DANA DIABECARE II INSULIN PUMP & SUPERLINE-EASYRELEASE, SOFT-RELEASE-ST, & SOFT-RELEASE-R INFUSION SETS
K022317 · Sooil Development Co., Ltd. · LZG · Aug 2, 2002 · General Hospital
Device Facts
Record ID
K022317
Device Name
DANA DIABECARE II INSULIN PUMP & SUPERLINE-EASYRELEASE, SOFT-RELEASE-ST, & SOFT-RELEASE-R INFUSION SETS
Applicant
Sooil Development Co., Ltd.
Product Code
LZG · General Hospital
Decision Date
Aug 2, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DANA Diabecare® II external programmable syringe infusion pump and the SUPERLINE-EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets are indicated for the subcutaneous delivery of insulin for the treatment of diabetes mellitus. The pump and administration sets are not intended for use with blood or blood products.
Device Story
DANA Diabecare® II is an external programmable syringe infusion pump with associated infusion sets (EasyRelease, Soft-Release-ST, Soft-Release-R). Device delivers insulin subcutaneously to treat diabetes mellitus. Modifications from predicate include software updates for basal/bolus delivery modes, password-protected lock-out, alarm features, fail-safe functions, and data storage/retrieval. Infusion sets feature a disconnection site allowing pump removal without needle extraction and optional indwelling soft cannulas. Operated by patients or clinicians to manage insulin dosing. Output consists of controlled insulin delivery; healthcare providers use stored data to monitor therapy. Benefits include enhanced convenience, ease of use, and comfort compared to previous models.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility per ISO 10993-1, burst testing, bioburden testing, insulin potency assays (stability up to 120 hours), and shelf-life testing (3 years).
Technological Characteristics
External programmable syringe infusion pump. Infusion sets include indwelling soft cannulas and detachment sites (sliding lock or turning screw lock). Materials meet ISO 10993-1 biocompatibility standards. Software-controlled delivery with password-protected lock-out, alarm system, and data logging. Fluid path remains sterile during detachment/reattachment.
Indications for Use
Indicated for patients with diabetes mellitus requiring subcutaneous insulin delivery.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
DANA Diabecare® insulin pump and SUPERLINE infusion set (K001604)
Related Devices
K063126 — DANA DIABECARE IIS · Sooil Development Co., Ltd. · Feb 2, 2007
K042873 — ANIMAS MODEL IR1250 INSULIN INFUSION PUMP · Animas Corp. · Dec 10, 2004
K032257 — MODEL IR 1200 INSULIN PUMP · Animas Corp. · Oct 16, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K022317
## AUG 0 2 2002 510(k) Summary
### for
## DANA Diabecare® II Insulin Pump and SUPERLINE-EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets
#### SPONSOR 1.
Sooil Development Co., Ltd. 111-1, Heukseokl-Dong, Dongjak-Ku Seoul, 156-071 KOREA
| Contact Person: | Soo Bong Choi |
|-----------------|-----------------------------|
| Telephone: | 82-441-845-2129 (Telephone) |
| | 82-441-845-2128 (Facsimile) |
July 16, 2002 Date Prepared:
#### 2. DEVICE NAME
| Proprietary Name: | DANA Diabecare® II Insulin Pump and SUPERLINE-<br>EasyRelease, Soft-Release-ST, and Soft-Release-R<br>Infusion Sets |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Insulin infusion pump and infusion set accessories |
| Classification Name: | Infusion pump |
#### 3. PREDICATE DEVICES
DANA Diabecare® insulin pump and SUPERLINE infusion set (Sooil Development Co., Ltd., K001604)
#### DEVICE DESCRIPTION 4.
The DANA Diabecare® II insulin pump, SUPERLINE-EasyRelease (EasyRelease) Infusion Set, and Soft-Release Infusion Sets are modifications of the DANA Diabecare® insulin pump and SUPERLINE infusion set that were described in K001604, the 510(k) premarket notification for the DANA Diabecare® insulin infusion pump.
{1}------------------------------------------------
#### ડ. INTENDED USE
The DANA Diabecare® II external programmable syringe infusion pump and the SUPERLINE-EasyRelease, Soft-Release-R, and Soft-Release-ST Infusion Sets are indicated for the subcutaneous delivery of insulin for the treatment of diabetes mellitus. The pump and administration sets are not intended for use with blood or blood products.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The modifications made to the parent DANA Diabecare® insulin pump and SUPERLINE Infusion Set to produce the proposed DANA Diabecare® II Insulin Pump and the EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets were implemented to enhance the convenience, ease of use, and comfort of the devices.
The modifications made to the parent DANA Diabecare® insulin pump to produce the modified DANA Diabecare® II Insulin Pump were limited to software modifications including changes to the basal and bolus insulin delivery modes. features accessed via the password-protected (lock-out) mode, and the addition of alarms, a fail-safe function, and the ability to store and retrieve information (alarm, bolus dose, daily insulin dose).
The overall design of the proposed EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets is identical to the design of the parent SUPERLINE Infusion Set. The design modifications are limited to the following:
- Addition of a site for disconnecting the insulin pump from the infusion set . without removing the needle from the abdomen
- Replacement of the infusion needle with an indwelling soft cannula (Soft-. Release-R and Soft-Release-ST only)
The two versions of the Soft-Release Infusion Sets differ in the design of the detachment site and cannula assembly. The Soft-Release-R Infusion Set has a sliding lock detachment site and a "rectangular" cannula oriented perpendicular to the infusion tubing axis. The Soft-Release-ST has a turning screw lock detachment site that uses a conical fitting and a straight cannula oriented parallel to the infusion tubing axis.
There are minor differences in the materials used for the construction of the proposed EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets as compared to the parent SUPERLINE. Sooil has conducted the necessary testing to satisfy the
{2}------------------------------------------------
biocompatibility requirements of ISO 10993-1 for use as an insulin administration set.
#### 7. PERFORMANCE TESTING
In addition to the biocompatibility testing mentioned in the previous section, design verification activities included burst testing, bioburden testing, insulin potency assay, and shelf-life testing. The results of the burst and bioburden tests confirm that the detachment sites incorporated in the EasyRelease and Soft-Release Infusion Sets do not leak or unexpectedly detach and the fluid path remains sterile following repeated detachment/reattachment procedures. The insulin potency assay indicates that insulin remains stable for up to 120 hours in the EasyRelease and Soft-Release Infusion Sets. The shelf-life testing supported a three-year shelf life for the EasyRelease and Soft-Release Infusion Sets.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the largest and the third profile being the smallest.
### Public Health Service
# AUG 0 2 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sooil Development Company Limited Ms. Cynthia J. M. Nolte Consultant Medical Device Consultanting, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760
Re: K022317
Trade/Device Name: DANA Diabecare® II Insulin Pump and SUPERLINE-Easy Release, Soft-Release-ST, and Soft-Release-R Infusion Sets Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: July 16, 2002 Received: July 17, 2002
Dear Ms. Nolte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Ms. Nolte
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device.complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
the world of the starting of the starting of the starting of the starting of the first of the first of the first of the first of the first of the first of the first of the fi
Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
K 02 2317 510(k) Number (if known):
Device Name: DANA Diabecare® II Insulin Pump and SUPERLINE-EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets
Indications for Use:
The DANA Diabecare® II external programmable syringe infusion pump and the SUPERLINE-EasyRelease, Soft-Release-ST, and Soft-Release-R Infusion Sets are used for the subcutaneous delivery of insulin for the treatment of diabetes mellitus. These devices are not intended for use with blood or blood products.
(Please do not write below this line - Continue on another page if necessary)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patricia Ciccerite
Division Sign-Off Division of Dental, Infection Control, and General Hospital Devi 510(k) Number _
V Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.