ACUITY STEERABLE LEAD SYSTEM
P050046 · Guidant Corp. · OJX · Apr 13, 2007 · Cardiovascular
Device Facts
| Record ID | P050046 |
| Device Name | ACUITY STEERABLE LEAD SYSTEM |
| Applicant | Guidant Corp. |
| Product Code | OJX · Cardiovascular |
| Decision Date | Apr 13, 2007 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
The Guidant ACUITY™ Steerable IS-1 coronary venous, steroid-eluting, dual electrode pace/sense leads are transvenous leads intended for chronic, left-ventricular pacing and sensing via the coronary veins when used in conjunction with a compatible pulse generator. Extended bipolar pacing and sensing is available using ACUITY™ Steerable with an RV pace/sense/defibrillation lead or a bipolar RV pace/sense lead.
Device Story
The ACUITY Steerable lead is a transvenous, steroid-eluting, bipolar pace/sense lead designed for chronic left-ventricular (LV) pacing. It features an over-the-wire design with a distal J-shaped fixation mechanism. The lead is inserted through the coronary sinus into a cardiac vein branch using a stylet or guidewire, which allows for tip deflection and steering. The device incorporates two dexamethasone acetate collars to suppress local inflammatory responses. It is used in conjunction with a compatible pulse generator (CRT-P or CRT-D). Physicians use the lead to provide cardiac resynchronization therapy, which can improve heart failure symptoms and cardiac function. The device is supplied sterile with accessories including stylets and a vein lifter.
Clinical Evidence
Prospective, multi-center, non-randomized study of 110 patients. Primary safety endpoint: 3-month lead-related complication-free rate was 91.8% (lower 95% CI: 86.2%), exceeding the 80% threshold. Primary effectiveness endpoint: 3-month mean LV pacing threshold was 1.1 ± 0.9 V (upper 95% CI: 1.3 V), meeting the <2.5 V requirement. Secondary endpoints: 3-month mean R-wave amplitude was 14.3 ± 7.4 mV (>3 mV threshold) and mean pacing impedance was 644 ± 207 Ohms (>300 Ohm threshold). Implant success rate was 91.8%.
Technological Characteristics
6 Fr diameter lead with coaxial coil design. Materials: Titanium/Titanium alloy terminal, silicone/polyurethane insulation, platinum-clad tantalum outer conductor, MP-tantalum inner conductor. Electrodes: IROX-coated platinum/iridium. Fixation: J-shaped distal body. Steroid elution: Two dexamethasone acetate/silicone collars. Connectivity: IS-1 bipolar. Sterilization: 100% Ethylene Oxide (ETO) per ANSI/AAMI 11135-1994. Standards: EN 45502-2-1, ASTM D 4169-04.
Indications for Use
Indicated for patients requiring chronic left-ventricular pacing and sensing via coronary veins. Contraindicated in patients with hypersensitivity to 1.0 mg dexamethasone acetate, mechanical tricuspid heart valves, or obstructed/inadequate vasculature for intravenous catheterization.
Regulatory Classification
Identification
Drug eluting permanent lv pacemaker electrodes are used with compatible cardiac rhythm therapy (crt) devices to deliver pacing pulses to the heart.
Predicate Devices
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
Device Generic Name: Permanent drug-eluting pacemaker electrode
Device Trade Name: ACUITY™ Steerable Lead Models 4554, 4555, and 4556
Applicant's Name and Address: Guidant Corporation
4100 Hamline Avenue North
St. Paul, MN 55112-5798
Date of Panel Recommendation: None
Pre-Market Application (PMA) Number: P050046
Date of Notice of Approval to Applicant: April 13, 2007
## II. INDICATIONS FOR USE
The Guidant ACUITY™ Steerable IS-1 coronary venous, steroid-eluting, dual electrode pace/sense leads are transvenous leads intended for chronic, left-ventricular pacing and sensing via the coronary veins when used in conjunction with a compatible pulse generator. Extended bipolar pacing and sensing is available using ACUITY™ Steerable with an RV pace/sense/defibrillation lead or a bipolar RV pace/sense lead.
## III. CONTRAINDICATIONS
Use of the ACUITY lead is contraindicated in patients with:
- patients with a hypersensitivity to a nominal dose of 1.0 mg (0.5 mg per electrode) of dexamethasone acetate drug
- mechanical tricuspid heart valves
- obstructed or inadequate vasculature for intravenous catheterization.
P050046 SSED
Page 1 of 28
{1}
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the ACUITY™ Steerable lead labeling.
# V. DEVICE DESCRIPTION
The Guidant ACUITY™ Steerable lead is a device/drug combination product made up of two regulated components: a device (the left ventricular ACUITY Steerable lead) device and a drug component (dexamethasone acetate). The characteristics for the lead appear in Table 1.
Table 1: Design of ACUITY Steerable Lead
| Characteristic | ACUITY Steerable Lead |
| --- | --- |
| Model number and Length | Model 4554, 80cm
Model 4555, 90 cm
Model 4556, 100 cm |
| Terminal Compatibility | IS-1, Bipolar |
| Terminal Pin / Ring Material | Titanium or Titanium Alloy |
| Terminal Seal Material | Silicone Rubber |
| Pulse Generator Compatibility | Guidant Pulse Generators that accept IS-1 Connectors |
| Serial Number Label | Laser-marked titanium dioxide polyurethane tubing |
| Lead Body Diameter | Polyurethane - 1.98mm (6F)
Silicone - 1.88mm (5.7F) |
| Conductor Coils | Coaxial coil design with distal two-dimensional J shaped bias.
Inner conductor: Quad-filar MP-tantalum core wire heat-treated at 275°F.
Outer Conductor: Trifilar platinum-clad tantalum wire coated with ETFE fluoropolymer. |
P050046 SSED
Page 2 of 28
{2}
P050046 SSED
Page 3 of 28
10
| Characteristic | ACUITY Steerable Lead |
| --- | --- |
| Electrical Insulation | Silicone tubing provides primary insulation between coaxial conductors and on the distal portion of the lead body. Polyurethane provides primary insulation on the proximal lead body. ETFE coating on the outer coil filars provides an additional barrier. |
| ETFE Filar Coating | Outer conductor coil filars are individually coated.
Coating terminates just proximal to the proximal electrode. |
| Lead Body Inner Insulation Material | Sil-glide coated silicone tubing provides primary insulation between coaxial conductors and on the distal portion of the lead body is 75A durometer. |
| Lead Body Outer Insulation Material (Polyurethane Lead Body) | 55D Polyurethane provides primary insulation from the environment on the proximal lead body (from terminal connector to approximately 2 cm proximal to tip of lead). |
| Lead Body Outer Insulation Material (Distal Silicone Lead Section) | Sil-glide coated silicone tubing (55A durometer) provides primary insulation from the environment on the distal lead body (from Polyurethane Lead Body Tubing to the proximal electrode).
A silicone rubber molded section (50A durometer) provides primary insulation from the environment on the distal lead body (from Proximal Electrode to the distal tip electrode). |
| Distal Silicone Section Length | 2.0 cm |
| Atraumatic Tip | Rounded Tip electrode with silicone rubber neck between electrodes.
OD of 0.074" tapers to 0.070". |
| Proximal Electrode | IROX coated platinum/iridium, 2 part (inner and outer) full circumference design;
Outer ring OD 0.074"
Length 0.060"
Calculated exposed surface area is 9.0mm². |
| Distal Electrode | IROX coated platinum/iridium, tip electrode design.
Tip OD 0.070"
Tip Length 0.050" with a hemi-spherical end
Calculated exposed surface area is 7.8 mm². |
| Stylet Stop | Stylet stop prevents it from exiting the distal tip |
{3}
| Characteristic | ACUITY Steerable Lead |
| --- | --- |
| End Ring | Not Applicable |
| Electrode Spacing | 8 mm longitudinal spacing between proximal and distal electrodes. |
| Fixation | Distal lead body J-shape: J-Radius 11.4 mm (centerline)
Length 40 mm |
| Steroid Eluting Collar | Two Dexamethasone acetate (DXA) and silicone rubber collars; one distal to the proximal electrode and one proximal to the distal electrode.
ID 0.058", OD 0.070"
Length 0.060". Nominal Dose 0.050 mg per collar |
| Suture Sleeve | A triple grooved, silicone rubber, TiO₂ pigmented, suture sleeve is assembled onto the lead body. |
## A. Device Component Description
Guidant ACUITY™ Steerable coronary venous pace/sense leads, Models 4554/4555/4556, provide chronic left ventricular bipolar pacing and sensing. Lead diameter is 6 Fr. The leads have an over-the-wire design with an IS-1 bipolar connector and are steroid-eluting at the proximal and tip electrodes. The lead is anchored with J-shaped fixation and the electrodes are IROX coated (iridium oxide). Placement is achieved by inserting the lead through the coronary sinus and placing it into a branch of the cardiac veins using a stylet or an over-the-wire delivery accessory. The stylet or guide wire can be retracted or advanced within the "J"-shape to deflect the tip and steer the lead into coronary venous branch veins. A step in the tip electrode lumen allows engagement of the stylet tip and stops it from exiting the distal tip of the lead. The tip of the lead is design to allow a guidewire to pass through for over-the-wire delivery. The ACUITY Steerable lead is used in conjunction with a compatible pulse generator.
The lead and accessories are supplied sterile. Each package contains one lead (with an anchoring sleeve already attached, a stylet/guidewire guide in-place), two stylets, and a vein lifter.
P050046 SSED
Page 4 of 28
{4}
# B. Drug Component Description
The active drug component in the ACUITY Steerable lead is dexamethasone acetate (DMA). This is an anti-inflammatory steroid that is a white crystalline solid with a melting point of 238-240°C and maximum UV of 239 nm (Merck Index, Twelfth edition 1996). The structural formula is shown in Figure 1.
The target dose of DMA on the steroid collar is 0.5 mg. The ACUITY Steerable lead contains two (2) drug components (steroid collars).
Each collar consists of DMA in a silicone rubber matrix, which is positioned adjacent and distal to each electrode. Figure 2 depicts the drug component on the lead.

Figure 1: Structural Formula of Dexamethasone Acetate
P050046 SSED
Page 5 of 28
12
{5}

Figure 2: Location of Drug Component (Steroid Collar)
## C. Dexamethasone Mechanism of Action
Steroids suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted leads. Dexamethasone is a synthetic steroid of the glucocorticoid family. Glucocorticoid steroids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. While the mechanism of action of glucocorticoids is not fully understood, it is known that glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as to a membrane-bound receptor. Binding to the cytoplasmic receptor, the receptor becomes activated and leads to translocation to the nucleus. The receptor interacts with specific DNA sequences (glucocorticoid responsive elements) within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production by multiple cells of factors that are critical in generating the inflammatory response, in particular via modulation of transcription factors.
## VI. ALTERNATIVE PRACTICES OR PROCEDURES
The present established therapies include the use of commercially available left ventricular leads.
P050046 SSED
Page 6 of 28
{6}
# VII. MARKETING HISTORY
The ACUITY Steerable lead is currently distributed commercially outside the United States. Specifically, this lead is approved for sale in the European Union.
The ACUITY Steerable lead has not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device.
# VIII. ADVERSE EFFECTS OF THE DEVICE ON HEALTH
## A. Potential Adverse Events
Based on the literature and lead implant experience, the following alphabetical list includes possible adverse events associated with implantation of an implantable cardioverter defibrillator and/or pacemaker lead system:
- Acceleration of arrhythmias
- Adverse reaction to procedure (e.g., bradycardia, general, respiratory, hypotension)
- Air embolism
- Allergic reaction
- Bleeding
- Cardiac tamponade
- Chronic nerve damage
- Conductor coil fracture
- Coronary venous spasm
- Death
- Elevated thresholds
- Erosion/extrusion
- Extracardiac stimulation (e.g., phrenic, diaphragm, chest wall)
- Fibrotic tissue formation (e.g., keloid formation)
- Fluid accumulation
- Formation of hematomas or cysts
- Heart block
- Inappropriate therapy (e.g., shocks, ATP, pacing)
- Incomplete lead connection with pulse generator
- Infection
- Lead displacement/dislodgement
- Lead fracture
- Lead insulation breakage or abrasion
- Lead tip deformation and/or breakage
- Local tissue reaction
- Muscle and nerve stimulation
P050046 SSED
Page 7 of 28
14
{7}
- Myocardial trauma (e.g., cardiac perforation, irritability, injury)
- Myopotential sensing
- Oversensing/undersensing
- Pacemaker-medicated tachycardia
- Pericardial rub, effusion
- Pneumothorax/hemothorax
- Random component failures
- Shunting current or insulating myocardium during defibrillation with internal or external paddles
- Thrombosis/thromboemboli
- Valve damage
- Venous occlusion
- Venous trauma (e.g., perforation, dissection, erosion)
In addition to the implantation of an implantable cardioverter defibrillator and/or pacemaker lead system, possible adverse events associated with implantation of a coronary venous lead system are listed below in alphabetical order:
- Allergic reaction to contract media
- Breakage/failure of implant tools
- Coronary venous occlusion
- Coronary venous trauma (e.g., perforation, dissection, erosion)
- Prolonged exposure to fluoroscopic radiation
- Renal failure from contract media used to visualize coronary veins
## B. Observed Adverse Events (AE)
The safety of the ACUITY Steerable lead was evaluated in 110 patients who underwent an implant procedure for the ACUITY Steerable lead in the ACUITY Steerable clinical study conducted in the United States. Safety and effectiveness endpoints were based on three-month follow-up data.
Investigators were required to report adverse events during the clinical study according to the definitions in the protocol. Investigators were required to provide a detailed description of the AE, with suspected cause, including information on the relation of the AE to the device, procedure or investigation, what corrective actions were taken and what the clinical outcome was for the patient.
ACUITY Steerable lead-related adverse events were defined as all lead-related or procedure-related adverse events attributed to the ACUITY Steerable lead by the investigator, or when the ACUITY Steerable lead could not be ruled out as the course of the adverse
P050046 SSED
Page 8 of 28
{8}
event. Those adverse events attributed to commercially available guide wires, guide catheters and diagnostic electrophysiology catheters were excluded from the ACUITY Steerable lead-related adverse events, and were categorized as procedure-related adverse events.
All adverse events were further classified as an observation or complication based on the following definitions:
- Observation: A clinical observation is a clinical event that does not result in invasive intervention, injury, or death, and is not an unanticipated adverse event. Corrective actions are simple adjustments such as reprogramming of the pulse generator or oral antibiotic treatment of a pocket infection.
- Complication: A clinical complication is a clinical event that results in invasive intervention after implant, injury, or death (e.g., surgical evacuation of a hematoma, lead dislodgment requiring lead repositioning, generator replacement, intravenous antibiotic treatment of a pocket infection). A complication is an event that results in an outcome classified as serious, unanticipated, or death.
Table 2 provides information on adverse events (AEs) reported from implant through the three-month follow-up visit. In the 110 patients attempted or implanted with the ACUITY Steerable lead, there were 30 lead-related events (27 patients), including 10 that were classified as complications and 20 that were classified as observations.
Table 10 summarizes all AEs reported in the study, including those that occurred beyond the three-month follow-up. Thus, it includes data beyond the 3-month endpoint time frame. This all inclusive list of AEs is located in Appendix A of this document.
P050046 SSED
Page 9 of 28
16
{9}
All patients implanted or attempted; N=110
Total device months = 313
Includes all adverse events reported through three months of follow-up
Table 2: Clinical Observation and Complications Summary
| | | Complications | | Observations | |
| --- | --- | --- | --- | --- | --- |
| Adverse Event | Number Of Events (Number of Patients) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) |
| Total Adverse Events | 103 (61) | 27.3 (30) | 11.8 (37) | 40.0 (44) | 21.1 (66) |
| ACUITY Steerable Related Events (N=101) | | | | | |
| Coronary venous dissection | 1 (1) | 0.0 (0) | 0.0 (0) | 1.0 (1) | 0.3 (1) |
| Dislodgment - Elevated threshold - LV | 2 (2) | 2.0 (2) | 0.7 (2) | 0.0 (0) | 0.0 (0) |
| Dislodgment - Extracardiac stimulation - LV | 4 (4) | 2.0 (2) | 0.7 (2) | 2.0 (2) | 0.7 (2) |
| Dislodgment - Multiple signs - LV | 2 (2) | 2.0 (2) | 0.7 (2) | 0.0 (0) | 0.0 (0) |
| Dislodgment - No reported signs - LV | 1 (1) | 1.0 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Dislodgment - Unable to capture - LV | 2 (2) | 2.0 (2) | 0.7 (2) | 0.0 (0) | 0.0 (0) |
| Elevated threshold - LV | 2 (2) | 0.0 (0) | 0.0 (0) | 2.0 (2) | 0.7 (2) |
| Extracardiac stimulation - LV | 16 (15) | 1.0 (1) | 0.3 (1) | 13.9 (14) | 5.1 (15) |
| Subtotal ACUITY Steerable Related Events | 30 (27) | 8.9 (9) | 3.4 (10) | 18.8 (19) | 6.8 (20) |
| PG Related Events (N=108) | | | | | |
| Elevated DFT - Defibrillation | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Extracardiac stimulation - Daily impedance testing | 2 (2) | 0.0 (0) | 0.0 (0) | 1.9 (2) | 0.6 (2) |
| Hematoma - Pocket (> 30 days post-implant) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Migration | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Pacemaker-mediated tachycardia (PMT) | 2 (2) | 0.0 (0) | 0.0 (0) | 1.9 (2) | 0.6 (2) |
| Psychological effect due to device therapy | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Subtotal PG Related Events | 8 (7) | 0.0 (0) | 0.0 (0) | 6.6 (7) | 2.6 (8) |
| RA Lead Related Events (N=106) | | | | | |
| Dislodgment - Unable to capture - RA | 2 (2) | 1.9 (2) | 0.6 (2) | 0.0 (0) | 0.0 (0) |
| Elevated threshold - RA | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Oversensing - RA | 2 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.6 (2) |
| Unable to capture - RA | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
P050046 SSED
Page 10 of 28
{10}
| | | Complications | | Observations | |
| --- | --- | --- | --- | --- | --- |
| Adverse Event | Number Of Events (Number of Patients) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) |
| Subtotal RA Lead Related Events | 6 (4) | 2.8 (3) | 1.0 (3) | 0.9 (1) | 1.0 (3) |
| RV Lead Related Events (N=106) | | | | | |
| Dislodgment - Elevated threshold - RV | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Elevated threshold - RV | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Subtotal RV Lead Related Events | 2 (2) | 1.9 (2) | 0.6 (2) | 0.0 (0) | 0.0 (0) |
| Procedure Related Events (N=110) | | | | | |
| Adverse reaction - Hypotension | 2 (2) | 0.9 (1) | 0.3 (1) | 0.9 (1) | 0.3 (1) |
| Chest pain | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Hematoma - Pocket (<=30 days post-implant) | 6 (6) | 0.9 (1) | 0.3 (1) | 4.5 (5) | 1.6 (5) |
| LV Lead Insulation Damaged During Procedure | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Post-surgical wound discomfort | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Psychological effect due to recall | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.6 (2) |
| Thrombus | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Subtotal Procedure Related Events | 14 (12) | 3.6 (4) | 1.3 (4) | 8.2 (9) | 3.2 (10) |
| Protocol Testing Related Events (N=110) | | | | | |
| Extracardiac stimulation - LV | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.6 (2) |
| Subtotal Protocol Testing Related Events | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.6 (2) |
| Cardiovascular Related Events (N=110) | | | | | |
| Atrial fibrillation (AF) | 4 (3) | 0.9 (1) | 0.3 (1) | 1.8 (2) | 1.0 (3) |
| Cerebrovascular accident (CVA) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Chest pain - Heart failure | 2 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.6 (2) |
| Chest pain - Ischemic | 2 (2) | 0.9 (1) | 0.3 (1) | 0.9 (1) | 0.3 (1) |
| Chest pain - Other | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Chronotropic incompetence | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Dizziness | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Dizziness - Heart failure | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Dyspnea - Heart failure | 2 (2) | 1.8 (2) | 0.6 (2) | 0.0 (0) | 0.0 (0) |
| Hypotension - Heart failure | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Multi-system failure - Heart failure | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Multiple heart failure symptoms | 9 (8) | 5.5 (6) | 1.9 (6) | 2.7 (3) | 1.0 (3) |
| Multiple symptoms | 2 (2) | 0.9 (1) | 0.3 (1) | 0.9 (1) | 0.3 (1) |
P050046 SSED
Page 11 of 28
{11}
| | | Complications | | Observations | |
| --- | --- | --- | --- | --- | --- |
| Adverse Event | Number Of Events (Number of Patients) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) |
| Myocardial infarction | 1 (1) | 0.9 (1) | 0.3 (1) | 0.0 (0) | 0.0 (0) |
| Other SVT (AVRT, AVNRT, EAT etc.) | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.6 (2) |
| Sinus tachycardia | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Ventricular fibrillation (VF) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.3 (1) |
| Subtotal Cardiovascular Related Events | 33 (26) | 10.9 (12) | 4.2 (13) | 13.6 (15) | 6.4 (20) |
| Subtotal Non-cardiovascular Related Events | 8 (7) | 3.6 (4) | 1.6 (5) | 2.7 (3) | 1.0 (3) |
As of March 21, 2006, a total of 4 deaths out of 106 implants were reported in the study. Of these deaths, 2 have been adjudicated by the Morbidity and Mortality Events Committee as Cardiac: Unknown and the second as Cardiac: Pump Failure. The deaths not yet adjudicated were reported by the principal investigator as Cardiac: Unknown and Cardiac: Pump Failure. None of these deaths have been attributed to the investigation or the ACUITY Steerable lead.
## IX. SUMMARY OF PRE-CLINICAL STUDIES
### A. Biocompatibility Studies
The materials used in the ACUITY Steerable leads that are directly exposed to body tissues or fluids are summarized in Table 3. Most of the materials are identical to the materials used on previous Guidant lead designs. Biocompatibility assessment was previously performed in accordance with ISO 10993-1, Biological Evaluation of Medical Devices: Evaluation and Testing. All materials were found to be biocompatible.
P050046 SSED
Page 12 of 28
{12}
Table 3: Biocompatibility Information
| Material (polymer/metal/ceramic/composite | Component Name Subject Device | Component Predicate Device | Predicate Device Name | Approval Date |
| --- | --- | --- | --- | --- |
| Tecothane TT 1075D-M | Molded Terminal Assembly | Molded Terminal Assembly | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Titanium-ASTM grade 5 | Terminal Pin | Terminal Pin | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Titanium-ASTM grade 5 | Terminal Ring | Terminal Ring | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Titanium-ASTM grade 5 | Terminal Outer Ring | Terminal Outer Ring | EASYTRAK 2 IS-1 | P010012/S024 06AUG-04 |
| Pellethane 2363-55D | Serial Number Label | Serial Number Label | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| TiO2 Pigment | | | | |
| Pellethane 2363-55D | Serialized Heat Bonded PU Tubing Subassembly | Serialized Heat Bonded PU Tubing Subassembly | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| TiO2 Pigmented Tubing | | | | |
| Liquid Silicone Rubber, DOW 7-6860 | Front Seal | Front Seal | EASYTRAK 2 IS-1 | P010012/S024 06AUG-04 |
| Liquid Silicone Rubber, DOW 7-6860 | Rear Seal | Rear Seal | EASYTRAK 2 IS-1 | P010012/S024 06AUG-04 |
| Platinum Clad Tantalum | Outer Coil Conductor Wire, ETFE coated | Outer Coil Conductor Wire, ETFE coated | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Ethylene tetrafluoroethylene (ETFE) | | | | |
| MP/35N with Tantalum | Inner Coil | Inner Coil | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Silglide Coating | Outer Silicone Tubing | Outer Silicone Tubing | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Silicone Rubber, Nusil MED-4755 | | | | |
| Silglide Coating | Inner Silicone Tubing | Inner Silicone Tubing | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Silicone Rubber, Nusil MED5-4770** | | | | |
| Platinum/Iridium (90/10) | Inner Proximal Electrode | Inner Proximal Electrode | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| IROX | IROX coated Proximal Ring | IROX coated Proximal Ring | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Platinum/Iridium (90/10) | | | | |
| Platinum/Iridium (90/10) | Tip Electrode Crimp Tube | Tip Electrode Crimp Tube | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| IROX | IROX coated Tip Electrode | IROX coated Tip Electrode | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Platinum/Iridium (90/10) | | | | |
| Liquid Silicone Rubber, DOW 7-6860 | Silicone Molded Neck | Silicone Molded Neck | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Silicone Adhesive, Nusil MED-1514 | Medical Adhesive | Medical Adhesive | EASYTRAK 2 IS-1 | P010012/S024 06AUG-04 |
| Polyurethane Adhesive, Polycin 937/Vorite 689 | Polyurethane Adhesive | Polyurethane Adhesive | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Dexamethasone Acetate | Steroid Collar | Steroid Collar | EASYTRAK 3 LV-1 | P010012/S025 06AUG-04 |
| Liquid Silicone Rubber, DOW 7-6860 | | | | |
| Dow Corning 1200 Primer Coat, Clear | Silicone Primer | Silicone Primer | RELIANCE G/SG | P910073/S041 04NOV03 P910073/S035 02NOV00 |
| TiO2 Pigment | Lead Stabilizer | Lead Stabilizer | EASYTRAK 2 LV-1 | P010012/S024 06AUG-04 |
| Platinum Cure Silicone Rubber, Dow Q7-4765 | | | | |
| White Silicone Paste, Ferro Corp 61-18000 and 61-18200 | | | | |
** Nusil MED5-4770 is a new durometer of the same silicone rubber family. Biocompatibility testing was performed on the Nusil MED-4755 material
P050046 SSED
Page 13 of 28
20
{13}
# B. Engineering Testing
The ACUITY Steerable lead was tested to examine the electrical and mechanical integrity. Testing is summarized in Table 4.
Table 4: Engineering Testing
| Summary of ACUITY Steerable Design Verification Testing | Sample Size | Test Results (Pass/Fail) |
| --- | --- | --- |
| MDVT (Package, Mechanical, Electrical, and Suture Sleeve Lead Testing) | | |
| Shipping Conditioning Subject the leads to simulated shipping conditions per ASTM D 4169-04 and ensure the lead meets visual criteria (ex. Ensure “J-shape” of lead is maintained) and withstands packaging stylet insertion/withdrawal forces | 10 | Pass |
| Packaging Visual Verify packaging literature, items properly positioned | 10 | Pass |
| Lead Visual – Insulation Structural Continuity Inspection: Verify the lead’s insulation structure continuity meet specification. All pre-molded silicone elastomer sections and all components bonded with Medical Adhesive shall not exhibit delaminating or joint separation | 20 | Pass |
| Lead Visual – Device Length verification: Verify the lead’s length. | 20 | Pass |
| Lead Visual – Device Diameter verification: Verify the lead’s diameter | 20 | Pass |
| Lead Visual – Electrode Spacing: Verify electrode spacing. | 20 | Pass |
| Resistance: Verify direct current resistance. For 100 cm lead, max is 82Ω. | 20 | Pass |
| Insulation Integrity (Dielectric): Verify the integrity of the lead insulation between the conductor for the proximal electrode and the conductor for the distal electrode per the requirements of the Device Spec of less than 0.6 mA. | 20 | Pass |
| Insulation Integrity (Pressure): Verify the integrity of the insulation and bonds; they must display a leak rate between 0.030 and 0.165 cc/min once pressurized. | 20 | Pass |
| Packaging Stylet Insertion/Withdrawal: Verify the withdrawal force does not exceed a maximum of 0.25 Lb. | 10 | Pass |
| System Compatibility: Verify the lead is compatible with components (RAPIDO system and Finishing wires). | 10 | Pass |
| Fixation Stability: Verify the fixation method meets minimum allowable peak extraction force of 2.7g. | 10 | Pass |
| Pacing Impedance: Measure and record the pacing impedance. There is no acceptance criterion. Results showed pacing impedances ranged from 1199 to 1395 ohms. | 10 | N/A |
| Sensing Impedance: Measure and record the sensing impedance. There is no acceptance criterion. Results showed sensing impedances ranged from 10.5 to 10.6 K-ohms. | 10 | N/A |
| Axial Load: Meet CEN/CENELEC EN 45502-2-1, which includes max permanent elongation 5%, leakage current ≤2mA, and resistance test (see above) | 10 | Pass |
P050046 SSED
Page 14 of 28
21
{14}
| Summary of ACUITY Steerable Design Verification Testing | Sample Size | Test Results (Pass/Fail) |
| --- | --- | --- |
| Suture Sleeve Stability on Lead Body: Without being tied to lead, suture sleeve must not slide down lead. | 10 | Pass |
| Suture Sleeve Retention: Verify when tied to lead (with 3Lb force), it does not move more than 1 cm with a 0.5Lb load on lead. | 10 | Pass |
| Terminal Assembly Strength: Verify durability of terminal connector when subjected to torsional, compressive, and tensile forces. Lead must meet electrical performance criteria in Resistance and Dielectric testing (see above) | 10 | Pass |
| Corrosion | | |
| Corrosion DVT: Verify inner coil is not functionally impacted by corrosion after 10-year pacing simulation. Leads must meet electrical & axial pull test. | 8 | Pass |
| Corrosion Assessment: Evaluate each component and metal-to-metal joint according to (1) environment the component will be exposed to, (2) determine possible corrosion modes, and (3) evaluate corrosion risk based on prior tests or well documented scientific justification when available. ACUITY Steerable has eight (8) metal components and six (6) metal-to-metal joints. | n/a -- no testing was required per the evaluation | Pass |
| Packaging and Label Integrity | | |
| Packaging: Verify the printed markings on the labels for lead and accessories meet applicable sections of EN 980:2003, EN 1041:1998, and EN 45502-1:1997 | 4 | Pass |
| Particulate | | |
| Particulate: Verify lead meets EN 45502 part 2-1 | 6 | Pass |
| Flex Fatigue and (Bell Mouth) | | |
| IS-1 Connector Flex Fatigue: Comply with EN 45502 Part 2-1 section 23.5 | 6 | Pass |
| Bell Mouth Flex Fatigue Test of Polyurethane (PU) lead body: Comply with EN 45502-2-1 at 85,000 cycles. | 8 | Pass |
| Intracardiac Buckle Flex Fatigue in PU body: This test has been created unique to Guidant. After flexing in the polyurethane lead body, verify lead body conductor withstands ten years (420 million cycles) of flexing. Test is based upon minimum in-vivo intracardiac bend radius conditions. Sample must meet DC resistance of 166Ω for a 100 cm lead, 350Ω maximum change in DC resistance, and maintain ability to properly pace and sense. | 8 | Pass |
| Intracardiac Buckle Flex Fatigue in PU to Silicone transition and J-Shape: This test has been created unique to Guidant. After flexing in the distal tubing and bias zone, verify lead body conductor withstands ten years (420 million cycles) of flexing. Test is based upon minimum in-vivo intracardiac bend radius conditions. Sample must meet DC resistance of 166Ω for a 100 cm lead, 350Ω maximum change in DC resistance, and maintain ability to properly pace and sense. | 8 | Pass |
P050046 SSED
Page 15 of 28
22
{15}
The ACUITY™ Steerable Lead contains several components that are identical to devices that are commercially available from Guidant. Generally, the tests listed in the table below are needed. However, for this submission, the sponsor referenced tests from previously marketed devices that are applicable to the ACUITY™ Steerable Lead. The PMA supplement number for each referenced test appears in the table below.
| Referenced Tests | | |
| --- | --- | --- |
| Steroid Dosage | Identical to ET3 LV-1
(P010012/S025, approved 06Aug04) | |
| Lead Visual Inspection | Identical to ET3 IS-1
(P010012/S032, approved 24Nov04) | |
| IS-1 Insertion/Withdrawal Force | Identical to ET3 IS-1
(P010012/S032, approved 24Nov04) | |
| IS-1 Pin/Setscrew Deformation¹ | Identical to ET3 IS-1
(P010012/S032, approved 24Nov04) | |
| IS-1 Dimension Inspection | Identical to ET3 IS-1
(P010012/S032, approved 24Nov04) | |
¹ Test was performed twice because of a bond design enhancement. Both reports were submitted with lead. (P010012/S032, approved 24Nov04).
P050046 SSED
Page 16 of 28
23
{16}
| Referenced Tests (continued) | | |
| --- | --- | --- |
| Low Voltage Seal Integrity | Identical to ET3 IS-1 (the enhanced bond report) (P010012/S032, approved 24Nov04) | |
| Lead Cap testing² | Identical to ET2 IS-1 for model 6623 (P010012/S024, approved 06AUG04) and ET 1 IS-1 for model 6829 (P010012/S003, approved 22Jan03) | |
| IS-1 Hemostasis Valve Insertion/Withdrawal Force (Dry Terminal) | Identical to ET2 IS-1 (P010012/S024, approved 06Aug04) | |
| IS-1 Hemostasis Valve Lead System Leak Test (Dry Terminal Pin) | Identical to ET2 IS-1 (P010012/S024, approved 06AUG04) | |
| IS-1 Hemostasis Valve Insertion/Withdrawal Force (Wet Terminal Pin) | Identical to IS-1 (P010012/S024, approved 06AUG04) | |
| IS-1 Hemostasis Valve Lead System Leak Test (wet Terminal Pin) | Identical to o ET2 IS-1 (P010012/S024, approved 06AUG04) | |
| Packaging and Lead Visual | Identical to RELIANCE G/SG tachy lead (P910073/S041, approved 04NOV03 P910073/S035, approved 02NOV00) | |
| Sterilization Assessment: A 100% Ethylene Oxide (ETO) process is used. An overkill approach based on ANSI/AAMI standard 11135-1994 “Medical Device Validation and Routine Control of ETO” is used because it ensures the probability of a microbial survivor is less than one in a million. | Identified product capable of being sterilized per the validated Guidant ethylene oxide sterilization process | |
| Shelf Life: Lead must meet 2-year shelf life requirement. Accelerated aging was submitted along with an assessment. | 10 | Pass |
| Safety and Risk Analysis: Hazard Analysis and Process FMEA were conducted. | Mitigations for potential hazards were reviewed and identified as acceptable | |
| Second Supplier | | |
| Axial load MDVT testing was performed to verify inner coil insulation silicone tubing extruded from Helix Medical inc. conformed to axial load requirements of maximum permanent elongation = 5% and leakage current ≤ 2mA. | 10 | Pass |
## C. Animal Studies
Guidant conducted a chronic, 90-day, study to demonstrate that the ACUITY Steerable lead pacing system is safe in an animal model. ACUITY Steerable differs from the market approved EASYTRAK family of left ventricular leads because it has a stylet delivery option and a "J" shaped fixation tip. The purpose of the animal study was to (1) evaluate in vivo the mechanical and electrical performance of the lead and (2) demonstrate that the various components of the pacing lead system are functional, compatible, and safe in an animal model. The animal model chosen was a female porcine because the veins are similar in size to human, which allowed Guidant to assess lead handling and fixation appropriately.
P050046 SSED
Page 17 of 28
24
{17}
The GLP animal study conclusion shows that the ACUITY Steerable lead can be successfully implanted with the lead delivery tools (including the soft and standard stylets) and remains fixed without dislodging or perforating and continues to provide acceptable chronic electrical performance. Chronic data further supports ACUITY Steerable system safety.
Acute animal GLP results at 28-days show all primary endpoints were successfully met:
- Mean pacing threshold was less than 3.5 volts (N=3)
- Mean pacing impedance was greater than 300 ohms (N=7)
- Mean R-wave amplitude was greater than 3mV (N=8)
Acute Observational data were also collected at 28 days:
1. Adverse Events involving the compatibility of the lead implant tools (guiding catheter, ET4 soft or standard stylet, standard guide wire or finishing wire, and handling characteristics (steerability with Stylet) of the ACUITY Steerable lead at implantation
2. Verification of the mean unipolar and bipolar impedance of ACUITY Steerable less than 2000 ohms at 28 days
And, successful Chronic Observations were collected at 90 days:
3. Lead integrity assessment
4. Gross tissue assessment
5. Histopathology assessment
6. Chronic Fixation that is qualitatively assessed; lead movement rate from original implant location at 90 days (±6days). Additionally, pacing thresholds, pacing impedances, R-wave amplitudes on Day 60 (± 6 days) and Day 90 (± 6 days) were measured.
The animal testing and test results were acceptable to support the safety and effectiveness of the ACUITY Steerable lead.
P050046 SSED
Page 18 of 28
{18}
# X. Clinical Study
## Study Design
This clinical investigation was a prospective, multi-center, non-randomized study designed to demonstrate the safety and effectiveness of the ACUITY Steerable Left Ventricular lead, models 4554/4555/4556, in humans. A total of 110 patients were enrolled at 26 centers in the United States.
In all patients implanted with the ACUITY Steerable lead, the lead was connected to a CONTAK RENEWAL® 3 or CONTAK RENEWAL 3 HE, cardiac resynchronization therapy defibrillator (CRT-D) or to a CONTAK RENEWAL TR cardiac resynchronization therapy pacemaker (CRT-P).
## Patient Selection
### Patient Inclusion Criteria
- Must receive a commercially available Guidant CRT-P or CRT-D device
- Creatinine < 2.5 mg/dL obtained no more than two weeks prior to enrollment
- Age 18 or above, or of legal age to give informed consent specific to state and national law
- Willing and capable of providing informed consent, undergoing a device implant, participating in all testing associated with this clinical investigation at an approved clinical investigation center and at the intervals defined by this protocol
- Geographically stable residents who are available for follow-up
### Exclusion Criteria
- Have a known hypersensitivity to a 1.0 mg (0.5 mg per electrode) nominal dose of dexamethasone acetate
- Have or had previous cardiac resynchronization therapy, a coronary venous pace/sense lead or attempted LV lead placement
- Have pre-existing cardioversion/defibrillation leads other than those specified in the investigational plan (unless the investigator intends to replace them with permitted cardioversion/defibrillation leads)
- Currently requiring dialysis
P050046 SSED
Page 19 of 28
{19}
- Have had a myocardial infarct, unstable angina, percutaneous coronary intervention, or coronary artery bypass graft during the preceding 30 days prior to enrollment
- Have hypertrophic obstructive cardiomyopathy or infiltrative cardiomyopathy (e.g., amyloidosis, sarcoidosis)
- Documented life expectancy of less than six months or expected to undergo heart transplant within the next six months
- Enrolled or participating in any concurrent study without Guidant written approval, that may confound the results of this study
- Have a pre-existing unipolar pacemaker that will not be explanted/abandoned
- Have a mechanical tricuspid heart valve
- Women who are pregnant or plan to become pregnant
Note: Women of childbearing potential must have had a negative pregnancy test within seven days of enrollment.
## Patient Enrollment and Follow-Up
Data was collected for a total of 110 patients enrolled in the study. All 110 patients enrolled in the study underwent an implant procedure to receive the Acuity steerable lead. 101 patients were successfully implanted. Follow-up was performed at pre-discharge, 1-month and 3-months to meet study endpoints. Patients will be followed once a quarter until study closure. Figure 3 outlines the Patient status of the trial.
P050046 SSED
Page 20 of 28
27
{20}

Figure 3: Patient Status
# Objectives
Primary Objective 1: Safety – Lead Related Complications
Hypothesis: The ACUITY Steerable lead will be equal-to or greater-than 80% at three-month, post implant for lead-related complication-free rate; where 80% is the pre-determined one-sided 95% acceptance boundary.
Primary Objective 2: Effectiveness – Pacing Thresholds
Hypothesis: The ACUITY Steerable lead will be equal-to or greater-than 2.5 Volts at three-month, post implant for left ventricular pacing threshold when measured at 0.5ms pulse width; where 2.5V is the pre-determined one-sided 95% acceptance boundary.
Secondary Objective 1: Effectiveness – Sensed Amplitudes
Hypothesis: The ACUITY Steerable lead will be programmed to the sensing amplitude capable of detecting R-waves that is equal-to or less-than 3.0 mVolts at three-month, post implant; where 3.0mV is the pre-determined one-sided 95% acceptance boundary.
P050046 SSED
Page 21 of 28
{21}
# Secondary Objective 2: Effectiveness – Pacing Impedances
Hypothesis: The ACUITY Steerable lead will be tested in each possible pacing configuration and programmed to the pacing impedance that is equal-to or less-than 300 Ω, at three-month, post implant; where 300 Ω is the pre-determined one-sided 95% acceptance boundary.
# Patient Demographic Information
Patient demographic information is provided in Table 6.
All patients enrolled: N=110
Table 6: Patient Demographics
| Characteristic | Measurement | Result |
| --- | --- | --- |
| Age at Implant (years) | N | 110 |
| | Mean ± SD | 68.0 ± 11.6 |
| | Range | 28.7 - 84.9 |
| Gender [N (%)] | Male | 72 (65) |
| | Female | 38 (35) |
| NYHA Class [N (%)] | II | 1 (1) |
| | III | 98 (89) |
| | IV | 11 (10) |
| LVEF (%) | N | 110 |
| | Mean ± SD | 23.5 ± 7.0 |
| | Range | 10.0 - 40.0 |
| QRS Duration† (ms) | N | 109 |
| | Mean ± SD | 153 ± 27 |
| | Range | 80 - 240 |
| Etiology [N (%)] | Ischemic | 70 (64) |
| | Nonischemic | 40 (36) |
| Conduction Disorder‡ [N (%)] | Left Bundle Branch Block | 76 (74) |
| | Nonspecific Intraventricular Delay | 15 (15) |
| | Right Bundle Branch Block | 12 (12) |
| Arrhythmias* [N (%)] | SVT | 37 (65) |
| | Nonsustained VT | 14 (25) |
| | Monomorphic VT (MVT) | 5 (9) |
| | Nonsustained VT with inducible MVT | 5 (9) |
| | Premature ventricular contractions | 4 (7) |
| | Heart block | 3 (5) |
| | Ventricular Fibrillation (VF) | 3 (5) |
P050046 SSED
Page 22 of 28
{22}
| Characteristic | Measurement | Result |
| --- | --- | --- |
| | Sick sinus syndrome | 2 (4) |
| | Conduction disorder | 1 (2) |
| | Other** | 4 (7) |
†Not available for 1 pacer dependent patient
‡Conduction disorder not present in 7 patients
*Patients may appear in more than one category
**Includes 1 patient with Bradycardia, 1 patient with nonsustained accelerated ventricular rhythm, 1 patient with recurrent ventricular tachycardia, and 1 patient with both junctional escape and premature atrial contractions
## Gender Bias
The gender selection in this clinical trial was completely random, and patient selection was solely based upon exclusion and inclusion criteria. Men represented 65% of the population. There was no difference in safety and effectiveness of the ACUITY Steerable lead with respect to gender.
## Data Analysis and Results
Results of the primary objectives are provided in Table 7.
Table 7: Data Analysis and Results
| Endpoint and Hypothesis | Result | | Met Endpoint |
| --- | --- | --- | --- |
| Primary Safety Endpoint | | | |
| Lower one-sided 95% confidence bound of the 3-month lead-related complication-free rate > 80% [% (LB*)] | N (pts) | 110 | Yes |
| | Number of pts with complications @ 3 months | 9 | |
| | Complication-Free Rate | 91.8% | |
| | Lower One-Sided 95% Confidence Bound | 86.2% | |
| Primary Effectiveness Endpoint | | | |
| Upper one-sided 95% confidence bound of the 3-month pacing threshold < 2.5 V [mean (UB*)] | N (pts) | 90 | Yes |
| | 3-month pacing threshold (Mean +/- SD) | 1.1 +/- 0.9 | |
| | Upper One-Sided 95% Confidence Bound | 1.3 | |
| Secondary Effectiveness Endpoints | | | |
| Lower one-sided 95% confidence bound of the 3-month R-wave amplitude > 3mV [mean (LB)] | N (pts) | 80 | Yes |
| | 3-month R-wave amplitude (Mean +/- SD) | 14.3 +/- 7.4 | |
| | Lower One-Sided 95% Confidence Bound | 12.9 | |
| Lower one-sided 95% confidence bound of the 3-month pacing impedance > 300 Ohms [mean (LB)] | N (pts) | 90 | Yes |
| | 3-month pacing impedance (Mean +/- SD) | 644 +/- 207 | |
| | Lower One-Sided 95% Confidence Bound | 608 | |
*LB: Lower Bound, UB: Upper Bound.
P050046 SSED
Page 23 of 28
{23}
# Implant Success Rate
Table 8 shows the Acuity Steerable lead implant success rates.
## Table 8: Implant Success Rate
All patients implanted or attempted with an LV lead; N=110
| Left Ventricular Lead | Number of Patients Undergoing Procedure | Number of Patients Successfully Implanted | Success Rate |
| --- | --- | --- | --- |
| ACUITY STEERABLE LV lead success rate | 110 | 101 | 91.8% |
| EASYTRAK family success rate* | 110 | 106 | 96.3% |
*The EASYTRAK family implant success included patients who received any lead in the EASYTRAK family (EASYTRAK, EASYTRAK 2, EASYTRAK 3 and ACUITY steerable).
## Lead Placement
The final implant positions of the Acuity Steerable lead are shown in Table 9
### Table 9: Acuity Steerable Lead Placement
All patients implanted with an Acuity Steerable lead; N=101
| Position from RAO View | Position from LAO View | | | | Total |
| --- | --- | --- | --- | --- | --- |
| | Anterior | Lateral | Posterior | Other* | |
| Basal | 1 (1.0%) | 9 (8.9%) | 1 (1.0%) | 1 (1.0%) | 12 (11.9%) |
| Mid | 2 (2.0%) | 74 (73.3%) | 5 (5.0%) | 2 (2.0%) | 83 (82.2%) |
| Apical | 0 (0.0%) | 3 (3.0%) | 1 (1.0%) | 1 (1.0%) | 5 (5.0%) |
| Other* | 0 (0.0%) | 0 (0.0%) | 1 (1.0%) | 0 (0.0%) | 1 (1.0%) |
| Total | 3 (3.0%) | 86 (85.1%) | 8 (7.9%) | 4 (4.0%) | 101 (100.0%) |
*Other RAO position reported as posterior/lateral (1); other LAO positions reported as posterior/lateral (3) and lateral/apical (1).
## Device Failures and Replacements
There were seven (7) ACUITY Steerable lead replacements and no device failures for the clinical trial. Summary of the replacements at data cutoff included two (2) that were replaced with another ACUITY Steerable lead and five (5) that were replaced with a different LV lead.
P050046 SSED
Page 24 of 28
{24}
# XI. Conclusions Drawn from the Studies
Safety of the ACUITY™ Steerable lead was characterized by evaluating the lead in bench and animal studies, and by examining survival from lead-related complications and events in a clinical trial. Survival from all complications and events at 3 months post-implant was determined to be acceptable in this study. All primary objectives were met.
Effectiveness of the ACUITY™ Steerable lead was characterized by evaluating the lead in bench and animal studies, and by examining pacing and sensing performance through three months post-implant in a clinical trial. All primary and secondary objectives were met.
Therefore, it is reasonable to conclude that the benefits of use of the device for the target population outweigh the risk of illness or injury when used as indicated in accordance with the directions for use.
# XII. Panel Recommendation
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory Systems Panel, and FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
# XIII. CDRH Decision
FDA issued an approval order on April 13, 2007.
The applicant's manufacturing facility was inspected and was found to be in compliance with the Quality System Regulation (21 CFR 820).
# XIV. Approval Specifications
Directions for Use: See labeling
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.
Postapproval Requirements and Restrictions: See approval order.
P050046 SSED
Page 25 of 28
32
{25}
# APPENDIX A: All Adverse Events
All patients implanted or attempted; N=110
Total device months = 984
Includes all adverse events reported through March 21, 2006
Table 10: Clinical Observation and Complications Summary
| Adverse Event | Number Of Events (Number of Patients) | Complications | | Observations | |
| --- | --- | --- | --- | --- | --- |
| | | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) |
| Total Adverse Events | 157 (65) | 34.5 (38) | 6.6 (65) | 44.5 (49) | 9.3 (92) |
| ACUITY Steerable Related Events (N=101) | | | | | |
| Coronary venous dissection | 1 (1) | 0.0 (0) | 0.0 (0) | 1.0 (1) | 0.1 (1) |
| Dislodgment - Elevated threshold - LV | 2 (2) | 2.0 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Dislodgment - Extracardiac stimulation - LV | 4 (4) | 2.0 (2) | 0.2 (2) | 2.0 (2) | 0.2 (2) |
| Dislodgment - Multiple signs - LV | 2 (2) | 2.0 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Dislodgment - No reported signs - LV | 2 (2) | 2.0 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Dislodgment - Unable to capture - LV | 2 (2) | 2.0 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Elevated threshold - LV | 2 (2) | 0.0 (0) | 0.0 (0) | 2.0 (2) | 0.2 (2) |
| Extracardiac stimulation - LV | 19 (17) | 1.0 (1) | 0.1 (1) | 15.8 (16) | 1.9 (18) |
| Subtotal ACUITY Steerable Related Events | 34 (29) | 9.9 (10) | 1.1 (11) | 19.8 (20) | 2.4 (23) |
| PG Related Events (N=106) | | | | | |
| Elevated DFT - Defibrillation | 2 (2) | 0.9 (1) | 0.1 (1) | 0.9 (1) | 0.1 (1) |
| Extracardiac stimulation - Daily impedance testing | 3 (3) | 0.0 (0) | 0.0 (0) | 2.8 (3) | 0.3 (3) |
| Hematoma - Pocket (> 30 days post-implant) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Infection (> 30 days post-implant) | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Migration | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Pacemaker-mediated tachycardia (PMT) | 3 (3) | 0.0 (0) | 0.0 (0) | 2.8 (3) | 0.3 (3) |
| Psychological effect due to device therapy | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Subtotal PG Related Events | 12 (9) | 0.9 (1) | 0.2 (2) | 7.5 (8) | 1.0 (10) |
| RA Lead Related Events (N=106) | | | | | |
| Dislodgment - Unable to capture - RA | 2 (2) | 1.9 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Elevated threshold - RA | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Oversensing - RA | 2 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.2 (2) |
P050046 SSED
Page 26 of 28
{26}
| | | Complications | | Observations | |
| --- | --- | --- | --- | --- | --- |
| Adverse Event | Number Of Events (Number of Patients) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) |
| Unable to capture - RA | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Subtotal RA Lead Related Events | 6 (4) | 2.8 (3) | 0.3 (3) | 0.9 (1) | 0.3 (3) |
| RV Lead Related Events (N=106) | | | | | |
| Dislodgment - Elevated threshold - RV | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Elevated threshold - RV | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Subtotal RV Lead Related Events | 2 (2) | 1.9 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Procedure Related Events (N=110) | | | | | |
| Adverse reaction - General | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Adverse reaction - Hypotension | 2 (2) | 0.9 (1) | 0.1 (1) | 0.9 (1) | 0.1 (1) |
| Chest pain | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Hematoma - Pocket (<=30 days post-implant) | 6 (6) | 0.9 (1) | 0.1 (1) | 4.5 (5) | 0.5 (5) |
| Inappropriate VF sensing - Noise | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| LV Lead Insulation Damaged | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| LV Lead Insulation Damaged During Procedure | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Post-surgical wound discomfort | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Psychological effect due to recall | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.2 (2) |
| RV and LV leads transposed on header | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Thrombus | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Subtotal Procedure Related Events | 18 (16) | 5.5 (6) | 0.6 (6) | 10.0 (11) | 1.2 (12) |
| Protocol Testing Related Events (N=110) | | | | | |
| Extracardiac stimulation - LV | 3 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.3 (3) |
| Subtotal Protocol Testing Related Events | 3 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.3 (3) |
| Cardiovascular Related Events (N=110) | | | | | |
| Atrial fibrillation (AF) | 5 (4) | 0.9 (1) | 0.1 (1) | 2.7 (3) | 0.4 (4) |
| Atrial flutter | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Cerebrovascular accident (CVA) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Chest pain - Heart failure | 3 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.3 (3) |
| Chest pain - Ischemic | 4 (3) | 0.9 (1) | 0.1 (1) | 1.8 (2) | 0.3 (3) |
| Chest pain - Other | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Chronotropic incompetence | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Dizziness | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
P050046 SSED
Page 27 of 28
34
{27}
| | | Complications | | Observations | |
| --- | --- | --- | --- | --- | --- |
| Adverse Event | Number Of Events (Number of Patients) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) | % of Patients (N Patients) | N Events/ 100 Device Months (N Events) |
| Dizziness - Heart failure | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Dyspnea - Heart failure | 4 (4) | 2.7 (3) | 0.3 (3) | 0.9 (1) | 0.1 (1) |
| Gastrointestinal - Heart failure | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Hypotension - Heart failure | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.2 (2) |
| Multi-system failure - Heart failure | 3 (3) | 2.7 (3) | 0.3 (3) | 0.0 (0) | 0.0 (0) |
| Multiple heart failure symptoms | 16 (14) | 10.0 (11) | 1.2 (12) | 3.6 (4) | 0.4 (4) |
| Multiple symptoms | 2 (2) | 0.9 (1) | 0.1 (1) | 0.9 (1) | 0.1 (1) |
| Myocardial infarction | 2 (2) | 1.8 (2) | 0.2 (2) | 0.0 (0) | 0.0 (0) |
| Nonsustained ventricular tachycardia (NSVT) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Other SVT (AVRT, AVNRT, EAT etc.) | 2 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.2 (2) |
| Peripheral edema - Heart failure | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Prophylactic treatment | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Pulmonary embolism (PE) | 1 (1) | 0.9 (1) | 0.1 (1) | 0.0 (0) | 0.0 (0) |
| Renal insufficiency - Heart failure | 2 (2) | 0.9 (1) | 0.1 (1) | 0.9 (1) | 0.1 (1) |
| Sinus tachycardia | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Syncope | 3 (2) | 0.0 (0) | 0.0 (0) | 1.8 (2) | 0.3 (3) |
| Ventricular fibrillation (VF) | 1 (1) | 0.0 (0) | 0.0 (0) | 0.9 (1) | 0.1 (1) |
| Ventricular tachycardia (VT) | 2 (2) | 0.9 (1) | 0.1 (1) | 0.9 (1) | 0.1 (1) |
| Subtotal Cardiovascular Related Events | 63 (36) | 17.3 (19) | 2.9 (29) | 20.9 (23) | 3.5 (34) |
| Subtotal Non-cardiovascular Related Events | 19 (11) | 6.4 (7) | 1.2 (12) | 4.5 (5) | 0.7 (7) |
P050046 SSED
Page 28 of 28
35