the anatomy when the user-selected electrode is within 4mm of the reconstructed surface. Graphical User Interface Refresh is a functionality that was previously accessed through dropdown menus but can now be additionally accessed through readily available icons within the user interface.
Table 1 provides a comparison of the modified AcQMap System classification and indications for use against the predicate and reference devices. Table 2 provides a comparison of the technological characteristics for the modified AcQMap System against the predicate device.
{6}------------------------------------------------
| Characteristics | AcQMap® High Resolution<br>Imaging and Mapping System | AcQMap® High Resolution<br>Imaging and Mapping System | Rationale for<br>Substantial<br>Equivalence |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Regulatory | | | |
| 510(k) Number | K181577 | K170948 | Identical |
| Classification/<br>Regulation<br>Number/<br>Regulation Name/<br>Product Code | Class II/ 21 CFR § 870.1425<br>/Programable diagnostic computer/<br>DQK<br>Class II/ 21 CFR § 892.1560/<br>Ultrasonic pulsed echo imaging<br>system/ IYO, ITX | Class II/ 21 CFR § 870.1425<br>/Programable diagnostic computer/<br>DQK<br>Class II/ 21 CFR § 892.1560/<br>Ultrasonic pulsed echo imaging<br>system/ IYO, ITX | Identical |
| Indications for Use | The AcQMap System is intended<br>for use in patients for whom<br>electrophysiology procedures have<br>been prescribed.<br>When used with the AcQMap<br>Catheters, the AcQMap System is<br>intended to be used to reconstruct<br>the selected chamber from<br>ultrasound data for purposes of<br>visualizing the chamber anatomy<br>and displaying electrical impulses<br>as either dipole density or voltage<br>maps of complex arrhythmias that<br>may be difficult to identify using<br>conventional mapping systems<br>alone.<br>AND<br>When used with the specified<br>Patient Electrodes, the AcQMap<br>System is intended to display the<br>position of AcQMap Catheters and<br>conventional electrophysiology<br>(EP) catheters in the heart. | The AcQMap System is intended<br>for use in patients for whom<br>electrophysiology procedures have<br>been prescribed.<br>When used with the AcQMap<br>Catheters, the AcQMap System is<br>intended to be used to reconstruct<br>the selected chamber from<br>ultrasound data for purposes of<br>visualizing the chamber anatomy<br>and displaying electrical impulses as<br>either dipole density or voltage<br>maps of complex arrhythmias that<br>may be difficult to identify using<br>conventional mapping systems<br>alone.<br>AND<br>When used with the specified<br>Patient Electrodes, the AcQMap<br>System is intended to display the<br>position of AcQMap Catheters and<br>conventional electrophysiology (EP)<br>catheters in the heart. | Identical |
| Patient Anatomy | Intracardiac Structures | Intracardiac Structures | Identical |
| Testing to Support<br>Substantial<br>Equivalence | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, Animal Testing; and Clinical Testing | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, Animal Testing; and Clinical Testing | Identical |
| System Safety<br>Standards | IEC 60601-1: 2005 /A1:2012 IEC 60601-2-37: 2007 IEC 60601-1-2: 2007 /AC:2010 IEC60601-1-6:2010 | IEC 60601-1: 2005 /A1:2012 IEC 60601-2-37: 2007 IEC 60601-1-2: 2007 /AC:2010 IEC60601-1-6:2010 | Identical |
| Characteristics | AcQMap® High Resolution<br>Imaging and Mapping System<br>(K181577) | AcQMap® High Resolution<br>Imaging and Mapping System<br>(K170948) | Rationale for<br>Substantial<br>Equivalence |
| Physical Characteristics | | | |
| System<br>Components | Console, Workstation,<br>Workstation Cable Patient<br>Interface Unit, Patient Interface<br>Unit Cable, Ablation Interface<br>Unit, Ablation Interface Unit<br>Cable, Auxiliary Catheter<br>Cable, Surface ECG Input<br>Cable and Patient Electrode Kit. | Console, Workstation,<br>Workstation Cable Patient<br>Interface Unit, Patient Interface<br>Unit Cable, Ablation Interface<br>Unit, Ablation Interface Unit<br>Cable, Auxiliary Catheter<br>Cable, Surface ECG Input<br>Cable and Patient Electrode Kit. | Identical |
| Visual/Mapping<br>Characteristics | CT-like cardiac chamber reconstructions imaging, Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters, Cardiac electrical activity as waveform traces, Remapping of the chamber at any time during the procedure; and Three-dimensional, dipole density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation. | CT-like cardiac chamber reconstructions imaging, Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters, Cardiac electrical activity as waveform traces, Remapping of the chamber at any time during the procedure; and Three-dimensional, dipole density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation. | Identical |
| Visualization<br>Device/Catheter | AcQMap Catheter (electrodes and<br>transducers) | AcQMap Catheter (electrodes and<br>transducers) | Identical |
| Physical Characteristics - Console/Amplifier Comparison | | | |
| Dimensions | 68 cm L x 48.3 cm W x 72.6 cm D | 68 cm L x 48.3 cm W x 72.6 cm D | Identical |
| Weight Maximum | 50 kg | 50 kg | |
| Power<br>Requirement | 110 - 240 V, 50/60 Hz | 110 - 240 V, 50/60 Hz | |
| Input Current | 5A | 5A | |
| Fuse protection | 250 V, 6.3 A, two high breaking<br>capacity fuses | 250 V, 6.3 A, two high breaking<br>capacity fuses | |
# Table 1: Comparison of Technological Characteristics with the Predicate Device
{7}------------------------------------------------
# Table 1: Comparison of Technological Characteristics with the Predicate Device (cont.)
{8}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE
The indications for use for the AcOMap High Resolution Imaging and Mapping System predicate device are identical to the proposed indications for use for the AcQMap High Resolution Imaging and Mapping System in this submission. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the AcQMap System is substantially equivalent to the predicate device.
#### PERFORMANCE DATA [807.92(b)]
All necessary bench and clinical testing was conducted on the AcQMap System to support a determination of substantial equivalence to the predicate device.
#### [807.92(b)(1)] Nonclinical Testing Summary
The necessary bench testing was performed on the AcQMap System to ensure that it conforms to the design specifications and to support a determination of substantial equivalence to the predicate device. The bench testing performed included the following:
- Transportation Testing ●
- AcQMap Verification Testing ●
- Software Verification and Validation ●
- System Accuracy Testing
- Electromagnetic Compatibility and Electrical Safety Testing ●
- o AcQMap Catheter Validation Testing-Animal Study
The AcQMap System device was tested to verify that the device met the established performance specifications.
The collective results of the testing demonstrate that the design of the AcQMap System meets its established performance specifications necessary for performance during its intended use.
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the AcQMap System meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the AcQMap System does not raise different questions of safety or effectiveness.
#### [807.92(b)(2)] Clinical Testing Summary
Although clinical testing is not required to demonstrate substantial equivalence to the predicate device for its intended use. Acutus performed a clinical study to demonstrate the safety and performance of the AcQMap High Resolution Imaging and Mapping System in gathering data to create right and/or left atrial dipole density activation maps in subjects with supraventricular tachycardia. The study entitled, "Dipole Density Right (and left) Atrial Mapping and Assessment of Therapy In Complex Supraventricular Tachycardia, (DDRAMATIC-SVT)" is a prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S. The results for 84 patients demonstrate that the AcQMap System is safe and effective for its intended use.
{9}------------------------------------------------
# CONCLUSIONS [807.92(b)(3)]
Extensive nonclinical performance testing was conducted on the AcQMap System to evaluate the overall performance of the device. The collective results demonstrate that the AcQMap System is safe and effective for its intended use.
### SUMMARY
The AcQMap System is substantially equivalent to the predicate device, the previously cleared AcQMap System.