← Product Code [DFC](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFC) · K982708

# SPQ TEST SYSTEM (K982708)

_DiaSorin, Inc. · DFC · Apr 2, 1999 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-f%E2%80%94immunological-test-systems/DFC/K982708

## Device Facts

- **Applicant:** DiaSorin, Inc.
- **Product Code:** [DFC](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFC.md)
- **Decision Date:** Apr 2, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5600
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The SPQTM Antibody Reagent Set for Lp(a) is designed for the quantitative determination of human lipoprotein(a) in human serum by immunoprecipitin analysis using a turbidimetric clinical analyzer. The measurement of Lp(a) is indicated for use in conjunction with clinical evaluation, patient risk assessment and other lipoprotein tests to evaluate disorders of lipid metabolism and to assess coronary heart disease (CHD) in male Caucasian populations.

## Device Story

SPQ Antibody Reagent Set performs quantitative determination of human lipoprotein(a) in serum; utilizes immunoprecipitin analysis on automated turbidimetric clinical analyzers (e.g., Cobas FARA II). Process: samples/controls/polymeric enhancer pipetted into cuvettes; initial incubation/blank measurement; antiserum added; reaction at 37°C forms insoluble antigen-antibody complexes; turbidity increases light scatter proportional to Lp(a) concentration. Absorbance measured after 10-minute incubation. Calibration curve generated via known standards; instrument data reduction interpolates sample concentrations. Used in clinical laboratory settings by trained personnel. Output assists clinicians in lipid metabolism evaluation and CHD risk assessment.

## Clinical Evidence

Clinical study compared Lp(a) levels in normal males (median 11.6 mg/dL) vs. males with atherosclerotic heart disease (median 26.3 mg/dL), showing statistically significant difference. Assay correlated with NWLRL reference method (r=0.962) and Macra assay (r=0.961). Recovery study confirmed equivalent recovery across Lp(a) polymorphs (419-796 kD) using dilution parallelism. Reproducibility tested at 3 sites over 2 weeks (n=42); total %CV ranged from 3.6% to 13.9% across assay range.

## Technological Characteristics

Immunoprecipitin turbidimetric assay; requires automated clinical analyzer (e.g., Cobas FARA II). Reaction temperature controlled at 37°C. Measures light scatter/absorbance of antigen-antibody complexes. Reagents include antibody set and polymeric enhancer. Not for manual use.

## Regulatory Identification

A low-density lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the low-density lipoprotein in serum and other body fluids. Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify young persons at risk from cardiovascular diseases.

## Predicate Devices

- Strategic Diagnostics Macra Lp(a)

## Reference Devices

- Northwest Lipid Research Laboratory (NWLRL) reference method

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR-31-99 WED 04:43 PM

AMERICANSTANDARDDIASORIN

31-99 WED 04:43

4/2/99

FAX NO. +301 497 1126 P. 04
K982708

## 510(K) SUMMARY

SUBMITTED BY:

Judith J. Smith DiaSorin, Inc. 9175 Guilford Rd. Suite 100 Columbia, MD 21046

NAME OF DEVICES: Trade Name:

SPQ Test System Antibody Reagent Set for Lp(a)

Immunoprecipitin assay for quantitative

Common Names/Descriptions:

Classification Names:

Strategic Diagnostics Macra Lp(a)

determination of Lp(a) levels

Lipoprotein a (Lp(a)) test

PREDICATE DEVICES:

DEVICE DESCRIPTION:

INTENDED USE: The SPQTM Antibody Reagent Set for Lp(a) is designed for the quantitative determination of human lipoprotein(a) in human serum by immunoprecipitin analysis using a turbidimetric clinical analyzer. The measurement of Lp(a) is indicated for use in conjunction with clinical evaluation, patient risk assessment and other lipoprotein tests to evaluate disorders of lipid metabolism and to assess coronary heart disease (CHD) in male Caucasian populations.

KIT DESCRIPTION: The SPQ™ Test System for Lp(a) provides the quantitative determination of human lipoprotein(a) by automated immunoprecipitin analysis. Standards, controls, and samples are pipetted undiluted into sample cups. Microvolumes of these samples and a polymeric enhancer are pipetted into individual cuvettes. Following an initial incubation and measurement of sample blank, undiluted antiserum is added to the cuvettes. The sample and antiserum are mixed in the reaction cuvettes. Reaction temperature is controlled at 37°C. Insoluble antibody complexes form immediately, producing turbidity in the mixture and increasing the amount of light scattered by the solution. The amount of antigen-antibody complex formed, and thus the amount of light scatter, is proportional to the amount of lipoprotein(a) in the initial sample. The solution absorbance is measured after a 10 minute incubation period.

A calibration curve is generated by analyzing a series of calibrators with known concentrations of lipoprotein(a) and using the instrument's data reduction capability or manually plotting the change in absorbance versus lipoprotein(a) concentration. Concentrations of lipoprotein(a) within the controls and samples are interpolated from the calibration curve.

The test is designed for use on instruments capable of immunoprecipitin analysis, as the Cobas FARA II. The instruments measure the amount of light scattering in the reaction cuvettes due to the formation of insoluble antigen-antibody complexes. The systems are capable of storing a calibration curve. This assay is not designed for manual use.

{1}------------------------------------------------

PERFORMANCE DATA: All performance data were collected using the Cobas FARA II. Clinical Correlation: A clinical study was conducted with the SPQ Test System to assess whether the assay could detect a difference in Lo(a) levels between a normal population with known cardiac risk factors and a population with known atherosclerotic heart disease. The median Lp(a) level for the normal male population was 11.6 mg/dL, while the median level for the diseased male population was 26.3 mg/dL. The populations were statistically significantly different.

The assay was compared to the Northwest Lipid Research Laboratory reference method and to the Strategic Diagnostics Macra assay. The correlation coefficient to the NWLRL assay was 0.962 (equivalent to the correlation of the Macra assay to the NWLRL method) and the correlation to the Macra assay was 0.961.

Recovery: A study was performed to evaluate the quantitative recovery of different Lp(a) polymorphs which spanned the molecular weight range from 419 to 796 kD. A model of dilution parallelism was used to test whether the different polymorphs were recovered equivalently. Statistical analysis of the data obtained from this study showed that no departure from parallelism was observed.

Reproducibility: Reproducibility studies were performed at 3 sites using one lot of SPQ reagents. Assay reproducibility was determined by testing 6 samples that spanned the range of the assay as well as the 2 control samples. Samples were tested in duplicate once a day for 7 days over a 2 week period. The results are summarized below.

| Sample          | Mean<br>mg/dL | Within<br>Run %CV | Total<br>%CV |
|-----------------|---------------|-------------------|--------------|
| Level 1 Control | 22.3          | 4.0               | 5.4          |
| Level 2 Control | 52.0          | 2.7               | 4.9          |
| RP 1            | 9.9           | 9.4               | 13.9         |
| RP 2            | 21.9          | 3.6               | 7.7          |
| RP 3            | 45.2          | 3.1               | 6.1          |
| RP 4            | 58.4          | 5.3               | 8.6          |
| RP 5            | 71.9          | 2.6               | 4.9          |
| RP 6            | 85.5          | 2.0               | 3.6          |
| N               | 42            |                   |              |

## REPRODUCIBILITY RESULTS FOR SPQ LP(A) ASSAY - COMBINED SITES

{2}------------------------------------------------

Public Health Service

APR 2 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Judith J. Smith DiaSorin, Inc. 9175 Guilford Road, Suite 100 Ouarry Park Place Columbia, Maryland 21046 Re: K982708 Trade Name: SPO Test System Antibody Reagent Set for Lp(a) Regulatory Class: II Product Code: DFC

> II I Dated: March 4, 1999 Received: March 5, 1999

Dear Ms. Smith:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

IIS

JJX

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Image /page/2/Picture/9 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

## 510(k) NUMBER: K982708

DEVICE NAME: Antibody Reagent Set for Lp(a) SPQ™ Test System

INDICATIONS FOR USE: The SPQTM Antibody Reagent Set for Lp(a) is designed for the quantitative determination of human lipoprotein(a) in human serum by immunoprecipitin analysis using a turbidimetric clinical analyzer. The measurment of Lp(a) is Indicated for use in conjunction with clinical evaluation, patient risk assessment and other lipoprotein tests to evaluate disorders of lipid metabolism and to assess coronary heart disease (CHD) in male Caucasian populations.

K982708 Jun Cooper

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K982708

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-f%E2%80%94immunological-test-systems/DFC/K982708](https://fda.innolitics.com/submissions/CH/subpart-f%E2%80%94immunological-test-systems/DFC/K982708)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
