K963678 · American Telecare, Inc. · DQD · Feb 24, 1997 · Cardiovascular
Device Facts
Record ID
K963678
Device Name
CARETONE II
Applicant
American Telecare, Inc.
Product Code
DQD · Cardiovascular
Decision Date
Feb 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Indications for Use
The intended use of American Telecare, Inc.'s CareTone II is to conduct a stethoscope examination of a patient by a nurse or physician at another location using a communications line.
Device Story
CareTone II is an electronic amplifying stethoscope for remote auscultation. System consists of a sending unit (patient side) and a receiving unit (clinician side). Patient places chest piece on body; microphone captures acoustic signals. Sending unit processes signal, converts to digital format, and transmits over standard telephone lines. Receiving unit receives digital signal, converts back to audio, and outputs to headphones for clinician. Used in clinical settings or remote locations to facilitate physical exams by physicians or nurses. Enables remote diagnostic assessment of heart/lung sounds; improves access to specialist care.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing (EMC/EMI) and a physician assessment of the quality of the digital signal received compared to the predicate device.
Technological Characteristics
Electronic amplifying stethoscope; digital signal transmission; utilizes commanding circuitry and data communication cables; connects via standard telephone lines; output via headphones.
Indications for Use
Indicated for patients requiring remote auscultation examinations conducted by a nurse or physician.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
CareTone Telephonic Stethoscope
Submission Summary (Full Text)
{0}
K963678
510(k) Summary
FEB 24 1997
American Telecare, Inc.'s
## CareTone II Telephonic Stethoscope
## Submitter's Name, Address, and Telephone Number
American Telecare, Inc.
7680 Golden Triangle Drive
Eden Prairie, MN 55344-3732
Telephone: (612) 897-0000
Facsimile: (612) 944-2247
## Contact Persons
C. Richard Abbruscato
or
Gerard J. Prud'homme, Esq.
American Telecare, Inc.
7680 Golden Triangle Drive
Eden Prairie, MN 55344-2247
Telephone: (612) 897-0000
Facsimile: (612) 944-2247
Hogan & Hartson L.L.P.
555 Thirteenth Street, N.W.
Washington, D.C. 20004-1109
Telephone: (202) 637-5735
Facsimile: (202) 637-5910
(as Regulatory Counsel to American Telecare, Inc.)
## Date Prepared
September 6, 1996
## Name of Device
CareTone II Telephonic Stethoscope
## Classification Name
Stethoscope (Electronic)
## Common Name
Electronic Amplifying Stethoscope
## Product Code
DQD
\DOC - 65798/1 - 0335863.01
{1}
# Predicate Device
## CareTone Telephonic Stethoscope
### Intended Use
The intended use of American Telecare, Inc.'s CareTone II is to conduct a stethoscope examination of a patient by a nurse or physician at another location using a communications line.
### Substantial Equivalence
To use the CareTone, a patient plugs the sending unit's telephone cable into a telephone wall outlet and the examiner plugs the receiving unit's telephone cable into a telephone wall outlet. The examiner dials the patient's telephone number to connect the receiving unit and sending unit over the telephone line. When the connection is made, the patient places the stethoscope chest piece on his/her chest. The sounds captured by the chest piece are transmitted to the patient end unit of the sending unit by a microphone located within the chest piece. The signal is processed within the circuitry of the patient end unit and transmitted over the telephone line to the receiving unit and through headphones to the examiner.
American Telecare is principally modifying the CareTone to permit the transmission and receipt of a digital signal. The safety and effectiveness of the modified device, the CareTone II, has been demonstrated by EMC/EMI testing and a physician assessment of the quality of the signal received. The CareTone II has the same intended use as the CareTone: to conduct a stethoscope examination of a patient by a nurse or physician at another location using a communications line. It also has very similar principles of operation and technological characteristics as they both transmit an auscultatory signal to a receiver. The minor technological differences between the CareTone II and the CareTone, primarily, use of commanding circuitry and the use of data communication cables, do not raise new questions of safety or effectiveness. Thus, the CareTone II is substantially equivalent to the CareTone.
\ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \
Page 2
\ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \