CARDIOVASCULAR SONOSPECTROGRAPHIC ANALYZER

K023759 · Sonomedica · DQD · May 6, 2003 · Cardiovascular

Device Facts

Record IDK023759
Device NameCARDIOVASCULAR SONOSPECTROGRAPHIC ANALYZER
ApplicantSonomedica
Product CodeDQD · Cardiovascular
Decision DateMay 6, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1875
Device ClassClass 2

Indications for Use

The Cardiovascular Sonospectrographic Analyzer (CSA) is intended for use in health care environments and setting where conventional stethoscopes (both electronically augmented and non-augmented) and phonocardiographs are used by health care professionals for the auscultation, graphical representation, measurement and archiving of cardiovascular, pulmonary, and other such sounds, both normal and abnormal, as produced by the human body. The CSA is a medical instrument intended for use by, or at the direction of, a qualified health care professional as a diagnostic aid for the clinical assessment and discrimination of acoustic patterns relating to or caused by clinically significant pathologies and medical conditions as known, defined, or specified in standard medical auscultation/phonocardiographic reference textbooks, peer review journals, and medically approved auscultation training materials. The CSA is further intended for the monitoring of interventional or prophylactic therapies that are susceptible to the level of medical efficacy provided by CSA and as is clinically indicated at the suscetion of the p physician. The CSA is also intended for use in telemedical applications where discreation of the p prysicial medical medical acoustic experts are indicated or desired. Limitations on Intended Use: The CSA is not intended for use on a singular basis for delivery of health care, not as a standalone diagnostic aid for evaluation of cardiovascular conditions and nethologies for which other more suitable diagnostic tools or medical procedures are available or indicated.

Device Story

CSA is a cardiovascular acoustic detection system; inputs physiologic acoustic signals via two sensors and one environmental noise cancellation sensor; utilizes digital signal processing and Fast Fourier Transform (FFT) to process signals; displays graphical representations of acoustic data. Used in healthcare environments by clinicians; aids in diagnosing cardiovascular/pulmonary conditions and monitoring therapies. Hardware includes medical cart, UPS, IBM-compatible computer, and I/O Tech DaqBoard 2000 series data acquisition module. Clinicians use output to assess acoustic patterns against established medical references. Benefits include enhanced sensitivity (detecting signals 1,000,000 times lower than peak heart sound energy) and automated analysis compared to manual methods.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on hardware identity, identical signal processing methodology (FFT), and software automation of previously manual analysis tasks.

Technological Characteristics

System includes medical cart, UPS, IBM-compatible PC, I/O Tech DaqBoard 2000 series, and acoustic sensors. Employs digital signal processing and FFT for spectral analysis. Connectivity via standard PC interfaces. Software developed in Visual Basic. Complies with CISPR-11, IEC-601-1-2, MIL-STD-461D, and IEC 801-2 standards.

Indications for Use

Indicated for use by healthcare professionals for auscultation, graphical representation, measurement, and archiving of cardiovascular, pulmonary, and other human body sounds to aid in clinical assessment and monitoring of therapies.

Regulatory Classification

Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K023759 SE 5.6.02 510(k) submission Cardiovascular Sonospectrographic Analyzer # APPENDIX G ## 510(K) Summary DATE: November, 2002 SonoMedica LLC Applicant: 8239 Leesburg Pike Vienna. Virginia 22182 703-448-4300 Fax: 703-734-0410 Sailor Mohler, Executive Vice President, Chief Technology Officer Contact: Phone: Fax: #### Predicate Device: Transmedica: Digital Electronic Stethoscope 510(k) 963418 910(k) 905410 CFR section: 米米米 CFR section: XXXX ## Device Description The Cardiovascular Sonospectrographic Analyzer (CSA) is a cardiovascular acoustic detection The CSA is a passive, non-invasive means of detecting, processing and displaying svstem. human physiologic acoustic data as an aide/adjunct to the clinician in diagnosing conditions. It is useable in all applications where clinicians have traditionally used the stethoscope. However, through the use of ultra-sensitive acoustic detection technology, digital signal processing and a Fast Fourier Transform (FFT), the CSA is able to detect, process and display physiologic acoustic signals from the human body that are as little as 1,000,000 times less than the peak acoustic energy level of a heart sound. #### Intended Use The Cardiovascular Sonospectrographic Analyzer (CSA) is intended for use in health care environments and setting where conventional stethoscopes (both electronically augmented and non-augmented) and phonocardiographs are used by health care professionals for the auscultation, graphical representation, measurement and archiving of cardiovascular, pulmonary, and other such sounds, both normal and abnormal, as produced by the human body. The CSA is a medical instrument intended for use by, or at the direction of, a qualified health care professional as a diagnostic aid for the clinical assessment and discrimination of acoustic patterns relating to or caused by clinically significant pathologies and medical conditions as known, defined, or specified in standard medical auscultation/phonocardiographic reference textbooks, peer review journals, and medically approved auscultation training materials. The CSA is further intended for the monitoring of interventional or prophylactic therapies that are Appendix G-I 453 {1}------------------------------------------------ ## 510(k) submission Cardiovascular Sonospectrographic Analyzer susceptible to the level of medical efficacy provided by CSA and as is clinically indicated at the suscetion of the p physician. The CSA is also intended for use in telemedical applications where discreation of the p prysicial medical medical acoustic experts are indicated or desired. Limitations on Intended Use: The CSA is not intended for use on a singular basis for delivery of health care, not as a standalone diagnostic aid for evaluation of cardiovascular conditions and nethologies for which other more suitable diagnostic tools or medical procedures are available or indicated. ## Hardware Components Medical cart with Uninterruptible Power Supply (UPS) See Figure 2.2. IBM Compatible Computer, monitor, mouse, keyboard (See Figure 2.2) Data Acquisition Module, I/O Tech DaqBoard 2000 series Conditioning Electronics amplifier(See Figures 2.3 and 2.4) Two physiological sensors (See Figure 2.5) One environmental noise cancellation sensor (See Figure 2.6) Audio reproduction Headphone (optional) #### Software Components: - Data Recording Application (the DRA) ● - Data Analysis Application (the DAA) . #### Performance Standards: ## Electromagnetic Capability | PERFORMANCE TEST | APPLICABLE STANDARD | |----------------------------------------------|----------------------------------------------------| | RF Emissions | CISPR-11/IEC-601-1-2 (modified) UL 2601 | | Magnetic Fields | RE101(mod)<br>MIL-STD-461D | | Electrostatic Discharges (Immunity) | IEC 801-2 | | Radiated EM Fields (Immunity) | IEC 601-1-2 (modified) | | AC Voltage Fluctuations Trasients and Surges | Guidelines per Para. (M) (7) and sub-para.C<br>(2) | #### Mechanical and Environmental | Performance Tests | Applicable Standard | |-----------------------------------|-------------------------------| | Controls Protection | Company design specifications | | Connector Protective | Company design specifications | | Mechanical Safety | Company design specifications | | Vibration and Shock<br>Resistance | Company design specifications | Appendix G-2 {2}------------------------------------------------ 510(k) submission Cardiovascular Sonospectrographic Analyzer | Fluid Spill Resistance | IEC-601 (44.6)<br>IEC 529 | |--------------------------|------------------------------------------------------------------------------------------------| | Temperature and Humidity | Company design specifications | | Surface Temperature | Company design specifications | | Toxic Materials | No toxic material from the device come into contact with the patient<br>or operator during use | ## Similarities and Differences to Predicate Devices The DES predicate device was developed in the 1990's based on basic technology and signal processing techniques well established in the scientific medical literature and in cleared auscultation and phonocardiography technology for the detection, processing, discrimination and display of the full range of cardiovascular sounds and murmurs. The DES development had required a design composed of the best features and properties of various phoncardiographic sensors for the purpose of detecting all of the physio-acoustic signals of interest and it also required novel digital signal processing and display software which utilized the well established and medically accepted Fast Fourier Transform technique as the basis for analyzing cardiovascular sounds and murmurs for their spectral content. However the DES was deficient as a viable medical device due to the requirement for operator training in the integral spectral analysis field and the time consuming visual analysis of the spectral signals and the manual measurement and calculations of the signals to discriminate the clinical information contained therein. SonoMedica, Inc. acquired all rights to the DES predicate device and its original USFDA clearance and commenced a product development effort to reduce the original DES into a viable medical product. This development caused some minor refinements in the data acquisition circuitry but focused primarily upon developing software that automated the time intensive process of analyzing the detected cardiovascular signals in order to present the derived data in an easily accessible and understandable graphical format for ease and rapidity of use by the clinician or other medical user. The CSA hardware is essentially the same as was used and cleared in the predicate device. The CSA software is fully equivalent to the DES software in the CSA simply automates the analysis process as was required by the DES predicate device. The CSA use the Fast Fourier Transform as was used in the DES for signal decomposition as well as all of the same basic displays. The CSA employs features in some of its displays which are improvements upon the DES and the software is much more processing efficient due to its use of improved programming techniques but it is still all based on the same Visual Basic language as was used in the DES. In fact, the clinical data acquired by the DES is 100% compatible with the CSA and data acquired by the CSA is fully backwards compatible and can be processed and analyzed using the DES software. Appendix G-3 {3}------------------------------------------------ 510(k) submission Cardiovascular Sonospectrographic Analyzer # Conclusion As the CSA hardware is essentially the same as the DES predicate device; the basic digital signal Its the CD7 marting to dolisplays being equivalent; both devices are based on the same probosing tochards and the full backwards and forward compatibility of the recorded cardiovascular physio-acoustic data, it is concluded that the Cardiovascular Sonospectrographic Analyzer (CSA) is substantially equivalent to the DES predicate device from which is derived. The CSA does not use any form of radiation, contains no physical properties or features that present a risk, use no chemicals and has no design elements that also manifest a risk, it is further concluded that it is safe and effective for its intended use. Appendix G-4 456
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...