K162440 · Cardioinsight Technologies, Inc. · DQK · Nov 4, 2016 · Cardiovascular
Device Facts
Record ID
K162440
Device Name
CardioInsight Cardiac Mapping System
Applicant
Cardioinsight Technologies, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Nov 4, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The Medtronic CardioInsight Cardiac Mapping System is intended for acquisition, analysis, display and storage of cardiac electrophysiological data and maps for analysis by a physician.
Device Story
Non-invasive system for beat-by-beat, multichamber, 3D cardiac mapping; inputs include real-time chest ECG signals from torso-worn sensor array and patient CT scan data; system uses mathematical algorithms to solve cardiac inverse problem, transforming body surface signals into epicardial signals; outputs include interactive 3D color maps (potential, activation, voltage, propagation, phase) and virtual electrograms; used in clinical settings (bedside or EP lab) by physicians; output assists in cardiac electrophysiological analysis and clinical decision-making; benefits include non-invasive visualization of cardiac electrical activity.
Clinical Evidence
Bench testing only. No clinical data required. Verification included hardware (dimensions, mechanical/electrical safety), software/firmware integration, and usability testing in a simulated environment. Compliance with ANSI/AAMI ES 60601-1, IEC 60601-1, and AAMI/ANSI/IEC 60601-1-2 verified.
Technological Characteristics
System includes cart, monitor, core processor, sensor array (vest), amplifier, and cabling. Sensing principle: body surface ECG potentials combined with CT-derived anatomical geometry. Software: moderate level of concern, includes OTS components (Fellow Oak DICOM, Boost C++, DevExpress, WPF, Parallel Studio XE). Connectivity: standalone workstation. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for individuals undergoing an EP study requiring acquisition, analysis, display, and storage of cardiac electrophysiological data and maps.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 4, 2016
Medtronic Inc. Laurie Lewandowski Regulatory Associate 8200 Coral Sea St NE Mounds View, MN 55112
Re: K162440
Trade/Device Name: Medtronic CardioInsight® Cardiac Mapping System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: October 6, 2016 Received: October 7, 2016
Dear Laurie Lewandowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162440
Device Name
Medtronic CardioInsight® Cardiac Mapping System
Indications for Use (Describe)
The Medtronic CardioInsight Cardiac Mapping System is intended for acquisition, analysis, display and storage of cardiac electrophysiological data and maps for analysis by a physician.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo consists of a gray circular graphic above the company name. The graphic shows a human figure in motion, rotating within the circle. The company name, "Medtronic," is written in a bold, blue font.
# 510(k) Summary per 21 CFR §807.92
| Date Summary<br>Prepared: | October 4, 2016 |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic (Owner/Operator)<br>CardioInsight Technologies<br>11000 Cedar Avenue Suite 210<br>Cleveland, OH 44106<br>Establishment Registration No.: 3006526833 |
| Primary Contact: | Laurie Lewandowski<br>Regulatory Affairs Specialist (consultant)<br>Telephone: 612-770-4038<br>Email: laurie.a.lewandowski@medtronic.com |
| Alternate Contact: | Kristyn Benson<br>Director, Regulatory Affairs<br>Telephone 763-526-2523<br>Email: Kristyn.m.benson@medtronic.com |
| | Medtronic Inc./AF Solutions<br>8200 Coral Sea Street NE<br>Mounds View, MN 55112<br>Fax: 763.367.9903 |
| Trade Name: | Medtronic CardioInsight® Cardiac Mapping System |
| Common Name: | Electrophysiological cardiac mapping system |
| Classification Name: | Programmable diagnostic computer, |
| Classification &<br>Panel: | Class II<br>Cardiovascular |
| Product Code: | DQK |
| Regulation: | 21 CFR 870.1425 |
| Predicate Device(s): | ECVUE MAPPING SYSTEM. (K150990) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the Medtronic logo. The logo features a circular graphic with three stylized human figures in different positions, suggesting movement or transformation. Below the graphic is the word "Medtronic" in a bold, blue sans-serif font. The overall design is clean and corporate, conveying a sense of innovation and healthcare focus.
### Device Description:
The Medtronic CardioInsight® Cardiac Mapping System is a non-invasive mapping system for beat-by-beat, multichamber, 3D mapping of the heart. The CardioInsight Cardiac Mapping System displays cardiac maps and virtual electrograms from real-time chest ECG signals (measured by a Sensor Array placed on the torso) and CT scan data. The ECG signals in concert with the CT scan information (geometrical information) are used in mathematical algorithms to transform the measured body surface signals into epicardial signals via solving the cardiac inverse problem. The CardioInsight Cardiac Mapping System software uses this data to provide various cardiac signal analyses and displays interactive 3D color maps including potential, activation, voltage, propagation, and phase maps. The CardioInsight Cardiac Mapping System system is mobile and can be used for mapping at the patient's bedside or in the EP lab.
| System<br>Component | or<br>Model Number |
|-----------------------------------------|--------------------------------------------|
| CardioInsight Cardiac<br>Mapping System | CIT 200 |
| Work Station<br>Computer | CIT 200-PC-0001 (w/SW) |
| Sensor Array (Vest) | CIT200-SA-0001 (S) |
| | CIT 200-SA-0002 (M) |
| | CIT 200-SA-0003 (L) |
| | CIT 200-SA-0004 (XL) |
| Cart | 200-WC-0001 |
| Amplifier | CIT 200-AMP-0001 |
| Connector Cables | CIT 200-AMP-0002 Signal Cable (Right) |
| | CIT 200-AMP-0003 Signal Cable (Left) |
| | CIT 200-AMP-0004 Signal Cable (Back Left) |
| | CIT 200-AMP-0005 Signal Cable (Back Right) |
| Patient/Ground<br>Reference Cable | CIT 200-AMP-0021 |
Model numbers of the system components are as follows:
Indications for Use: The Medtronic CardioInsight Cardiac Mapping System is intended for acquisition, analysis, display and storage of cardiac electrophysiological data and maps for analysis by a physician.
The Indications for Use statement is identical to the predicate device.
The CardioInsight Cardiac Mapping System uses the same technology, and Comparison of has the same intended use, fundamental technology, principal of operation and Technological Characteristics performance as the predicate device. Modifications were made for processing capacity and usability.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the Medtronic logo. The logo features a stylized graphic of a human figure in motion, depicted in a circular frame. Below the graphic is the word "Medtronic" in a bold, sans-serif typeface, with the letters in a dark blue color.
| Characteristic | Medtronic<br>CardioInsight Cardiac<br>Mapping System<br>(Proposed System) | CardioInsight<br>ECVUE Mapping<br>System<br>(Predicate System) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Identical | For individuals<br>undergoing an EP Study |
| Indications for Use | Identical | The CardioInsight<br>ECVUE Mapping<br>System is intended for<br>acquisition, analysis,<br>display and storage of<br>cardiac<br>electrophysiological data<br>and maps for analysis by<br>a physician |
| System | Cart,<br>Monitor,<br>Core Processor,<br>Keyboard,<br>Mouse,<br>Isolation Transformer,<br>Cabling,<br>Sensor Array,<br>Second Monitor connection | Cart,<br>Monitor,<br>Core Processor,<br>Keyboard,<br>Mouse,<br>Isolation Transformer,<br>Cabling,<br>Sensor Array,<br>Printer,<br>UPS |
| Principal of Operation | Identical | Electrocardiographic<br>potentials are<br>measured from the<br>torso sensors on the<br>surface of the body. A<br>CT scan is segmented<br>to obtain the 3-<br>dimensional location<br>of each sensor and the<br>detailed anatomy of the<br>epicardial surface of<br>the heart. From these<br>data, the system uses<br>mathematical<br>algorithms to use the<br>geometrical<br>information to<br>transform the<br>measured body surface<br>signals into<br>epicardial signals via<br>solving the cardiac<br>inverse problem |
| Characteristic | Medtronic<br>CardioInsight Cardiac<br>Mapping System<br>(Proposed System) | CardioInsight<br>ECVUE Mapping<br>System<br>(Predicate System) |
| Software / Firmware /<br>Algorithm | Equivalent functionality –<br>improved usability,<br>reducing manual input<br>required and fixing bugs<br>OTS software | Create patient records<br>Segment heart and vest<br>electrodes<br>Acquire sensor array<br>signals<br>Create and review maps |
| Off the Shelf Software | Updated to latest versions.<br>Added additional Off the<br>Shelf Software to support<br>functionality<br>• Fellow Oak DICOM<br>• Boost C++ Libraries<br>• DevExpress<br>• WPF Components<br>Library<br>• Parallel Studio XE<br>Composer | Entity Framework<br>SQL Server Express<br>OpenInventor 64-bit<br>.NET Framework |
| Performance testing was completed on the CardioInsight Cardiac Mapping<br>System which verified that the System complies with the safety and<br>specifications and performs as designed; it is suitable for its intended use. | | |
| Performance testing for the proposed system included the following: | | |
| Hardware Verification compliance with dimensions, cart functionality and<br>mechanical safety as well as the amplifier mechanical and electrical<br>functionality and electrical safety.<br>Labeling Verification to applicable requirements<br>Electrical Safety compliance per<br>ANSI/AAMI ES 60601-1:2005/A1:2012<br>IEC 60601-1:2005/A1:2012<br>EN 60601 1:2006/A1:2013<br>All excluding: excluding biocompatibility (cl. 11.7), usability (cl. 1<br>and 15.1), and EMC (cl. 17)<br>EMC/EMI compliance per<br>AAMI / ANSI / IEC 60601-1-2:2007/(R)2012 | | |
| Mechanical Safety compliance per<br>ANSI/AAMI ES 60601-1:2005/A1:2012<br>IEC 60601-1:2005/A1:2012<br>EN 60601 1:2006/A1:2013<br>All excluding: excluding biocompatibility (cl. 11.7), usability (cl. 1<br>and 15.1), and EMC (cl. 17) | | |
| | with FDA's Guidance for the Content of Premarket Submissions for Software<br>Contained in Medical Devices" and AAMI / ANSI / IEC 62304:2006, Medical<br>Device Software - Software Life Cycle Processes. The software in this system<br>is considered moderate level of concern as failure could contribute to a hazard<br>leading to non-serious injury.<br>Algorithm Testing and Integration – verified the algorithms met requirements<br>and functioned as intended and when integrated performed as expected. | |
| | Firmware Verification – demonstrated that the firmware met requirements<br>Packaging Validation was performed and demonstrated that the packaging and<br>the system met the environmental conditioning and simulated shipping per the<br>applicable sections of ASTM D4332-14 and ASTM D4169.<br>Usability Testing<br>System Verification and Validation testing for functionality and performance<br>in a simulated environment. | |
| | Preclinical and clinical testing were not required for the CardioInsight Cardiac<br>Mapping System.<br>Testing demonstrated that the CardioInsight Cardiac Mapping System met the | |
| | requirements and functioned as intended. | |
| Discussion | Minor differences in technology, update of hardware due to obsolescence,<br>removal of the printer that was not used, removal of the UPS for portability<br>and not necessary for function of current system and the addition of an<br>optional second monitor do not change the fundamental technology.<br>Modifications to the software, firmware and algorithms were made for<br>usability, reducing the manual interaction required. Updates were made to<br>firmware for bug fixes, off the shelf software to update to the latest versions<br>and additions to support functionality. Indications for use, principals of<br>operations and fundamental technology do not change with these<br>modifications. | |
| Conclusion | The data presented in this submission demonstrate that the Medtronic<br>CardioInsight Cardiac Mapping System is substantially equivalent to the<br>predicate device identified in intended use, device design, fundamental<br>technology and performance. | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular graphic above the word "Medtronic" in a bold, blue font. The graphic features a stylized human figure in motion, depicted in a light gray color. The figure appears to be rotating or moving through different positions within the circle.
Performance Data
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the Medtronic logo. The logo features a circular graphic with three stylized human figures in different poses, suggesting movement or transformation. Below the graphic is the word "Medtronic" in a bold, blue sans-serif font. The overall design is clean and corporate, reflecting the company's presence in the medical technology industry.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.