PICCOX PERIPHERALLY INSERTED PRESSURE INJECTABLE, CENTRAL VENOUS OXIMETRY CATHETER

K140129 · Icu Medical, Inc. · DQE · Jul 2, 2014 · Cardiovascular

Device Facts

Record IDK140129
Device NamePICCOX PERIPHERALLY INSERTED PRESSURE INJECTABLE, CENTRAL VENOUS OXIMETRY CATHETER
ApplicantIcu Medical, Inc.
Product CodeDQE · Cardiovascular
Decision DateJul 2, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1230
Device ClassClass 2

Indications for Use

The Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter (PICCOx), is indicated for access to the central venous system for continuous in vivo measurement of the oxyhemoglobin saturation of blood (ScvO2), intravenous therapy, blood sampling, pressure injection of contrast media, and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the PICCOx may not exceed 300 psi.

Device Story

PICCOx is a peripherally inserted, pressure-injectable, central venous oximetry catheter. It features fiber optics for continuous in vivo measurement of central venous oxygen saturation (ScvO2) via reflection spectrophotometry. The device includes single or multi-lumen configurations for IV therapy, blood sampling, and central venous pressure monitoring. It supports pressure injection of contrast media up to 300 psi. Used in clinical settings, the catheter is inserted by a clinician using the Seldinger technique; position is verified via fluoroscopy or X-ray using radiopaque markers. The fiber optic lumen connects to an external Optical Module for data processing. The output (ScvO2) is displayed on a monitoring computer, assisting clinicians in assessing patient oxygenation status and guiding hemodynamic management. The device benefits patients by providing real-time, continuous monitoring of venous oxygen saturation while maintaining standard central venous access capabilities.

Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993-1), dynamic response, fiber optic performance, flow rates, leakage, force at break, radiopacity, and needlefree connector compatibility. Results met pre-determined acceptance criteria.

Technological Characteristics

Materials: Barium sulfate-filled polyurethane, ABS, PCTG. Principle: Reflection spectrophotometry for ScvO2. Energy: Optical light transmission. Form factor: 5 Fr, 40-55 cm lengths, single/multi-lumen. Connectivity: Fiber optic connection to external Optical Module. Sterilization: Gamma irradiation.

Indications for Use

Indicated for patients requiring central venous access for continuous ScvO2 monitoring, IV therapy, blood sampling, pressure injection of contrast media, and central venous pressure monitoring.

Regulatory Classification

Identification

A fiberoptic oximeter catheter is a device used to estimate the oxygen saturation of the blood. It consists of two fiberoptic bundles that conduct light at a desired wavelength through blood and detect the reflected and scattered light at the distal end of the catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUL 0 2 2014 # Section 5 510(k) Summary human connections icur The following information is being submitted in accordance with the requirements of 21 CFR 807.92. | Company Name: | ICU Medical, Inc | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 4455 Atherton Drive | | Address: | Salt Lake City, UT 84123 | | Contact Person: | Kristin Casper | | | Regulatory Affairs Specialist | | | ICU Medical, Inc. | | | Phone: (801)264-1313 | | | Fax: (801)264 1755 | | Preparation Date: | January 15, 2013 | | Device (Trade Name): | PICCOx Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter | | Common/Usual Name: | Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter | | Classification Names: | 21 CFR 870.1230. Fiberoptic Oximeter Catheter. Product Code: DQE | | Predicate Device: | K061585 ICU Medical, Inc. Opticath Central Venous Oximetry Catheter (also known as TriOx ScvO₂ Central Venous Catheter} | | | K113277 Teleflex ArrowADVANTAGE™ Pressure Injectable PICC | | Device Description: | The Peripherally Inserted, Pressure Injectable Central Venous Oximetry Catheter (PICCOx) is a single or multi-lumen catheter capable of intravenous infusion including pressure infusion up to 300psig, central venous pressure monitoring, venous blood sampling, and allows for continuous measurement of central venous oxygen saturation (ScvO2) in the superior vena cava. The device includes fiber optics in a dedicated lumen for light transmission and single or multiple fluid lumens. Constructed of barium sulfate filled polyurethane, these catheters are radiopaque enabling the use of fluoroscopy or X-ray, to guide insertion and verify position. The maximum recommended infusion rate for the pressure injectable lumen is 5 mL/sec. | | | Distance markings on the catheters provide a visual indication of insertion depth. Every 5cm the insertion depth is printed with an actual number and in between the numbers are black dots which are printed every 1cm. The catheter is non-trimmable, and offered in usable lengths of 40cm, 45cm, 50cm, and 55cm. | | | Individual lumen hubs are ISO standard female luers which are adaptable to NeedleFree Connectors, syringes or IV tubing compliant with the ISO standard. The fiber optic lumen which is terminated in a housing, can be plugged into a compatible Optical Module. | | | PICCOx is intended to be used with a user defined convenience kit. These "kits" are cleared under 510(k) K052865. All kits are terminally sterilized. The contents of the kit are sterile, medical devices intended for single patient use only with fluid path and invasive surfaces non-pyrogenic (or as indicated on the labeling of internal component packaging.) | | Intended Use: | The Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter (PICCOx), is indicated for access to the central venous system for | 949.366.2183 л 800.824.7890 A > www.icumed.com {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for ICU Medical. The word "icu" is in bold, black font, while "medical" is in a thinner, outlined font. Below the word "medical" are the words "human connections" in a smaller font. continuous in vivo measurement of the oxyhemoglobin saturation of blood (ScvO2), intravenous therapy, blood sampling, pressure injection of contrast media, and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the PICCOx may not exceed 300 psi. - Technology: The PICCOx Peripherally-Inserted, Pressure-Injectable, Central Venous Oximetry Catheter employs the same fundamental scientific technology as its predicate devices. Technological characteristics of the subject PICCOx Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter are equivalent to those of the predicate ICU Medical, Inc. Opticath Central Venous Oximetry Catheter with respect to the basic catheter design and functionality including continuous in vivo measurement of the oxyhemoglobin saturation of blood (ScvO2), intravenous therapy, blood sampling, and central venous pressure monitoring. Technical characteristics based on the subject device's peripheral insertion and power injection capabilities are comparable to the predicate Teleflex ArrowADVANTAGE™ Pressure Injectable PICC. Differences do not raise any new questions regarding safety and effectiveness. Please see Table 1 for more information. | Feature | PICCOX | Opticath® Central<br>Venous Oximetry<br>Catheter | ArrowADVANTAGE™ | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510K NUMBER | TBD | K061585 | K113277 | | GENERAL CHARACTERISTICS | | | | | Product Code | DQE | Same | LIS | | Sterility | Supplied within a sterile<br>package with non-pyrogenic<br>fluid path. | Same | Same | | Sterilization<br>Method | Catheter is gamma irradiated. | Same | EO (Arrow Pressure Injectable PICC<br>Promotional Materials) | | Materials of Composition | | | | | Shaft | Yellow polyurethane compound | Same | Polyurethane (K113277) | | Junction | Yellow polyurethane resin | Same | Not known | | Fiber Optics | Black coated fibers | Same | N/A | | Optical | Polyurethane | Same | N/A | | Coupler | Black ABS | Same | N/A | | Lead Tubing | Polyurethane | Same | Not known | | Feature | PICCOX | Opticath® Central<br>Venous Oximetry<br>Catheter | ArrowADVANTAGETM | | Luer | PCTG | Same | Not known | | Clamps | ABS | Same | Not known | | Technology | | | | | Operating<br>Principles | Multilumen catheters are<br>capable of continuous<br>measurement of central venous<br>oxygen saturation (ScvO₂) in<br>the superior vena cava. The<br>catheter includes fiber optics<br>for light transmission, multiple<br>lumens with access hubs and<br>shutoff clamps. Catheters are<br>radiopaque enabling the use of<br>fluoroscopy or X-ray to guide<br>insertion and verify position.<br><br>The distal end of the catheter is<br>constructed of polyurethane<br>that is radiopaque providing<br>greater deflection when coming<br>into contact with vessels,<br>minimizing trauma to the<br>vessels during use. | Same | The Arrow® Pressure Injectable<br>PICC is a peripherally inserted<br>central venous catheter (PICC)<br>manufactured with medical grade,<br>flexible polyurethane. The Arrow®<br>PICC has a non-tapered catheter<br>body with either a blunt tip or a<br>Blue FlexTip® that is softer than a<br>cut tip with a contour design to<br>enhance maneuverability. The Blue<br>FlexTip® also provides visual<br>confirmation of an intact catheter<br>upon removal. The kit components<br>assist the clinician in maintaining<br>maximal sterile barrier<br>precautions. (Arrow Pressure<br>Injectable PICC Promotional<br>Materials) | | Principle of<br>Operation | Reflection Spectrophotometry | Same | N/A | | Pressure<br>Injection | Pressure injection of contrast<br>media may not exceed 300 psi | N/A | Same (K113277) | | Number of<br>Pressure<br>Injection | Ten | N/A | Same (Arrow Pressure Injectable<br>PICC Promotional Materials) | | Pressure<br>Injectable<br>Lumen Flow | Maximum flow rate of 5<br>mL/sec. | N/A | Same (K113277) | | Manufacturing<br>Process | Trained personnel using<br>documented and validated<br>procedures within a Quality | Same | Not known | Table 1: Summary Comparison of Technological Characteristics Table {2}------------------------------------------------ icur human connections > > 800.824.7890 ハ {3}------------------------------------------------ . Image /page/3/Picture/1 description: The image shows the logo for ICU Medical. The letters "icu" are in bold black font, while "medical" is in a lighter, non-bold font. Below the company name, there is a tagline that is partially visible. human connections | Feature | PICCOX | Opticath® Central Venous Oximetry Catheter | ArrowADVANTAGE™ | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------| | Wavelengths | Three Wavelengths | Same | N/A | | Calibration | In vitro using included calibration block or | Same | N/A | | Preferred Insertion | Seldinger Technique | Same | Same (Arrow Pressure Injectable PICC Promotional Materials) | | Central Venous System Access | Peripheral Access | Not indicated for peripheral placement. | Same (K113277) | | Optical Module | 2 Optic Fiber Mechanical Coupler | Same | N/A | | Size | 5 Fr<br>40-55 cm<br>Single and multilumen | 8 Fr<br>20 cm<br>Multilumen | 4, 5, & 6 Fr<br>40-55 cm<br>Single and multilumen (K113277) | | Needlefree Connector | MicroClave | Same | Same (Arrow Pressure Injectable PICC Promotional Materials) | | Packaging | Tray and Pouch | Same | Not known | | Distinct Labeling | Pressure injection capabilities are marked on the extension line offering a clear | N/A | Same | | Connection to External Equipment | -Connected to the monitoring computer by inserting the optical connector into the optical module when measuring central venous oxygen saturation.<br>-Administration sets are connected to the PICCOx catheter when delivering I.V. therapy to the patient.<br>-Syringes are connected when obtaining blood samples or delivering drugs.<br>-A manometer or an electronic pressure transducer with any | Same EXCEPT for pressure injector equipment. | Same EXCEPT for connected to the monitoring computer and optical module when measuring central venous oxygen saturation. | » 800.824.7890 A {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for ICU Medical. The word "icu" is in bold black font, while "medical" is in a thin, light font. Below the company name is the tagline "human connections" in a smaller font. . | Feature | PICCOX | Opticath® Central<br>Venous Oximetry<br>Catheter | ArrowADVANTAGE™ | |------------|------------|--------------------------------------------------|-------------------------| | Duration | Short-term | Same | Short-term or long-term | | Single Use | Yes | Same | Same | 951 Calle Amanecer, San Clemente, CA 92673 ﺮ 949.366.2183 > 800.824.7890 ・ - A . , . · {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for ICU Medical. The word "icu" is in bold black font, while "medical" is in a thinner, lighter font. Below the logo is the tagline "human connections" in a smaller font. Determination of Substantial Non Clinical Performance Data Equivalence: Summary The demonstration of substantial equivalence is based on a comparison of features to the predicate devices and an assessment of non-clinical performance data. Information is included with this 510(k) submission that supports this determination. The PICCOx, Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter has the same indications for use, principles of operation and technological characteristics making it substantially equivalent to the predicate devices. No new issues of safety and efficacy have been raised. No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. The PICCOx, Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter has been tested in accordance with its performance specification which accommodates known functional requirements. The PICCOx has been tested and passed the required testing for biocompatibility for external communicating devices, circulating blood, with prolonged exposure of > 24 hours and < 30 days and the Clear MicroClave has been tested and passed the required testing for biocompatibility for external communicating devices, blood path, indirect, prolonged contact > 24 hours and < 30 days according to the requirements of ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing. The successful results of the key tests involving dynamic response, fiber optics, flow rates, leakage, force at break, lead tubing, markings, priming volumes, radiopacity, needlefree connector compatibility, and catheter to introducer fitment demonstrate that the proposed PICCOx Peripherally Inserted Central Venous Oximetry Catheter has met the pre-determined acceptance criteria applicable to the safe use of the devices. Conclusion: Performance testing included within this 510(k) demonstrates that the Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter (PICCOx) is safe, effective and performs in an equivalent manner to the predicate devices and in accordance with its intended use. 951 Calle Amanecer, San Clemente, CA 92673 949.366.2183 800.824.7890 www.icumed.com Page 6 of 6 {6}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Image /page/6/Picture/13 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines extending from its body, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. July 2, 2014 Icu Medical, Inc. Kristin Casper Regulatory Affairs Specialist 4455 Atherton Drive Salt Lake City, UT 84123 Re: K140129 > Trade/Device Name: PICCOx peripherally inserted, pressure injectable, central venous oximetry catheter Regulation Number: 21 CFR 870.1230 Regulation Name: Fiberoptic Oximeter Catheter Regulatory Class: Class II Product Code: DQE Dated: June 5, 2014 Received: June 6, 2014 Dear Ms. Casper, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device- "> {7}------------------------------------------------ # Page 2 - Ms. Kristin Casper related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address ﺎ ﻣﻌﻤﺪ ﺳ http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Image /page/7/Picture/10 description: The image shows the closing of a letter with the phrase "Sincerely yours,". Below the closing is a logo that appears to be the FDA logo. The logo is stylized with thick lines and a geometric design. The signature of the sender is written over the logo. for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K140129 ### Device Name PICCOx Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter ### Indications for Use (Describe) The Peripherally Inserted, Pressure Injectable, Central Venous Oximetry Catheter (PICCOx), is indicated for access to the central venous system for continuous in vivo measurement of the oxyhemoglobin saturation of blood (ScvO2), intravenous therapy, blood sampling, pressure injection of contrast media, and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the PICCOx may not exceed 300 psi. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) e: 2014.07.02 14:45:15 # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR A Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Mitchell Stein This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." SL Publishing Services (10); 417-6740 Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...