MODIFICATION TO: DIRECT ENZYMATIC HBA1C ASSAY KIT, MODEL# DZ168A, DZ168A-CAL, DZ168A-CON
Applicant
Diazyme Laboratories
Product Code
LCP · Hematology
Decision Date
Apr 9, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.7470
Device Class
Class 2
Indications for Use
Diazyme Direct Enzymatic Hemoglobin A1c (glycated hemoglobin A1c; A1c; HbAlc) reagents are intended for use in the quantitative determination of stable HbAlc in human whole blood samples. Measurement of hemoglobin A1c is a valuable indicator for long-term diabetic control. For in-vitro diagnostic use only.
Device Story
The Diazyme Direct Enzymatic HbA1c Assay is an in-vitro diagnostic reagent kit used for the quantitative measurement of stable HbA1c in human whole blood. The assay utilizes an enzymatic method to determine HbA1c levels, which serves as a clinical indicator for long-term glycemic control in diabetic patients. The device is intended for professional use in clinical laboratory settings. By providing accurate HbA1c measurements, the assay assists healthcare providers in assessing a patient's average blood glucose levels over the preceding weeks to months, thereby informing clinical decisions regarding diabetes management and treatment adjustments.
Clinical Evidence
No clinical data provided in the document; the submission relies on the demonstration of substantial equivalence to legally marketed predicate devices for in-vitro diagnostic use.
Technological Characteristics
In vitro diagnostic reagent system. Modifications: reagent count reduced from 4 to 3; component concentration adjusted; calculation algorithm updated to NGSP format. Design controls per 21 CFR 820.30 applied.
Indications for Use
Indicated for quantitative determination of stable HbA1c in human whole blood samples to monitor long-term diabetic control. For in-vitro diagnostic use.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
Related Devices
K031042 — ATAC HEMOGLOBIN A1C REAGENT KIT · Clinical Data, Inc. · Nov 3, 2003
K020532 — POLYMEDCO HBA1C TEST · Polymedco, Inc. · Apr 24, 2002
K091711 — S- TEST HEMOGLOBIN ALC (HBA1C), MODEL RC0023 · Alfa Wassermann Diagnostic Technologies, Inc. · Dec 11, 2009
K031539 — HEMOGLOBIN A1C REAGENT SET · Pointe Scientific, Inc., · Aug 6, 2003
K160762 — Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis), Diazyme Direct HbA1c Assay Calibrator Set, Diazyme Direct HbA1c Assay Control Set · Diazyme Laboratories · Nov 21, 2016
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER k070734
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k050178
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for change in the number of reagents supplied (from 4 to 3), change in concentration of component in reagent, change in calculation equation to direct reporting in %HbA1c values in the NGSP format, and labeling changes to reflect the modifications.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and precision testing, accuracy testing, linearity testing.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis - FMECA
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.