K233313 · Lapsi Health Holding B.V. · DQD · Apr 10, 2024 · Cardiovascular
Device Facts
Record ID
K233313
Device Name
Keikku Electronic stethoscope
Applicant
Lapsi Health Holding B.V.
Product Code
DQD · Cardiovascular
Decision Date
Apr 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Keikku is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation data of the patient (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on a network can listen to the auscultation data of a patient on site or at a different location on the network. The Keikku is intended for use on pediatric and adult patients. The Keikku is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.
Device Story
Keikku is a digital stethoscope for auscultation of heart, lungs, bowel, arteries, and veins. System comprises a hardware stethoscope and mobile application. Hardware captures body sounds via MEMS microphone, converts to digital data, and transmits via Bluetooth to mobile device. App provides visualization, annotation, secure storage, playback, and sharing. Used by healthcare professionals in clinical settings or lay users in non-clinical environments under provider supervision. Enables remote monitoring, second opinions, and telemedicine. Healthcare providers review audio/visual output to assist in clinical assessment. Benefits include enhanced sound amplification (up to 42dB) and remote access to auscultation data.
Clinical Evidence
Bench testing only. Performance testing compared Keikku to Eko CORE and 3M Littmann 3200, verifying audio frequency and noise-to-signal ratio response met predefined acceptance criteria. Biocompatibility evaluated per ISO 10993-1. Electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and usability (IEC 60601-1-6) testing passed. Software verification and validation conducted per FDA guidance.
Technological Characteristics
Electronic stethoscope using MEMS microphone sensor. Digital signal processing with up to 42dB amplification. Frequency response 50-800Hz. Connectivity via Bluetooth 5 LE to iOS/Android mobile app. Rechargeable battery power. IP44 rating. Diaphragm material: silicon rubber. Software-based visualization and annotation.
Indications for Use
Indicated for pediatric and adult patients for amplification, filtering, and transmission of auscultation data (heart, lungs, bowel, arteries, veins). Used by clinicians in clinical settings or lay users in non-clinical settings under provider supervision. Not for self-diagnosis.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
3M LITTMANN ELECTRONIC STETHOSCOPE, MODEL 3200 (K083903)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 10, 2024
Lapsi Health Holding B.V % Orly Maor Consultant 25 Sirkin Street Kfar Saba, 4442157 Israel
Re: K233313
Trade/Device Name: Keikku Electronic stethoscope Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: September 29, 2023 Received: September 29, 2023
Dear Orly Maor:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices
{2}------------------------------------------------
Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
## K233313
Device Name
Keikku Electronic Stethoscope Indications for Use (Describe)
The Keikku is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation data of the patient (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on a network can listen to the auscultation data of a patient on site or at a different location on the network. The Keikku is intended for use on pediatric and adult patients. The Keikku is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human
Services Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA)
Staff PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1 FDA
PSC Publishing Services (301) 443-6740 ER
{4}------------------------------------------------
# Traditional Premarket Notification Submission - 510(k) Keikku Electronic Stethoscope 510(k) Number K233313
## Date Prepared: April 5, 2024
### I. SUBMITTER
Lapsi Health Holding B.V. Laarderhoogtweg 25, Amsterdam The Netherlands www.lapsihealth.com
## Regulatory Correspondent:
Katja Koskinen, Regulatory Officer Lapsi Health Holding B.V. Laarderhoogtweg 25, Amsterdam -The Netherlands katja@lapsihealth.com
## DEVICE II.
Name of Device: Keikku Electronic Stethoscope Common or Usual Name: Keikku Electronic Stethoscope Classification Name: 21 CFR 870.1875; Stethoscope, Electronic Regulatory Class: II Product Code: DQD.
## III. PREDICATE DEVICE
Lapsi Health Holding B.V. believes that the Keikku Electronic Stethoscope is substantially equivalent to the following predicate device:
- Eko CORE manufactured by Eko Devices, Inc. cleared under K200776, Classification name Electronic stethoscope, Product code: DQD, Regulation: 21 CFR 870.1875
The following device is used as a reference device:
- 3M LITTMANN ELECTRONIC STETHOSCOPE, MODEL 3200 manufactured by Eko ● Devices, Inc. cleared under K083903, Classification name Electronic stethoscope, Product code: DQD, Regulation: 21 CFR 870.1875
### DEVICE DESCRIPTION IV.
The Keikku (Rx) is a digital stethoscope device designed for use by health care professionals in clinical settings and by lay users in non-clinical environments under healthcare provider supervision. The Keikku electronically amplifies, filters and transfers body sounds through the accompanying mobile application and is used for storage, sharing and transmitting the data for
{5}------------------------------------------------
telemedicine use. It also enables lay users, under supervision from a healthcare provider, to listen to their body sounds (lungs, heart, arteries, veins, gastrointestinal tract, etc.), record and share it with their physicians during telehealth sessions.
The Keikku consists of two primary components:
- 1. The Keikku device is an electronic stethoscope. The Keikku device is used for recording audio, converting it to digital data, and transmitting the data to a mobile device via Bluetooth®. It includes volume adjustment via rotation, tap feature for starting and ending the recording, and an LED light indicator for indicating the status of the device.
- 2. The Keikku App. The app captures audio data from the Keikku device and provides data visualization and annotation, secure data storage, audio playback, and sharing features. These features enable a healthcare professional to monitor patients, seek second opinions from a specialist or use the device for telemedicine use.
## INDICATIONS FOR USE V.
The Keikku is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation data of the patient (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on a network can listen to the auscultation data of a patient on site or at a different location on the network. The Keikku is intended for use on pediatric and adult patients. The Keikku is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.
#### OF TECHNOLOGICAL CHARACTERISTICS VI. COMPARISON WITH THE PREDICATE DEVICE
The Keikku Electronic Stethoscope has the same intended use as the predicate device. Its indications for use are identical to that of the predicate device.
The Keikku Electronic Stethoscope has similar technological characteristics as the predicate device as demonstrated in the table below:
| Specification/<br>Feature | Keikku electronic<br>stethoscope<br>K | Eko CORE<br>K200776 | Littmann 3200<br>K083903 | Comparison |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Regulation number and<br>Product Code | DQD , 870.1875 | DQD , 870.1875 | DQD, 870.1875 | Same |
| Intended use | Electronic<br>Stethoscope | Electronic<br>Stethoscope | Electronic<br>Stethoscope | Same |
| Indications for<br>use | The Keikku is an<br>electronic<br>stethoscope that<br>enables<br>amplification,<br>filtering, and<br>transmission of<br>auscultation data of<br>the patient (heart,<br>lungs, bowel, | The Eko CORE is<br>an electronic<br>stethoscope that<br>enables<br>amplification,<br>filtering, and<br>transmission of<br>auscultation sound<br>data (heart, lungs,<br>bowel, arteries, and | The 3MTM<br>Littmann®<br>Electronic<br>Stethoscope Model<br>3200 is intended for<br>medical diagnostic<br>purposes only. It<br>may be used for the<br>detection and<br>amplification of | Same as EKO<br>core |
| Specification/<br>Feature | Keikku electronic<br>stethoscope<br>K | Eko CORE<br>K200776 | Littmann 3200<br>K083903 | Comparison |
| | arteries, and veins),<br>whereby a clinician<br>at one location on<br>network can listen to<br>the auscultation data<br>of a patient on site<br>or at a different<br>location on the<br>network. The<br>Keikku is intended<br>for use on pediatric<br>and adult patients.<br>The Keikku is<br>intended to be used<br>by professional<br>users in a clinical<br>environment or by<br>lay users in a<br>nonclinical<br>environment. The<br>device is not<br>intended for self-<br>diagnosis. | veins), whereby a<br>clinician at one<br>location on network<br>can listen to the<br>auscultation sounds<br>of a patient on site<br>or at a different<br>location on the<br>network. Eko<br>CORE is intended<br>for use on pediatric<br>and adult patients.<br>The Eko CORE is<br>intended to be used<br>by professional<br>users in a clinical<br>environment or by<br>lay users in a<br>nonclinical<br>environment. The<br>device is not<br>intended for self-<br>diagnosis. | sounds from the<br>heart, lungs,<br>arteries, veins, and<br>other internal<br>organs with the use<br>of selective<br>frequency ranges. It<br>can be used on any<br>person undergoing<br>a physical<br>assessment | |
| Intended patient<br>population | Adults and pediatric | Adults and<br>pediatric | Adults and<br>pediatric | Same |
| Prescribed | Rx | OTC | Rx | Same as the<br>reference |
| Chest-piece | Yes | Yes | Yes | Same |
| Sound<br>Processing | Digital signal<br>processing | Digital signal<br>processing | Digital signal<br>processing | Same |
| Amplification | Increases the<br>volume to reach a<br>constant value, up to<br>42dB | Increases the<br>volume up to 40x<br>amplification | Up to 28dB<br>acoustic gain,<br>equivalent to 24<br>times amplification | Same as the<br>predicate |
| Auscultation<br>position<br>guide | Yes, using the video<br>session | Yes, using the<br>Software<br>Application | NA | Similar |
| Signal<br>Transmission | Wireless Network,<br>Bluetooth | Wireless Network,<br>Bluetooth | NA | Same |
| Display | Smartphone (iOS or<br>Android) device | Smartphone (iOS<br>or Android) | liquid crystal<br>display (LCD) | Same as<br>predicate |
| Form Factor | Wireless headsets | Similar to<br>traditional<br>stethoscope +<br>optional wireless<br>headsets | Similar to<br>traditional<br>stethoscope | Same as<br>predicate |
| Specification/<br>Feature | Keikku electronic<br>stethoscope<br>K | Eko CORE<br>K200776 | Littmann 3200<br>K083903 | Comparison |
| Environment of<br>use | Professional users in<br>a clinical<br>environment or by<br>lay users in a<br>nonclinical<br>environment. | Professional users<br>in a clinical<br>environment or by<br>lay users in a<br>nonclinical<br>environment. | Professional users<br>in a clinical<br>environment | Same as<br>predicate |
| Application | Real time | Real time | Real time | Same |
| Bluetooth | 5 LE | 4.2 LE | Class 2 Bluetooth | Same |
| Diaphragm | Silicon rubber | Epoxy/Fiberglass | Polyurethane<br>coated silicon | Different<br>however has<br>no impact on<br>performance<br>as was<br>determined by<br>testing. |
| Pickup sensor | MEMS microphone | MEMS microphone | Unknown | Same as EKO<br>core |
| Frequency<br>Response | Extended: 50-800Hz | Bell: 20 - 200Hz.<br>Diaphragm: 100 -<br>500Hz<br>Extended: 50 -<br>500Hz | Bell: 20-200Hz<br>Diaphragm: 100-<br>500Hz Extended<br>range: 50-500 Hz. | Similar |
| Power Source | Rechargeable<br>battery | Rechargeable<br>battery | AA battery | Same as EKO<br>core |
| IP | IP44 | IP22 | IPX4 (chest piece<br>only) | Similar |
| Signal input<br>method | Sound waves<br>collected via a<br>digital microphone | Toggle between<br>analog and<br>amplified listening<br>modes | Sound waves<br>collected via a<br>transducer | Similar |
| Signal storage | Keikku app (iOS<br>and Android) | EKO software (iOS<br>and Android) | via Bluetooth link<br>to a PC | Same as EKO<br>core |
{6}------------------------------------------------
{7}------------------------------------------------
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
# Biocompatibility
Biocompatibility evaluation in compliance with ISO 10993-1 was evaluated.
{8}------------------------------------------------
## -Cleaning and Disinfection Testing
Cleaning and disinfection validation were performed. All tests were successfully completed.
### Performance Testing -
Performance testing included comparison testing of the Keikku Electronic Stethoscope to its predicate and reference device. The main purpose of this test was to verify the Keikku's performance is similar to that of its predicate and reference devices, Eko Core and 3M Littmann electronic stethoscope, in terms of audio frequency and NSR response. The test passed and met the predefined acceptance criteria.
## Software Validation -
The Keiiku Electronic Stethoscope is categorized as Enhanced. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions".
## Electrical Safely and EMC -
Electrical Safety per IEC 60601-1, Electromagnetic compatibility (EMC) per IEC 60601-1-2 and usability per IEC 60601-1-6 were conducted on the Keikku Electronic Stethoscope. The tests passed.
### Usability Evaluation -
Usability study was conducted with passing results.
## VIII. CONCLUSION
The Keikku Electronic Stethoscope was determined to be substantially equivalent to the predicate and reference device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.