VASCULAR TOOLS 5, CONSISTING OF BRACHIAL ANALYZER 5, CAROTID ANALYZER 5, VASCULAR IMAGER 5 AND INTEGRATED VASCULAR TOOLS

K033266 · Medical Imaging Applications, LLC · DQK · Dec 16, 2004 · Cardiovascular

Device Facts

Record IDK033266
Device NameVASCULAR TOOLS 5, CONSISTING OF BRACHIAL ANALYZER 5, CAROTID ANALYZER 5, VASCULAR IMAGER 5 AND INTEGRATED VASCULAR TOOLS
ApplicantMedical Imaging Applications, LLC
Product CodeDQK · Cardiovascular
Decision DateDec 16, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Vascular Tools 5 software program has been developed to aid in quantitative analysis of longitudinal vascular ultrasound images, particularly to determine vascular diameter and intima-media thickness, as well as their changes as depicted in brachial and carotid arterial ultrasound images.

Device Story

Vascular Tools 5 is a software-based quantification system for vascular ultrasound images. It accepts longitudinal ultrasound image loops as input. The software uses computer-aided detection to identify vessel wall borders within a user-defined region of interest. It calculates vascular diameter and intima-media thickness (IMT) for individual frames or subsets. The system is installed on IBM-PC-compatible computers running Windows 2000 or XP. A human expert analyst reviews, edits, and approves the identified vessel borders. Output includes quantitative morphological data, which can be exported to spreadsheets or databases. The device is used in clinical settings to assist clinicians in objective, reproducible vascular assessment, potentially improving diagnostic accuracy and monitoring of vascular health.

Clinical Evidence

Bench testing only. The manufacturer performed internal verification and validation tests to confirm that device performance satisfies design intent and conforms to system performance specifications. No clinical trial data was provided.

Technological Characteristics

Software-based quantification system for IBM-PC-compatible hardware. Operates on Windows 2000 or XP. Utilizes computer-aided detection algorithms for vessel wall border identification in longitudinal ultrasound images. Standalone software application. No specific materials or energy sources as it is a software-only device.

Indications for Use

Indicated for quantitative analysis of longitudinal vascular ultrasound images to determine brachial and carotid artery diameter and carotid intima-media thickness (IMT). Intended for use by human expert analysts to aid in obtaining quantitative vascular morphology information.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K033266 p.1/2 Medical Imaging Applications, LLC 832 Forest Hill Dr., Coralville IA 52241 phone: (319) 358-1529, fax: (319) 688-5296 e-mail: mia-llc@mia-llc.com URL: http://www.mia-llc.com Medical Imaging Applications # DEC 1 € 2004 ## 510(k) SUMMARY Submission in accordance with the requirements of 21 CFR Part 807.87(h) | 1) Submitter | Medical Imaging Applications LLC<br>832 Forest Hill Dr.<br>Coralville IA 52241<br>phone: (319) 358-1529<br>fax: (319) 688-5296<br>e-mail: mia-llc@mia-llc.com | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact person<br>Prepared | Milan Sonka<br>October 10, 2004 | | 2) Device Name: | "Vascular Tools 5" consisting of modules: "Brachial Analyzer 5"<br>"Carotid Analyzer 5" | | Common name | Picture Archival and Communications System | <sup>3)</sup> Predicate Device: QLAB - Philips Ultrasound, Inc. 892.2050 #### 4) Description of the device: Classification name: Regulation Number: "Vascular Tools 5" quantification software system is a state-of-the-art product for computer-aided assessment of vascular diameter, intima-media thickness derived from existing ultrasound images. Based on computer-aided detection of vessel wall borders within a region of interest of a single longitudinal ultrasound image or a sequence of longitudinal ultrasound images, vascular diameter and/or IMT are reported for each ultrasound image frame or a user-defined subset of image frames. Picture Archival and Communications System, Class II "Vascular Tools 5" is a proprietary software program for vascular ultrasound image quantification designed for installation on an IBM-PC-compatible personal computer and use with Windows 2000 r XP operating platform. It provides tools for display, selection, evaluation, and storage of multiple loops of vascular ultrasound image data. Furthermore, "Vascular Tools 5" facilitate impge on namprt of complete studies in "Vascular Tools 5" format, and the export of results into general spread sheets, databases, or in ASCII formats for further statistical analysis of study results by the users, #### 5) Intended use: "Vascular Tools 5" have been developed for the objective and reproducible analysis of vascular ultrasound images. The "Vascular Tools 5" data analysis program package components are intended to: - Brachial Analyzer: Determine brachial artery diameter . - Carotid Analyzer: Determine carotid artery diameter . - Carotid Analyzer: Determine carotid artery intima-media thickness . In each of the above listed intended use items, the Vascular Tools software aids the human expert to obtain quantitative information about the vascular morphology. Same as when using standard clinical {1}------------------------------------------------ methods, the human expert analyst is responsible for inspection and - if needed - editing of the identified vessel wall borders prior to their approval. #### 6) Substantial equivalence Information: The "Vascular Tools 5" software is believed substantially equivalent to the predicate devices of Philips Medical Systems' K040227"QLAB QUANTIFICATION SOFTWARE" by using the same technological characteristics and identical intended use. Technological differences between the subject and predicate devices are minor and presented above. lt is concluded that the differences in technical characteristics between the two products are not significant in that there are no features in the subject device that raise new questions regarding safety nor that could result in a decrease of effectiveness as compared to the predicate device. ### 7) Conclusions respecting safety and performance - Level of Concern: It is the opinion of Medical Imaging Applications LLC that "Vascular Tools 5" is a safe image quantification software and potential hazards are controlled by a risk management plan for the software development process (see Appendix B), including hazard analysis (see Appendix C), verification and validation tests (see Appendix D). In Medical Imaging Applications LLC's opinion, the level of concern for the stand-alone software to view images is moderate. The use of "Vascular Tools 5" quantification software neither changes the intended use of ultrasound scanners in practice, nor does the use of software result in any new potential hazards. #### 8) Validation and test conclusions Test and validation results support the conclusion that the actual device performance satisfies the design intent. Actual device performance as tested internally conforms to the system performance specifications. Detailed assessment of "Vascular Tools 5" performance is given in Appendix D. #### 9) Release version number The software release described in this document has a release version number: Version 5.1.1, release date June 1, 2004 10) Revision Level History No revisions were recorded to date. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle or bird with three curved lines representing its wings or body. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # DEC 1 6 2004 Medical Imaging Applications, LLC c/o Mr. Milan Sonka 832 Forest Hill Dr. Coralville, IA 52241 Re: K033266 Trade Name: Vascular Tools 5 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: II (two) Product Code: DQK Dated: October 10, 2004 Received: October 13, 2004 ### Dear Mr. Sonka: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 mmerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Mr. Milan Sonka comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally proficated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Brimmer for Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use | 510(k) Number (if known): | K033266 | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | "Vascular Tools 5" consisting of modules:<br>"Brachial Analyzer 5", "Carotid Analyzer 5" | | Indications For Use: | "Vascular Tools 5" software program has been<br>developed to aid in quantitative analysis of<br>longitudinal vascular ultrasound images,<br>particularly to determine vascular diameter and<br>intima-media thickness, as well as their<br>changes as depicted in brachial and carotid<br>arterial ultrasound images. | X _ AND/OR Over-The-Counter Use ______________________________________________________________________________________________________________________________________________ Prescription Use __ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K23326 Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...