← Product Code [DQK](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK) · K051471

# CARDIODAY (K051471)

_Getemed Medizin- Und Informationstechnik AG · DQK · Jun 29, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK/K051471

## Device Facts

- **Applicant:** Getemed Medizin- Und Informationstechnik AG
- **Product Code:** [DQK](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK.md)
- **Decision Date:** Jun 29, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

CardioDay" is a software package that allows a trained physician or health care professional knowledgeable in Holter interpretation, after having performed a long-term continuous electrocardiographic (ECG) recording on digital flash memory Holter recorder, to download and analyze the data from the recorder, review it and produce printed reports. This device is available only upon the order of a physician or other licensed medical professional and not intended for any ambulatory or home applications. Federal law restricts CardioDay® to use on order of a physician.

## Device Story

CardioDay is a software package for clinical use by trained physicians/healthcare professionals to analyze long-term continuous ECG data recorded on digital flash memory Holter recorders. The device imports data from the recorder, performs high-speed analysis, and displays ECG morphologies, heart rate trends, RR variability, and statistical values. It does not perform automated diagnosis; the physician reviews, edits, and prints reports to support clinical decision-making. The device is used in clinical settings (not home/ambulatory) to assist in evaluating cardiac functionality and arrhythmias. Benefits include providing organized, reviewable data for diagnostic assessment of cardiac symptoms.

## Clinical Evidence

No clinical data. Performance evaluation conducted using standard American Heart Association (AHA) and Massachusetts Institute of Technology (MIT) databases to validate analysis modes and parameters.

## Technological Characteristics

Software package delivered on CD; runs on IBM PC AT compatible hardware (Pentium III 500MHz+, 128MB RAM, Windows 98 SE/NT/2000/XP). Complies with IEC 60601-1-4, IEC 60601-2-47, and ANSI/AAMI EC38. Performs high-speed ECG analysis and graphical display of statistical values.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- Holter for Windows® ([K930564](/device/K930564.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 29 2005

| 510(k) Notification | getemed              |
|---------------------|----------------------|
| CardioDay®          | Project ID: 0429S1   |
| 510(k) - Summary    | Section 1-0001-Rev D |

#### 1 510(k) - Summary

| Submitted By:             | getemed Medizin- und Informationstechnik AG<br>Oderstr. 59<br>14513 Teltow<br>Germany<br>Tel.: +49 3328 - 3942-70<br>Fax: +49 3328 - 3942-99 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                  | Dr. Bert Schadow<br>Regulatory Affairs Manager                                                                                               |
| Manufacturing Facility:   | getemed Medizin- und Informationstechnik AG<br>Oderstr. 59<br>14513 Teltow<br>Germany                                                        |
| Date of Preparation:      | 2005-06-13                                                                                                                                   |
| Trade Name:               | CardioDay®                                                                                                                                   |
| Common Name:              | Holter ECG                                                                                                                                   |
| Classification Name:      | Computer, Diagnostic, Programmable                                                                                                           |
| Product Classification:   | 21 CFR 870.1425, Class II                                                                                                                    |
| Product Code:             | DQK                                                                                                                                          |
| Legally Marketed Devices: | Holter for Windows® (K930564, Northeast<br>Monitoring Inc.)                                                                                  |

#### Reason for Submission

Premarket notification for CardioDay® (Version 1.9.5), a New Device, seeking authority to market the device under Section 510(k) as a device that is substantially equivalent to the Holter for Windows® evaluation software (K930564, Northeast Monitoring Inc.).

#### Intended Use

CardioDay" is a software package that allows a trained physician or health care professional knowledgeable in Holter interpretation, after having performed a long-term continuous electrocardiographic (ECG) recording on digital flash memory Holter recorder, to download and analyze the data from the recorder, review it and produce printed reports.

This device is available only upon the order of a physician or other licensed medical professional and not intended for any ambulatory or home applications.

Federal law restricts CardioDay® to use on order of a physician.

<sup>0429</sup>S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc CONFIDENTIAL AND PROPRIETARY

{1}------------------------------------------------

| 510(k) Notification | getemed              |
|---------------------|----------------------|
| CardioDay®          | Project ID: 0429S1   |
| 510(k) - Summary    | Section 1-0001-Rev D |

#### Indications for Use

CardioDay is a Holter software which is indicated for patients who may benefit from a long-term continuous electrocardiographic (ECG) recording, but not limited to, those with complaints of palpitations, syncope, chest pain, shortness of breath, or those that need to be monitored to judge their current cardiac functionality such as patients who have recently received pacemakers.

#### WARNING:

CardioDay® does not perform any diagnosis of data by itself but only displays ECG morphologies and associated, calculated graphs such as heart rate trends, RR variability, and other statistical values.

#### Device Description

CardioDay® does not perform any diagnosis of data by itself but only displays ECG morphologies and associated, calculated graphs such as heart rate trends, RR variability, and other statistical values in graphical form. The physician will be able to review, edit, and print the data collected.

| Description  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Holter for Windows® is Holter<br>monitoring software, designed to<br>perform high-speed data analysis<br>of continuous, long-term<br>electrocardiograms saved in the<br>form of direct recordings on a<br>cassette tape or flash memory<br>card. Holter for Windows® is<br>intended for use as an analysis<br>tool and produces printed reports<br>to be reviewed by a person<br>knowledgeable in Holter<br>interpretation.<br><br>Federal law restricts Holter for<br>Windows® to use on order of a<br>physician. | CardioDay® is a software package<br>that allows a trained physician or<br>health care professional<br>knowledgeable in Holter<br>interpretation, after having<br>performed a long-term continuous<br>electrocardiographic (ECG)<br>recording on digital flash memory<br>Holter recorder, to download and<br>analyze the data from the recorder,<br>review it and produce printed<br>reports.<br><br>This device is available only upon<br>the order of a physician or other<br>licensed medical professional and<br>not intended for any ambulatory or<br>home applications.<br><br>Federal law restricts CardioDay® to<br>use on order of a physician. |

## Comparison to Legally Marketed Device

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc CONFIDENTIAL AND PROPRIETARY

Page 2 of 11

{2}------------------------------------------------

| 510(k) Notification |                                          | getemed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |  |
|---------------------|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| CardioDay®          |                                          | Project ID: 0429S1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |
| 510(k) - Summary    |                                          | Section 1-0001-Rev D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |
| Description         | Facility's CardioDay®<br>Predicate Index |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Indications for use | Same as CardioDay®                       | CardioDay® is a Holter software<br>which is indicated for patients who<br>may benefit from a long-term<br>continuous electrocardiographic<br>(ECG) recording, including, but not<br>limited to, those with complaints of<br>palpitations, syncope, chest pain,<br>shortness of breath, or those that<br>need to be monitored to judge their<br>current cardiac functionality such as<br>patients who have recently received<br>pacemakers.<br><br>WARNING:<br>CardioDay® does not perform any<br>diagnosis of data by itself but only<br>displays ECG morphologies and<br>associated, calculated graphs such<br>as heart rate trends, RR variability,<br>and other statistical values. |  |
| Device description  | Same as CardioDay®                       | CardioDay® does not perform any<br>diagnosis of data by itself but only<br>displays ECG morphologies and<br>associated, calculated graphs such<br>as heart rate trends, RR variability,<br>and other statistical values in<br>graphical form. The physician will be<br>able to review, edit, and print the<br>data collected.                                                                                                                                                                                                                                                                                                                                                          |  |
| Target population   | Same as CardioDay®                       | Patients who may benefit from a<br>long-term continuous<br>electrocardiographic (ECG)<br>recording, including, but not limited<br>to, those with complaints of<br>palpitations, syncope, chest pain,<br>shortness of breath, or those that<br>need to be monitored to judge their<br>current cardiac functionality such as<br>patients who have recently received<br>pacemakers.                                                                                                                                                                                                                                                                                                       |  |
| Design / materials  | Same as CardioDay®                       | CardioDay® is a software package<br>which is delivered on CD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |
| Sterility           | This section does not apply.             | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Biocompatibility    | This section does not apply.             | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Mechanical safety   | This section does not apply.             | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Chemical safety     | This section does not apply.             | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Anatomical sites    | This section does not apply.             | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

·

Page 3 of 11

ાં ડા

{3}------------------------------------------------

| 510(k) Notification                                    |                                                                                                                                       | getemed                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CardioDay®                                             |                                                                                                                                       | Project ID: 0429S1                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 510(k) - Summary                                       |                                                                                                                                       | Section 1-0001-Rev D                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Description                                            | Holter for Windows®                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Human factors                                          | Same as CardioDay®                                                                                                                    | Human factors engineering:<br>- device technology (see<br>comparison of technology<br>characteristics),<br>- the users (see target<br>population),<br>- environment in which the<br>technology will be used (see<br>intended use),<br>- how dangerous is the use of the<br>device (see level of concern in<br>section 2),<br>- how critical is the device for<br>patient care (see product<br>classification and level of<br>concern in section 2). |
| Energy used and/or<br>delivered                        | This section does not apply.                                                                                                          | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Compatibility with<br>environment and<br>other devices | See comparison of technology<br>characteristics and miscellaneous                                                                     | See comparison of technology<br>characteristics and miscellaneous                                                                                                                                                                                                                                                                                                                                                                                   |
| Where used:<br>hospital, home,<br>ambulance, etc.      | Holter for Windows® is available<br>for sale only upon the order of a<br>physician or other related licensed<br>medical professional. | CardioDay® is available for sale only<br>upon the order of a physician or<br>other related licensed medical<br>professional and not intended for<br>any home use applications.                                                                                                                                                                                                                                                                      |
| Electrical safety                                      | This section does not apply.                                                                                                          | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Thermal safety                                         | This section does not apply.                                                                                                          | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Radiation safety                                       | This section does not apply.                                                                                                          | This section does not apply.                                                                                                                                                                                                                                                                                                                                                                                                                        |

Comparison of Technology Characteristics Compared to Legally Marketed Device:

| Specifications               |                                           |                                                           |
|------------------------------|-------------------------------------------|-----------------------------------------------------------|
| Type                         | IBM PC AT compatible                      | IBM PC AT compatible                                      |
| CPU                          | Pentium II Processor                      | Pentium III, 500MHz or greater                            |
| RAM                          | 32 Mbytes minimum                         | 128 Mbytes minimum,<br>256 Mbytes minimum for XP          |
| Free hard disk<br>space      | 200 MB minimum                            | 5 GB minimum<br>20 GB for 12-lead recordings              |
| Display                      | VGA, 14", 1024 x 768 pixel, 256<br>colors | 17" CRT or 15" TFT, 1024 x 768<br>pixel (XGA), 256 colors |
| Disc drive / floppy<br>drive | 1.4 Mbytes                                | Not required                                              |

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

.

{4}------------------------------------------------

| 510(k) Notification<br>CardioDay® |                                        | o getemed                                                                                                                                 |
|-----------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Project ID: 0429S1                |                                        |                                                                                                                                           |
|                                   | 510(k) - Summary                       | Section 1-0001-Rev D                                                                                                                      |
| Specifications                    | Hotter for Win                         |                                                                                                                                           |
| CD-ROM drive                      | For installation only                  | For installation only.<br>Needs to be replaced by a CD-ROM<br>writer or DVD writer if archive option<br>implemented                       |
| Operating system                  | Windows 98                             | Windows 98 SE, NT (SP6a), 2000<br>or XP                                                                                                   |
| Ports                             | One free parallel port                 | 1 parallel port sufficient; if USB<br>devices used for printing, software<br>key or card reader, then<br>corresponding USB ports required |
| Printer                           | Printer as any Windows ™<br>compatible | Printer as any Windows ™<br>compatible, 300 dpi                                                                                           |
| Keyboard                          | Standard device                        | Standard device                                                                                                                           |
| Mouse                             | Standard, 2-button device              | Standard, 2 or 3-button device                                                                                                            |
| Installation media                | (1) CD                                 | (1) CD                                                                                                                                    |
| Further periphery                 | Compact Flash card reader              | CompactFlash (Type I) memory<br>card reader                                                                                               |

# Comparison of Software Characteristics:

| Patient Screen                                              | FORM DESIGNATION<br>Section Windows |     |
|-------------------------------------------------------------|-------------------------------------|-----|
| Patient Identification Number                               | Yes                                 | Yes |
| Patient Name, Address, Telephone                            | Yes                                 | Yes |
| Patient Personal Data (Age, Gender, Date of Birth,<br>etc.) | Yes                                 | Yes |
| Medication                                                  | Yes                                 | Yes |
| Indication                                                  | Yes                                 | Yes |
| Physician's Name                                            | Yes                                 | Yes |
| Date of Recording                                           | Yes                                 | Yes |

| Analysis Options             |     |     |
|------------------------------|-----|-----|
| Analysis Duration            | Yes | Yes |
| Primary Channel Selection    | Yes | Yes |
| Sensitivity / Signal Quality | Yes | Yes |
| Tachycardia Threshold [bpm]  | Yes | Yes |
| Bradycardia Threshold [bpm]  | Yes | Yes |
| Pause Duration [ms]          | Yes | Yes |
| SV Prematurity [%]           | Yes | Yes |
| V Prematurity [%]            | Yes | Yes |
| R on T [ms]                  | Yes | Yes |
| Pacemaker Type               | Yes | Yes |
| Minimum Pulse Rate [bpm]     | Yes | Yes |
| Maximum Pulse Rate [bpm]     | Yes | Yes |
| Superimposition / QuickScan  | Yes | Yes |

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

·

Page 5 of 11

长0.51471/51

75/11

{5}------------------------------------------------

| 510(k) Notification                         |                    | getemed                 |  |
|---------------------------------------------|--------------------|-------------------------|--|
| CardioDay®                                  |                    | Project ID: 0429S1      |  |
| 510(k) - Summary                            |                    | Section 1-0001-Rev D    |  |
| Analysis Options                            | Holter for Windows | New Device<br>CardioDay |  |
| 12-Lead ECG Module                          | Yes                | Yes                     |  |
| Holter Data Transfer                        | Yes                | Yes                     |  |
| Events Detected                             | Holter for Windows | New Device<br>CardioDay |  |
| Burst                                       | Yes                | Yes                     |  |
| VES / VPB (Extrasystole)                    | Yes                | Yes                     |  |
| SVES / SVPB (Supraventricular Extrasystole) | Yes                | Yes                     |  |
| Couplet / VPB Pair                          | Yes                | Yes                     |  |
| Triplet                                     | Yes                | Yes                     |  |
| VTACH / VTAC (Ventricular Tachycardia)      | Yes                | Yes                     |  |
| Bigeminy                                    | Yes                | Yes                     |  |
| R on T                                      | Yes                | Yes                     |  |
| ST-Analysis                                 | Yes                | Yes                     |  |
| SVTACH / SVT (Supraventricular Tachycardia) | Yes                | Yes                     |  |
| Arrhythmia / Irregular R-R                  | Yes                | Yes                     |  |
| Bradycardia                                 | Yes                | Yes                     |  |
| VTACH (with 4 Beats) / Burst                | Yes                | Yes                     |  |
| V. STIM / V. Paced                          | Yes                | Yes                     |  |
| A. STIM / A. Paced                          | Yes                | Yes                     |  |
| AV. STIM / AV Paced                         | Yes                | Yes                     |  |
| Undersense / Sense Failure                  | Yes                | Yes                     |  |
| Exitblock / Capture Failure                 | Yes                | Yes                     |  |
| Oversense / Inhibition                      | Yes                | Yes                     |  |
| Pause / Arrest                              | Yes                | Yes                     |  |
| Event Marker                                | Yes                | Yes                     |  |
| HR Stripes                                  | Yes                | Yes                     |  |
| Artifact                                    | Yes                | Yes                     |  |
| Normal                                      | Yes                | Yes                     |  |

| Functionality Available                                                                                                                  |     |     |
|------------------------------------------------------------------------------------------------------------------------------------------|-----|-----|
| The comparison between the legally marketed de-<br>vice and the NEW DEVICE is based on their func-<br>tions only and not on their names. |     |     |
| Start                                                                                                                                    | Yes | Yes |
| Read Tape                                                                                                                                | Yes | Yes |
| Read Digital Recorder                                                                                                                    | Yes | Yes |
| Import                                                                                                                                   | Yes | Yes |
| Analyze New                                                                                                                              | Yes | Yes |
| Open                                                                                                                                     | Yes | Yes |
| Edit Patient Data                                                                                                                        | Yes | Yes |
| Print Preview                                                                                                                            | Yes | Yes |
| Print                                                                                                                                    | Yes | Yes |
| Close Recording                                                                                                                          | Yes | Yes |
| Close                                                                                                                                    | Yes | Yes |

0429S1-Sec01-0001-RevD-CardioDay-Execulive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

Page 6 of 11

{6}------------------------------------------------

| 510(k) Notification                                                                                                                                                                                                                                                 |                                               | o getemed                     |  |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|-------------------------------|--|
| CardioDay®                                                                                                                                                                                                                                                          |                                               | Project ID: 0429S1            |  |
| 510(k) - Summary                                                                                                                                                                                                                                                    | Section 1-0001-Rev D                          |                               |  |
| Functionality Available                                                                                                                                                                                                                                             | Legally Marketed Device<br>Holter for Windows | Predicate Device<br>CardioDay |  |
| Delete Recording                                                                                                                                                                                                                                                    | Yes                                           | Yes                           |  |
| Archive                                                                                                                                                                                                                                                             | Yes                                           | Yes                           |  |
| Diagnosis                                                                                                                                                                                                                                                           | Yes                                           | Yes                           |  |
| View ECG                                                                                                                                                                                                                                                            | Yes                                           | Yes                           |  |
| Print Preview                                                                                                                                                                                                                                                       | Yes                                           | Yes                           |  |
| Screen Scale Calibration                                                                                                                                                                                                                                            | Yes                                           | Yes                           |  |
| Screen: Colour Setup                                                                                                                                                                                                                                                | Yes                                           | Yes                           |  |
| FFT Setup                                                                                                                                                                                                                                                           | Yes                                           | Yes                           |  |
| Report Setup                                                                                                                                                                                                                                                        | Yes                                           | Yes                           |  |
| Classes Display                                                                                                                                                                                                                                                     | Yes                                           | Yes                           |  |
| PM Events Display                                                                                                                                                                                                                                                   | Yes                                           | Yes                           |  |
| Events Display                                                                                                                                                                                                                                                      | Yes                                           | Yes                           |  |
| HR Min./Max. Display                                                                                                                                                                                                                                                | Yes                                           | Yes                           |  |
| Statistics Display                                                                                                                                                                                                                                                  | Yes                                           | Yes                           |  |
| Diagnosis Display                                                                                                                                                                                                                                                   | Yes                                           | Yes                           |  |
| Overview Display                                                                                                                                                                                                                                                    | Yes                                           | Yes                           |  |
| Context sensitive Help                                                                                                                                                                                                                                              | Yes                                           | Yes                           |  |
| Keyboard Shortcuts Help                                                                                                                                                                                                                                             | Yes                                           | Yes                           |  |
| Menu Entries Help                                                                                                                                                                                                                                                   | Yes                                           | Yes                           |  |
| Help: About                                                                                                                                                                                                                                                         | Yes                                           | Yes                           |  |
| Help: Version                                                                                                                                                                                                                                                       | Yes                                           | Yes                           |  |
| Icons/Buttons Available                                                                                                                                                                                                                                             | Legally Marketed Device (Hotar or Window)     | NEW DEVICE (Display)          |  |
| The following comparison between the legally<br>marketed device and the NEW DEVICE is based on<br>their functions only. The label and form of the icons<br>/ buttons, however, are different. It might even take<br>more than one click to initiate a given action. |                                               |                               |  |
| Start: Read Digital Recorder                                                                                                                                                                                                                                        | Yes                                           | Yes                           |  |
| Start: Read Tape Recorder                                                                                                                                                                                                                                           | Yes                                           | Yes                           |  |
| Start: Open Existing Record                                                                                                                                                                                                                                         | Yes                                           | Yes                           |  |
| Digital Recorder                                                                                                                                                                                                                                                    | Yes                                           | Yes                           |  |
| Tape Recorder                                                                                                                                                                                                                                                       | Yes                                           | Yes                           |  |
| Open: List of Patients                                                                                                                                                                                                                                              | Yes                                           | Yes                           |  |
| Print                                                                                                                                                                                                                                                               | Yes                                           | Yes                           |  |
| Rhythm Analysis                                                                                                                                                                                                                                                     | Yes                                           | Yes                           |  |
| Print Preview on Screen                                                                                                                                                                                                                                             | Yes                                           | Yes                           |  |
| Options Available                                                                                                                                                                                                                                                   | Legally Marketed Device (Note on Wind)        | New Device (Note on Wind)     |  |
| The following comparison between the legally mar-<br>keted device and the NEW DEVICE is based on their<br>functions only. The name of those options may vary.                                                                                                       |                                               |                               |  |
| Classes                                                                                                                                                                                                                                                             | Yes                                           | Yes                           |  |
| Events                                                                                                                                                                                                                                                              | Yes                                           | Yes                           |  |
| Heart Rate Min/Max                                                                                                                                                                                                                                                  | Yes                                           | Yes                           |  |

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

Page 7 of 11

051471/51

{7}------------------------------------------------

| 510(k) Notification                                                                                                                                                               | getemed                                        |                          |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|--------------------------|
| CardioDay®                                                                                                                                                                        | Project ID: 0429S1                             |                          |
| 510(k) - Summary                                                                                                                                                                  | Section 1-0001-Rev D                           |                          |
| Options Available                                                                                                                                                                 | Legally Marketed Device Holter for Windows     | New Device CardioDay     |
| Average Heart Rate                                                                                                                                                                | Yes                                            | Yes                      |
| Statistics: FFT Analysis                                                                                                                                                          | Yes                                            | Yes                      |
| Statistics: ST Diagrams                                                                                                                                                           | Yes                                            | Yes                      |
| Report                                                                                                                                                                            | Yes                                            | Yes                      |
| Overview                                                                                                                                                                          | Yes                                            | Yes                      |
| Heart Variability: RR Delay                                                                                                                                                       | Yes                                            | Yes                      |
| Heart Rate Variability: RR FFT                                                                                                                                                    | Yes                                            | Yes                      |
| Heart Rate Variability: 24h RR FFT                                                                                                                                                | Yes                                            | Yes                      |
| Heart Rate Variability: RR Histograms                                                                                                                                             | Yes                                            | Yes                      |
| Graphics & Displays Available                                                                                                                                                     | Legally Marketed Device:<br>Holter for Windows | New Device:<br>CardioDay |
| The following comparison between the legally mar-<br>keted device and the NEW DEVICE is based on their<br>functions only. The label and appearance of those<br>displays may vary. |                                                |                          |
| Classified Beats Grouped into Morphology Bins                                                                                                                                     | Yes                                            | Yes                      |
| Zoomed Version of Selected Beat                                                                                                                                                   | Yes                                            | Yes                      |
| Context of Selected Beat                                                                                                                                                          | Yes                                            | Yes                      |
| Events                                                                                                                                                                            | Yes                                            | Yes                      |
| Heart Rate Trend in Recording Period                                                                                                                                              | Yes                                            | Yes                      |
| Average RR Interval                                                                                                                                                               | Yes                                            | Yes                      |
| Y-T Distribution                                                                                                                                                                  | Yes                                            | Yes                      |
| RR > 50ms Distribution                                                                                                                                                            | Yes                                            | Yes                      |
| FFT Analysis                                                                                                                                                                      | Yes                                            | Yes                      |
| ST Diagrams                                                                                                                                                                       | Yes                                            | Yes                      |
| Overview 2 channels at Different Scaling Factors                                                                                                                                  | Yes                                            | Yes                      |
| Printout Options                                                                                                                                                                  | Legally Marketed Device<br>Hoferator Window    | New Device<br>CardioDay  |
| The following comparison between the legally mar-<br>keted device and the NEW DEVICE is based on their                                                                            |                                                |                          |
| functionality only. The commands to generate a<br>given printout as well as its appearance do vary.                                                                               |                                                |                          |
| Full Disclosure 2 Channels, 1 h/Page                                                                                                                                              | Yes                                            | Yes                      |
| Full Disclosure 2 Channels, 15 min./Page                                                                                                                                          | No                                             | Yes                      |
| Full Disclosure 2 Channels, 30 min./Page                                                                                                                                          | Yes                                            | Yes                      |
| Marked Events: 8 Events/Page                                                                                                                                                      | Yes                                            | Yes                      |
| Marked Events: 32 Events/Page                                                                                                                                                     | Yes                                            | Yes                      |
| Marked Events: Analysis Channel 25 mm/s                                                                                                                                           | Yes                                            | Yes                      |
| Marked Events: Analysis Channel 2.5 min +<br>25mm/s                                                                                                                               | No                                             | Yes                      |
| Selected Channels 25 mm/s                                                                                                                                                         | Yes                                            | Yes                      |
| Selected Channels 1 min. + 25 mm/s                                                                                                                                                | Yes                                            | Yes                      |
| Selected Channels 2.5 min. + 25 mm/s                                                                                                                                              | No                                             | Yes                      |
| Selected Channels 10 min. + 25 mm/s                                                                                                                                               | No                                             | Yes                      |
| Event Table                                                                                                                                                                       | Yes                                            | Yes                      |

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

Page 8 of 11

051471/51

가이디

{8}------------------------------------------------

| 510(k) Notification                                                                                                                                                     | getemed                                    |                       |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-----------------------|
| CardioDay®                                                                                                                                                              | Project ID: 0429S1                         |                       |
| 510(k) - Summary                                                                                                                                                        | Section 1-0001-Rev D                       |                       |
| Printout Options                                                                                                                                                        | Legally Marketed Device Holter for Window  | New Device CardioDay  |
| Event Histogram                                                                                                                                                         | Yes                                        | Yes                   |
| Heart Rate and ST Diagrams                                                                                                                                              | Yes                                        | Yes                   |
| HR diagram + Min/Max per Minute                                                                                                                                         | No                                         | Yes                   |
| RR Intervals                                                                                                                                                            | Yes                                        | Yes                   |
| RR Delay                                                                                                                                                                | No                                         | Yes                   |
| RR Histograms                                                                                                                                                           | No                                         | Yes                   |
| RR Interval Spectra                                                                                                                                                     | No                                         | Yes                   |
| Pacemaker Event Histogram                                                                                                                                               | Yes                                        | Yes                   |
| Pacemaker Function Analysis                                                                                                                                             | Yes                                        | Yes                   |
| Report                                                                                                                                                                  | Yes                                        | Yes                   |
| Print to File (PDF)                                                                                                                                                     | Yes                                        | Yes                   |
| Save as Default Option                                                                                                                                                  | Yes                                        | Yes                   |
| Editing & Reviewing Options                                                                                                                                             | Legally Marketed Device to Windows         | New Device CardioDay  |
| The following comparison between the legally mar- keted device and the NEW DEVICE is based on their functions only. The label and appearance of those options may vary. |                                            |                       |
| Scroll through all Beats in a Morphology Bin                                                                                                                            | Yes                                        | Yes                   |
| Edit all Beat Labels in a Morphology Bin                                                                                                                                | No                                         | Yes                   |
| Scroll through ECG and Edit Single Beat Labels                                                                                                                          | No                                         | Yes                   |
| Scroll through Events of the Same Type                                                                                                                                  | Yes                                        | Yes                   |
| Edit Event Marker                                                                                                                                                       | Yes                                        | Yes                   |
| View Patient Event Markers                                                                                                                                              | Yes                                        | Yes                   |
| Jump from any Statistics Diagram to the corresponding ECG                                                                                                               | No                                         | Yes                   |
| Jump from ECG Overview to the corresponding ECG                                                                                                                         | Yes                                        | Yes                   |
| Select Tme Interval for Time Domain RR Parameters                                                                                                                       | Yes                                        | Yes                   |
| Edit Report                                                                                                                                                             | Yes                                        | Yes                   |
| Miscellaneous                                                                                                                                                           | Legally Marketed Device Holter for Windows | New Device Cardiology |
| RZ153+ Digital Recorder Supported (K022540)                                                                                                                             | Yes                                        | Yes                   |
| CardioMem® CM 3000 (SMA) Supported                                                                                                                                      | Yes                                        | Yes                   |
| CardioMem® CM 3000-12 Supported                                                                                                                                         | No                                         | Yes                   |
| CD Installation Medium                                                                                                                                                  | Yes                                        | Yes                   |
| RZ151 Analog Recorder Supported                                                                                                                                         | No                                         | No                    |
| CD Installation Medium                                                                                                                                                  | Yes                                        | Yes                   |

# Standards Comparison

| Standard                                             | Device Master Record<br>Device History Record | CardioDay |
|------------------------------------------------------|-----------------------------------------------|-----------|
| 21 CFR 820 (FDA cGMP Good Manufacturing<br>Practice) | Yes                                           | Yes       |

042951-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

1/5/

4/

{9}------------------------------------------------

| 510(k) Notification                                                                                                                      | o getemed                                  |                          |  |
|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|--------------------------|--|
| CardioDay®                                                                                                                               |                                            | Project ID: 0429S1       |  |
| 510(k) - Summary                                                                                                                         | Section 1-0001-Rev D                       |                          |  |
| Standard                                                                                                                                 | Comply Marked Device<br>Holter for Windows | Non-Device (C-Code Only) |  |
| ISO 9001:2000 / EN 46001 / ISO 13485:2003<br>Quality Management Systems                                                                  | Yes                                        | Yes                      |  |
| IEC 60601-1-4 + A1 Programmable Electrical<br>Medical Systems                                                                            | Yes                                        | Yes                      |  |
| ANSI/AAMI EC38 Ambulatory Electrocardiographs                                                                                            | Yes                                        | Yes                      |  |
| IEC 60601-2-47 Particular Requirements for the<br>Safety, including Essential Performance, of<br>Ambulatory Electrocardiographic Systems | Yes                                        | Yes                      |  |
| ISO 14971 +A1 Application of the Risk<br>Management to Medical Devices                                                                   | Yes                                        | Yes                      |  |
| EN 980 Graphic Symbols for the Marking of<br>Medical Devices                                                                             | Yes                                        | Yes                      |  |
| EN 1041 Supply of Information by the Manufacturer<br>of a Medical Device                                                                 | Yes                                        | Yes                      |  |
| ISO 15223 + AMD1 + AMD2 Symbols to be Used<br>with Medical Devices                                                                       | Yes                                        | Yes                      |  |

# Differences between the Legally Marketed Device (Holter for Windows®) and the New Device (CardioDay®):

# Where Used: Hospital, Home, Ambulance, etc.:

CardioDay® clearly excludes home use applications. This has no influence on safety and effectiveness.

## Performance

CardioDay and Holter for Windows® have the following different technology specifications:

| Specifications            | Legally Marketed Device<br>Hot for Windows | New Device<br>OrtotoBay                                                                                             |
|---------------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| CPU                       | Pentium II Processor                       | Pentium III, 500MHz or greater                                                                                      |
| RAM                       | 32 Mbytes minimum                          | 128 Mbytes minimum,<br>256 Mbytes minimum for XP                                                                    |
| Free hard disk space      | 200 MB minimum                             | 5 GB minimum<br>20 GB for 12-lead recordings                                                                        |
| Display                   | VGA, 14", 1024 x 768 pixel,<br>256 colors  | 17" CRT or 15" TFT, 1024 x 768 pixel<br>(XGA), 256 colors                                                           |
| Disc drive / floppy drive | 1.4 Mbytes                                 | Not required                                                                                                        |
| CD-ROM drive              | For installation only                      | For installation only.<br>Needs to be replaced by a CD-ROM<br>writer or DVD writer if archive option<br>implemented |
| Operating system          | Windows 98                                 | Windows 98 SE, NT (SP6a), 2000 or XP                                                                                |

0429S1-Sec01-0001-RevD-CardioDay-Executive_Summary.doc
CONFIDENTIAL AND PROPRIETARY

Page 10 of 11

{10}------------------------------------------------

K051471/SI
\$\frac{11}{11}\$

|       | 510(k) Notification    | getemed                                                                                                                                   |
|-------|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
|       | CardioDay®             | Project ID: 0429S1                                                                                                                        |
|       | 510(k) - Summary       | Section 1-0001-Rev D                                                                                                                      |
| Ports | One free parallel port | 1 parallel port sufficient; if USB devices<br>used for printing, software key or card<br>reader, then corresponding USB ports<br>required |

The minimum hardware requirements, such as CPU, RAM, free Hard disk space, display, CD-ROM drive and ports and supported operating system of CardioDay® correspond to the current state of the art. No influence on safety and effectiveness is expected.

# Software characteristics

The new device CardioDay offers a number of editing and printing options not available in the legally marketed device. No influence on safety and effectiveness is expected.

# Miscellaneous

Holter for Windows® does not support the Holter recorder CardioMem® CM 3000-12. No influence on safety and effectiveness is expected.

CardioDay® does not support the RZ 151 analog tape recorder. This technology is no longer state of the art and therefore it is not supported. No influence on safety and effectiveness is expected.

# Conclusion

CardioDay and Holter for Windows® software packages are both used in clinical applications to allow trained physicians or other health care providers to download, review and print electrocardiographic (ECG) data recorded on digital Holter recorders. Both are computerized programs which run under a computer operating system. Both perform a high-speed analysis of the recorded ECG data and use the computer operating system to access the displayed data. CardioDay® has the same analysis modes as Holter for Windows®. The impact of all parameters, especially those specified above, is evaluated with the help of the data bases from American Heart Association (AHA) and Massachusetts Institute of Technology (MIT) and performance evaluations.

# WARNING:…

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK/K051471](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQK/K051471)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
