← Product Code [MRZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ) · K193654

# BlueStar Rx (K193654)

_Welldoc, Inc. · MRZ · Apr 7, 2020 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ/K193654

## Device Facts

- **Applicant:** Welldoc, Inc.
- **Product Code:** [MRZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ.md)
- **Decision Date:** Apr 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Software as a Medical Device, Therapeutic

## Indications for Use

The BlueStar® Rx is indicated for use by healthcare providers (HCPs) and their patients – aged 18 years and older - who have type 1 or type 2 diabetes. The BlueStar® Rx is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The BlueStar® Rx analyzes and reports blood glucose test results and supports medication adherence. In addition, the BlueStar® Rx provides coaching messages (motivational, behavioral, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information. - For bolus insulin users with type 1 and type 2 diabetes, BlueStar® Rx includes an insulin dose calculator to allow patients to use their prescribed regimen to calculate a dose of bolus insulin for a given amount of carbohydrates and/or blood glucose value. - For basal insulin users with type 2 diabetes, BlueStar® Rx includes an Insulin Adjustment Program (IAP) which calculates appropriate long-acting basal insulin doses for titrating insulin levels based on configuration by a healthcare provider knowledgeable in the care and management of diabetes. The healthcare provider must activate the Insulin Adjustment Program and configure it for patient-specific parameters. The BlueStar® Rx is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.

## Device Story

BlueStar® Rx is a prescription-only software system for diabetes self-management; operates on mobile phones and PCs. Inputs include user-entered blood glucose data, carbohydrate intake, and data from connected glucose meters (e.g., OneTouch Verio Flex) via Bluetooth. Device features an Insulin Adjustment Program (IAP) for basal insulin titration and a bolus insulin calculator. HCPs configure titration parameters via a web portal; patients follow generated dosing guidance on the mobile app. System provides motivational, behavioral, and educational coaching based on glucose trends. CGM data may be visualized but is not used for dose calculations. Device assists patients in managing insulin regimens and medication adherence; output informs patient dosing decisions. Intended for home and professional healthcare settings.

## Clinical Evidence

No clinical trials were conducted. Substantial equivalence is supported by human factors usability studies demonstrating that representative patients and healthcare providers can safely perform critical tasks associated with the new basal insulin titration feature.

## Technological Characteristics

Standalone software system; mobile and web-based interfaces. Connectivity via Bluetooth for glucose meter data and API for CGM data visualization. Software-based insulin dose calculation and titration logic. Major level of concern software. Operates on mg/dL glucose values.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- WellDoc® BlueStar® and WellDoc® BlueStar® Rx System ([K190013](/device/K190013.md))
- Glooko Mobile Insulin Dosing System (MIDS) ([K171450](/device/K171450.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

INSTRUMENT ONLY

## I Background Information:

A 510(k) Number

K193654

B Applicant

WellDoc, Inc

C Proprietary and Established Names

BlueStar® Rx

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  MRZ | Class II | 21 CFR 880.5725 - Infusion Pump | General Hospital  |
|  NDC | Class II | 21 CFR 868.1890 - Predictive pulmonary-function value calculator | Clinical Chemistry  |
|  LNX | Unclassified |  |   |

## II Submission/Device Overview:

A Purpose for Submission:

Addition of a new functionality to an existing device.

B Type of Test:

Insulin dose (basal and bolus) calculation and titration.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

III Intended Use/Indications for Use:

A Intended Use(s):
See Indications for Use below.

B Indication(s) for Use:
The BlueStar® Rx is indicated for use by healthcare providers (HCPs) and their patients – aged 18 years and older - who have type 1 or type 2 diabetes. The BlueStar® Rx is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The BlueStar® Rx analyzes and reports blood glucose test results and supports medication adherence. In addition, the BlueStar® Rx provides coaching messages (motivational, behavioral, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.

- For bolus insulin users with type 1 and type 2 diabetes, BlueStar® Rx includes an insulin dose calculator to allow patients to use their prescribed regimen to calculate a dose of bolus insulin for a given amount of carbohydrates and/or blood glucose value.
- For basal insulin users with type 2 diabetes, BlueStar® Rx includes an Insulin Adjustment Program (IAP) which calculates appropriate long-acting basal insulin doses for titrating insulin levels based on configuration by a healthcare provider knowledgeable in the care and management of diabetes. The healthcare provider must activate the Insulin Adjustment Program and configure it for patient-specific parameters.

The BlueStar® Rx is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.

C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only

BlueStar® Rx is not indicated for people with gestational diabetes or who use an insulin pump. Basal Insulin Titration is not for people who are treated with bolus insulin.

The BlueStar® Insulin Adjustment Program is designed to work with glucose results reported in mg/dL and cannot safely be used with mmol/L values. To avoid harm please ensure that any glucose meter used with the BlueStar® Insulin Adjustment Program system is set for blood glucose values in mg/dL.

D Special Instrument Requirements:
Not applicable

K193654 - Page 2 of 6

{2}

K193654 - Page 3 of 6

## IV Device/System Characteristics:

### A Device Description:

BlueStar® Rx is a modified version of the primary predicate (WellDoc BlueStar® Rx cleared under K190013). BlueStar® Rx maintains all of the features of the primary predicate and adds the titration of long-acting basal insulin doses for type 2 diabetes patients who are not using bolus insulin. WellDoc BlueStar® Rx, a prescription only device cleared under K190013, is a stand-alone software system intended to be used by healthcare providers (HCPs) and their patients – aged 18 years and older – who have type 1 or type 2 diabetes. The system is intended to assist type 1 and type 2 diabetes patients to self-manage their disease. Patients receive guidance on diabetes self-management and are encouraged to reach out to their healthcare team when needed. An over-the-counter (OTC) version of the device, WellDoc BlueStar®, was also cleared in K190013; this submission includes no changes to the OTC device. Like the bolus insulin dose calculator already included in the predicate (WellDoc BlueStar® Rx), the new long-acting basal insulin titration feature is also limited to the prescription use (Rx) only device version.

The new basal titration feature is only for patients who are taking basal insulin without any bolus insulin. The basal titration feature is for use with the following long-acting insulins: Lantus and Basaglar (U100 glargine), Toujeo (U300 glargine), Levemir (detemir), and Tresiba (degludec). The basal titration feature can also be used to titrate the following long-acting insulin / GLP-1 agonist products: Soliqua (glargine/lixisenatide), and Xultrophy (degludec/liraglutide).

The bolus insulin calculator feature will therefore be disabled for patients using the long-acting basal insulin titration feature in the BlueStar® Rx. Health care providers (HCPs) will be required to initiate and manage the basal insulin titration feature for type 2 diabetes patients using the following two interfaces:

- Web based HCP interface for use by the providers to prescribe long-acting basal insulin doses for type 2 diabetes patients.
- Web and mobile patient interface for use by patients to follow provider’s basal insulin titration plan.

The BlueStar® Rx provides directions to the patients based on prescription by their HCP for titrating long-acting insulin doses only.

BlueStar® Rx will also maintain the following features of the predicate (WellDoc BlueStar® Rx cleared under K190013): (1) ability to connect to the One Touch Verio Flex Blood Glucose Meter via Bluetooth which allows users to send data from their BG meter to the BlueStar® Rx app, which will provide coaching messages (motivational, behavioral, and educational) based on the real-time blood glucose values and trends. (2) The BlueStar® Server will also have the ability to transmit data to the OneTouch Reveal Server. With this application, 3-hour delayed continuous glucose monitoring (CGM) device data can be accessed via API and uploaded for data visualization purposes only. CGM data is not used by the BlueStar® Rx device for insulin dose calculation or titration.

{3}

B Instrument Description Information:

|  Modes of Operation | Yes | No  |
| --- | --- | --- |
|  Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐  |
|  Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐  |
|  Software  |   |   |
|  FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐  |

1. Instrument Name:
BlueStar® Rx

2. Specimen Identification:
Not Applicable.

3. Specimen Sampling and Handling:
Not Applicable.

4. Calibration:
Not Applicable.

5. Quality Control:
Not Applicable.

V Substantial Equivalence Information:

A Predicate Device Name(s):
WellDoc BlueStar® Rx

B Predicate 510(k) Number(s):
K190013

K193654 - Page 4 of 6

{4}

K193654 - Page 5 of 6

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K193654 | K190013  |
| --- | --- | --- |
|  Device Trade Name | BlueStar® Rx | WellDoc BlueStar® Rx  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Same | Determination of insulin dose based on user entered data  |
|  Environment of use | Same | Home use  |
|  User Interface | Same | Mobile App and HCP Portal  |
|  User Group | Same | Persons with type 1 or type 2 diabetes  |
|  General Device Characteristic Differences |  |   |
|  Type of insulin dose calculation | Bolus dose calculation or basal insulin titration | Bolus dose calculation  |
|  Type of insulin | Short acting or long acting insulins (Lantus, Basaglar, Toujeo, Levemir, Tresiba, Soliqua, or Xultrophy) | Short acting insulin  |

VI Standards/Guidance Documents Referenced:

FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)

FDA Guidance for Off-the-Shelf Software Use in Medical Devices (September 27, 2019)

FDA Guidance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2, 2014)

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Not Applicable.

2. Linearity:

{5}

Not Applicable.

3. Analytical Specificity/Interference:
Not Applicable.

4. Accuracy (Instrument):
Not Applicable.

5. Carry-Over:
Not Applicable.

## B Other Supportive Instrument Performance Characteristics Data:

### Usability:
The sponsor provided protocols and results from human factors studies to demonstrate that users can perform all critical tasks associated with the new device features. Subjects were representative of the device’s intended use population, including patients over 18 years of age and healthcare providers. The results of these studies were adequate to demonstrate safe use of the device and support substantial equivalence to the predicate.

### Software:
The firm provided software documentation consistent with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005), and consistent with software with a major level of concern. Software documentation was acceptable.

## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K193654 - Page 6 of 6

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ/K193654](https://fda.innolitics.com/submissions/CH/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ/K193654)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
