← Product Code [LCP](/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/LCP) · K040872

# VARIANT II HEMOGLOBIN A1C (K040872)

_Bio-Rad Laboratories, Inc. · LCP · Apr 15, 2004 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-h%E2%80%94hematology-kits-and-packages/LCP/K040872

## Device Facts

- **Applicant:** Bio-Rad Laboratories, Inc.
- **Product Code:** [LCP](/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/LCP.md)
- **Decision Date:** Apr 15, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7470
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Bio-Rad VARIANT II TURBO Hemoglobin A₁c Program is intended for the percent determination of hemoglobin A₁c in human whole blood using ion-exchange high performance liquid chromatography (HPLC). For In Vitro Diagnostic Use. The Bio-Rad VARIANT II TURBO Hemoglobin A₁c Program is intended for Professional Use Only. For In Vitro Diagnostic Use. Measurement of percent hemoglobin A₁c is effective in monitoring long-term glucose control in individuals with diabetes mellitus.

## Device Story

The VARIANT II TURBO Hemoglobin A1c Program is an in vitro diagnostic system for professional use in clinical settings. It utilizes ion-exchange high-performance liquid chromatography (HPLC) to analyze human anticoagulated whole blood (EDTA) samples. The system separates hemoglobin A1c from other hemoglobin components via a cation exchange cartridge. Detection occurs at 415 nm. The device provides a quantitative measurement of percent hemoglobin A1c, which healthcare providers use to monitor long-term glucose control in diabetic patients. The system is standardized to the Diabetes Control and Complications Trial (DCCT) reference method and the IFCC, and is certified by the National Glycohemoglobin Standardization Program (NGSP).

## Clinical Evidence

No clinical trials were performed. Substantial equivalence was established via bench testing, including precision (n=80 per sample type), linearity (n=2 per standard), and method comparison (n=201 patient samples) against the predicate. Regression analysis showed r²=0.9946, slope=0.9792, and intercept=0.185. Interference studies confirmed no impact from bilirubin, lipids, or excess EDTA.

## Technological Characteristics

Ion-exchange HPLC system; uses polymeric cation exchange resin-packed analytical and guard cartridges. Energy source: electrical. Connectivity: computer-based data management. Software: Clinical Data Management. Standards: NGSP/IFCC traceable; electrical safety per EN61010-1, EN61010-2-010, EN61010-2-101, EN61326. Dimensions: modular 3-unit system (Sampling Station, Chromatographic Station, Computer).

## Regulatory Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

## Predicate Devices

- VARIANT II Hemoglobin A₁c Program (k984268)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 10

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number:
k040872

B. Purpose of the Submission:
Improvements over the predicate chromatographic device known as the VARIANT II Hemoglobin A₁c Program [k984268]

C. Analyte:
Glycosylated hemoglobin (Hemoglobin A₁c or HbA₁c)

D. Type of Test:
Quantitative

E. Applicant:
Bio-Rad Laboratories, Inc.

F. Proprietary and Established Names:
VARIANT II TURBO Hemoglobin A₁c Program

G. Regulatory Information:
1. Regulatory Section: 21 CFR 864.7470 [Assay, glycosylated hemoglobin]
2. Classification: II
3. Product Code: LCP
4. Panel: 81 - Hematology

H. Intended Use:
1. Intended Use(s): The Bio-Rad VARIANT II TURBO Hemoglobin A₁c Program is intended for the percent determination of hemoglobin A₁c in human whole blood using ion-exchange high performance liquid chromatography (HPLC). For In Vitro Diagnostic Use.
2. Indication(s) for Use: The Bio-Rad VARIANT II TURBO Hemoglobin A₁c Program is intended for the percent determination of hemoglobin A₁c in human whole blood using ion-exchange high performance liquid chromatography (HPLC). The Bio-Rad VARIANT II TURBO Hemoglobin A₁c Program is intended for Professional Use Only. For In Vitro Diagnostic Use. Measurement of percent hemoglobin A₁c is effective in monitoring long-term glucose control in individuals with diabetes mellitus.
3. Special condition for use statement(s): NA
4. Special Instrument Requirements: VARIANT II TURBO Hemoglobin Testing System

{1}

Page 2 of 10

# I. Device Description :

This VARIANT II TURBO Hemoglobin A₁c Program is a testing system that uses the principles of HPLC for the chromatographic separation of Hemoglobin A₁c present in human blood. The kit contains supplies for 2000 tests and consists of an analytical cartridge, guard cartridge, elution buffers, wash/diluent Solution, calibrators, calibrator diluent and whole blood primer.

# J. Substantial Equivalence Information:

1. Predicate Device Name: VARIANT II Hemoglobin A₁c Program
2. Predicate K number(s) k984268
3. Comparison with predicate:

## Similarities

|  Characteristic Item or Parameter | VARIANT II TURBO Hemoglobin A₁c | VARIANT II Hemoglobin A₁c  |
| --- | --- | --- |
|  Analyte Measured: Reported | % Hemoglobin A₁c | % Hemoglobin A₁c  |
|  Intended Use | The Bio-Rad VARIANT II TURBO Hemoglobin A₁c Program is intended for percent determination of HbA₁c in whole blood using ion-exchange high performance liquid chromatography (HPLC). The Bio-Rad VARIANT II TURBO HbA₁c Program is intended for Professional Use only. For In Vitro Diagnostic Use. | The Bio-Rad VARIANT II Hemoglobin A₁c (HbA₁c) Program is intended for percent determination of HbA₁c in human whole blood using ion-exchange high performance liquid chromatography (HPLC).  |
|  Assay Principle with Visible Light Absorbance Detection | Cation exchange high performance liquid chromatography (HPLC) with detection by Absorbance at 415 nm wavelength | Cation exchange high performance liquid chromatography (HPLC) with detection by Absorbance at 415 nm wavelength  |
|  Sample Type | Human anticoagulated whole blood (EDTA) | Human anticoagulated whole blood (EDTA)  |
|  Sample Volume | 23 μL | 23 μL  |
|  Use of Controls | Bilevel Control for A₁c per run. | Bilevel Control for A₁c per run.  |
|  Standardization | Traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC. Certified via the National Glycohemoglobin Standardization Program (NGSP). | Traceable to Diabetes Control and Complications Trial (DCCT) reference method and IFCC. Certified via the National Glycohemoglobin Standardization Program (NGSP).  |
|  Measurement Type | Quantitative area percent | Quantitative area percent  |
|  Design | Modular system that is designed for a medium size laboratory. | Modular system that is designed for a medium size laboratory.  |
|  Instrumentation | VARIANT II TURBO is one system with 3 discrete units; VARIANT Sampling Station (VSS), VARIANT Chromatographic Station (VCS) and computer. | VARIANT II is one system with 3 discrete units; VARIANT Sampling Station (VSS), VARIANT Chromatographic Station (VCS) and computer.  |

{2}

Differences

|  Characteristic Item/ Parameter | VARIANT II TURBO Hemoglobin A1c | VARIANT II Hemoglobin A1c  |
| --- | --- | --- |
|  Analytes Reported | Hemoglobin A1a, A1b, F, AO, S, C, CHb LA1c, P3 and VARIANT window. | Hemoglobin A1a, A1b, F, AO, C, LA1c, P3 and E, D, S window.  |
|  Calibration | 2 point calibration, once every new analytical or guard cartridge | 2 point calibration, once every new analytical cartridge.  |
|  Reagents | Analytical cartridge, Guard cartridge, Elution buffers, Wash/Diluent Solution, calibrators, calibrator diluent and whole blood primer. | Analytical cartridge, Elution buffers, Wash/Diluent Solution, calibrators, calibrator diluent, whole blood primer and pre-filter.  |
|  Temperature | 36°C | 28°C  |
|  Standards met & Electrical Safety issues met | Specific NGSP, IFCC Standards and FDA [09/30/1991] Guidance Document for 510(k) Submission of Glycohemoglobin (Glycated or Glycosylated) Hemoglobin for IVDs; and FDA Guidance for the Software Contained in Medical Devices[05/29/1998]; New: Off-the-Shelf Software Use in Medical Devices [09/09/1999]; Electrical: EN61010-1: 2001, EN61010-2-010, EN61010-2-101, EN61326/A2:2001, EN61326:1997. | Specific NGSP, IFCC Standards & FDA [09/30/1991] Guidance Document for 510(k) Submission of Glycohemoglobin (Glycated or Glycosylated) Hemoglobin for IVDs; and FDA Guidance for the Software Contained in Medical Devices [05/29/1998]; Electrical: EN61010-1: 2001, EN61010-2-010, EN61010-2-101, EN61326/A2:2001, EN61326:1997.  |
|  Analysis Medium | Bis-Tris/Phosphate Buffers A & B and modified polymeric cation exchange based resin packed analytical and guard cartridges. | Bis-Tris/Phosphate Buffer A & B, polymeric cation exchange based resin packed cartridge.  |
|  New design feature | New column holder for the VCS that accepts analytical and guard cartridges to give 2000 injections per analytical cartridge. | Cartridge holder has both a prefilter assembly and holder. The cartridge gives 1000 injections per analytical cartridge.  |
|  New design feature | Redesigned column heater to hold new cartridge holder and to give better temperature calibration. | Column heater holds Brownlee design cartridge holder and prefilter.  |
|  New design feature | Coil mixer has replaced the T-mixer to improve mixing and reduce gradient delay. | T- mixer and dynamic mixer.  |
|  New design feature | 2-groove sample probe allows simultaneous venting and sampling of patient tubes for faster speed. | Stainless Steel Sample Probe.  |
|  New design feature | No mixing of patient sample on the TURBO required. | Patient sample mixing is performed by the instrument.  |
|  Time to process sample | 97 seconds, providing for a doubling of throughput of blood samples | 3 minutes  |
|  Media to Update HbA1c kit information | CD ROM operating at 24x-48x and using software compatible for the VARIANT II TURBO Hemoglobin Testing System. | CD ROM operating at 24x-48x and using software compatible for the VARIANT II Hemoglobin Testing System. [K984268]  |

{3}

Page 4 of 10

## K. Standard/Guidance Document Referenced (if applicable):

Standards and Guidelines used to evaluate the VARIANT II TURBO Hemoglobin A₁c Program device included: a) NGSP (National Glycohemoglobin Standardization Program), IFCC HbA₁c Calibrators; b) FDA [09/30/1991] Guidance Document for 510(k) Submission of Glycohemoglobin (Glycated or Glycosylated) Hemoglobin for IVDs; c) FDA Guidance for Software Contained in Medical Devices [05/29/1998]; d) New: Off-the-Shelf Software Use in Medical Devices [09/09/1999].

## L. Test Principle

This VARIANT II TURBO Hemoglobin A₁c Program utilizes principles of ion-exchange high-performance liquid chromatography (HPLC). The samples are automatically diluted on the Sampling Station and injected into the analytical cartridge. The Chromatographic Station dual pumps deliver a programmed buffer gradient of increasing ionic strength to the cartridge, where the hemoglobins are separated based on their ionic interactions with the cartridge material. The absorbance at 415 nm is measured. An additional filter at 690 nm corrects the background absorbance. The Clinical Data Management software performs reduction of raw data collected from each analysis. Two-level calibration is used for adjustment of the calculated values

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. Precision/Reproducibility - The following table provides comparison data on the precision between VARIANT II TURBO Hemoglobin A₁c and VARIANT II Hemoglobin A₁c Programs, each utilizing EDTA whole blood patient samples, and both tested against samples with moderate (5.4-6.2%) and high (12.5-13.7%) Hemoglobin A₁c content. The method of precision analysis was performed using a protocol based on the NCCLS Evaluation protocol, Vol.12, No. 4, EP5-A (Feb. 1999) applied to the VARIANT II TURBO Hemoglobin A₁c, and NCCLS Evaluation protocol, Vol.12, No. 4, EP5-T2 (Mar.1992) applied to the VARIANT II Hemoglobin A₁c Program. The protocols for both the VARIANT II TURBO Hemoglobin A₁c and VARIANT II Hemoglobin A₁c Programs are similar. Using these protocols, 40 runs (2 per day) were performed on one VARIANT II TURBO (or VARIANT II) Hemoglobin Testing System over 20 working days. In each duplicate daily run, one aliquot of low HbA₁c and one aliquot of high HbA₁c patient samples were each analyzed per run. Although the precision samples are different, since they were run at different time periods, the precision results between the VARIANT II TURBO Hemoglobin A₁c and the VARIANT II Hemoglobin A₁c Program are substantially equivalent. A summary of these combined comparative precision results is presented in the precision table below.

{4}

Page 5 of 10

VARIANT II TURBO Hemoglobin A₁c and VARIANT II Hemoglobin A₁c Precision

|   | VARIANT II TURBO Hemoglobin A1c |   | VARIANT II Hemoglobin A1c  |   |
| --- | --- | --- | --- | --- |
|   | Low Patient (HbA₁c) | High Patient (HbA₁c) | Low Patient (HbA₁c) | High Patient (HbA₁c)  |
|  n= (number of samples) | 80 | 80 | 80 | 80  |
|  Mean | 6.2 | 12.5 | 5.4 | 13.7  |
|  Within run | 0.82% CV | 0.54% CV | 1.46 % CV | 0.65 % CV  |
|  Total Precision | 1.94% CV | 2.58 % CV | 2.14 % CV | 1.68 % CV  |

b. Linearity/assay reportable range:

The following table provides comparison data on the linearity and recovery analyses between VARIANT II TURBO Hemoglobin A₁c and VARIANT II Hemoglobin A₁c Programs, each utilizing eight EDTA-based blood standards (n=2 for each standard). The % Recovery for Hemoglobin A₁c by the VARIANT II TURBO Hemoglobin A₁c Program is essentially the same as the VARIANT II Hemoglobin A₁c Program. The standard values chosen for the linearity studies represent % Hb A₁c recovery values for normal [non-diabetic] patients, who might range from low [i.e., between 3.5–3.8 %] to moderate [i.e., between 4.7 -5.0 %] values of HbA₁c, as well as recovery values for abnormal [diabetic] patients, who might range from slightly high [i.e., above the range 5.8-6.3 %] to extremely high [i.e., between 14.1-17.3 %] values of Hb A₁c. These comparative recovery results are presented in the following linearity table.

VARIANT II TURBO Hemoglobin A₁c and VARIANT II Hemoglobin A₁c Linearity

|   |   | VARIANT II TURBO Hemoglobin A₁c |   |   | VARIANT II Hemoglobin A₁c  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  % Contribution |   | Theoretical % HbA₁c | Observed % HbA₁c | % Recovery | Theoretical % HbA₁c | Observed % HbA₁c | % Recovery  |
|  Normal | Diabetic |  |  |  |  |  |   |
|  100 | 0 | 3.8 | 3.8 | 100 | 3.5 | 3.5 | 100  |
|  90 | 10 | 5.0 | 5.0 | 100 | 4.7 | 4.7 | 100  |
|  80 | 20 | 6.3 | 6.1 | 96.8 | 5.9 | 5.8 | 98.3  |
|  67 | 33 | 8.0 | 7.9 | 98.8 | 7.6 | 7.4 | 97.4  |
|  50 | 50 | 10.2 | 10.0 | 97.9 | 9.8 | 9.6 | 98.0  |
|  33 | 67 | 12.5 | 12.4 | 98.0 | 12.1 | 11.9 | 98.3  |
|  20 | 80 | 14.4 | 14.3 | 99.3 | 14.1 | 13.8 | 97.9  |
|  0 | 100 | 17.3 | 17.3 | 100 | 17.2 | 17.2 | 100  |

c. Traceability (controls, calibrators, or method):

The Hemoglobin A₁c Calibrators provided with the VARIANT II TURBO Hemoglobin A₁c Program are traceable to the approved IFCC reference method. The IFCC reference method is used to assign IFCC values to the secondary reference materials. These secondary reference materials are used to value assign product calibrators and determine the product calibration parameters though use of the IFCC/NGSP master equation.

Master equation: (NGSP A₁c) = 0.9148 (IFCC A₁c) + 2.152
(IFCC A₁c) = 1.093 (NGSP A₁c) - 2.350

{5}

The NGSP values were assigned by the relationship between IFCC and NGSP. The VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program is certified by the NGSP.

# d. Detection Limit:

Based on pooled and individual patient-derived blood standards, the lower limit of linear detection for this VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program was  $4.1\%$ $\mathrm{HbA_{1c}}$  as listed in the Instructions for Use (IFU). The upper limit of linear detection for this VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program was  $16.8\%$ $\mathrm{HbA_{1c}}$ . This linear or detection range was chosen to span all typical diagnostic testing that would occur for normal, diabetic, and extremely diabetic patients.

# e. Analytical Specificity:

Three closely related but chemical derived analogs of  $\mathrm{HbA_{1c}}$  where analytically evaluated using this VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program as part of a detailed analytical specificity study. The influence of carbamylated hemoglobin was studied by spiking specimens with sodium cyanate until the carbamylated hemoglobin levels increased to a range of  $1 - 1.5\%$ . The influence of acetylated hemoglobin was studied by spiking specimens with acetylsalicylic acid until acetylated hemoglobin levels increased to a range of  $1 - 2\%$ . The influence of labile hemoglobin  $\mathrm{A_{1c}}$  was studied by spiking samples with glucose until labile A1c in hemoglobin reached  $1 - 3\%$ . Thus, final measurement of  $\mathrm{HbA_{1c}}$  in these blood-based human specimens was not influenced by carbamylated, acetylated or labile hemoglobin  $\mathrm{A_{1c}}$  at the above indicated limits.

Additional low, moderate and high blood samples were obtained as patient bloods that were anticoagulated with EDTA in the standard manner. In three separate trials of patient pools or individual blood samples, concentrated bilirubin was added to a final level of  $20\mathrm{mg / dL}$ ; concentrated lipids were added to a final level between 5680 and  $6000\mathrm{mg / dL}$ ; and additional dipotassium EDTA was added to a concentration of  $\sim 1980\mathrm{mg / dL}$  (11x the normal level) to determine the effect of high concentrations of EDTA that can occur in cases of "short draws." The table below shows that, for the final measurement of  $\mathrm{Hb - A_{1c}}$ , neither the TURBO VARIANT II nor cleared predicate VARIANT II device was influenced by these excess biochemicals or excess EDTA anticoagulant, as illustrated in the following table.

|  Interfering Substance | VARIANT II TURBO Hemoglobin A1c | VARIANT II Hemoglobin A1c  |
| --- | --- | --- |
|  Bilirubin | No interference up to 20 mg/dL | No interference up to 20 mg/dL  |
|  Lipids (Triglycerides) | No interference up to 5680 mg/dL | No interference up to 6000 mg/dL  |
|  EDTA | No interference up to 11X EDTA | No interference up to 11X EDTA  |

# f. Assay Cut-off/Decision Level:

This VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program spans the typical cut-off limits (generally at  $&lt; 6.0\%$ $\mathrm{HbA_{1c}}$  for normal, and well above the suggested diabetic action level at  $&gt;8.0\%$ $\mathrm{HbA_{1c}}$ ).

{6}

# 2. Comparison studies:

a. Method Comparison &amp; Correlation Analyses with Predicate for Accuracy:

Method correlation between VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program and VARIANT II Hemoglobin  $\mathrm{A_{1c}}$  Program was evaluated using  $n = 201$  EDTA whole blood patient samples that ranged from  $3.9\%$  to  $17.5\%$ $\mathrm{HbA_{1c}}$ . The coded patient blood specimens used in this study were blood samples that remained after earlier clinical results were reported. The results of this non-clinical correlation study are presented in the following regression table.

|  Regression Method | n | r2 | Slope | Intercept  |
| --- | --- | --- | --- | --- |
|  Least Squares | 201 | 0.9946 | 0.9792 | 0.185  |

By this comparison, a linear regression equation was obtained as:  $\mathrm{Y} = 0.9792\mathrm{X} - 0.185$ , and an  $\mathrm{r}^2 = 0.9946$ , showing close correlation and substantial equivalence between the new VARIANT II TURBO (Y) device and the predicate VARIANT II (X) device.

b. Matrix Comparison:

EDTA whole blood is the only sample type indicated.

# 3. Clinical studies:

a. Clinical sensitivity:

NA

b. Clinical specificity

NA

c. Other clinical supportive data (when  $a$  and  $b$  are not applicable)

NA

# 4. Clinical cut-off

NA

# 5. Expected value/Reference Range:

The Expected Value range was established from the article Rohlfing et al, published in the reference article entitled "Use of GHb (HbA1c) in screening for Undiagnosed Diabetes in the U.S. Population". This study used the Bio-Rad DIAMAT Hemoglobin  $\mathrm{A_{1c}}$  Program. The recommended weighted mean  $\mathrm{HbA_{1c}}$  for patients with normal fasting plasma glucose  $(n = 5,694)$  was  $5.17\%$  with the standard deviation of  $0.45\%$ , while the  $95\%$  confidence limits (mean  $\pm 2\mathrm{SD}$ ) were  $4.27\% - 6.07\mathrm{HbA_{1c}}$ . The VARIANT II TURBO Hemoglobin  $\mathrm{A_{1c}}$  Program matched this expected result range and was so certified by the NGSP.

# N. Conclusion:

Based upon a Third Party Review of the information provided in this 510(k), the submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-h%E2%80%94hematology-kits-and-packages/LCP/K040872](https://fda.innolitics.com/submissions/CH/subpart-h%E2%80%94hematology-kits-and-packages/LCP/K040872)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
