SYMPHONY GLUCOSE MANAGEMENT SYSTEM

K080639 · Animas Corp. · LZG · Jun 24, 2008 · General Hospital

Device Facts

Record IDK080639
Device NameSYMPHONY GLUCOSE MANAGEMENT SYSTEM
ApplicantAnimas Corp.
Product CodeLZG · General Hospital
Decision DateJun 24, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Symphony System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood. The Symphony Pump is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes. The Symphony Meter Remote is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood, and as a wireless (RF) remote control to deliver insulin from the Symphony Pump. The Symphony Meter Remote is intended for use for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home and by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The Symphony Meter Remote is specifically indicated for use on the finger, forearm or palm. It should not be used for the diagnosis of diabetes or testing of newborns.

Device Story

Symphony Glucose Management System combines insulin pump and blood glucose meter. Pump delivers insulin subcutaneously; meter measures capillary blood glucose. Devices communicate via bi-directional RF. Meter remote controls bolus insulin delivery, displays pump status, and confirms alerts. Used by patients at home or healthcare professionals in clinical settings. Meter includes food database and insulin calculator. Pump functions as standalone device if RF communication is unavailable. System aids diabetes management by integrating glucose monitoring with insulin delivery control.

Clinical Evidence

Performance evaluation included both bench and clinical testing. No new issues of safety or efficacy were raised. Specific clinical metrics (sensitivity, specificity, AUC) were not provided in the summary.

Technological Characteristics

System comprises an insulin infusion pump and a blood glucose meter. Features bi-directional RF communication. Pump includes color display and insulin calculator. Meter uses OneTouch Ultra Test Strips and Control Solution. Supports data download to PC for management systems. Standalone operation capability for both components.

Indications for Use

Indicated for insulin-requiring diabetes patients for continuous subcutaneous insulin infusion and quantitative blood glucose monitoring. Meter remote indicated for finger, forearm, or palm testing. Contraindicated for newborn testing and diabetes diagnosis.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary # JUN 2 4 2008 | SUBMITTER: | Animas Corporation<br>200 Lawrence Drive<br>West Chester, PA 19380<br>Contact: Amy Smith; Director, Regulatory Affairs<br>& Change Management | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | DEVICE NAME: | Symphony Glucose Management System | | COMMON OR USUAL NAME: | Insulin infusion pump and glucose test system | | DEVICE CLASSIFICATION: | Class II per CFR 21 §880.5725, Infusion Pump,<br>product code LZG<br>Class II per CFR 21 §862.1345, Glucose Test<br>System, product code NBW<br>Devices used with the Glucose Test System are classified as follows:<br><ul><li>Class I, OneTouch® Ultra Control Solution, per 21 CFR § 862.1660.</li><li>Class I (exempt) OneTouch® Lancing Device with OneTouch® AST ClearCap™ and OneTouch® UltraSoft® Sterile Lancets per 21 CFR § 878.4800</li><li>Class II, OneTouch® Ultra Test Strips, per 21 CFR § 862.1345</li></ul> | | PREDICATE DEVICES: | Animas® IR 1250 insulin pump (K042873)<br>OneTouch® UltraSmart® Blood Glucose Monitoring System (K021819, K043197)<br>OneTouch® Ultra® 2 Blood Glucose Monitoring System (K053529)<br>Medtronic MiniMed Paradigm Model 512 Insulin Pump and Paradigm Link Glucose Meter (K030531, K040603)<br>Insulet OmniPod iXL-II Diabetes Management System (K042792) | Animas Corporation {1}------------------------------------------------ #### DEVICE DESCRIPTION: The Symphony System combines the use of an insulin pump and a blood glucose meter for the treatment of insulin requiring diabetes. The insulin pump, Symphony Pump, operates as a stand alone device to deliver insulin through an infusion set placed subcutaneously. The blood glucose meter, Symphony Meter Remote, operates as a stand alone device to measure blood glucose concentrations. When paired through radio frequency (RF) communication the Symphony Meter Remote can be used to control bolus insulin delivery, review the status of the Symphony Pump and view and confirm selected pump alerts and warnings. #### INTENDED USE: The Symphony System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood. The Symphony Pump is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes. The Symphony Meter Remote is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood, and as a wireless (RF) remote control to deliver insulin from the Symphony Pump. The Symphony Meter Remote is intended for use for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home and by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The Symphony Meter Remote is specifically indicated for use on the finger, forearm or palm. It should not be used for the diagnosis of diabetes or testing of newborns. #### SUBSTANTIAL EQUIVALENCE: The Symphony System is similar to the Paradigm Model 512 Insulin Pump and Paradigm Link Glucose Meter in that both systems are comprised a pump and meter that can be operated independently as stand alone devices. The main difference between the Symphony System and the Paradigm Model 512 Insulin Pump and Paradigm Link Glucose Meter is that the communication for the Symphony System is bi-directional which allows for the display of pump information on the meter remote screen and allows for the meter remote to control bolus insulin delivery of the pump. In the Paradigm Model 512 Insulin Pump and Paradigm Link Glucose Meter the only communication is one-direction (meter to pump) for the communication of a blood glucose value. The Symphony System is similar to the Insulet OmniPod iXL-II Diabetes Management System in that both systems allow for bi-directional RF communication. This allows pump information to be displayed on the meter remote and for the control of insulin delivery. These systems also have the same intended use. The Symphony System is different to the Insulet OmniPod iXL-II Diabetes Management System in that the pump in the Symphony System can be used as a stand alone device if the meter remote is not available or RF communication is not possible (such as on an aircraft). In the Insulet {2}------------------------------------------------ OmniPod iXL-II Diabetes Management System the pump is a pod that contains no user interface. The Symphony Pump is most similar to the Animas IR 1250. The Symphony Pump is not a new design, but a modification to the Animas IR1250 insulin pump. The differences between the Symphony Pump and the Animas IR1250 insulin pump are the color display, increased number of insulin calculator settings, RF capability, size, weight and the additional alarms associated with RF capability. The Symphony Meter Remote is a modification to the OneTouch UltraSmart Blood Glucose Monitoring System (K021819). The modifications included both hardware and software changes to implement the requirements for its intended use as a remote control for the Symphony insulin pump. It is also similar to the OneTouch® Ultra®2 Blood Glucose Monitoring System (K053529). A food database and insulin calculator feature have been added to the meter. Like the OneTouch UltraSmart and OneTouch Ultra 2 Blood Glucose monitoring systems, the Symphony Meter Remote uses the OneTouch Ultra Test Strips and Control Solution. Both the subject and the predicate devices store blood glucose results and allow for download to a PC for use with a data management system. #### RESULTS OF PERFORMANCE EVALUATION: Performance evaluations, both bench and clinical, of the Symphony System were completed and did not raise any new issues of safety and efficacy. #### CONCLUSION: The Symphony Glucose Management System is substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the top half of the circle. Inside the circle is an abstract image of an eagle. JUN 2 4 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Amy Smith Dircctor, Regulatory Affairs & Change Management Animas Corporation 200 Lawrence Drive West Chester, Pennsylvania 19380 Re: K080639 Trade/Device Name: Symphony Glucose management System Regulation Number: 880,5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: June 3, 2008 Reccived: June 4, 2008 Dear Ms. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Smith Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Runner fd Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): Device Name: Symphony Glucose Management System Indications for Use: The Symphony System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood. The Symphony Pump is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes. The Symphony Meter Remote is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood, and as a wireless (RF) remote control to deliver insulin from the Symphony Pump. The Symphony Meter Remote is intended for use for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home and by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The Symphony Meter Remote is specifically indicated for use on the finger, forearm or palm. It should not be used for the diagnosis of diabetes or testing of newborns. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Clint Verr (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K980639 Page_1_of_1_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...