MEDTRONIC MINIMED PARADIGM MODEL 512 INSULIN PUMP AND THE BD PARADIGM LINK GLUCOSE METER
K030531 · Minimed, Inc. · LZG · Jun 17, 2003 · General Hospital
Device Facts
Record ID
K030531
Device Name
MEDTRONIC MINIMED PARADIGM MODEL 512 INSULIN PUMP AND THE BD PARADIGM LINK GLUCOSE METER
Applicant
Minimed, Inc.
Product Code
LZG · General Hospital
Decision Date
Jun 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Model 512 Insulin Pump: The Medtronic MiniMed Paradigm Model 512 Insulin Pump is indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. BD Paradigm Link Glucose Monitor: The BD Paradigm Link Blood Glucose Monitor is intended to be used for the quantitative measurement of glucose in whole blood. It is intended for use by people with diabetes mellitus in the home as an aid to monitor the effectiveness of diabetes control. It is not intended for use in the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. The BD Paradigm Link Blood Glucose Monitor is specifically indicated for the quantitative measurement of glucose in whole blood samples obtained from the fingertip. Model 512 Insulin Pump and BD Paradigm Link Glucose Monitor: When used together, the BD Paradigm Link Glucose Monitor can automatically telemeter glucose values to the Model 512 Insulin Pump using radio frequency communication. The glucose value received by the Model 512 Insulin Pump is used as the default glucose value by the pump's Bolus Wizard feature if the Bolus Wizard is used within 12 minutes of the glucose value transmission. The BD Paradigm Link Glucose Monitor can also serve as a radio frequency interface to allow communication between the Model 512 Insulin Pump and a personal computer running the appropriate Medtronic MiniMed communications software.
Device Story
Paradigm Model 512 is an ambulatory, battery-operated, rate-programmable microinfusion pump for continuous subcutaneous insulin delivery. It delivers basal profiles and bolus amounts via a motor-driven reservoir. BD Paradigm Link Glucose Monitor measures capillary blood glucose (0.3 µL sample, 5-second result). The meter transmits glucose values to the pump via radio frequency (RF) telemetry. The pump's 'Bolus Wizard' feature uses the received glucose value as a default if used within 12 minutes. The meter also acts as an RF interface for data transfer between the pump and a PC running Medtronic software. Used by patients in home settings; pump requires physician prescription. Benefits include automated data transfer to reduce manual entry errors and facilitate diabetes management.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Ambulatory microinfusion pump (battery-operated, motor-driven) and in-vitro diagnostic glucose meter. Features RF telemetry for device-to-device communication and PC interface. Software-controlled insulin delivery with Bolus Wizard functionality. Materials and basic design consistent with predicate devices.
Indications for Use
Indicated for persons with diabetes mellitus requiring insulin (pump) and for home use by people with diabetes to monitor glucose control (meter). Meter not for neonates, diagnosis, or screening. Meter indicated for fingertip capillary blood samples.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Medtronic MiniMed Paradigm Model 511 Insulin Pump (K030531)
K080639 — SYMPHONY GLUCOSE MANAGEMENT SYSTEM · Animas Corp. · Jun 24, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K03053/
# 510(k) Summary
In accordance with the requirements of SMDA 1990, and 21 CFR 807.92, this 510(k) Summary is provided:
Submitter: Medtronic MiniMed 18000 Devonshire St., Northridge, CA 91325
Contact: Mark Faillace (818) 576-5616
Name of Devices: Medtronic MiniMed Paradigm Model 512 Insulin Pump and BD Paradigm Link Glucose Monitor
Predicate Devices: Medtronic MiniMed Paradigm Model 511 Insulin Pump, BD Logic Glucose Monitor and Medtronic MiniMed Model 7304 ComLink
#### Description of the Devices:
The Paradigm Model 512 external insulin pump is an ambulatory, battery operated, rateprogrammable microinfusion pump, designed for continuous delivery of insulin. A reservoir is driven by a drive motor to deliver preset basal profiles and patient programmed bolus amounts of insulin through infusion sets into subcutaneous tissue. This pump incorporates new software improvements over the predicate model and is designed to allow installation of upgraded software with new features through radio frequency telemetry without the need to replace any pump components.
The BD Paradigm Link Blood Glucose Monitor is an in-vitro diagnostic device for use in the quantitative measurement of glucose in capillary blood collected from fingertips. The BD Paradigm Link™ is designed to be simple and easy to use. It provides accurate blood glucose test results in 5 seconds while requiring a very small (0.3 µL) sample.
When the BD Paradigm Link Glucose Monitor is used together with the Medtronic MiniMed 512 External Insulin Pump, blood glucose test values obtained using the BD Paradigm Link Glucose Monitor can be automatically transmitted to the insulin pump for use in the pump's Bolus Wizard feature. Consumers or healthcare practitioners may also use the meter as a communication device to facilitate transfer of information between the pump and a personal computer.
The Paradigm Model 512 pump is restricted to sale by, or on the order of, a physician. The Paradigm Link Glucose Monitor is intended for over-the-counter use.
{1}------------------------------------------------
# 510(k) Summary (Continued)
## Intended Use of the Devices:
#### Model 512 Insulin Pump:
The Medtronic MiniMed Paradigm Model 512 Insulin Pump is indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
## BD Paradigm Link Glucose Monitor:
The BD Paradigm Link Blood Glucose Monitor is intended to be used for the quantitative measurement of glucose in whole blood. It is intended for use by people with diabetes mellitus in the home as an aid to monitor the effectiveness of diabetes control. It is not intended for use in the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.
The BD Paradigm Link Blood Glucose Monitor is specifically indicated for the quantitative measurement of glucose in whole blood samples obtained from the fingertip.
## Model 512 Insulin Pump and BD Paradigm Link Glucose Monitor:
When used together, the BD Paradigm Link Glucose Monitor can automatically telemeter glucose values to the Model 512 Insulin Pump using radio frequency communication. The glucose value received by the Model 512 Insulin Pump is used as the default glucose value by the pump's Bolus Wizard feature if the Bolus Wizard is used within 12 minutes of the glucose value transmission.
The BD Paradigm Link Glucose Monitor can also serve as a radio frequency interface to allow communication between the Model 512 Insulin Pump and a personal computer running the appropriate Medtronic MiniMed communications software.
# Comparison of the Technological Features of the New Devices and Predicate Devices:
The Model 512 and Model 511 have similar materials and basic design. The new device contains additional features compared to the predicate device. These additional features include a Bolus Wizard function and the ability to receive glucose values transmitted by the BD Paradigm Link Glucose Monitor.
{2}------------------------------------------------
# 510(k) Summary (Continued)
The BD Paradigm Link and BD Logic Glucose Monitors also have similar materials and basic design. The new device has additional circuitry and software modifications that allow it to transmit glucose values to the Model 512 Insulin Pump and to facilitate transfer of data between the Model 512 Insulin Pump and a personal computer running the appropriate Medtronic MiniMed communications software.
Arnold Faillace
5/12/03
Date
Mark Faillace Senior Director, Device Regulatory Affairs and Quality Assurance Medtronic MiniMed
TM Paradigm 512 is a Trademark of Medtronic MiniMed
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. The eagle is facing to the right.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Mark J. Faillace Senior Director, Device RA/QA Medtronic MiniMed 18000 Devonshire Street Northridge, California 91325
Re: K030531
Trade/Device Name: Medtronic MiniMed Paradigm Model 512 Insulin Pump and the BD Paradigm Link™ Glucose Meter Regulation Number: 880.5725, 862.1345 Regulation Name: Infusion Pump, Glucose Test System Regulatory Class: II Product Code: LZG, NBW Dated: May 13, 2003 Received: May 14, 2003
Dear Mr. Mark J. Faillace:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JUN 1 7 2003
{4}------------------------------------------------
Page 2 - Mr. Mark J. Faillace
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Suser Runos
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number:
K030531
Device Names:
Medtronic MiniMed Paradigm Model 512 Insulin Pump and BD Paradigm Link Glucose Monitor
Indications for Use:
## Model 512 Insulin Pump:
The Medtronic MiniMed Paradigm Model 512 Insulin Pump is indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
or
Prescription Use V (Per 21 CFR 801.109)
Over-the-Counter Use ------
d
#### BD Paradigm Link Glucose Monitor:
The BD Paradigm Link Blood Glucose Monitor is intended to be used for the quantitative measurement of glucose in whole blood. It is intended for use by people with diabetes mellius in the home as an aid to monitor the effectiveness of diabetes control. It is not intended for use in the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.
The BD Paradigm Link Blood Glucose Monitor is specifically indicated for the quantitative measurement of glucose in whole blood samples obtained from the fingertin.
Prescription Use (Per 21 CFR 801.109) or
Over-the-Counter Use
# Model 512 Insulin Pump and BD Paradigm Link Glucose Monitor:
When used together, the BD Paradigm Link Glucose Monitor can automatically telemeter glucose values to the Model 512 Insulin Pump using radio frequency communication, The glucose value received by the Model 512 Insulin Pump is used as the default glucose value by the pump's Bolus Wizard feature if the Bolus Wizard is used within 12 minutes of the glucose value transmission.
{6}------------------------------------------------
# INDICATIONS FOR USE (CONTINUED)
.. .. : "
> The BD Link Glucose Monitor can also serve as a radio frequency interface to allow communication between the Model 512 Insulin Pump and a personal computer running the appropriate Medtronic MiniMed communications software.
| Prescription Use<br>(Per 21 CFR 801.109) | |
|------------------------------------------|----|
| | or |
| Over-the-Counter Use | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
1
Patrizia Quenti
(Division Sign-Off) Division of Anesthesiology, General Ho Infection Control, Dental De
510(k) Number. K030537
# 000228
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.