VARIANT II TURBO HBA1C KIT-2.0 VARIANT II TURBO HEMOGLOBIN TESTING SYSTEM

K122472 · Bio-Rad Laboratories, Inc., Clinical Systems Divis · LCP · Oct 26, 2012 · Hematology

Device Facts

Record IDK122472
Device NameVARIANT II TURBO HBA1C KIT-2.0 VARIANT II TURBO HEMOGLOBIN TESTING SYSTEM
ApplicantBio-Rad Laboratories, Inc., Clinical Systems Divis
Product CodeLCP · Hematology
Decision DateOct 26, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Indications for Use

The Bio-Rad VARIANT™ II TURBO HbA1c Kit-2.0 is intended for the quantitative determination of hemoglobin Ate in human whole blood using ionexchange high performance liquid chromatography (HPLC) on the VARIANT™ II TURBO Hemoglobin Testing System. Measurement of hemoglobin A1c is effective in monitoring long term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT™ II TURBO HbA1c Kit-2.0 is intended for Professional Use Only.

Device Story

The VARIANT™ II TURBO HbA1c Kit-2.0 is an in vitro diagnostic assay used with the VARIANT™ II TURBO Hemoglobin Testing System. The system utilizes ion-exchange high-performance liquid chromatography (HPLC) to separate and quantify hemoglobin A1c in EDTA-anticoagulated human whole blood. The system consists of a chromatographic station and a sampling station, controlled by a PC running Clinical Data Management (CDM) software. It is a fully automated, high-throughput analyzer operated by laboratory professionals in clinical settings. The system processes samples, performs chromatographic separation on a cation exchange cartridge, and calculates HbA1c concentrations. Results are reported in %HbA1c (NGSP), mmol/mol (IFCC), or %HbA1c (JDS). These quantitative results assist healthcare providers in assessing long-term glycemic control for patients with diabetes mellitus.

Clinical Evidence

Bench testing only. Method comparison study (CLSI EP9-A2) compared CDM software v5.1 vs v4.03 using 100 EDTA whole blood samples and 16 spiked samples. Results showed high correlation (R²=0.999) and acceptable bias at decision points (6.0% and 9.0% HbA1c). Interference testing (CLSI EP7-A2) confirmed no significant interference from HbS (≤67%), HbC (≤72%), HbD (≤55%), HbE (≤41%), or β-thalassemia trait (HbA2 up to 10%).

Technological Characteristics

Ion-exchange HPLC; cation exchange matrix. Components: analytical cartridge, prefilters, elution buffers, calibrators, whole blood primer. Electrical safety: BS EN 61010; EMC: BS EN 61326. Software: Clinical Data Management (CDM) v5.1.1. Connectivity: PC-controlled system. NGSP and IFCC certified.

Indications for Use

Indicated for the quantitative determination of hemoglobin A1c in human whole blood to monitor long-term glycemic control in individuals with diabetes mellitus. For professional use only.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k122472 B. Purpose for Submission: Modification to a marketed device to revise hemoglobin variant interference claim and updated software, hardware and firmware. C. Measurand: Hemoglobin A₁c (HbA₁c) D. Type of Test: Quantitative ion-exchange high-performance liquid chromatography (HPLC) E. Applicant: Bio-Rad Laboratories, Inc. F. Proprietary and Established Names: VARIANT II TURBO HbA₁c Kit – 2.0 and VARIANT II TURBO Hemoglobin Testing System. G. Regulatory Information: | Regulation Description | Product Code | Device Class | Regulation | Panel | | --- | --- | --- | --- | --- | | Glycosylated Hemoglobin Assay | LCP | II | 21 CFR § 864.7470 | Hematology, 81 | H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: The Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 is intended for the quantitative determination of hemoglobin A1c in human whole blood using ion-exchange high performance liquid chromatography (HPLC) on the VARIANT II TURBO Hemoglobin Testing System. Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 is intended for Professional Use Only. 3. Special conditions for use statement(s): {1} For prescription use only. ## 4. Special instrument requirements: For use with the Bio-Rad VARIANT II TURBO Hemoglobin Testing System (cleared under k040872 and k090699; modified in the current submission). ## I. Device Description: The VARIANT II TURBO HbA₁c Kit – 2.0 contains an analytical cartridge, five prefilters, Elution Buffers A and B, Calibrator Level 1, Calibrator Level 2, Whole Blood Primer, sample vials and a CD-ROM with test parameters. The Calibrators and the Whole Blood Primer contain lyophilized human red blood cell hemolysate with gentamicin, tobramycin, and EDTA as preservatives. Each unit of whole blood used in the manufacture of the calibrators and whole blood primer was tested by FDA accepted methods and found non-reactive for HIV-1, HIV-2, Hepatitis B (HBV), Hepatitis C (HCV), and syphilis. The items in the VARIANT II TURBO HbA₁c Kit – 2.0 are described in the table below. | Component | Description | | --- | --- | | Analytical Cartridge | One cation exchange cartridge (2500 tests), 4.6 ID x 27.5 mm. Five prefilter elements (500 tests each) are included with the cartridge. | | Elution Buffer A | Five bottles containing 2500 mL of a sodium perchlorate buffer. Contains <0.05% sodium azide as preservative. | | Elution Buffer B | One bottle containing 2500 mL of a sodium perchlorate buffer. Contains <0.05% sodium azide as preservative. | | Calibrator/Diluent Set | One set consisting of two vials of Calibrator Level 1, two vials of Calibrator Level 2, and one bottle of Calibrator Diluent. The calibrator vials contains lyophilized human red blood cell hemolysate with gentamicin, tobramycin, and EDTA as preservatives. Reconstituted volume is 7 mL per vial. Calibrator Diluent contains 100 mL of deionized water with <0.05% sodium azide as a preservative. | | Whole Blood Primer | Two vials of lyophilized human red blood cell hemoglysate with gentamicin, tobramycin, and EDTA as preservatives. Reconstituted volume is 1.0 mL per vial. | | Sample Vials | 100 polypropylene vials with pierceable caps, 1.5 mL. | | CD-ROM | VARIANT II TURBO HbA1c Kit – 2.0 parameters. | The VARIANT II TURBO Hemoglobin Testing System provides an integrated method for sample preparation, separation and the percent determination of HbA1c in EDTA human whole blood. The VARIANT II TURBO Hemoglobin Testing System is a fully automated, high-throughput hemoglobin analyzer. It consists of two modules - the VARIANT II Chromatographic Station (VCS) and the VARIANT II Sampling Station (VSS). There have been hardware updates due to obsolescence of parts and firmware updates to support the {2} replacement hardware components. A personal computer is used to control the VARIANT II TURBO Hemoglobin Testing System using Clinical Data Management (CDM) software version 5.1.1. J. Substantial Equivalence Information: | Predicate device name | Predicate 510(k) number | | --- | --- | | Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 | k090699 | Comparison with predicate: | Item | Candidate Device | Predicate Device | | --- | --- | --- | | Intended Use | Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. | Same | | Test Principle | Ion-exchange high performance liquid chromatography | Same | | Sample Type | Anticoagulated whole blood (EDTA) | Same | | Detection Wavelength | 415 nanometers | Same | | Reporting Units | % HbA1c (NGSP) mmol/mol HbA1c (IFCC) | Same | | Expected Range | 3.5 to 19.0% HbA1c (NGSP) | Same | | Interference from HbF | HbF up to 25% had no significant effect on HbA1c determination. | Same | | Interference from variants HbC, HbD, HbE, HbS | No significant interference was observed at the following concentrations: • HbC <72% • HbD <55% • HbE <41% • HbS <67% | Hemoglobin variants: Two out of 7 hemoglobin AD-trait, 2 out of 11 hemoglobin AS-trait, 1 out of 12 hemoglobin AE-trait, and 3 out of 9 hemoglobin AC-trait patient samples at the clinically significant levels of 6% and 9% HbA1c exhibited differences of more than +10% from values obtained using boronate affinity reference method. | | Interference from HbA2 | β-thalassemia trait, as indicated by increased HbA2 concentrations up to 10%, does not interfere with the assay. | No claim previously | | Instrument | VARIANT II TURBO Hemoglobin Testing System | Same | | Software | CDM Version 5.1.1 | CDM Version 4.03 | | Calibrator | 2 levels | Same | {3} | Calibration frequency | Once every 500 injections/ 2500 injections total column life | Same | | --- | --- | --- | K. Standard/Guidance Document Referenced (if applicable): CLSI EP7-A2: Interference Testing in Clinical Chemistry. CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples. L. Test Principle: The VARIANT II TURBO HbA1c Kit- 2.0 is based on chromatographic separation of HbA1c on a cation exchange cartridge. The various forms of hemoglobin exhibit charge differences (positive) at the acidic pH of the mobile phase, and thus can be separated on a support that is negatively charged (cation exchange). The use of ion-exchange chromatography allows molecules to be separated based upon a molecule's charge. Separation is optimized to eliminate interferences from hemoglobin variants (HbS, HbC, HbD and HbE trait), labile A1c, hemoglobin F and carbamylated hemoglobin. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Previously established in submission k090699. b. Linearity/assay reportable range: The reportable range of 3.5 to 19.0% HbA₁c was previously established in submission k090699. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrators were previously cleared in k070452. The recommended control materials were cleared in k070546 and k052838. The modified device is certified by the National Glycohemoglobin Standardization Program (NGSP) as having documented traceability to the Diabetes Control and Complications Trial (DCCT) reference method. d. Detection limit: Previously established in submission k090699. e. Analytical specificity: The interference claims for icterus (bilirubin), lipemia (triglyceride), EDTA, hemoglobin F, labile A₁c and carbamylated hemoglobin were previously established in submission k090699. The possible interferant effects from hemoglobin variants C, D, E, F, S and A2 (HbC, HbD, HbE, HbF, HbS and HbA2) were evaluated. Two fresh, EDTA non-variant {4} human blood sample pools at 6.5% and 8.0-9.0% HbA₁c were collected. Fresh, homozygous EDTA-whole blood samples containing each of the hemoglobin variants (E, D, S, and C) were obtained. EDTA-whole blood samples were prepared by spiking purified HbF or HbA2. For the interference testing of each variant, a series of test sample pools (0, 25, 50, 75, 100% and additional intermediate pools) were prepared by the dilution of a homozygous variant patient sample (as interferant) in the non-variant patient sample pool. The samples were run in duplicate on two VARIANT II TURBO Hemoglobin Testing Systems. Non-significant interference was defined by the sponsor as less than ±10% bias from control. Based upon the data in the study, the sponsor claims no interference at the following concentrations: HbC ≤72%, HbD ≤55%, HbE ≤41%, HbF ≤25%, HbS ≤67% and HbA2 ≤10%. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: To demonstrate the correlation between samples analyzed on devices using CDM Software Versions 5.1 (candidate device) and samples run on devices using CDM Software Version 4.03 (predicate device), 40 EDTA whole blood patient samples, 60 commercially purchased EDTA whole blood patient samples and 16 contrived samples prepared by mixing EDTA whole blood samples with either purified HbA0 or purified HbA1c (8 samples cover 3.5% to 4.5% and 8 samples cover 14.2% to 19.5%) were analyzed by the VARIANT II TURBO HbA1c Kit - 2.0 run on the VARIANT II TURBO Hemoglobin Testing System using both software versions. The samples were run in singlet on the subject device and on the predicate device. The results of the linear regression analysis are presented in the table below. | Reporting Unit | Regression Data | Sample Range | | --- | --- | --- | | %HbA1c (NGSP) | y= 0.99x + 0.06, R²= 0.9996 | 3.5% to 19.5% | | mmol/mol (IFCC) | y= 0.99x + 0.48, R²= 0.9996 | 15 to 190 mmol/mol | b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. {5} c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Hemoglobin A₁c expected value range was cited from American Diabetes Association Standards of Medical Care in Diabetes 2012, 35 (Supplement 1), S11-S63. | Hemoglobin A₁c (%) | Glycemic Goal | | --- | --- | | <8 | Less Stringent | | <7 | General Goal | | <6.5 | More Stringent | | <5.7 | Non-Diabetic Goal | N. Instrument Name: Bio-Rad VARIANT II TURBO Hemoglobin Testing System O. System Descriptions: 1. Modes of Operation: A completely closed, fully automated, high-throughput analyzer consisting of a chromatographic and a sampling station. 2. Software: FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types: Yes ☐ X ☐ or No ☐ 3. Specimen Identification: Samples are identified from the barcode on blood collection test tube. The code is read automatically by the analyzer. The system software (CDM 5.1.1) detects and prevents mismatch of results with sample ID. If a mismatch is detected, the sample result will not be sent to the Laboratory Information System (LIS) and the operator will be notified. 4. Specimen Sampling and Handling: Blood is drawn into normal vacuum test tube. The blood collection tube is placed directly into a sample rack, which is part of the analyzer. A volume of sample is automatically drawn from the tube through its rubber seal. {6} 5. Calibration: The calibrators were previously cleared in k070452. 6. Quality Control: The recommended control materials were cleared in k052838 and k070452. P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: The software documentation was reviewed and it supports that the device was developed under good software cycle processes. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...