K043564 · Mennen Medical , Ltd. · DQK · Jan 27, 2005 · Cardiovascular
Device Facts
Record ID
K043564
Device Name
HORIZON ANGIO CATHLAB
Applicant
Mennen Medical , Ltd.
Product Code
DQK · Cardiovascular
Decision Date
Jan 27, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The Horizon Angio is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature. Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
Device Story
Horizon Angio is a computerized laboratory system for hemodynamic monitoring during vascular catheterization. Inputs include multi-lead ECG, invasive/non-invasive blood pressure, pulse oximetry (SpO2), respiration, cardiac output, and body temperature sensors. The system uses a Sun Blade 150 workstation running a UNIX-based Cathlab program to digitize, process, and display real-time vital signs and waveforms. It provides graphical presentations of abdominal, cranial, and peripheral vascular systems to support catheterization reporting. Output includes real-time waveform/numeric displays on local LCDs and printed textual/graphical case history reports via laser printer. Used in clinical settings by healthcare professionals to document procedures and monitor patient status. Benefits include continuous hemodynamic oversight and automated case documentation.
Clinical Evidence
Bench testing only. Performance testing verified functional requirements and specifications. Electrical safety and EMC testing conducted by an independent laboratory to ensure compliance with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, and IEC 60601-2-49 standards.
Indicated for patients requiring continuous monitoring of vital signs including ECG, heart rate, invasive/non-invasive blood pressure, pulse oximetry, respiration, cardiac output, and body temperature in a clinical laboratory setting.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings and head.
Public Health Service
FEB 2 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mennen Medical Ltd. c/o Ms. Micha Oestereich Regulatory Affairs 4 Ha-Yarden Street, Yavne P.O. Box 102, Rehovot 76100 ISRAEL
Re: K043564
Trade Name: Horizon Angio Cathlab Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: II (two) Product Code: DQK Dated: December 20, 2005 Received: December 27, 2005
Dear Ms. Oestereich:
This letter corrects our substantially equivalent letter of January 27, 2005 regarding the trade name.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Micha Oestereich
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0293. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Brimmimer for
Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# 043564
Food and Drug Administration Device Modification - Horizon 9000WS Cathlab: Special 510(k) for Horizon Angio
#### INDICATIONS FOR USE
The Horizon Angio is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ・
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of
Blanuma
Slan-Om (Divis Cardlovascylar Devic 510(k) Number
{3}------------------------------------------------
Food and Drug Administration Device Modification - Horizon 9000WS Cathlab: Special 510(k) for Horizon Angio
K043564
p 1/8
JAN 2 7 2005
Image /page/3/Picture/3 description: The image shows the logo for Mennen Medical. The logo consists of a lowercase letter 'm' inside of a black circle. The words 'mennen medical' are written in lowercase letters to the right of the circle.
Partners in Patient Care
Date: December 14, 2004
# Topic: 510(k) Safety and Effectiveness Summary as per 21 CFR Section 807.92(c) Special 510(k): Device Modification - Horizon Angio Cathlab
#### Establishment Name, Registration Number and Address:
| Name: | Mennen Medical Ltd. |
|----------------------|-----------------------------------------|
| Registration Number: | 9611022 |
| Operator Number: | 9011766 |
| Address: | 4 Hayarden Street, Yavne, 81228, Israel |
| Postal Address: | PO Box 102,<br>Rehovot, 76100, Israel |
| Tel: | +972-8-9323333 |
| Fax: | +972-8-9328510 |
To: Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville MD, 20850
Attn .: Document Control Clerk From: Micha Oestereich, Regulatory Affairs
#### Product Name:
Horizon Angio Proprietary: Horizon Angio Common: Mennen Medical Part Number:
960-100-410 (115V) 960-100-420 (230V)
#### FDA Classification of Cathlab:
| Classification Name: | Programmable diagnostic computer |
|------------------------|----------------------------------|
| Classification Number: | 21 CFR 870.1425 |
| Classification: | Class II |
| Product Code: | DXG |
{4}------------------------------------------------
Food and Drug Administration Device Modification - Horizon 9000WS Cathlab: Special 510(k) for Horizon Angio
K043564
P 2/8
#### None promulgated Performance Standards:
# Voluntary Standards:
IEC 60601-1: General Requirement for Safety for Medical Electrical Systems - part 1, (1988); Amendment 1 - 1991-11 Amendment 2 - 1995-03
# IEC 60601-1-1 (2000)
Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems
# IEC 60601-1-2 (2001):
Medical electrical equipment. General requirements for Safety. Electromagnetic Compatibility Requirements and Tests.
# IEC 60602-2-27 (1994):
Medical electrical equipment, Part 2, Requirements for safety of electrocardiograph monitoring equipment.
# IEC 60601-2-30 (1995):
Medical electrical equipment, Part 2 - requirements for safety of automatic cycling indirect blood pressure monitoring equipment
# IEC 60601-2-34 (1994):
Medical electrical equipment, Part 2 - Particular requirements for the safety of direct blood pressure monitoring equipment
# IEC 60601-2-49 (2001):
Particular Requirements for the safety of multifunction patient monitoring equipment
# Terminology:
Horizon 9000WS Cathlab = The predicate device The Horizon 9000WS Cathlab was approved for marketing by the FDA (K940415 and K991775 and K032997)
Horizon Angio Cathlab = Subject of this Special 510(k). The Horizon Angio is a modified device, a sub-system of the Horizon 9000WS Cathlab
# Predicate Device:
- · Horizon 9000WS Cathlab (K032997)
{5}------------------------------------------------
The Horizon Angio is capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output, and body temperature. Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, analyzed and presented in a variety of report formats.
The Horizon Angio software includes graphic presentation of the abdominal, cranial and peripheral vascular system, to support reporting of vascular catheterization.
The Sun Blade 150 workstation is a uni-processor system that runs the Cathlab program on a UNIX operating system. The workstation receives the digitized signals from the CFE via the Ethernet hub, displays real-time vital signs, analyzes, processes, and calculates the vital sign data and waveforms, cardiac status in real time during the catheterization process, creating a fully documented case history. The workstation continuously displays the vital signs waveforms and data on the local LCD displays.
The hemodynamic data, waveform and numerical, can be stored, analyzed and presented in a variety of report formats.
A Laser Printer is provided in the system. This provides printouts of textual and graphical summaries of all patient data and catheterization procedures.
#### Base Configuration: Cathlab parameters
- . 2 Invasive Blood Pressure channels
- . Diagnostic 7 or 12 Lead ECG
- . Non-invasive Blood Pressure
- Pulse Oximetry (SpO2) .
#### Horizon Angio Options:
- Full Disclosure .
- Off-line workstations ●
- Remote Interactive terminal .
- . Angiography Analysis Package
- CDR, DVD or Optomagnetic drive .
- Choice of Console Table regular, enhanced, compact or without consol .
04:3564 > 3/8
{6}------------------------------------------------
# Intended Use of the Horizon Angio
The Horizon Angio is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuous/v presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
# Summary of the technological characteristics of the Horizon Angio
The following tables summarize data on the Horizon Angio (modified device) of the Horizon 9000WS Cathlab:
| | Horizon Angio |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Part/Option Number | 960-100-410 (115V) |
| Dimensions (H x W x D) | 24 x 22 x 10 cm (9 x 11 x 4") |
| Input Circuit Parameters | |
| Chassis Leakage Current | All patient signal inputs fully isolated (<50 μA). |
| | Meets or exceeds ANSI standard: "Safe Current Limits for Electromedical<br>Aparatus," (SCLE) Dec, 1978 item 2.1.1. |
| ECG | 7 or 12 leads |
| Frequency Response | Monitor Mode: 0.5 to 40 Hz |
| | Diagnostic: 0.05 to 150 Hz, |
| | Exercise: 1 to 25 Hz, -3 dB |
| Input Impedance: | Typical 20 MΩ |
| | Minimum greater than: |
| | 5 MΩ differential, DC to 10 Hz; 2.5 MΩ differential 10 to 100 Hz. 3 MΩ<br>differential at 10 Hz |
| Common Mode Rejection: | At least 100 dB at 50/60 Hz Without lead misbalance<br>86 db with lead misbalance |
| | The common mode rejection ratio is in accordance with ANSI/AAMI EC11(9)<br>Para. 3.2.11. |
| Input Dynamic Range: | ±5mV p-p at a rate up to 320mV/sec, as per ANSI/AAMI EC13(8) Para.<br>3.2.9.1. |
| Input offset | ± 300mV, as per ANSI/AAMI |
| | EC13 Para. 3.2.9.1. |
| | Horizon Angio |
| ECG | |
| Gain: | Manual selection of 250, 500, 1000, 2000, 4000 and 8000 x ECG. Signal<br>impressed across selected lead |
| Noise: | Less than 30 μ V p-p referenced to input |
| Pacemaker Pulse<br>Rejection: | Reject pulses from:<br>2.0 mV to 700 mV pulses of 0.2 to 2.0 mSec pulse widths and<br>≥3.0mV for 0.1mSec pulse width |
| Defibrillator Protection: | Up to 5 KV. Amplifier |
| | Recovery time: < 3 seconds |
| Lead Fault Sense: | On any ECG electrode |
| QRS Detection: | 0.25 to 5.0 mV, 70-120 msec width |
| Synchronous Defibrillation<br>Signal: | Pulse Width: 100 ms.<br>Amplitude: 5 Vdc amplitude into 500Ω, short-circuit proof |
| ECG Analog Output: | 1 Volt / mVolt |
| Heart Rate | |
| Range: | 20 to 350 bpm |
| Accuracy: | Within 2 bpm. |
| Response Time: | Less than 7 sec for step change of 60 bpm from a base of 60 bpm |
| Blood Pressure | |
| Input Sensitivity: | 5 μvolts/volt/mmHg |
| Transducer Excitation: | 5 Volt |
| Ranges: | -50 to +300 mmHg |
| Maximum variation during<br>zero: | ± 2 mmHg |
| Zero Accuracy: | ± 0.2 mmHg |
| Zero Drift: | Less than +/- 0.2 mmHg in 24 hours |
| Transducer Load<br>Impedance: | 300 – 600Ω |
| Linearity: | Better than 1% of full scale |
| Common Mode Rejection: | 80 dB minimum (reference to chassis 50/60Hz) |
| | Horizon Angio |
| Blood Pressure | |
| Frequency Response: | DC to 12 Hz<br>(DC to 40 Hz optional) |
| | |
| Temperature | |
| Range: | 27 °C to 45°C. |
| Accuracy: | ± 0.2°C. |
| | |
| Respiration | |
| Frequency Response: | 0.13 to 2.5 Hz., 3 dB bandwidth. |
| Range: | 8 to 150 bpm. |
| Excitation: | 65 kHz |
| | |
| Pulse Oximetry (SpO2) | |
| Probe Type: | Masimo™ reusable or disposable |
| Range: | 0% to 100% |
| Pulse Rate Range: | 20-250 bpm, below 20 displays zero |
| Rate Accuracy: | ± 3 bpm |
| SpO2 Accuracy: | Determined by specific sensor: |
| | Adult: ±2 digits between 70% and 100% |
| | ±3 digits between 50% and 70%.<br>Neonatal: ±3 digits between 70% and 95% |
| | |
| Auxiliary Inputs | |
| Input Voltage: | +/-5 Volt |
| Frequency Response: | DC to 120 Hz |
| | |
| Non-Invasive Blood<br>Pressure (NIBP) | |
| Method: | Oscillometric |
| | Horizon Angio |
| Non-Invasive Blood<br>Pressure (NIBP) | |
| Initial Inflation: | 150 mmHg (adult)<br>120 mmHg (pediatric). |
| Pressure Accuracy: | Overall ± 3 mmHg, full scale. |
| Defib. Pulse Protection | 5KV as per ANSI/AAMI EC13 (9), clause 3.2.2.2 and per<br>IEC 60601-2-27 (12), clauses 17,101 and 102 |
| Degree of protection<br>against electrical shock | Type CF and BF.<br>ECG, IBP and CO = CF<br>NIBP and SpO2 = BF |
| Electrosurgical<br>Interference Suppression | Yes |
| Displayed Waveforms | |
| ECG | Up to 12 lead |
| BP | 2 separate or superimposed |
| Respiration | 1 |
| SpO2 | 1 |
| Displayed Numeric<br>Parameters | 1 |
| Heart Rate | Yes |
| Respiration Rate | Yes |
| SpO2 | Yes |
| BP - Systolic, Diastolic,<br>Mean | Yes |
| Temperature | 2 |
| | Horizon Angio |
| Alarm Indications | No |
| | |
| Display Functions | |
| Change ECG Lead<br>Selection | YES |
| Display of Arrhythmia<br>Information | YES |
| Data Review: Trends | YES |
| Data Review: Tabular | YES |
| User defined Configuration<br>Setup | YES |
| User defined Default<br>Settings | YES |
| Accessories | Compatible with Mennen Medical Envoy patient monitor |
K043564-
P4/8
{7}------------------------------------------------
# Food and Drug Administration
Device Modification - Horizon 9000WS Cathlab: Special 510(k) for Horizon Angio
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### Conclusion of technological characteristics:
We consider the Horizon Angio to be substantially equivalent to the Horizon 9000WS Cathlab. We submit that any differences between the two modules:
- fall within the scope of a Special 510(k) Device Modification and .
- do not raise any new issues of safety and effectiveness .
#### Testing
Final testing for the whole Horizon Angio system included various performance tests designed to ensure that the device meets all functional requirements and performance specifications. Electrical Safety testing and EMC testing were performed by an independent testing laboratory to ensure that the device complies to applicable industry and safety standards.
#### Indications for Use
There is no change to the Indications for Use for the Horizon Angio. Mennen intends to use the Horizon Angio for "acquiring and displaying essential patient data" according to the approved "Indications for Use" in K032997. The full "Indications for Usc" appear on page 9 below.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.