← Product Code [DQK](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK) · K101419

# ENSITE VELOCITY SYSTEM, MODEL EE3000 (K101419)

_St Jude Medical · DQK · Jun 11, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK/K101419

## Device Facts

- **Applicant:** St Jude Medical
- **Product Code:** [DQK](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK.md)
- **Decision Date:** Jun 11, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated. When used with the EnSite Array™ Catheter, the EnSite System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. OR When used with the EnSite NavX Surface Electrode Kit, the EnSite System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.

## Device Story

EnSite Velocity System is an electrophysiology mapping workstation; includes display console, amplifier, EnSite Array™ diagnostic catheter, and EnSite NavX Surface Electrode Kit. Used in clinical settings by physicians to perform cardiac mapping. System processes electrical signals from catheters and surface electrodes to visualize catheter position and cardiac electrical activity. EnSite Courier™ software feature enables networking and DICOM-compliant data transfer to PACS. Output provides real-time visualization of catheter location and cardiac mapping to assist physicians in identifying complex arrhythmias and guiding diagnostic procedures. Benefits include improved identification of complex arrhythmias compared to conventional mapping systems.

## Clinical Evidence

No clinical trials performed. Evidence consists of cross-vendor interoperability testing sessions conducted between EnSite Courier and 15 different PACS models/versions across seven vendors to confirm data communication functionality.

## Technological Characteristics

Programmable diagnostic computer (21 CFR 870.1425). System components: display workstation, amplifier, EnSite Array™ catheter, EnSite NavX Surface Electrode Kit. Connectivity: DICOM-compliant networking via EnSite Courier™ software. Software-based signal processing for cardiac mapping and catheter localization.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- EnSite Velocity System v.1.3 ([K093942](/device/K093942.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K101419
P1/3

# 12 510(k) Summary for Public Disclosure

JUN
11 2010

### 12.1 Submitter's Name/Contact Person

Donna R. Lunak St. Jude Medical One St. Jude Medical Drive St. Paul, MN 55117 USA

The Establishment Registration Number is 2184149.

#### 12.2 Common or Usual Name

Electrophysiology Mapping System with console and catheter

#### 12.3 Proprietary Name

EnSite Velocity System Consisting of: EnSite Multi-electrode Diagnostic Catheter (EnSite Array™ - Model EC1000) EnSite Electrophysiology Workstation - Model EE3000 EnSite NavX Surface Electrode Kit - Model EN0010

### 12.4 Classification Name

DQK, Programmable diagnostic computer (21 CFR 870.1425), Class II, Circulatory Systems Device Panel

#### 12.5 Hardware Description

The EnSite Velocity System consists of the following:

- Display Workstation .
- Amplifier ●

{1}------------------------------------------------

## 12.6 Indications for Use

The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.

- When used with the EnSite Array™ Catheter, the EnSite System is . intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
OR

- When used with the EnSite NavX Surface Electrode Kit, the EnSite . System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.
#### 12.7 Device Comparison to the Predicate Device

The EnSite Velocity System with EnSite Courier™ has the same intended use and fundamental scientific technology as the predicate device, EnSite Velocity System v.1.3 (K093942). All technological characteristics of the EnSite Velocity System with EnSite Courier™ are substantially equivalent to the predicate device, EnSite Velocity System v. 1.3 (K093942).

#### 12.8 DICOM Conformance Statement

The EnSite Velocity System with EnSite Courier ™ conforms to the ACR-NEMA Digital Imaging and Communications in Medicine (DICOM) standard. A DICOM Conformance Statement is presented in Appendix R.

#### 12.9 Summary of Non-Clinical Testing

Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product.

#### 12.10 Summary of Clinical Testing

Cross-vendor interoperability testing sessions were conducted between EnSite Courier and 15 different PACS models/versions, seven vendors. This clinical test report can be found in Appendix S.

## 12.11 Summary of Design Control Activities

The development of the EnSite Networking Software Feature - EnSite Courier™ was performed in accordance with St. Jude Medical's Quality System requirements, and in compliance with Quality System Regulation design controls

{2}------------------------------------------------

K101419
P 3/3

requirements documented in 21 CFR 820.30. A Declaration of Conformity with Design Controls follows in section 13.4.

#### 12.12 Conclusion

The EnSite Velocity System with EnSite Courier ™ has the same indications for use, intended use and fundamental scientific technology as the predicate device. All technological characteristics of the EnSite Velocity System with EnSite Courier ™ are substantially equivalent to the predicate device, EnSite Velocity System v.1.3 (K093942).

Where operational and performance differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness.

Therefore, St. Jude Medical considers the EnSite Velocity System to be substantially equivalent to the predicate device, EnSite Velocity System v.1.3 (K093942).

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with flowing lines, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

JUN 1 1 2010

St. Jude Medical c/o Ms. Donna Lunak Regulatory Specialist II One St. Jude Medical Drive Saint Paul, MN 55117

K101419 Re:

> Trade/Device Name: EnSite Velocity System Regulatory Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: II (two) Product Code: DQK Dated: May 19, 2010 Received: May 20, 2010

Dear Ms. Lunak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Ms. Donna Lunak

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Bram D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Directo Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# 7 Indications for Use

Device Name: EnSite Velocity System

Indications for Use:

The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.

- When used with the EnSite Array™ Catheter, the EnSite System is intended to . be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
- OR
- When used with the EnSite NavX Surface Electrode Kit, the EnSite System is � intended to display the position of conventional electrophysiology (EP) catheters in the heart.

Prescription Use X
(Part 21 CFR 801 Subpart D) AND/OR   Over-The-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign Off

(Division Sia Devices

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK/K101419](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK/K101419)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
