FREESTYLE GLUCOSE METER INCORPORATED INTO THE OMNIPOD INSULIN MANAGEMENT SYSTEMS
K111669 · Insulet Corporation · LZG · Dec 15, 2011 · General Hospital
Device Facts
Record ID
K111669
Device Name
FREESTYLE GLUCOSE METER INCORPORATED INTO THE OMNIPOD INSULIN MANAGEMENT SYSTEMS
Applicant
Insulet Corporation
Product Code
LZG · General Hospital
Decision Date
Dec 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared. Abbott FreeStyle test strips are used with the built in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott FreeStyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.
Device Story
OmniPod Insulin Management System consists of a Personal Diabetes Manager (PDM) and a disposable Pod. PDM acts as the controller; user inputs insulin profiles, checks status, and initiates bolus doses via PDM interface. PDM integrates a FreeStyle glucose meter; receives blood glucose readings via internal serial interface; stores and displays data. Pod is filled with U-100 rapid-acting insulin (Novolog/NovoRapid, Humalog, or Apidra); adheres to patient skin; uses automated needle/cannula deployment for subcutaneous delivery. Once initiated, Pod operates independently of PDM. System used by patients for diabetes management; provides automated insulin delivery and glucose monitoring to assist in glycemic control.
Clinical Evidence
Bench testing included verification of meter functionality per ISO 15917:2003, cleaning/disinfection validation, software verification, EMC testing (60601-1-2), and hardware verification. Clinical evaluation involved 75 subjects at two sites comparing subject-measured capillary blood glucose against YSI 2300 Stat Plus reference device. Subjects also provided usability feedback.
Technological Characteristics
System comprises PDM and disposable Pod. PDM features color LCD, USB port, and integrated FreeStyle glucose meter. Pod is a wearable, adhesive-backed pump for subcutaneous insulin delivery. Connectivity via internal serial interface between meter and PDM. Software performs insulin profile management and glucose data processing. Sterilization and material specs not detailed beyond general device description.
Indications for Use
Indicated for subcutaneous insulin delivery at set/variable rates for diabetes mellitus management in patients requiring insulin; includes integrated glucose meter for quantitative measurement of fresh whole capillary blood glucose from finger, upper arm, or palm. Not for diabetes diagnosis or screening. Single-patient use only.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
i
## Section 7- 510(k) Summary of Safety and Effectiveness
| 7.1<br>Statement | This summary of 510(k) safety and effectiveness information is being submitted<br>in accordance with the requirements of SMDA 1990 and CFR 807.92 | |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 7.2<br>Submitter/<br>Date | Insulet Corporation<br>9 Oak Park Drive<br>Bedford, MA. 01730<br><br>Date Prepared: December 14, 2011 | |
| 7.3<br>Company<br>Contact | Susan Finneran<br>Senior Director, Clinical and Regulatory Affairs<br>9 Oak Park Drive<br>Bedford, MA.<br>781-457-5234 | |
| 7.4<br>Device Name | Proprietary Name: OmniPod® Insulin Management System<br>Common Name: OmniPod System<br>Classification Name: Infusion pump, 21CFR 880.5725 | |
| 7.5<br>Predicate<br>Legally<br>Marketed<br>Devices | The Insulet OmniPod Diabetes Management System that is the subject of this<br>submission is substantially equivalent to the IXL-ii System that was cleared via<br>K042972. | |
| 7.6<br>Device<br>Description | The OmniPod® Insulin Management System is intended for<br>subcutaneous delivery of insulin at set and variable rates for the<br>management of diabetes mellitus in persons requiring insulin<br>and for the quantitative measurement of glucose in fresh whole | |
| | 7.7<br>Indications<br>for Use | The OmniPod Insulin Management System is intended for subcutaneous<br>delivery of insulin at set and variable rates for the management of<br>diabetes mellitus in persons requiring insulin and for the quantitative<br>measurement of glucose in fresh whole capillary blood (in vitro) from the<br>finger. |
| | | The glucose measurements should not be used for the diagnosis of or<br>screening for diabetes. The PDM glucose meter is intended for single-<br>patient use and should not be shared. |
| | | Abbott FreeStyle test strips are used with the built in FreeStyle meter for<br>the quantitative measurement of blood glucose in fresh whole capillary<br>blood from the finger, upper arm and palm. Abbott FreeStyle Control<br>Solutions are used to verify that the meter and test strips are working<br>together properly and that the test is performed correctly. |
{1}------------------------------------------------
Capillary blood (in vitro).
The OmniPod Insulin Management system consists of two main components:
- . The Personal Diabetes Manager (PDM)
- The OmniPod (Pod) .
The Personal Diabetes Manager (PDM) handles all processes for the operation of the OmniPod System. The display uses full text language to prompt the user through the set up process. It is also utilized to initialize and program the Pod with the user's custom based insulin profile, to check the pod status, and initiate a bolus dose of insulin. Using the PDM the user can also make temporary changes to the insulin delivery profile. After set-up and initiation, the Pod will run independently of the PDM.
The PDM device incorporates the Freestyle glucose meter. The PDM receives blood glucose readings that are measured using the Freestyle blood glucose meter via an internal serial interface. The blood glucose data is stored and can be displayed by the PDM. The integrated glucose test meter is intended to be used with Abbott Freestyle test strips.
The Pod is the disposable component of the system. The Pod is intended to be filled with rapid acting insulin U-100. The brands of insulin that can be utilized with the Pod are as follows: Novolog@/NovoRapid®, Humalog®, or Apidra®.
Once the Pod has been filled with a minimum volume of insulin the tubing in the pump is primed. Once primed the user is instructed to remove the adhesive backing and place the Pod adhesive side down on the body. Once in place, the needle punctures the injection site and the cannula is deployed for insulin delivery.
{2}------------------------------------------------
| 7.8<br>Summary of<br>changes to the<br>predicate device | Several changes were made to the OmniPod System since it was originally cleared via K042792. Following is a summary of the changes that were made to the PDM component of the system since it was originally cleared via K042792.<br><br>Replacement of the LCD with a color version. A modified device shell and combined function of the Home and Power On/Off buttons into one button. Addition of USB port Removal of two “hard” buttons – Blood Glucose Records button and Food Database button. Changes to the Software User Interface: New home screen. Customized BG tags New status screen. New confirmation screen Update to labeling to indicate that the OmniPod System may be used with Abbott Freestyle test strips. |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
7.9 Summary of supporting data
The intent of this premarket notification was to submit data to support that the glucose meter that is incorporated into the device can be used with the GDH-FAD glucose test strips that have been recently cleared by Abbott Diabetes Care. This premarket notification also includes verification data to support the changes to the device since it was cleared via K042792.
Table 7.1 provides a summary of the tested that was completed to verify performance of the GDH-FAD strip with the OmniPod System and to support changes that were made to the OmniPod System since it was initially cleared.
Table 7.1 Summary of Supporting Verification/ Clinical Testing
| Test Report<br>Reference | Summary of Testing |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| TP/TR09-217 | Verification of meter functionality was completed in<br>accordance with ISO 15917:2003. |
| TP/TR10-067 | 75 subjects at two sites were evaluated. Whole<br>capillary blood was collected and measured by the<br>subject and by a trained user. The results were also<br>compared to a glucose reference sample measured<br>with the YSI 2300 Stat Plus device. Subject's were<br>also asked to rate the use of the device |
{3}------------------------------------------------
| TP/TR10-326 | Cleaning/ Disinfection testing of the meter was<br>completed to demonstrate that the device could be<br>disinfected appropriately and remain functional after<br>multiple cycles. |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| TP/TR08-078<br>TP/TR08-233<br>TP/TR09-21 | Software verification, unit testing, code reviews. |
| TP/TR07-273 | EMC testing to demonstrate compliance to 60601-1-2 |
| TP/TR07-176<br>TP/TR07-177<br>TP/TR07-178<br>TP/TR07-179 | Hardware verification testing |
The data presented above demonstrates that the device is safe and effective and is substantially equivalent to the predicate device cleared via K042792.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered horizontally and vertically in the image. The background of the image is white.
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or lines extending upwards, representing health and well-being. The figure is positioned within a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Susan Finneran Senior Director Clinical and Regulatory Affairs Insulet Corporation 9 Oak Park Drive Bedford. Massachusetts 01730
DEC 1 5 2011
Re: K111669
Trade/Device Name: OmniPod® Insulin Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: December 9, 2011 Received: December 12, 2011
Dear Ms. Finneran:
We have reviewed your Section 510(k) premarket notification of intent-to-market-the-devicereferenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Finneran
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Anton V. m.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):_K111669
Device Name:_OmniPod® Insulin Management System
Indications For Use:
The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger,
The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared.
Abbott FreeStyle test strips are used with the built in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott FreeStyle Control Solutions are used to verify that the meter and test aring are working together properly and that the test is performed correctly.
Prescription Use ✓ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kathleen E. Abbott fakcc
. Wision of Anesmesiology, General Hostifal
Page 1 of **__**
510(k) Number: K111669