The MINICAP HbA1c kit is designed for separation and quantification of the HbA1c glycated fraction of hemoglobin in human whole blood, by capillary electrophoresis in alkaline buffer with the MINICAP FLEX-PIERCING instrument. Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. Results are provided in IFCC (mmol/mol) and NGSP (%Hb A1c) units. The MINICAP Hb A1c kit is designed for Professional Use Only. For In Vitro Use. The Hb A1c CAPILLARY Controls are designed for the quality control of human glycated hemoglobin A1c quantification with SEBIA capillary electrophoresis procedures using the MINICAP HbA1c kit with the MINICAP FLEX-PIERCING automated instrument. The Hb A1c CAPILLARY Controls are designed for Professional Use Only. For In Vitro Diagnostic Use. The HbA1c CAPILLARY Calibrators are designed for the calibration and migration control of human glycated hemoglobin A1c quantification with SEBIA capillary electrophoresis procedures using the MINICAP HbA1c kit with the MINICAP FLEX-PIERCING automated instrument. The HbA1c CAPILLARY Calibrators are designed for Professional Use Only. For In Vitro Diagnostic Use.
Device Story
MINICAP FLEX-PIERCING is an automated analyzer for HbA1c quantification in human whole blood (K2EDTA anticoagulant). System uses capillary electrophoresis in alkaline buffer; separates hemoglobin fractions based on electrophoretic mobility; detects hemoglobins via absorbance at 415 nm. Instrument features 2 parallel silica capillaries; performs automated sample dilution and hemolysis. Used in clinical laboratories by trained personnel. Output includes HbA1c fraction quantification in IFCC (mmol/mol) and NGSP (%) units. High resolution allows quantification in presence of variants (HbS, HbC, HbD, HbE, HbF) and common interferents. Results assist clinicians in monitoring glycemic control in diabetic patients.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated per CLSI EP5-A2; total CVs ranged 0.0-4.0%. Linearity confirmed across 4.8-13.8% HbA1c range. Interference studies showed no significant impact (≤0.3% HbA1c) from triglycerides, bilirubin, ascorbic acid, urea, rheumatoid factor, glybenclamide, carbamylated Hb, labile HbA1c, acetylated Hb, or variants (HbS, E, D, C, F). Method comparison (n=367 samples across internal/external studies) showed high correlation (r=0.998) with predicate.
Technological Characteristics
Capillary electrophoresis in free solution; alkaline buffer; silica capillaries; 415 nm absorbance detection; deuterium lamp and LED light source. Automated sample dilution/hemolysis. Standalone instrument with PC interface. Software-controlled. Standardized to NGSP/IFCC. Dimensions: 44cm x 41.5cm x 58cm; Weight: 32kg.
Indications for Use
Indicated for the separation and quantification of HbA1c in human whole blood to monitor long-term glycemic control in diabetic patients. For professional in vitro diagnostic use.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
Predicate Devices
CAPILLARYS Hb A1c kit (k122101)
CAPILLARYS 2 FLEX-PIERCING instrument (k122101)
Hb A1c CAPILLARY Controls (k122101)
Hb A1c CAPILLARY Calibrators (k122101)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k133344
B. Purpose for Submission:
New Device
C. Measurand:
Whole blood glycated hemoglobin A1c (HbA1c)
D. Type of Test:
Capillary Electrophoresis
E. Applicant:
Sebia, Inc.
F. Proprietary and Established Names:
MINICAP Hb A1c kit
Hb A1c CAPILLARY Calibrators
Hb A1c CAPILLARY Controls
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LCP | II | 21 CFR 864.7470
Glycosylated
hemoglobin assay | Hematology (81) |
| JIS | II | 21 CFR 862.1150
Calibrator | Chemistry (75) |
| JJX | Class I, Reserved | 21 CFR 862.1660
Quality control
material (assayed
and unassayed) | Chemistry (75) |
{1}
H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
The MINICAP HbA1c kit is designed for separation and quantification of the HbA1c glycated fraction of hemoglobin in human whole blood, by capillary electrophoresis in alkaline buffer with the MINICAP FLEX-PIERCING instrument. Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. Results are provided in IFCC (mmol/mol) and NGSP (%Hb A1c) units. The MINICAP Hb A1c kit is designed for Professional Use Only. For In Vitro Use.
The Hb A1c CAPILLARY Controls are designed for the quality control of human glycated hemoglobin A1c quantification with SEBIA capillary electrophoresis procedures using the MINICAP HbA1c kit with the MINICAP FLEX-PIERCING automated instrument.
The Hb A1c CAPILLARY Controls are designed for Professional Use Only. For In Vitro Diagnostic Use.
The HbA1c CAPILLARY Calibrators are designed for the calibration and migration control of human glycated hemoglobin A1c quantification with SEBIA capillary electrophoresis procedures using the MINICAP HbA1c kit with the MINICAP FLEX-PIERCING automated instrument.
- The HbA1c CAPILLARY Calibrators are designed for Professional Use Only. For In Vitro Diagnostic Use.
3. Special conditions for use statement(s):
For Prescription Use Only
- Should not be used for the diagnosis of diabetes mellitus
- Should not be used in monitoring daily glucose control
- Should not be used to replace daily home testing of urine and blood glucose levels.
- Should not be used for analyzing samples from patients with conditions causing shortened red cell blood cell survival such as hemolytic diseases, pregnancy and significant acute or chronic blood loss.
- Not for Point of Care Use
4. Special instrument requirements:
MINICAP Flex-Piercing Instrument
{2}
# I. Device Description:
The MINICAP Hb A1c kits, Hb A1c CAPILLARY Controls and Hb A1c CAPILLARY Calibrators are used with the MINICAP FLEX-PIERCING system. The configurations of the components are summarized:
MINICAP Hb A1c kits in Table I.
- Hb A1c CAPILLARY Calibrators in Table II.
- Hb A1c CAPILLARY Controls in Table III.
- Reagents that are required to perform the test but are sold separately in Table IV.
TABLE I. REAGENTS AND MATERIALS SUPPLIED IN THE MINICAP Hb A1c KIT (PN 2215)
| ITEMS | PN 2215 |
| --- | --- |
| Buffer (ready to use) | 2 vials, 250 mL each |
| Hemolysing solution (ready to use) | 1 vial, 225 mL |
| Wash solution (stock solution) | 1 vial, 25 mL |
| Reagent Cups | 1 pack of 125 |
| Filters | 3 filters |
| Bins for used cups | 4 bins |
| Hemolysing solution bar code labels | 5 sheets of 4 labels |
TABLE II. REAGENTS AND MATERIALS SUPPLIED WITH Hb A1c CAPILLARY CALIBRATORS (PN 4755)
| ITEMS | PN 4755 |
| --- | --- |
| Hb A1c CAPILLARY Calibrator 1 (green cap) | 1 vial of each, 600μL each |
| Hb A1c CAPILLARY Calibrator 2 (red cap) | |
| Barcode label Hb A1c CAPILLARY Calibrator 1 | 1 |
| Barcode label Hb A1c CAPILLARY Calibrator 2 | 1 |
TABLE III. REAGENTS AND MATERIALS SUPPLIED WITH Hb A1c CAPILLARY CONTROLS (PN 4774)
| ITEMS | PN 4744 |
| --- | --- |
| Hb A1c CAPILLARY Control 1 ( white cap ) | 1 vial of each, 600μL each |
| Hb A1c CAPILLARY Control 2 ( black cap) | |
| Barcode label HbA1c CAPILLARY Control 1 | 2 |
| Barcode label HbA1c CAPILLARY | 2 |
{3}
4
| Control 2 | |
| --- | --- |
| White dilution segments* | 4 |
| Grey dilution segments* | 4 |
TABLE IV. REAGENTS AND MATERIALS REQUIRED BUT NOT SUPPLIED IN THE MINICAP HbA1c KIT, Hb A1c CAPILLARY CONTROLS OR Hb A1c CAPILLARY CALIBRATORS
| ITEMS | PN | COMPONENTS |
| --- | --- | --- |
| CAPICLEAN | 2058 | 1 vial, 25 mL |
| CAPILLARYS / MINICAP Wash Solution | 2052 | 2 vials, 75 mL |
| Tubes and caps for controls | 9202, 9205 | 200 per box, 500 per box |
| MINICAP Reagent Cups | 2280 | 250 per box |
| Lids for bins for used reagent cups | 2286 | 12 per box |
| "AUTOMATIC LOW VOLUME" bar code labels | 9208 | 20 per box |
| "MANUAL LOW VOLUME" bar code labels | 9209 | 20 per box |
| MINICAP FLEX-PIERCING centering rings | 1612 | 27 per box |
| PHORESIS software | 1110 | |
| MINICAP FLEX-PIERCING INSTRUMENT | 1232 | |
| Update HbA1c kit for MINICAP FLEX-PIERCING | 1238 | |
J. Substantial Equivalence Information:
1. Predicate device name(s):
CAPILLARYS Hb A1c kit
CAPILLARYS 2 FLEX-PIERCING instrument
Hb A1c CAPILLARY Controls
Hb A1c CAPILLARY Calibrators
2. Predicate K number(s):
k122101
{4}
3. Comparison with predicate:
| Similarities and Differences: Reagent | | |
| --- | --- | --- |
| Item | Predicate Device: CAPILLARYS Hb A1c kit (k122101) | Candidate Device: MINICAP Hb A1c kit |
| Intended Use | Measurement of hemoglobin A1c for monitoring long-term glycemic control in individuals with diabetes mellitus. For Professional Use Only. | Same |
| Method | Free solution capillary electrophoresis | Same |
| Sample type | Whole blood | Same |
| Measuring Range (%HbA1c) | 4.0-14.7% | 4.8-13.8% |
| Collection Tubes | Tubes with K2 and K3 EDTA anticoagulant | Same |
| Absorbance Wavelength | 415 nm | Same |
| Similarities and Differences: Calibrators | | |
| --- | --- | --- |
| Item | Predicate Device: Hb A1c CAPILLARY Calibrators (k122101) | Candidate Device: Hb A1c CAPILLARY Calibrators |
| Intended Use/Indications for Use | The HbA1c CAPILLARY Calibrators are designed for the calibration and migration control of human glycated hemoglobin A1c quantification with SEBIA electrophoresis procedure. | Same |
| Format | 2 levels; 1 vial (0.6 mL) per level | Same |
| Storage Temperature | Before reconstitution, store at -30°C/-18°C (-22°F/0°F). Stable until expiration date on vial. | 5 yrs between -30°C and -18°C |
| In Use Storage | After reconstitution, store at 2-8°C (36-46°F) up to 8 hours, up to 6 months at -22°C/-18°C (-8°F/0°F). Do not freeze and thaw more than 3 times. | CAPILLARYS 2 FLEX-PIERCING: After reconstitution, store at 2-8°C (36-46°F) up to 8 hours, up to 22 months at -22°C/-18°C (-8°F/0°F). Do not freeze and thaw more than 3 times. MINICAP FLEX-PIERCING: After reconstitution, store |
{5}
| Similarities and Differences: Calibrators | | |
| --- | --- | --- |
| Item | Predicate Device: Hb A1c CAPILLARY Calibrators (k122101) | Candidate Device: Hb A1c CAPILLARY Calibrators |
| | | at 2-8°C (36-46°F) up to 8 hours, up to 22 months at -22°C/-18°C (-8°F/0°F). Do not freeze and thaw more than 5 times. |
| Similarities and Differences: Controls | | |
| --- | --- | --- |
| Item | Predicate Device: Hb A1c CAPILLARY Controls (k122101) | Candidate Device: Hb A1c CAPILLARY Controls |
| Intended Use/Indications for Use | The HbA1c CAPILLARY Controls are designed for the quality control of human glycated hemoglobin A1c quantification with SEBIA electrophoresis procedure. | Same |
| Format | 2 levels; 1vial (0.6mL) per level | Same |
| Storage Temperature | Before reconstitution, store refrigerated at 2-8°C (36-46°F). Stable until expiration date on vial. | Same |
| In Use Storage | After reconstitution, store at 2-8°C (36-46°F) up to 8 hours, up to 6 months at -22°C/-18°C (-8°F/0°F). Do not freeze and thaw more than 30 times. After hemolysis with the CAPILLARYS 2 FLEX-PIERCING instrument, store the dilution segments with controls at 2-8°C (36-46°F) up to 8 hours; stable for 1 month at -22°C/-18°C (-8°F/0°F). Do not freeze and thaw a dilution segment with hemolyzed control more than 3 times. | Same |
{6}
K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP09-A2-IR: Method Comparison and Bias Estimation using patient samples, Approved guidelines 2nd edition (Interim Revision) – July 2010
- CLSI EP6-A: Evaluation of the linearity of Quantitative Measurement Procedures: A Statistical Approach, Approved guideline - April 2003
- CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measure Measurement Methods, Approved guideline 2nd edition – August 2004
- CLSI EP7-A2: Interference Testing in Clinical Chemistry, approved guideline 2nd edition – November 2005
- Guidance for Industry and FDA staff- Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use
- Guidance for Industry and FDA staff- General Principals of Software Validation, 2002
- Guidance for Content of Premarket Submissions for Software Contained in Medical Devices, 2005
L. Test Principle:
The MINICAP FLEX-PIERCING instrument uses the principle of capillary electrophoresis in free solution. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Separation also occurs according to the electrolyte pH and electroosmotic flow.
The MINICAP FLEX-PIERCING instrument has silica capillaries functioning in parallel allowing 2 simultaneous analyses for HbA1c quantification from whole blood sample. A sample dilution with hemolysing solution is prepared and injected by aspiration at the anodic end of the capillary. A high voltage protein separation is then performed and direct detection of the hemoglobins is made at the cathodic end of the capillary at 415 nm, which is the absorbance wave length specific to hemoglobins. Before each run, the capillaries are washed with a wash solution and prepared for the next analysis with buffer.
The high resolution of MINICAP Hb A1c procedure allows the quantification of HbA1c, even in the presence of labile HbA1c, carbamylated and acetylated hemoglobins, and major hemoglobin variants such as HbS, HbC, HbD, HbE and HbF and common interfering factors such as Triglycerides, Bilirubin, Ascorbic Acid, Urea, Rheumatoid factor and Glybenclamide as outline in the package insert labeling.
By using alkaline pH buffer, normal and abnormal (or variant) hemoglobins are detected in the following order, from cathode to anode: A2/C, E, S/D, F, A0, other Hb (including minor Hb A1) and then A1c.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{7}
a. Precision/Reproducibility:
Reproducibility within the same capillary and between capillaries from the same instrument
Four different pooled blood samples were analyzed using the CAPILLARYS HbA1c assay in both capillaries of the same MINICAP FLEX-PIERCING instrument. The analyzed EDTA whole blood samples included a pool of HbA1c samples at the normal (~5.2%), cut-off value (~6.4%) and elevated HbA1c (~9.0 and 11.9%)
Each sample was assayed on both capillaries from the same instrument to include 2 runs per day over 20 working days. Samples were analyzed in duplicate within each run. The results are shown below:
Results in NGSP units (%HbA1c)
| Sample | N | Mean (%A1c) | Within-run (%CV) | Within Capillary (%CV) | Between-run (%CV) | Between-day (%CV) | Total (%CV) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample No.1 | 80 | 5.2 | 1.1 | 1.4 | 0.3 | 0.0 | 1.1 |
| Sample No.2 | 80 | 6.4 | 0.9 | 1.1 | 0.0 | 0.6 | 1.1 |
| Sample No.3 | 80 | 9.0 | 0.9 | 0.8 | 0.0 | 0.4 | 1.0 |
| Sample No.4 | 80 | 11.9 | 1.1 | 0.9 | 0.0 | 0.1 | 1.1 |
Results in IFCC units (mmol/mol)
| Sample | N | Mean (mmol/mol) | Within-run (%CV) | Within Capillary (%CV) | Between-run (%CV) | Between-day (%CV) | Total (%CV) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample No.1 | 80 | 34 | 1.9 | 2.8 | 1.4 | 0.0 | 2.3 |
| Sample No.2 | 80 | 46 | 1.1 | 1.3 | 0.5 | 0.5 | 1.3 |
| Sample No.3 | 80 | 75 | 1.2 | 1.4 | 0.0 | 0.6 | 1.3 |
| Sample No.4 | 80 | 106 | 1.2 | 0.9 | 0.0 | 0.2 | 1.2 |
Reproducibility between lots and instruments
Six different EDTA whole blood samples were analyzed using the CAPILLARYS HbA1c assay in both capillaries of 3 different MINICAP FLEX-PIERCING instruments and with 3 lots of MINICAP HbA1c kits. The analyzed blood samples included a sample at the normal (~5.4%), a sample at the cut-off value (~6.4%), and two elevated HbA1c samples (9.1 and 12.3%). Two whole blood control samples were also analyzed (~5.2 and ~7.9% HbA1c).
Each EDTA whole blood sample was analyzed on both capillaries from each instrument, including 60 runs over 10 days, 2 runs per day. Samples were analyzed in duplicate. The results are shown below:
{8}
Results in NGSP units (%HbA1c)
| Sample | Mean (%A1c) | Within-Run reproducibility | | Total reproducibility | |
| --- | --- | --- | --- | --- | --- |
| | | % CV min | % CV max | % CV min | % CV max |
| Sample No.1 | 5.4 | 0.9 | 2.1 | 0.9 | 2.1 |
| Sample No.2 | 6.4 | 0.5 | 1.9 | 0.8 | 1.9 |
| Sample No.3 | 9.1 | 0.0 | 1.1 | 0.0 | 1.1 |
| Sample No.4 | 12.3 | 0.4 | 1.2 | 0.6 | 1.8 |
| Sample No.5 | 5.2 | 0.9 | 2.0 | 0.9 | 2.2 |
| Sample No.6 | 7.9 | 0.7 | 1.4 | 0.8 | 1.6 |
Results in IFCC units (mmol/mol)
| Sample | Mean (mmol/mol) | Within-Run reproducibility | | Total reproducibility | |
| --- | --- | --- | --- | --- | --- |
| | | % CV min | % CV max | % CV min | % CV max |
| Sample No.1 | 36 | 1.1 | 4.0 | 1.2 | 4.0 |
| Sample No.2 | 47 | 1.2 | 2.6 | 1.3 | 2.6 |
| Sample No.3 | 76 | 0.0 | 1.4 | 0.0 | 1.4 |
| Sample No.4 | 110 | 0.4 | 1.4 | 0.6 | 2.2 |
| Sample No.5 | 33 | 0.9 | 3.5 | 0.9 | 3.5 |
| Sample No.6 | 63 | 0.7 | 1.6 | 0.7 | 2.2 |
b. Linearity/assay reportable range:
The linearity study was performed according to the CLSI EP6-A guidelines.
Two EDTA whole blood samples, including a normal sample with HbA1c concentration at 4.8% HbA1c and an elevated HbA1c level sample with HbA1c concentration at 13.8% HbA1c were mixed within different proportions and the dilutions were electrophoresed with the MINICAP Hb A1c procedure for a total of 11 samples. For each dilution, samples were analyzed in duplicate. The tests were determined to be linear within the entire ranges studied for HbA1c hemoglobin fraction (between 4.8 and 13.8% of HbA1c).
NGSP units (%HbA1c)
The 1st order linear regression generated is:
$$
Y = 0.08982x + 4.764, r = 0.999
$$
Linearity range: 4.8-13.8% Hb A1c
IFCC units (mmol/mol)
$$
Y = 0.9855x + 28.41, r = 0.999
$$
Linearity range: 29-127 mmol/mol HbA1c
In addition, linearity of the MINICAP Hb A1c procedure performed with the MINICAP FLEX-PIERCING instrument was verified with different total
{9}
hemoglobin concentrations. 3 different EDTA whole blood samples, including a normal sample with HbA1c concentration at 5.0% HbA1c, a sample with HbA1c level close to the cut-off value at 6.3% HbA1c and an elevated HbA1c level sample with HbA1c concentration at 9.3% HbA1c, were all serially diluted in hemolysing solution and electrophoresed with the MINICAP Hb A1c procedure. The tests were determined to be linear within the entire ranges studied from 2.5 to 31.1 g/dL total hemoglobin and HbA1c fraction concentration and percentage were not affected by the hemoglobin concentration of the samples. 12 mixtures were used per sample.
Results in NGSP units (%HbA1c)
| Sample | Linear Regression | 95%CI |
| --- | --- | --- |
| 1 | Y = 0.001438x+4.778 | 0.000817 to 0.002060
4.722 to 4.834 |
| 2 | Y = 0.001671x+6.039 | 0.000960 to 0.002382
5.974 to 6.104 |
| 3 | Y = 0.00277 x+ 9.004 | 0.00209 to 0.00345
8.937 to 9.070 |
Results in IFCC units (mmol/mol)
| Sample | Linear Regression | 95%CI |
| --- | --- | --- |
| 1 | Y = 0.01698x+28.63 | 0.01036 to 0.02361
28.03 to 29.23 |
| 2 | Y = 0.01969x+42.27 | 0.01219 to 0.02719
41.58 to 42.95 |
| 3 | Y = 0.03035 x+ 74.98 | 0.02204 to 0.03865
74.17 to 75.80 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The MINICAP HbA1c assay is certified with the National Glycohemoglobin Standardization Program (NGSP). The NGSP certification expires in one year. See NGSP website for current certification at http://www.ngsp.org.
Value Assignment: Calibrator and control value assignments are lot-specific. Calibrator and control values are assigned by multiple measurements using the MINICAP FLEX-PIERCING instrument and IFCC traceable calibrators.
Stability: The stability protocols and acceptance criteria were reviewed and determined to be adequate.
The expiration dates of the different components of the CAPILLARYS HbA1c kit are indicated as follows:
{10}
| Kit component | Shelf-Life | On Board Stability |
| --- | --- | --- |
| MINICAP HbA1c buffer: | 2 years at 2-8°C (36-46°F) | 20 days at 15-30°C (59-86°F) |
| MINICAP HbA1c hemolysing solution: | 2 years at 2-30°C (36-86°F) | N/A* |
| MINICAP wash solution: | 3 years at 15-30°C (59-86°F) | 3 months at 37°C (99°F) |
* Hemolysing solution vial not on board.
The expiration date of the Hb A1c CAPILLARY Calibrators is stated as 5 years between $-30^{\circ}\mathrm{C}$ and $-18^{\circ}\mathrm{C}$ $(-22^{\circ}\mathrm{F} / 0^{\circ}\mathrm{F})$ . When reconstituted, the in use storage stability is 8 hours at $2 - 8^{\circ}\mathrm{C}$ $(36 - 46^{\circ}\mathrm{F})$ , up to 22 months at $-22^{\circ}\mathrm{C} / -18^{\circ}\mathrm{C}$ $(-8^{\circ}\mathrm{F} / 0^{\circ}\mathrm{F})$ . The labeling states that the Hb A1c CAPILLARY Calibrators should not be frozen and thawed more than 5 times.
The stability of the Hb A1c CAPILLARY Controls is demonstrated to be stable for 8 hours at $2 - 8^{\circ}\mathrm{C}$ (36-46°F), up to 6 months at $-22^{\circ}\mathrm{C} / -18^{\circ}\mathrm{C}$ (-8°F/0°F). The labeling states that the HbA1c CAPILLARYS controls should not be frozen and thawed more than 30 times.
# d. Detection limit:
The Limit of Blank (LoB) was determined by assaying five blood samples without HbA1c. The Limit of Detection (LoD) was determined by assaying six blood samples with low HbA1c. Both LoB and LoD were tested according to CLSI guideline EP17-A. The results are as follows: $\mathrm{LoB} = 0.3\%$ , $\mathrm{LoD} = 1.1\%$
The claimed measuring range, $4.8 - 13.8\%$ , is based on linearity. See 1b above.
# e. Analytical specificity:
i.) Studies were performed to assess common or known substances that could interfere with the CAPILLARYS HbA1c assay kit. The interfering substances were evaluated in whole blood samples that contained four different concentrations of A1c ( $\sim 4.9\%$ , $\sim 6.5\%$ , $\sim 8.8\%$ and $\sim 12.1\%$ ). Samples containing various concentrations of potential interferents were tested and the results compared to those obtained from control samples containing no potential interfering substances. The sponsor's definition of non-significant interference is $\leq 0.3\%$ HbA1c between the tested and the control samples.
The results are as follows:
| Potential interfering substance | Concentration at which no significant interference (≤0.3%) was observed |
| --- | --- |
| Bilirubin | ≤25.8 mg/dL (442 μM) |
{11}
| Triglycerides | ≤3070 mg/dL (35.1 mM) |
| --- | --- |
| Rheumatoid Factor | ≤2178 IU/mL |
| Urea | ≤291 mg/dL (48.5 mM) |
| Ascorbic Acid | ≤60 mg/dL (3.41 mM) |
| Glybenclamide | ≤3 mg/dL |
ii.) To study interference from Carbamlyated hemoglobin, four blood samples with A1c concentrations at $\sim 5.7\%$ , $\sim 6.9\%$ , $\sim 8.9\%$ and $12.4\%$ were split into two aliquots. One aliquot, at each A1c level, was spiked with $1\mathrm{mmol / L}$ of Potassium Cyanate and incubated for 3 hours at $37^{\circ}\mathrm{C}$ . Another aliquot, at each A1c level, was incubated for 3 hours at $37^{\circ}\mathrm{C}$ . Samples were then analyzed on the MINICAP FLEX-PIERCING instrument using the CAPILLARYS HbA1c assay kit. Samples were analyzed in triplicate. The sponsor's definition of non-significant interference is $\leq 0.3\mathrm{HbA1c}\%$ between the tested and the control samples.
The sponsor concluded that Carbamylated hemoglobin (at $\leq 8.11\mathrm{mg / dL}$ (1mmol/L) does not interfere with this assay.
iii.) To study interference from labile hemoglobin, four blood samples with A1c concentrations at $\sim 4.7\%$ , $\sim 6.8\%$ , $\sim 8.8\%$ and $\sim 12.7\%$ were split into two aliquots. One aliquot, at each A1c level, was spiked with glucose (0.5mol/L) and incubated for 3 hours at $37^{\circ}\mathrm{C}$ . Another aliquot, at each A1c level, was incubated for 3 hours at $37^{\circ}\mathrm{C}$ . Samples were then analyzed on the MINICAP FLEX-PIERCING instrument using the CAPILLARYS HbA1c assay kit. Samples were tested in triplicate. The sponsor's definition of non-significant interference is $\leq 0.3$ HbA1c%.
The sponsor concluded that labile A1c (at $\leq 1800\mathrm{mg / dL}$ (0.5mol/L) does not interfere with this assay.
iv.) To study interference from acetylated hemoglobin, four blood samples with A1c concentrations at $\sim 5.2\%$ , $\sim 6.6\%$ , $\sim 9.4\%$ , and $\sim 11.7\%$ were split into two aliquots. One aliquot, at each A1c level, was used as the control sample and the other aliquot was spiked with acetylated hemoglobin (10 mmol/L) and incubated for 4 hours at $37^{\circ}\mathrm{C}$ . All aliquots were tested on the MINICAP FLEX-PIERCING instrument using the CAPILLARYS HbA1c assay kit. Samples were tested in triplicate. The sponsor's definition of non-significant interference is $\leq 0.3$ HbA1c%.
The sponsor concluded that acetylated hemoglobin (at $\leq 180\mathrm{mg / dL}$ (10mmol/L)) does not interfere with this assay.
v) A hemoglobin variant interference study was carried out using samples known to contain Hemoglobin variants S, E, D and C. These variant samples were tested on the MINICAP FLEX-PIERCING instrument using the
{12}
13
CAPILLARYS HbA1c assay kit. The sponsor’s definition of non-significant interference is ±10% difference between the candidate method and the comparative method.
The testing results show there is no significant interference for HbS (≤ 40.5%), HbE (≤ 24.7%), HbD (≤ 41.0%) and HbC (≤ 37.0%).
vi) An additional variant interference study was carried out to study the variant interference from Hemoglobin F. The study was performed on 16 different samples with HbA1c values between 5.3 and 11.6%. The samples contained concentrations of Hb F from 2.3 to 19.7% and were tested on the MINICAP FLEX-PIERCING instrument using the CAPILLARYS HbA1c assay kit.
The testing results show there is no significant interference for Levels of HbF ≤ 15%, where significant interference is defined by ±10% difference between the candidate method and the comparative method. The following limitation regarding Hb F interference is included in the labeling:
Levels of Hb F up to 15% in the blood sample do not interfere with HbA1c fraction quantification, but no HbA1c result will be reported by the software when Hb F level is higher than 15%. Samples that contain high amounts of Hb F (> 15%), usually found in some people with thalassemia, infants, and some pregnant women, may yield a lower than expected HbA1c result with this assay.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
An internal and external method comparison study was conducted using a total of 367 EDTA whole blood samples ranging from 4.7-14.6%HbA1c. Testing was completed at three sites, with the following samples tested at each site:
Site 1 (Internal): 101 blood samples (35 normal and 66 with elevated HbA1c
Site 2 (external): 126 blood samples (71 normal and 55 with elevated Hb A1c and
Site 3 (external): 140 blood samples (68 normal and 72 with elevated Hb A1c)
Normal Hb A1c was defined as < 6.5%. No samples were spiked or diluted. Samples were analyzed in singlicate using the MINICAP HbA1c procedure on the MINICAP FLEX-PIERCING instrument (candidate device) and the CAPILLARYS 2 FLEX-PIERCING instrument (predicate device). Method comparison studies were performed according to CLSI EP9-A2 guideline. The
{13}
linear regression correlation was calculated as follows:
Site 1 (Internal):
| HbA1c | Correlation Coefficient | y-intercept (95%CI) | Slope (95%CI) | HbA1c range |
| --- | --- | --- | --- | --- |
| Percentage (%) | 0.998 | 0.165 (0.053 to 0.276) | 0.982 (0.968 to 0.996) | 4.8 – 13.3 |
| Concentration (mmol/mol) | 0.998 | 1.262 (0.375 to 2.149) | 0.985 (0.972 to 0.999) | 29 - 122 |
Site 2:(External)
| HbA1c | Correlation Coefficient | y-intercept (95%CI) | Slope (95%CI) | HbA1c range |
| --- | --- | --- | --- | --- |
| Percentage (%) | 0.998 | -0.032 (-0.107 to 0.044) | 0.997 (0.987 to 1.007) | 4.8 – 13.6 |
| Concentration (mmol/mol) | 0.998 | -0.396 (-1.027 to 0.235) | 0.996 (0.986 to 1.006) | 29 - 125 |
Site 3 (External)
| HbA1c | Correlation Coefficient | y-intercept (95%CI) | Slope (95%CI) | HbA1c range |
| --- | --- | --- | --- | --- |
| Percentage (%) | 0.998 | -0.057 (-0.127 to 0.013) | 1.019 (1.009 to 1.029) | 4.8 – 13.1 |
| Concentration (mmol/mol) | 0.998 | -0.316 (-0.848 to 0.216) | 1.023 (1.013 to 1.032) | 29 - 119 |
# b. Matrix comparison:
A total of 41 different blood samples (17 normal and 24 with elevated HbA1c) were collected in parallel on K2 EDTA and K3 EDTA tubes and tested on the MINICAP FLEX-PIERCING instrument using the CAPILLARYS HbA1s assay kit. No samples were spiked or diluted. The linear regression is presented in the table below:
| Fraction | Number of Samples | Correlation Coefficient | y-intercept | Slope | Range of Hb A1c fractions % values (test) |
| --- | --- | --- | --- | --- | --- |
| 1 | 1 | 0.998 | 0.000 | 0.999 | 0.999 to 1.000 |
| 2 | 2 | 0.998 | 0.000 | 0.999 | 0.999 to 1.000 |
{14}
| HbAc1% | 41 | 0.999 | 0.002 | 1.004 | 5.0-13.4 |
| --- | --- | --- | --- | --- | --- |
| HbA1c concentration mmol/mol | 41 | 0.999 | 0.022 | 1.003 | 31 - 123 |
The sponsor concluded that whole blood samples can be collected in tubes containing K2 EDTA or K3 EDTA for analysis with MINICAP Hb A1c procedure performed on MINICAP FLEX-PIERCING instrument.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Reference Range
| | NGSP | IFCC |
| --- | --- | --- |
| Increased risk for diabetes | ≥6.5% | ≥48mmol/mol |
$^{1}$ American Diabetes Association. Standards of medical care in diabetes-2012. Diabetes Care. 2012 Jan;35 Suppl 1:S11-63
N. Instrument Name:
MINICAP FLEX-PIERCING Instrument
O. System Descriptions:
1. Modes of Operation:
The MINICAP FLEX-PIERCING instrument is a multiparameter instrument for hemoglobins analysis on parallel capillaries. The hemoglobins assay uses 2 capillaries to run the samples.
The sequence of automated steps is as follows:
- Bar code reading of rotating sampler and sample tubes (for up to 26 tubes);
- Mixing of blood samples before analysis;
{15}
- Sample hemolysis and dilution from primary tubes into reagent cups;
- Capillary washing;
- Injection of hemolyzed samples;
- Hemoglobin separation and direct detection of the separated hemoglobins on capillaries.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
☐ Yes ☑ or ☐ No
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
☐ Yes ☑ or ☐ No
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
☑ Yes ☐ or ☐ No
3. Specimen Identification:
Barcode reader
4. Specimen Sampling and Handling:
Use EDTA whole blood samples. Check that all the tubes contain a minimum sample of 1 mL and are closed with their corresponding caps designed for the MINICAP Hb A1c procedure with the MINICAP FLEX-PIERCING. Vortex for 5 seconds blood samples stored at 2 – 8 °C for one week or stored at – 80 °C.
5. Calibration:
6. Quality Control:
It is necessary to perform a quality control analyses with Hb A1c CAPILLARY Controls 1 or 2, after capillaries activation, after each calibration of the instrument performed with the Hb A1c CAPILLARY Calibrators, after a capillary cleaning sequence with CAPICLEAN and before starting a new analysis sequence. In addition, for high-volume testing laboratories, it is advised to analyze one of the two controls after the analysis of a complete rotating sampler, by alternating Hb A1c CAPILLARY Control 1 and Control 2.
16
{16}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.