BIOSENSE WEBSTER NAVISTAR/CELSIUS THERMO COOL DIAGNOSTIC/ABLATION DEFLECTABLE TIP CATHETERS

P030031 · Biosense Webster, Inc. · OAD · Nov 5, 2004 · Cardiovascular

Device Facts

Record IDP030031
Device NameBIOSENSE WEBSTER NAVISTAR/CELSIUS THERMO COOL DIAGNOSTIC/ABLATION DEFLECTABLE TIP CATHETERS
ApplicantBiosense Webster, Inc.
Product CodeOAD · Cardiovascular
Decision DateNov 5, 2004
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The Biosense Webster NaviStar™/Celsius™ ThermoCool® Diagnostic/Ablation Deflectable Tip Catheters and related accessory devices are indicated for catheter-based cardiac electrophysiological mapping (stimulating and recording), and when used with the Stockert 70 generator, for the treatment of Type I atrial flutter in patients age 18 or older. The NaviStar™ ThermoCool® catheter provides location information when used with the Carto EP/XP Navigation System.

Device Story

ThermoCool catheters are steerable, multi-electrode deflectable tip catheters used for cardiac electrophysiological mapping and RF ablation of Type I atrial flutter. The device delivers RF current from a Stockert 70 generator to cardiac tissue; an internal lumen provides saline irrigation to cool the tip electrode, preventing impedance rise and coagulum formation. NaviStar models include a location sensor (three orthogonal coils) for use with the Carto EP/XP Navigation System to provide real-time catheter location/orientation. Celsius models lack the location sensor. Used in clinical settings by physicians; the device is inserted into the heart via the vascular system. Irrigation is managed by a commercially available infusion pump. Output includes electrical signals for mapping and thermal energy for ablation. Benefits include deeper, larger lesions and reduced risk of char formation compared to non-irrigated catheters.

Clinical Evidence

Prospective, non-randomized, unblinded, multi-center study (22 sites, 190 ablated subjects). Primary endpoints: procedural safety (major adverse events within 7 days) and acute procedural success (complete bi-directional conduction block). Acute success achieved in 85.3% (162/190) of subjects, meeting the OPC of 88% (lower bound 80.2% vs 80%). Major adverse event rate was 15.8%; events were attributed to concomitant procedures and patient co-morbidities. Six-month follow-up data (n=147) showed 93% freedom from recurrence, supporting acute success as a surrogate endpoint.

Technological Characteristics

7F/7.5F deflectable tip catheters; 3.5mm platinum-iridium tip electrode; three ring electrodes. Materials: Pellethane (tip/shaft), PEBAX (shaft), PEEK (sensor housing). Irrigation: internal lumen for saline infusion. Sensing: thermistor or thermocouple for temperature; NaviStar includes electromagnetic location sensor. Sterilization: Ethylene Oxide. Connectivity: interfaces with Stockert 70 RF generator and Carto EP/XP system (NaviStar only).

Indications for Use

Indicated for catheter-based cardiac electrophysiological mapping (stimulating and recording) and treatment of Type I atrial flutter in patients age 18 or older. Contraindicated in patients with active systemic infection, intracardiac mural thrombus, or those who have had a ventriculotomy or atriotomy within the preceding four weeks.

Regulatory Classification

Identification

For the treatment of atrial flutter.

Reference Devices

Submission Summary (Full Text)

{0} # SUMMARY OF SAFETY AND EFFECTIVENESS DATA ## I. GENERAL INFORMATION Device Generic Name: Irrigated Diagnostic/Ablation Catheter and accessories Device Trade Names: Biosense Webster NaviStar™/Celsius™ ThermoCool® Diagnostic/Ablation Deflectable Tip Catheters Applicant's Name and Address: Biosense Webster Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765 Date of Panel Recommendation: None Premarket Approval Application (PMA) Number: P030031 Date of Notice of Approval to Applicant: November 5, 2004 ## Device and Accessory Model Numbers: Table 1 – List of Catheter Families and Models | Family Name | Model Number | | --- | --- | | NaviStar™ ThermoCool® | NS75T-BCT-252-HS | | | NS75T-CCT-252-HS | | | NS75T-DCT-252-HS | | | NS75T-FCT-252-HS | | | NS75TC-BCT-252-HS | | | NS75TC-CCT-252-HS | | | NS75TC-DCT-252-HS | | | NS75TC-FCT-252-HS | | Celsius™ ThermoCool® | D7IT-BL-252-RT | | | D7IT-DL-252-RT | | | D7IT-FL-252-RT | | | D7ITC-BL-252-RT | | | D7ITC-DL-252-RT | | | D7ITC-FL-252-RT | ## Explanation of model numbers: The NaviStar™ ThermoCool® catheters are available in the B, C, D, and F curves, and the Celsius™ ThermoCool® catheters are available in the B, D, and F curves, each being determined by the angle between the tip and shaft of the catheter, and by the radius of the PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 1 of 22 {1} curve. Both catheter families are available with either a thermistor or thermocouple for sensing the tip temperature during an ablation procedure. For the NaviStar™ ThermoCool® catheter, NS75T signifies the thermistor, and NS75TC signifies the thermocouple. Furthermore, the letter located after the "NS75T(C)-" in the model number signifies the type of curve for the catheter. For example, the NS75TC-FCT-252-HSI model has curve type F. Similarly, for the Celsius™ ThermoCool® catheter, D7IT signifies the thermistor, and D7ITC signifies the thermocouple; the letter located after the "D7IT(C)-" in the model number signifies the type of curve for the catheter. ## Related Premarket Applications: The NaviStar™ ThermoCool® catheter is derived from the NaviStar™ catheter approved under P990025 and the NaviStar™ DS catheter approved under P010068. The Celsius™ ThermoCool® catheter is derived from the Celsius™ catheter approved under P950005 and the Celsius™ DS catheter approved under P010068. The major modifications include a difference in tip electrode length (3.5 mm for Thermocool versus 4 and 8 mm) and addition of an internal irrigation lumen in the ThermoCool catheters. Wherever testing is not required in the present PMA, please refer to the SSEDs for previous PMAs for detailed information. ## II. INDICATIONS FOR USE The Biosense Webster NaviStar™/Celsius™ ThermoCool® Diagnostic/Ablation Deflectable Tip Catheters and related accessory devices are indicated for catheter-based cardiac electrophysiological mapping (stimulating and recording), and when used with the Stockert 70 generator, for the treatment of Type I atrial flutter in patients age 18 or older. The NaviStar™ ThermoCool® catheter provides location information when used with the Carto EP/XP Navigation System. ## III. CONTRAINDICATIONS Do not use this device: - in patients with active systemic infection; and - if the patient has intracardiac mural thrombus or has had a ventriculotomy or atriotomy within the preceding four weeks. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 2 of 22 {2} # IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the NaviStar™ ThermoCool® Diagnostic/Ablation Catheter instructions for use, the Celsius™ ThermoCool® Ablation Catheter instructions for use, and the Stockert 70 Radiofrequency Generator User Manual. # V. DEVICE DESCRIPTION The device components which are the subject of the PMA are as follows: A. NaviStar™ ThermoCool® Diagnostic/Ablation Deflectable Tip Catheters; and B. Celsius™ ThermoCool® Diagnostic/Ablation Deflectable Tip Catheters. Hereinafter, the two catheters are collectively referred to as the "ThermoCool catheter" unless otherwise specified. For catheter ablation procedures, the ThermoCool catheter requires the use of the following: - Stockert 70 RF Generator previously approved under P990071 and P010068; - grounding pad (dispersive pad) previously approved under P990071; - catheter interface cables (models D-1195 and D-1170) previously approved under P950005 and P990071; and - a commercially available infusion pump that provides irrigation at rates indicated in the ThermoCool catheter instructions for use. For additional aid in navigation, the NaviStar™ ThermoCool® catheter requires the use of these legally marketed devices: - RefStar reference patch -- originally cleared under K954390 and K982415; and - Carto EP/XP Navigation System -- originally cleared under K954395, K013083 and K020863. # Description A. NaviStar™ ThermoCool® and Celsius™ ThermoCool® Diagnostic/Ablation Catheters The NaviStar™ and Celsius™ ThermoCool® Diagnostic/Ablation catheters are two families of steerable, multi-electrode catheters with a deflectable tip. The ThermoCool catheter is an electrophysiology electrode catheter with a 3.5-mm tip electrode, three ring electrodes, and a temperature sensor incorporated into the 7F deflectable tip. The tip electrode serves to deliver RF current from the RF generator to the desired PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 3 of 22 {3} ablation site, and incorporates six small holes through which normal saline is passed for irrigation and cooling. The tip electrode and ring electrodes are platinum-iridium with 2-5-2 spacing of the ring electrodes. The deflectable tip is extruded from polyurethane and is composed of three lumens. One lumen (0.022") contains a coil spring and a puller-wire, the second lumen (0.033") is used for irrigation, and the third lumen (0.036") contains the location sensor (for the NaviStar ThermoCool catheter only) and the lead wires. The catheter body is single lumen high-torque 7.5F shaft for the NaviStar ThermoCool (the shaft diameter is 7 F for the Celsius ThermoCool) extruded from biocompatible PEBAX with a handpiece at the proximal end. A puller wire is anchored in the tip electrode and runs through the catheter shaft to a piston in the handpiece. A saline tube also extends from the tip through the shaft to an irrigation port on the handpiece. The irrigation port terminates in a standard luer fitting to permit the injection of normal saline to irrigate the tip electrode. The usable length of the ThermoCool catheter is 115 centimeters. The catheter is provided sterile and for single patient use only. ## B Differences between NaviStar™ and Celsius ThermoCool® Catheters The NaviStar™ ThermoCool® catheter includes a location sensor, which consists of three orthogonally arranged sensor coils. The NaviStar™ ThermoCool® catheter is used with the Carto EP/XP Navigation System to generate catheter location and orientation information by interacting with a small alternating current (AC) magnetic field propagating from the location pad placed under the patient table. The Celsius does not include the location sensor. In addition, the NaviStar™ ThermoCool® catheter features a 7F tip electrode, 8 F ring electrodes, and a 7.5 F shaft, whereas the Celsius™ ThermoCool® catheter features 7F tip electrode, ring electrodes and shaft. ## VI. ALTERNATIVE PRACTICES OR PROCEDURES Alternative therapy for Type I atrial flutter includes direct surgical ablation, RF catheter ablation with other approved diagnostic/ablation catheters, use of drugs for arrhythmia control, and antiarrhythmia pacing. ## VII. MARKETING HISTORY The NaviStar™ ThermoCool® and Celsius™ ThermoCool® Diagnostic/Ablation Catheters and accessories are have been marketed in Canada, Australia, China, Hong Kong, Singapore, Pakistan, Malaysia, Argentina, Brazil, Mexico, and the European Union since February 1999. The NaviStar™ ThermoCool® and Celsius™ ThermoCool® catheter has not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 // {4} PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 5 of 22 # VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Potential Adverse Events associated with cardiac ablation for treatment of Type I atrial flutter include the following: | • Adult Respiratory Distress Syndrome (ARDS) | • Myocardial infarction | | --- | --- | | • Air embolism | • Obstruction or perforation or damage to the vascular system | | • Anemia | • Pericardial effusion | | • Anesthesia reaction | • Pericarditis | | • Arrhythmias | • Phrenic nerve damage | | • AV fistula | • Pleural effusion | | • Cardiac perforation/tamponade | • Pneumonia | | • Cardiac thromboembolism | • Pneumothorax | | • Cerebrovascular accident (CVA) | • Pseudoaneurysm | | • Chest pain/discomfort | • Pulmonary edema | | • Complete heart block | • Pulmonary embolism | | • Component damage to ICD or implantable pacemaker | • Respiratory Depression | | • Congestive heart failure | • Seizure | | • Coronary artery spasm | • Temporary complete heart block | | • Death | • Thrombi | | • Dislodgement of implantable cardioverter defibrillator or permanent pacing leads | • Thromboembolism | | • Endocarditis | • Transient ischemic attack (TIA) | | • Exacerbation of pre-existing atrial fibrillation | • Unintended (in)complete AV, sinus node or other heart block or damage | | • Expressive aphasia | • Valvular damage/insufficiency | | • Heart Failure | • Vascular bleeding | | • Hemothorax | • Vasovagal reactions | | • Increased phosphokinase level | • Volume overload | | • Infections | • Ventricular tachycardia | | • Laceration | • Worsening chronic obstructive pulmonary disease | | • Leakage of air or blood into the lungs or other organs due to perforation | | | • Local hematomas/ecchymosis | | For actual adverse events observed during the clinical study (within 7 days post-ablation), please refer to Table 16. # IX. SUMMARY OF PRECLINICAL STUDIES The sponsor conducted preclinical and animal studies on the NaviStar™ and Celsius™ ThermoCool® catheters. These tests are summarized below. {5} In the tests described below, the catheters were single-sterilized prior to testing. Visual examinations noted below were conducted using a microscope. ## A. NaviStar™ ThermoCool® Catheter ### A.1. – Mechanical Performance of the NaviStar™ ThermoCool® Catheter The sponsor performed mechanical testing involving the NaviStar™ ThermoCool catheter. Except for the deflection test, the catheters were soaked for 5 hours in a 37° C saline bath prior to testing. Table 2 summarizes the mechanical performance testing involving the NaviStar™ ThermoCool® catheter. Besides one sample that did not pass the deflection test, all remaining test samples met the established acceptance criteria for mechanical performance testing. Table 2 - Mechanical Testing of NaviStar™ ThermoCool® Catheter | Test | Sample Size | Acceptance Criteria | Results* | | --- | --- | --- | --- | | Deflection | 30 | No mechanical failures before 50 cycles | passed with deviation** | | Gas pressure/joint seal | 30 | No leaks under 4.6 psi | passed | | Pull force of full length catheter | 15 | No failure before 4 lbs | passed | | Pull force of irrigation lumen extension | 15 | No failure before 4 lbs | passed | | Entire catheter torque | 15 | No failure before 1 in and 2 turns | passed | * In documenting results, "passed" means that all samples passed. ** One sample failed after 28 deflection cycles. Failure investigation identified the root cause as an assembly error. Omitting the sample which failed due to an assembly error, 29 devices passed with 0 failures, meeting the requirement of 95% confidence and 90% reliability. ### Tests not conducted The main differences between the NaviStar™ ThermoCool® catheter in this PMA and NaviStar™ catheter approved under P990025 and the NaviStar™ DS catheter approved under P010068 are the length of the tip electrodes and the additional lumen for irrigation. Due to these similarities, the following mechanical tests were not repeated in this PMA: - Torque Barrel to Connector - Torque Plot - Pull Test Barrel to Connector - Pull Test Shaft to Piston Joint ### A.2. - Electrical Performance of the NaviStar™ ThermoCool® Catheter Electrical performance testing was conducted on catheters sterilized prior to the qualification test (pre- and post-simulated ablation) to evaluate whether electrical PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 6 of 22 {6} performance was compromised during the test cycle. Table 3 summarizes the electrical performance testing involving the NaviStar™ ThermoCool® catheter. All test samples met the established acceptance criteria for electrical performance testing. Table 3 - Electrical Performance Testing of NaviStar™ ThermoCool Catheters | Test | Sample Size | Acceptance Criteria | Results | | --- | --- | --- | --- | | DC Lead Resistance | 30 | <10 Ω | passed | | DC Leakage Current | 30 | <1μAmp | passed | | RF Lead Impedance @ 5 kHz | 30 | <10 Ω | passed | | RF Lead Impedance @ 500 kHz | 30 | <25 Ω | passed | | RF Isolation Impedance @ 5 kHz | 30 | >100 kΩ | passed | | RF Isolation Impedance @ 500 kHz | 30 | >1 kΩ | passed | | Temperature Accuracy (idle) | 30 | 37 ± 2°C and 60 ± 2°C | passed | | Verify Calibration | 30 | Calibration results read “OK” | passed | One note is that the temperature accuracy of the NaviStar catheter used in the present study has a margin of error of +/- 5 degrees C, while the usual accuracy for other catheters is +/- 3 degrees C. The larger margin of error is acceptable since the NaviStar and Celsius ThermoCool Catheters are used with the Stockert RF generator operating in the power control mode, so that temperature accuracy is less needed. This margin of error is indicated in the ThermoCool catheter labeling. ## A.3. - Simulated Use of NaviStar™ ThermoCool® catheter Functional performance testing and simulated use testing involving the NaviStar™ ThermoCool® catheter are summarized in Table 4. Besides one sample that did not pass the deflection test as mentioned in Table 2, all remaining test samples met the established acceptance criteria for functional performance testing. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 7 of 22 {7} Table 4 – Functional Performance Testing of NaviStar™ ThermoCool | Test | Sample Size | Acceptance Criteria | Results | | --- | --- | --- | --- | | Catheter soak test | 30 | No anomalies or defects after soaking at 37°C for ≥ 5 hours | passed | | Steering through vascular model | 30 | No deformation, kink and/or electrode detachment after 10 insertions | Passed | | Tip side load (5 runs) | 30 | Force required to bend tip 90 degrees > 4 grams | Passed | | In bath rotation and deflection (100 times) | 30 | No mechanical failure | passed with deviation* | | Catheter flow rate | 30 | 28.5 – 34.5 ml | Passed | | Ablation test – 10 burns** | 30 | Consistent lesions should be created | Passed | * All units passed except one sample, which failed the deflection test (see Table 2). ** Simulated use ablation using a beef heart in a 37°C saline bath was conducted to evaluate the functional performance of the catheter and accessories. A.4. - Electromagnetic Compatibility of NaviStar™ ThermoCool® catheter Electromagnetic compatibility (EMC) testing was conducted on the NaviStar™ ThermoCool® catheter, as part of an integrated system test with the CARTO System, Stockert generator and CoolFlow pump. One note is that the EMC tests are considered passing based on the premise that the device is intended for use inside an X-ray shielded room, in order to reduce the acceptance level. This restriction in locating the device is now reflected in the labeling. The affected test results concerned radiated immunity level and radiated emissions. A.5. - Electrical Safety of NaviStar™ ThermoCool® catheter Electrical safety testing was tested on 30 catheters: Sterilization, aging, simulated shipping and simulated use (soak, steering, side load, repeated deflection, irrigation flow, RF ablation), followed by DC leakage test (less than 1.0 microampere between wires at 300 V). A.6. - Biocompatibility and Sterilization of NaviStar™ ThermoCool® catheter Biocompatibility - Since the NaviStar™ ThermoCool catheter and the NaviStar™ 4 mm catheter approved under P990025 have the same patient contacting materials, biocompatibility was not re-validated, except for the polyimide tubing used inside ThermoCool catheter to deliver saline to the patient. The biocompatibility tests for the polyimide tube are the following (all tested samples passed): PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 8 of 22 {8} ISO MEM elution using L-929 mouse fibroblast cells ISO intracutaneous reactivity test USP acute systemic injection test Hemolysis test (NIH method) -- Direct contact method Partial thromboplastin time (PTT) Platelet &amp; leukocyte counts Complement activation C3a and SC5b-9 assay ISO Guinea pig maximization sensitization test method Thrombogenicity test Sterilization – The devices are sterilized by ethylene oxide, using the same methods as the NaviStar™ 4 mm catheter approved under P990025. Table 5 lists the patient and user contacting materials that were tested in accordance with ISO 10993 and submitted under P990025. All materials are classified as short duration, direct blood path, and externally communicating per ISO 10993-1. Table 5 - Patient Contacting Materials of the ThermoCool Catheter | Material Spec No. | Component | Material Description | | --- | --- | --- | | M-5439-03 | Deflectable Tip | 55D Pellethane | | M-5013-04 | Ring and Tip Electrodes | 90%/10% Platinum/Iridium | | M-5203-111 | | | | M-5439-05 | Flexible Shaft | 75D Pellethane | | P-9749-03 | Sensor Housing | Polyetherether ketone | | P-9231 | Shaft to Tip Adhesive | Polyurethane | ## A.7. - ThermoCool® Catheter Upper Allowable RF Application Limit A study was conducted using a beef heart (non-beating) in a 37 °C saline bath to determine the upper allowable lesion limits for the NaviStar™ ThermoCool® catheter. Thirty (30) catheters were tested for a total of 250 minute burn duration each in order to assure that 125 lesions, given for 120 second each, could be delivered safely and effectively. Ablation was followed by a full suite of mechanical and electrical testing. Based on the evaluation of the first five and last five lesions, the functionality of the catheter remained the same before and after the 125 lesions. Thus, the test results support the sponsor's ability to label the catheter for up to 125 two-minute applications. ## A.8. - Animal Data of NaviStar™ ThermoCool® Catheter The sponsor provided five animal studies in the canine model as verification of proof-of-principle, as summarized below: ### Study One (11 canines) The conclusion of this study was that saline irrigation maintains a low electrode-tissue interface temperature during radiofrequency application at high power, preventing an impedance rise and allowing for deeper and larger lesions. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 9 of 22 {9} Concern (for Study One) It was noted in 6/75 applications (8%) that an abrupt impedance rise with an audible pop and without coagulum occurred. The sponsor attributed this finding to steam release from below the surface, likely related to superheating of the tissue and subsequent steam formation. In addition, it was noted that tissue temperatures during radiofrequency ablation with saline irrigation were significantly higher than in the non-irrigated catheter ablation treatment groups. To resolve this issue, the sponsor has adopted the following controls in the use of the ThermoCool catheter: * Use power titration, and an irrigation scheme consistent with applied power. * Limit maximum power with orientation: Power not to exceed 50 W when the catheter is parallel to the tissue and 35 W if the catheter is perpendicular to the tissue. * Each RF application will not exceed 120 seconds in duration. Study Two (11 canines) In this study, lesion size was found to be significantly smaller for the 5 mm tip electrode compared to the 3.5 mm tip. The number of steam pops was found to be higher in the 4 mm conventional catheter compared to the irrigated catheter. Study Three (11 canines) In this study, smaller electrode sizes (2 mm vs. 5 mm) resulted in the transmission of a greater fraction of the radiofrequency power to the tissue and resulted in higher tissue temperature and larger lesions. Study Four (5 canines) In this study, an irrigation flow rate of 30 cc/min was found to be adequate to prevent an impedance rise and char (coagulum) formation for RF applications at 50 watts. At lower flow rates, impedance rises and coagulum formation was noted. Study Five (11 canines) The study results demonstrated that lower saline irrigation rates (10-30 cc/min) are sufficient to prevent an impedance rise and coagulum formation for applications of radiofrequency current at 30 watts for 60 seconds. For higher energy application (50 watts), an irrigation flow rates of 30 cc/min was adequate in preventing the incidence of impedance rise. A.9. - Shelf Life of NaviStar™ ThermoCool® Catheter The NaviStar™ ThermoCool® Catheter is validated for a one year shelf life, as follows: PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 10 of 22 {10} The NaviStar ThermoCool catheter is packaged in a straight configuration in a single package while the Celsius ThermoCool catheter is double pouched in a round configuration. The sponsor exposed the packaging to 3x EtO during packaging validation. Packages were aged adequately for 26.2 days when stored at 60°C, or 38°C above ambient temperature. Both Bubble Leak (in accordance with ASTM F2096) and Peel Strength tests were performed at one year for both packages. Sample size for each test ran between 30 and 60 packaged products each (packages were not used for more than one test type). Simulated shipping tests were performed in accordance with the International Safe Transit Association (ISTA) shipping protocols. After storage, the catheters were run through mechanical and electrical tests, including ablations. Only one out of 30 catheters failed. Tests for one year have been completed, and 3 year tests are ongoing. ## B. Celsius™ ThermoCool® Catheter The Celsius™ ThermoCool® catheter is claimed to be functionally equivalent to the NaviStar™ ThermoCool® catheter in most aspects, with the exception that the NaviStar™ ThermoCool® catheter has the additional ability to interface with a mapping system (CARTO EP/XP) and map the target areas of the heart. Design differences between the two catheter families are predominantly the result of the need to accommodate additional electrical connections and navigation electrodes. The therapeutic modality for both catheters is identical. The Celsius™ catheter is essentially a simplified version of the NaviStar™ Catheter. In addition, the Celsius™ ThermoCool® catheter is similar to the Celsius™ catheter approved under P950005 and the Celsius™ DS catheter approved under P010068. As a result, many tests performed on the NaviStar™ ThermoCool® catheter (described above), and previously conducted on the Celsius™ Catheter (reference P950005) or Celsius™ DS catheter are not repeated on the Celsius™ ThermoCool® catheter. The verification test results for the Celsius™ ThermoCool® catheter are summarized below: ### B.1. – Mechanical Performance of the Celsius™ ThermoCool® catheter The sponsor performed mechanical testing for the Celsius™ ThermoCool catheter. Except for the deflection test, the catheters were soaked for 5 hours in a 37°C saline bath prior to testing. Table 6 summarizes the mechanical performance testing involving the Celsius™ ThermoCool® catheter. Besides two samples that did not pass the deflection test, all remaining test samples met the established acceptance criteria for mechanical performance testing. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 11 of 22 {11} Table 6 - Mechanical Testing of Celsius™ ThermoCool® Catheters | Test | Sample Size | Acceptance Criteria | Results* | | --- | --- | --- | --- | | Deflection | 33 | No mechanical failures before 50 cycles | pass with deviation** | | Gas pressure/joint seal | 33 | No leaks under 4.6 psi | Passed | | Pull force of full length catheter | 17 | No failure before 4 lbs | Passed | | Pull force of irrigation lumen extension | 17 | No failure before 4 lbs | Passed | | Entire catheter torque | 17 | No failure before 1 in-oz and 2 turns | Passed | * In documenting results, "passed" means that all samples passed. ** Two samples failed with less than 50 deflection cycles. Failure investigation identified the root cause as an assembly error. Omitting the samples which failed due to an assembly error, 31 devices passed with 0 failures, meeting the requirement of 95% confidence and 90% reliability. ## B.2. - Electrical Performance of the Celsius™ ThermoCool® catheter Electrical performance testing was conducted on catheters sterilized prior to the qualification test (pre- and post-simulated ablation) to evaluate whether electrical performance was compromised during the test cycle. Table 7 summarizes the electrical performance testing involving the Celsius™ ThermoCool® catheter. All test samples met the established acceptance criteria for electrical performance testing. Table 7 - Electrical Performance Testing of Celsius™ ThermoCool® Catheters | Test | Sample Size | Acceptance Criteria | Results | | --- | --- | --- | --- | | DC Lead Resistance | 33 | <10 Ω | Passed | | DC Leakage Current | 33 | <1 μAmp | Passed | | RF Lead Impedance @ 5 kHz | 33 | <10 Ω | Passed | | RF Lead Impedance @ 500 kHz | 33 | <25 Ω | Passed | | RF Isolation Impedance @ 5 kHz | 33 | >100 kΩ | Passed | | RF Isolation Impedance @ 500 kHz | 33 | >1 kΩ | Passed | | Temperature Accuracy (idle) | 33 | 37 ± 2°C and 60 ± 2°C | Passed | ## B.3. - Simulated Use of the Celsius™ ThermoCool® catheter Functional performance testing and simulated use testing of the Celsius™ ThermoCool® catheter are summarized in Table 8. All test samples met the established PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 12 of 22 {12} acceptance criteria for functional performance testing, except that the test used an 8 F introducer. This compatibility to an 8 F introducer is now indicated in the labeling. Table 8 – Functional Performance Testing of Celsius™ ThermoCool® Catheters | Test | Sample Size | Acceptance Criteria | Results | | --- | --- | --- | --- | | Catheter soak test | 33 | No anomalies or defects after soaking at 37° C for > 5 hours | Passed | | Steering through vascular model | 33 | No deformation, kink and/or electrode detachment after 10 insertions | Passed | | Tip side load (5 runs) | 33 | Force required to bend tip 90 degrees > 4 grams | Passed | | In bath rotation and deflection (100 times) | 31* | No mechanical failure | Passed | | Catheter flow rate | 33 | 28.5 – 34.5 ml | Passed | | Ablation test – 10 burns** | 33 | Consistent lesions should be created | Passed | * Two units that failed the deflection test were excluded from this test. ** Simulated use ablation using a beef heart in a 37°C saline bath was conducted to evaluate the functional performance of the catheter and accessories. B.4. - Electromagnetic Compatibility of Celsius™ ThermoCool® catheter Electromagnetic compatibility (EMC) testing was not conducted on the Celsius™ ThermoCool® catheter, due to similarity with the NaviStar™ 4 mm catheter approved under P990025. B.5. - Electrical Safety of Celsius™ ThermoCool® catheter Electrical safety testing was tested on 33 catheters: Sterilization, aging, simulated shipping and simulated use (soak, steering, side load, repeated deflection, irrigation flow, RF ablation), followed by DC leakage test (less than 1.0 microampere between wires at 300 V). B.6. - Animal Testing of Celsius™ ThermoCool® catheter Due to the similarity of the Celsius™ ThermoCool® with the NaviStar™ ThermoCool® catheter, no animal testing was performed on the Celsius™ ThermoCool® catheter. B.7. - Shelf Life of Celsius™ ThermoCool® catheter The Celsius™ ThermoCool® Catheter is approved for a one year shelf life, based on similarity to the NaviStar ThermoCool catheter. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 13 of 22 {13} # X. SUMMARY OF CLINICAL STUDIES The clinical testing described below was performed with the NaviStar™ ThermoCool® catheter, and not with the Celsius™ ThermoCool® catheter. Since the ablation capabilities of both NaviStar™ and Celsius™ ThermoCool® catheters were shown with pre-clinical testing to be similar, clinical testing results from the NaviStar™ ThermoCool® study, as reported below, may be extrapolated to what would be expected when using the Celsius™ ThermoCool® catheter. ## A. Objective The objective of the study was to determine if the NaviStar™ ThermoCool® catheter, when used in conjunction with Carto™ EP/XP Navigation System, Stockert 70 RF Generator and related accessories, is safe and effective for the treatment of Type I atrial flutter in patients age 18 or older. ## B. Study Design The study was a prospective, non-randomized, unblinded, multi-center study conducted at 22 investigational sites (21 sites in US; 1 in Canada). ### B.1. - Study Endpoints: The endpoints for the study were as follows: - **procedural safety** – defined by the absence of serious complication associated with the use of the investigational device within seven days of the ablation procedure; and - **acute procedural success** – defined as complete bi-directional conduction block (BDB) across the isthmus, and the inability to induce Type I atrial flutter post-procedure. Long-term freedom from atrial flutter recurrence was not specifically identified as a study endpoint. Instead, acute procedural success was used as a surrogate endpoint for this parameter. Long-term (defined as 6 months post-treatment) freedom from atrial flutter recurrence information was also collected, in order to enable FDA to assess whether the surrogate endpoint was reasonable. ### B.2. - Objective Performance Criteria (OPC): Objective performance criteria (OPC) were prospectively established. The OPC for the safety endpoint used for this study was derived from the FDA guidance document “Cardiac Ablation Catheters Generic Arrhythmia Indications for Use; Guidance for Industry, July 2002 1998 NASPE Registry.” The OPC for the effectiveness endpoint was based on an extensive literature search involving acute success rates associated with radiofrequency ablation of atrial flutter. The OPCs are defined below: PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 14 of 22 {14} - Safety: major adverse events within 7 days of the procedure occur at a rate of 2.7% or less with a 7% one-sided 95% upper confidence bound; - Acute success: 88% with an 80% one-sided 95% lower confidence bound. ## B.3. - Subject Accountability The table below documents the accountability and disposition of enrolled subjects. | Subjects enrolled in study | 198 | | --- | --- | | Subjects not ablated with the NaviStar ThermoCool catheter | 8 | | Excluded Subjects - enrolled but in whom the investigational catheter was not inserted | 3 | | Discontinued Subjects - either (1) in whom the investigational catheter was inserted but did not receive RF energy because of non-investigational equipment failure, or (2) for whom the arrhythmia was determined to be non-study arrhythmia at the time of electrophysiologic study (e.g., atypical atrial flutter). | 5 | | Subjects ablated with NaviStar ThermoCool catheter | 190 | | Subjects ablated with NaviStar ThermoCool catheter and non-investigational catheter* | 19 | | Subjects ablated only with NaviStar ThermoCool catheter | 171 | | Subjects in whom BDB was not assessable | 4 | * This category includes enrolled subjects who received ablation therapy with the investigational catheter at the start of the procedure and for whom the investigator then switched to a non-protocol catheter to complete the procedure. Further, subjects who could not receive ablation due to investigational device failure are included in this category. These subjects were considered acute effectiveness failures. **Effectiveness Analysis Population** (n=190) was defined as all subjects who received ablation therapy with the investigational catheter and for whom a valid assessment of BDB at the acute endpoint could be made OR if 6 month follow-up data were available. **Safety Analysis Population** (n=190) was defined as all enrolled subjects in whom the investigational catheter was inserted and received ablation therapy. Additionally, the rate of major adverse events is also reported for subjects in whom the investigational catheter was inserted and used for either mapping and/or ablation and for discontinued subjects. This additional category is referred to as the Inserted Patient Cohort (n=195). ## B.4. - Subject Demographics The table below summarizes the demographic information of all study subjects who received ablation therapy. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 15 of 22 {15} Table 10 – Subject Demographics (All Subjects who Received Ablation Therapy - n=190) | Gender | N | % | | --- | --- | --- | | Female | 44 | 23.2 | | Male | 146 | 76.8 | | | | | | Age (years) | | | | Mean ± standard deviation | 59.8 ± 12.6 | | | Range | 18-90 | | Additionally, for the Inserted Patient Cohort of 195 subjects, 72 subjects (36.9%) did not have a concomitant arrhythmia reported in addition to Type I atrial flutter. One-hundred and sixty-five (165) concomitant arrhythmias were reported for 123 subjects. The most common concomitant arrhythmias were atrial fibrillation (n=104) and atypical atrial flutter (n=27). ## C. Results ### C.1. - Intraprocedural Data Tables 10 and 11 describe the procedural data. Twenty-eight (28) subjects received ablation therapy for an arrhythmia other than Type I atrial flutter during the same index ablation procedure. The additional arrhythmias ablated were: 14 atrial fibrillation, 9 atrial tachycardia, 3 AVNRT, 1 intra-atrial tachycardia, 1 nonisthmus atrial flutter and 1 macro-reentry around the SVC eustachian ridge. One subject had more than one concomitant arrhythmia ablated. Table 11 - Power, Temperature and Impedance Data | Description | Mean ± Standard Deviation | Range | | --- | --- | --- | | # RF applications/procedure^{1} (n=188 procedures) | 19 ± 16 | 1-86 | | Total saline infused by ThermoCool Catheter (ml)^{2} (n=169 procedures) | 999.7 ± 605.5 | 60-3750 | | Maximum power (Watts)/application^{3} (n=3502 RF applications) | 35.0 ± 9.5 | 2-59 | | Maximum temperature (°C)/application^{3} (n=3476 RF applications) | 39.6 ± 5.1 | 14-87 | | Maximum impedance (Ohms)/application^{3} (n=3431 RF applications) | 112.1 ± 21.0 | 13-251 | 1 One subject had missing RF information; one subject did not undergo ablation with the NaviStar ThermoCool catheter. 2 Some procedural data are missing. 3 Power, temperature, and impedance not documented for several RF applications. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 16 of 22 {16} Table 12 – Overall Fluoroscopy/Procedure Time (minutes) | Description | Mean ± Standard Deviation | Range | | --- | --- | --- | | Total fluoroscopy time/procedure^{1} (n=189 procedures) | 50.2 ± 32.4 | 8-174 | | Total procedure time^{1} (n=190 procedures) | 341.6 ± 166.9 (5.7±2.8 hours) | 96-925 | | Total fluoroscopy time/procedure for subjects with additional rhythms ablated during index procedure (n= 28 procedures) | 58.8 ± 24.7 | 18-115 | | Total fluoroscopy time/procedure for subjects without concomitant ablation (n=161 procedures) | 48.7 ± 33.4 | 8- 174 | | Total procedure time for subjects with additional rhythms ablated during index procedure (n= 28 procedures) | 503.8±193.0 (8.4±3.2 hours) | 158-804 | | Total procedure time for subjects without concomitant ablation (n= 162 procedures) | 313.5±145.2 (5.1±2.4 hours) | 96-925 | 1 Incomplete fluoroscopy time was reported for one (1) subject and incomplete procedure time was reported for one (1) subject. ## C.2. - Acute Procedural Success Acute success, defined as complete bi-directional conduction block across the isthmus at a minimum of 60 minutes following application of the last RF application, was analyzed. Acute success evaluation was based on the Efficacy Population, which was defined as all subjects who received ablation therapy with the investigational catheter and in whom a valid assessment of BDB could be made (n = 190 - 4 = 186). Table 13 describes the acute ablation outcomes. Table 13 - Acute Ablation Outcomes (n=186) | | # Success / # Subjects Ablated | Percentage (one-sided 95% confidence bound) | | --- | --- | --- | | Acute Study Results | 159/186 | 85% (81%) | | OPC | | 88% (80%) | ## C.3. - Composite Assessment of Atrial Flutter Ablation Success As noted in the above section, 159 subjects had BDB confirmed acutely after the ablation procedure. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 17 of 22 {17} In addition, of the four subjects in whom BDB was not measured acutely after the ablation procedure, 3 subjects were free of recurrence of atrial flutter at 6 months follow-up and one could not be validated. For the composite assessment, the 3 subjects were considered a success and the 1 subject a failure. Table 14 summarizes the composite results. Table 14 – Composite Assessment of Atrial Flutter Success | | # Success / # Subjects Ablated | Percentage (one-sided 95% confidence bound) | | --- | --- | --- | | Study Results | 162/190 | 85.3% (80.2%) | | OPC | | 88% (80%) | C.4. - Freedom from Type I Atrial Flutter Recurrence at Six-Month Follow-Up As indicated in section B.1 above, long-term freedom from atrial flutter recurrence was not a study endpoint. The long-term results are presented here in order to assess the suitability of the surrogate endpoint BDB. Freedom from Type I atrial flutter recurrence was evaluated in subjects in whom BDB was achieved (as measured acutely) and for whom 6-month post-ablation information was available. Based on these criteria, information was available on a total of 147 subjects. Results are described in the table below. Table 15 - Freedom from Type I atrial flutter at 6 months (Results based on 147 subjects) | Description | N | Percent | | --- | --- | --- | | Subjects in whom BDB was achieved acutely and for whom 6-month information was available | 147 | 100% | | Subjects free from recurrence | 136 | 93% | | Subjects free from recurrence and anti-arrhythmic drug change | 118 | 80% | | | | | | Subjects with recurrence of atrial flutter | 11 | | | Subjects with AAD changes to treat atrial fibrillation | 15 | | | Subjects with AAD changes to treat atrial or supraventricular tachycardias | 3 | | These results provide reasonable evidence that acute procedural success serves as an appropriate surrogate for long-term freedom from atrial flutter recurrence. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 18 of 22 {18} PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 19 of 22 ## C.5. - Adverse Events A major adverse event was defined as any clinical event that occurred within seven days post-ablation and which resulted in (1) death, (2) a life-threatening complication, or (3) a persistent or significant disability/incapacity that required inpatient hospitalization or prolonged hospitalization or required intervention to prevent a permanent impairment of a body function or damage to a body structure. A minor adverse event was defined as any adverse event resulting in minimal transient impairment of a body function or damage to a body structure, or which did not require any intervention other than monitoring or events occurring more than 7 days post-ablation. ## Major Adverse Events Of the 190 subjects who received ablation therapy with the investigational catheter, 33 major adverse events were reported in 30 subjects. The overall percentage of subjects who experienced a major adverse event was 15.8%. The one-sided 95% confidence bound rate was 20.9%. For subjects who had the investigational catheter inserted and used for mapping and/or ablation (n = 195), the major adverse event rate was 15.4%, and the one-sided 95% confidence bound rate was 20.4%. Table 16 summarizes the major adverse events. | Total Number Subjects with a Major AE n=30 | | --- | | Cardiovascular total = 15 subjects | | Arrhythmia complications = 5 subjects | | complete atrioventricular block during procedure | | bradycardia requiring pacemaker implant | | ventricular tachycardia | | atrial fibrillation | | atrial fibrillation & atypical atrial flutter | | Pericardial effusion/tamponade = 4 subjects | | pericardial tamponade | | pericardial tamponade after mapping only | | pericarditis with effusion | | RA thrombus, LV thrombus and pericardial effusion | | Intracardiac thrombus = 2 subjects | | RAA thrombus | | RA thrombus, LV thrombus and pericardial effusion | | myocardial infarction = 1 subject | | congestive heart failure = 4 subjects | | pedal edema | | dyspnea, rales requiring furosemide | | dyspnea treated with one dose furosemide | | pulmonary edema by PE treated with one dose furosemide | {19} | Pulmonary acute respiratory distress syndrome = 2 subjects aspiration pneumonia = 2 subjects pneumonia = 3 subjects asthma = 1 subject | total = 8 subjects | | --- | --- | | Anesthesia related sedation induced apnea (intubation not required) sedation induced co2 retention with lethargy (intubation not required) | total = 2 subjects | | Vascular arteriovenous fistula/femoral artery-saphenous vein pseudoaneurysm/right femoral artery | total = 2 subjects | | Urologic urinary tract infection urinary retention | total = 2 subjects | | Cholecystitis | 1 subject | | Neurologic parkinson’s disease transient extremity numbness/possible tia | 2 subjects | * Note: Some subjects are listed more than once in the above table. Three subjects died during the course of the study. One subject died due to cardiac arrest caused by cardiomyopathy and chronic obstructive pulmonary disease (COPD) complications 11 days post-ablation, one subject died following pulmonary valve replacement surgery 2 months post-ablation, and the third death was due to lung cancer more than 2 years following the ablation procedure. All deaths were determined to be unrelated to the procedure and device. An overall risk benefit evaluation of these adverse events was performed and a detailed review of each adverse event was completed. The adverse event rate described above was assessed to be specifically correlated to (1) the concomitant ablation procedures performed during the index procedure and (2) the increased number of co-morbid conditions present in the subject population enrolled relative to patient population from which the OPCs were derived. See section C.1 for a list of concomitant ablation procedures. ## C.6. - Statistical Analysis Table 17 summarizes the safety and effectiveness of the device when compared to the control group OPC established for safety and acute success. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 20 of 22 {20} Table 17 - Comparison of Endpoints between NaviStar™ ThermoCool Study and OPC | Endpoint | OPC | | NaviStar™ ThermoCool Study | | | --- | --- | --- | --- | --- | | | % | One-sided 95% Confidence Bound | % (N) | One-sided 95% Confidence Bound | | Acute Success | 88% | 80% | 85.3% (162/190) | 80.2% (Lower bound) | | Major Complications | 2.7% | 7% | 15.8% (30/190) | 20.9% (upper bound) | With comparison of the lower bounds of the acute success endpoints (80.2% vs. 80%), the results demonstrate that the NaviStar™ ThermoCool catheter met the OPC for acute success. As previously explained in section C.5, although the device exceeded the upper bound of major complications, review of the specific events showed that they were related to the concomitant ablation procedures performed in addition to atrial flutter ablation and the subject population co-morbid conditions. Accordingly, study results demonstrate a reasonable assurance of the safety profile of the device. ## XI. CONCLUSIONS DRAWN FROM THE STUDIES Pre-clinical testing demonstrates that the NaviStar™ ThermoCool® catheter, Celsius™ ThermoCool® catheter and accessories (Stockert 70 RF generator, cables and junction box) will maintain mechanical and electrical integrity under the proposed conditions of use. Additionally, biocompatibility testing of the patient-contacting materials demonstrates that the devices are biocompatible under the proposed conditions of use. Clinical testing and analysis demonstrate that the NaviStar™ ThermoCool® and Celsius™ ThermoCool® catheters when used with the Stockert 70 RF generator are reasonably safe and effective for the treatment of Type I atrial flutter. ## XII. PANEL RECOMMENDATION In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA is similar to information previously reviewed by this panel. ## XIII. CDRH DECISION CDRH issued an approval order on November 5, 2004. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 21 of 22 {21} The applicant's manufacturing facilities were inspected on March 20, 2003 and found to be in compliance with the device Quality System Regulation (Part 820). Hence, no new inspection is needed. ## XIV. APPROVAL SPECIFICATIONS Directions for use: See the labeling. Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the labeling. Post-approval Requirements and Restrictions: See approval order. PMA P030031: Summary of Safety and Effectiveness Data - version 5.1 Page 22 of 22 29
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...