VITROS Chemistry Products hsCRP Reagent

K160712 · Ortho Clinical Diagnostics · DCK · Jul 14, 2017 · Immunology

Device Facts

Record IDK160712
Device NameVITROS Chemistry Products hsCRP Reagent
ApplicantOrtho Clinical Diagnostics
Product CodeDCK · Immunology
Decision DateJul 14, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5270
Device ClassClass 2

Intended Use

VITROS Chemistry Products hsCRP Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System to quantitatively measure C-reactive protein (CRP) in human serum and plasma. CRP is used to evaluate conditions thought to be associated with inflammation in otherwise healthy individuals.

Device Story

VITROS Chemistry Products hsCRP Reagent is an immunoturbidimetric assay for quantitative measurement of C-reactive protein (CRP) in human serum and plasma. The device utilizes two liquid reagents: Reagent 1 (buffering solution) and Reagent 2 (latex microparticles coated with anti-CRP mouse monoclonal antibodies). Samples are mixed with Reagent 1, followed by Reagent 2, triggering an immunochemical reaction that forms CRP antigen/antibody complexes. The resulting turbidity is measured spectrophotometrically at 660 nm. The system calculates CRP concentration using a stored calibration curve and measured absorbance. The assay is performed on VITROS 5,1 FS, 4600, and 5600 Chemistry Systems. Healthcare providers use the quantitative output to assess inflammatory conditions. The device is intended for in vitro diagnostic use in clinical laboratory settings.

Clinical Evidence

Bench testing only. Precision was evaluated using three human serum pools (approx. 1, 10, 14 mg/L) across three systems, showing total imprecision <3% CV. Linearity was confirmed from 0.34 to 15.0 mg/L. LoB (0.21 mg/L), LoD (0.26 mg/L), and LoQ (0.34 mg/L) were established per CLSI EP17-A2. Method comparison with the predicate (N=119) yielded a correlation coefficient of 0.99 (Y = 1.02X + 0.26).

Technological Characteristics

Latex-enhanced immunoturbidimetric assay; two-point rate measurement. Reagents are liquid, ready-to-use, dual-chambered. Sensing via spectrophotometry at 660 nm. Traceable to ERM-DA474/IFCC. Designed for use on VITROS 5,1 FS, 4600, and 5600 automated chemistry analyzers.

Indications for Use

Indicated for the quantitative measurement of C-reactive protein (CRP) in human serum and plasma to evaluate conditions associated with inflammation in otherwise healthy individuals.

Regulatory Classification

Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a cascading manner. The profiles are black and are meant to represent the department's focus on people and health. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 14, 2017 # ORTHO-CLINICAL DIAGNOSTICS, INC. MARLENE HANNA, SENIOR MANAGER, REGULATORY AFFAIRS 100 INDIGO CREEK DRIVE ROCHESTER, NY 14626 Re: K160712 Trade/Device Name: VITROS Chemistry Products hsCRP Reagent Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: II Product Code: DCK Dated: June 9, 2017 Received: June 12, 2017 Dear Marlene Hanna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, **Kellie B. Kelm -S** for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K160712 Device Name VITROS Chemistry Products hsCRP Reagent Indications for Use (Describe) VITROS Chemistry Products hsCRP Reagent: For in vitro diagnostic use only. Rx Only. VITROS Chemistry Products hsCRP Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System to quantitatively measure C-reactive protein (CRP) in human serum and plasma. CRP is used to evaluate conditions thought to be associated with inflammation in otherwise healthy individuals. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary: k160712 | VITROS Chemistry Products hsCRP Reagent: A summary of 510(k) safety and effectiveness | |---------------------------------------------------------------------------------------| | information in accordance with the requirements of 21 CFR 807.92. | | Submitter Information | | |------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | Ortho-Clinical Diagnostics, Inc. | | Address | 100 Indigo Creek Drive<br>Rochester, New York 14626 | | Phone number | 585-453-4041 | | Fax number | 585-453-3368 | | Establishment<br>Registration | 1319681 | | Name of contact person | Marlene Hanna | | Date prepared | July 13, 2017 | | Trade or proprietary<br>name | VITROS Chemistry Products hsCRP Reagent | | Common or usual name | C-reactive protein immunological test system | | Classification name | Cardiac c-reactive protein, antigen, antiserum | | Classification panel | Immunology | | Regulation | 21 CFR 866.5270: C-Reactive Protein immunological test system<br>Classification: Class II | | Product Code(s) | DCK | | Legally marketed<br>device(s) to which<br>equivalence is claimed | The Diazyme high sensitivity C-reactive protein (hsCRP) assay<br>(k103557) | | Device description | The quantitative measurement of C-reactive protein (CRP) is<br>performed using the VITROS Chemistry Products hsCRP Reagent in<br>conjunction with the VITROS Chemistry Products Calibrator Kit 17<br>and VITROS Chemistry Products FS Calibrator 1 on the VITROS 5,1<br>FS/4600 Chemistry System and the VITROS 5600 Integrated<br>System. The VITROS Chemistry Products hsCRP Reagent is a dual<br>chambered package containing ready-to-use liquid reagents. Samples,<br>calibrators and controls are mixed with Reagent 1 containing a<br>buffer. Addition of anti-CRP antibodies coupled to latex<br>microparticles (Reagent 2) produces an immunochemical reaction<br>yielding CRP antigen/antibody complexes. The turbidity is measured<br>spectrophotometrically at 660 nm. Once a calibration has been<br>performed for each reagent lot, the CRP concentration in each<br>unknown sample can be determined using the stored calibration curve<br>and the measured absorbance obtained in the assay of the sample. | {4}------------------------------------------------ | Intended Use | VITROS hsCRP Reagent: | |--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | For in vitro diagnostic use only. Rx ONLY. | | | VITROS Chemistry Products hsCRP Reagent is used on the<br>VITROS 5,1 FS Chemistry System, the VITROS 4600<br>Chemistry System and the VITROS 5600 Integrated System to<br>quantitatively measure C-reactive protein (CRP) in human serum and<br>plasma. CRP is used to evaluate conditions thought to be associated<br>with inflammation in otherwise healthy individuals. | ## Comparison with Predicate Devices: ### Table 1: VITROS Chemistry Products hsCRP Reagent | Characteristic | Predicate (Diazyme high sensitivity<br>C-Reactive Protein: k103557) | New Device (VITROS hsCRP<br>Reagent (Modified)) | |------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For in vitro diagnostic use only. Rx<br>ONLY.<br>The Diazyme high sensitivity C-<br>reactive protein (hsCRP) assay is for<br>the in vitro quantitative determination<br>of C-reactive protein (CRP) in human<br>serum and plasma on automated<br>clinical chemistry analyzers.<br>Measurement of CRP is of use for the<br>detection and evaluation of<br>inflammatory disorders and<br>associated diseases, infection and<br>tissue injury. | For in vitro diagnostic use only. Rx<br>ONLY.<br>VITROS Chemistry Products hsCRP<br>Reagent is used<br>to quantitatively measure C-reactive<br>protein (CRP) in human serum and<br>plasma. CRP is used to evaluate<br>conditions thought to be associated with<br>inflammation in otherwise healthy<br>individuals. | | Basic Principle | Latex enhanced immunoturbidimetric<br>assay | Two-point rate, 2mmune-turbidimetric<br>assay | | Sample type | Serum, plasma | Serum, plasma | | Assay Range Serum,<br>Plasma | 0.20 to 20.0 mg/L | 0.34 to 15.00 mg/L | ### Performance Summary: Substantial Equivalence was demonstrated by testing several performance characteristics including method comparison, precision, reference interval, linearity and detection limit. ### Method Comparison: Method Comparison testing followed CLSI Protocol EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples. Approved Guideline - Third Edition. A total of 119 human serum samples were tested. The 119 samples were measured with VITROS hsCRP assay on the VITROS 5600, 4600, and 5,1 Systems and a predicate method, Diazyme {5}------------------------------------------------ hsCRP assay on the Hitachi 917 Systems. Of the 119 samples tested 113 were within the measuring range of both the VITROS hsCRP assay (0.34-15.0 mg/L) and the Diazyme hsCRP assay (0.20 - 20.0 mg/L). The relationship between the VITROS hsCRP method and the Predicate method (Diazyme hsCRP) based on data from 113 samples on the VITROS 5600 Chemistry System is as follows: VITROS 5600 Integrated System = 1.02 x Diazyme hsCRP Method + 0.26 The relationship between the VITROS hsCRP method on the VITROS 4600 System and the VITROS 5600 System based on data from 110 samples is as follows; VITROS 4600 Integrated System = 1.02 x VITROS 5600 System + 0.01 The relationship between the VITROS hsCRP method on the VITROS 5.1 FS System and the VITROS 5600 System based on data from 109 samples is as follows; VITROS 5,1 FS System = 1.07 x VITROS 5600 System + 0.11 The data demonstrates acceptable correlation was obtained between the VITROS hsCRP assay (traceable to European Reference Material ERM®-DA474/IFCC) versus the predicate method (traceable to European Reference Material ERM®-DA472/IFCC) and between VITROS Systems. # Precision: The method was based upon the Clinical and Laboratory Standards Institute (CLSI) Protocol EP05-A3, Evaluation of Precision of Ouantitative Measurement Procedures; Approved Guideline-Third Edition, October 2014. Five replicates of three human serum sample pools were run once per day for five days using two hsCRP Reagent lots and two Calibrator Kit 17 lots across three VITROS Chemistry Systems. Testing was conducted using VITROS Chemistry Products hsCRP Reagent calibrated with VITROS Chemistry Products Calibrator Kit 17 and VITROS Chemistry Products FS Calibrator 1 traceable to ERM-DA474. Precision study results for the VITROS hsCRP assay are listed in table 1. {6}------------------------------------------------ Table 1: Precision study | | Reagent<br>Mean | | Within-run | | Between-run | | Total | | |-----------------|-----------------|-------|------------|------|-------------|------|-------|------| | Analyzer | Lot | Conc. | SD | %CV | SD | %CV | SD | %CV | | 5,1 FS<br>PM675 | 35-5544 | 0.96 | 0.067 | 6.96 | 0.000 | 0.00 | 0.067 | 6.96 | | | | 9.65 | 0.104 | 1.08 | 0.080 | 0.83 | 0.132 | 1.37 | | | | 13.41 | 0.156 | 1.16 | 0.081 | 0.60 | 0.176 | 1.31 | | | 36-5726 | 1.02 | 0.060 | 5.90 | 0.024 | 2.36 | 0.065 | 6.39 | | | | 9.89 | 0.151 | 1.53 | 0.124 | 1.25 | 0.196 | 1.98 | | | | 13.86 | 0.266 | 1.92 | 0.134 | 0.97 | 0.298 | 2.15 | | 4600<br>PM102 | 35-5544 | 0.97 | 0.063 | 6.53 | 0.000 | 0.00 | 0.063 | 6.53 | | | | 9.52 | 0.188 | 1.97 | 0.000 | 0.00 | 0.188 | 1.97 | | | | 13.12 | 0.269 | 2.05 | 0.067 | 0.51 | 0.277 | 2.11 | | | 36-5726 | 1.01 | 0.047 | 4.66 | 0.021 | 2.08 | 0.052 | 5.16 | | | | 9.79 | 0.253 | 2.58 | 0.144 | 1.47 | 0.291 | 2.97 | | | | 13.64 | 0.400 | 2.93 | 0.078 | 0.57 | 0.408 | 2.99 | | 5600<br>PM109 | 35-5544 | 0.92 | 0.053 | 5.77 | 0.000 | 0.00 | 0.053 | 5.77 | | | | 9.34 | 0.081 | 0.87 | 0.081 | 0.87 | 0.115 | 1.23 | | | | 12.78 | 0.126 | 0.99 | 0.110 | 0.86 | 0.167 | 1.31 | | | 36-5726 | 1.00 | 0.054 | 5.40 | 0.000 | 0.00 | 0.054 | 5.40 | | | | 9.54 | 0.113 | 1.18 | 0.121 | 1.27 | 0.165 | 1.73 | | | | 13.26 | 0.278 | 2.10 | 0.148 | 1.12 | 0.315 | 2.38 | # Expected values/Reference range: The reference interval was verified according to CLSI EP28-A3 Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition; a 50% male and 50% female population was utilized. Reference intervals may differ for each population studied and therefore each laboratory should confirm the validity of these intervals for the population it serves. Increases in CRP values are non-specific and should be interpreted together with a complete clinical history. Follow-up testing of patients with elevated CRP values should be performed. | Conventional and SI Units (mg/L) | |----------------------------------| | $<5.0$ | {7}------------------------------------------------ # Linearity: Linearity was determined following CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: a Statistical Approved Guideline. A high CRP serum pool was intermixed with a low serum pool to generate 13 concentration levels each tested in three replicate determinations. Linear results were compared to 2m and 3w order polynomial fits against a pre-specified allowable error. The linearity range was found to extend across the measuring range of 0.34 to 15 mg/L. ## Limits of blank, detection, and quantitation: The Limit of Blank (LoB) for VITROS Chemistry Products hsCRP Reagent is 0.21 mg/L. The Limit of Detection (LOD) is 0.26 mg/L. The Limit of Quantitation (LOQ) is 0.34 mg/L. All results were determined according to the guidelines listed in CLSI EP17-A2 and the guidance for Industry and FDA Staff Review Criteria for Assessment of High Sensitivity C-Reactive Protein (hsCRP) issued September 22, 2005. | LoB* | LoD** | LoQ*** | |-----------|-----------|-----------| | 0.21 mg/L | 0.26 mg/L | 0.34 mg/L | * Limit of Blank: The highest result to be reported for a blank sample with a 95% level of confidence (o. = 0.05). ** Limit of Detection: The amount of analyte where 95% of measurement results (ß = 0.05) exceed the Limit of Blank. *** Limit of Quantitation: The minimum amount of analyte that can be quantitatively determined based on acceptable precision. # Conclusion: The conclusions drawn from the nonclinical tests (discussed above) demonstrate the modified VITROS Chemistry Products hsCRP Reagent is as safe, effective, and perform as well as the predicate device. The information submitted in the premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...