XTAG CYP2D6 KIT V3, MODEL I030C0300 (96 TESTS/KIT), TDAS CYP2D6 SOFTWARE (2030-0254)

K093420 · Luminex Molecular Diagnostics, Inc. · NTI · Aug 26, 2010 · Clinical Toxicology

Device Facts

Record IDK093420
Device NameXTAG CYP2D6 KIT V3, MODEL I030C0300 (96 TESTS/KIT), TDAS CYP2D6 SOFTWARE (2030-0254)
ApplicantLuminex Molecular Diagnostics, Inc.
Product CodeNTI · Clinical Toxicology
Decision DateAug 26, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3360
Device ClassClass 2

Indications for Use

The xTAG® CYP2D6 Kit v3 is a device used to simultaneously detect and identify a panel of nucleotide variants found within the highly polymorphic CYP2D6 gene located on chromosome 22 from genomic DNA extracted from an EDTA or citrate anticoagulated whole blood sample. This kit can also identify gene rearrangements associated with the deletion (*5) and duplication genotypes. The xTAG® CYP2D6 Kit v3 is a qualitative genotyping assay which can be used as an aid to clinicians in determining therapeutic strategy for therapeutics that are metabolized by the CYP2D6 gene product. This kit is not indicated for standalone diagnostic purposes. This test is not intended to be used to predict drug response or non-response.

Device Story

The xTAG® CYP2D6 Kit v3 is an in vitro diagnostic assay for genotyping the CYP2D6 gene. It processes genomic DNA extracted from EDTA or citrate anticoagulated whole blood. The workflow involves multiplex PCR amplification, followed by multiplex allele-specific primer extension (ASPE) using universally-tagged primers. The resulting products are hybridized to fluorescing microparticles and detected via flow cytometry on Luminex 100 or 200 IS instruments. The xTAG® Data Analysis Software (TDAS) processes the fluorescence signals to identify specific nucleotide variants, gene deletions (*5), and duplications. The output provides the patient's CYP2D6 genotype, which clinicians use to inform therapeutic strategies for drugs metabolized by the CYP2D6 enzyme. The device is intended for use in clinical laboratory settings by trained personnel.

Clinical Evidence

No clinical trials were performed. Substantial equivalence was established via analytical performance studies. Method comparison against bi-directional dideoxy sequencing using 459 clinical samples showed 98.4% overall agreement. Precision/reproducibility was assessed across multiple sites, operators, and lots, achieving 94.44% agreement. Limit of detection was established at 2 ng/μL. Analytical specificity was confirmed with common blood interferents (hemoglobin, bilirubin, albumin, triglycerides).

Technological Characteristics

Multiplex PCR and allele-specific primer extension (ASPE) assay. Detection via flow cytometry using Luminex 100/200 IS instruments. Analyzes genomic DNA from whole blood. Components include PCR mixes, ASPE mix, bead mix, and enzymes (Platinum TFI Exo(-) DNA Polymerase, LA Taq). Software-based analysis (TDAS).

Indications for Use

Indicated for qualitative genotyping of the CYP2D6 gene from genomic DNA extracted from EDTA or citrate anticoagulated whole blood. Used as an aid to clinicians in determining therapeutic strategy for drugs metabolized by CYP2D6. Not for standalone diagnostic use; not intended to predict drug response or non-response.

Regulatory Classification

Identification

A drug metabolizing enzyme genotyping system is a device intended for use in testing deoxyribonucleic acid (DNA) extracted from clinical samples to identify the presence or absence of human genotypic markers encoding a drug metabolizing enzyme. This device is used as an aid in determining treatment choice and individualizing treatment dose for therapeutics that are metabolized primarily by the specific enzyme about which the system provides genotypic information.

Special Controls

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping System."

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping Test System.” See § 862.1(d) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k093420 B. Purpose for Submission: New device C. Measurand: Genotype of Cytochrome P450 2D6 (CYP2D6) D. Type of Test: Multiplex PCR followed by multiplex allele specific primer extension for genotyping, hybridized to multiplexed fluorescing microparticles, detected by flow cytometry E. Applicant: Luminex Molecular Diagnostics, Inc. F. Proprietary and Established Names: xTAG® CYP2D6 Kit G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | NTI | II | 21 CFR §862.3360 - Drug Metabolizing Enzyme Genotyping System | Toxicology | H. Intended Use: 1. Intended use(s): The xTAG® CYP2D6 Kit v3 is a device used to simultaneously detect and identify a panel of nucleotide variants found within the highly polymorphic CYP2D6 gene located on chromosome 22 from genomic DNA extracted from an EDTA or citrate anticoagulated whole blood sample. This kit can also identify gene rearrangements associated with the deletion (*5) and duplication genotypes. The xTAG® CYP2D6 Kit v3 is a qualitative genotyping assay which can be used as an aid to clinicians in determining therapeutic strategy for therapeutics that are metabolized by the CYP2D6 gene product. This kit is not indicated for standalone diagnostic purposes. This test is not intended to be used to predict drug response or non-response. 2. Indication(s) for use: See Intended use above. {1} 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Luminex 100 or 200 IS instruments I. Device Description: The xTAG® CYP2D6 Kit v3 includes the following components: - PCR Mix A (includes dNTPs) - PCR Mix B (includes dNTPs) - ASPE Mix (includes dNTPs) - Bead Mix - 10x Buffer - Shrimp Alkaline Phosphatase - Exonuclease I - Streptavidin, R-Phycoerythrin conjugate - Platinum TFI Exo(-) DNA Polymerase and associated MgCl₂ and PCR buffer - xTAG® Data Analysis Software (TDAS) CYP2D6 - LA Taq DNA Polymerase J. Substantial Equivalence Information: 1. Predicate device name(s): Amplichip CYP450 Test 2. Predicate K number(s): k042259 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Predicate device (k042259) | Proposed device | | Indications for use | Intended to identify a patient's CYP2D6 genotype from genomic DNA extracted from a whole blood sample. | Same | | Limitation | This test is not intended to be used to predict drug response or non-response. | Same | | CYP450 gene genotyped by the test | 2D6 | Same | {2} | Differences | | | | --- | --- | --- | | Item | Predicate device (k042259) | Proposed device | | Number of 2D6 star genotypes detected by the device | 20 | 16 | | CYP450 2D6 star genotypes detected by the test | *1, *2, *3, *4, *5, *6, *7, *8, *9, *10, *11, *15, *17, *19, *20, *29, *35, *36, *40, *41 | *1, *2, *3, *4, *5, *6, *7, *8, *9, *10, *11, *15, *17, *29, *35, *41 | | Sample types | EDTA anticoagulated whole blood | EDTA and citrate anticoagulated whole blood | | Instrument | Affymetrix GeneChip Microarray Instrumentation System | Luminex 100 and 200 IS Instrument | | Test principle | PCR amplification of purified DNA, fragmentation and labeling of amplified products, hybridization of amplified products to a microarray and staining of the bound products, and scanning of the microarray | Multiplex PCR followed by multiplex allele specific primer extension for genotyping, hybridized to multiplexed fluorescing microparticles, detected by flow cytometry | K. Standard/Guidance Document Referenced (if applicable): - Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems - Guidance for Industry and FDA Staff - Guidance for Industry and FDA Staff: Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests - Guidance for Industry and FDA Staff: Pharmacogenetic Tests and Genetic Tests for Heritable Markers - Guidance for Industry and FDA Staff: Content of Pre-Market Submission for Software Contained in Medical Devices - CLSI Guideline MM01-A2: Molecular Diagnostic Methods for Genetic Diseases - CLSI Guideline MM13-PE: Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods - CLSI Guideline EP05-A2: Evaluation of Precision Performance of Clinical Chemistry Devices - CLSI Guideline EP07-A2E: Interference Testing in Clinical Chemistry - CLSI Guideline EP12-A: User Protocol for Evaluation of Qualitative Test Performance - CLSI Guideline EP17-A: Protocols for Determining Limits of Detection and Limits of Quantitation {3} L. Test Principle: For each genomic sample being tested, two separate PCR reactions are performed. PCR-A produces an alpha fragment used to detect the mutations in the table below, as well as a duplication amplimer which indicates the presence of the duplication genotype. PCR-B produces a beta fragment used to detect the mutations in the table below, as well as a deletion amplimer indicative of the deletion genotype. Following PCR amplification, the two reactions (PCR-A and PCR-B) are pooled. To enable efficient incorporation of biotin-dCTP during the ASPE reaction, the pooled PCR product is treated with Shrimp Alkaline Phosphatase to dephosphorylate any remaining nucleotides (particularly dCTP), and with Exonuclease I to degrade any primers left over from the PCR reaction. ASPE is then carried out using universally-tagged primers supplied in the ASPE primer mix. An aliquot of the ASPE reaction is hybridized with the universal array (Bead Mix) in the presence of the hybridization buffer and incubated with Streptavidin, R-Phycoerythrin conjugate (reporter solution). Samples are read on the Luminex 100 IS or 200 IS instrument and a signal is generated for each of the loci as well as for the duplication and deletion amplimers, if present. These fluorescence values are then analyzed to determine whether the wild-type/mutant allele for each of the loci has been detected, or whether the samples carry an allele(s) with the deletion or duplication. | Star Genotype | Mutations and Polymorphisms detected by xTAG® CYP2D6 | | | --- | --- | --- | | | PCR A | PCR B | | *1 | None | None | | *2 | -1584C>G, 1661G>C, | 2850C>T, 4180G>C | | *3 | | 2549A>del | | *4 | 100C>T, 1661G>C, 1846G>A, | 2850C>T, 4180G>C | | *5 | | Deletion | | *6 | 1707T>del, | 4180G>C | | *7 | | 2935A>C | | *8 | 1661G>C, 1758G>T, | 2850C>T, 4180G>C | | *9 | | 2613delAGA | | *10 | 100C>T, | 1661G>C, 4180G>C | | *11 | 883G>C, 1661G>C, | 2850C>T, 4180G>C | | *15 | 138insT | | {4} | Star Genotype | Mutations and Polymorphisms detected by xTAG® CYP2D6 | | | --- | --- | --- | | | PCR A | PCR B | | *17 | 1023C>T, 1661G>C, | 2850C>T, 4180G>C | | *29 | 1659G>A, 1661G>C, | 2850C>T, 3183G>A, 4180G>C | | *35 | -1584C>G, 31G>A, 1661G>C, | 2850C>T, 4180G>C | | *41 | 1661G>C, | 2850C>T, 2988G>A, 4180G>C | | DUP | Duplication | | Nucleotide changes that define the star genotype (*) are shown in bold font. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Reproducibility of the xTAG® CYP2D6 Kit v3 was assessed using a multiple-center, multiple-operator, multiple-lot, blinded study design. A subset of samples was extracted at each site with a different extraction method, therefore three different DNA extraction methods were included in the study. Reproducibility of the complete assay, including the extraction step, was assessed on 8 EDTA anticoagulated whole blood samples (genotypes provided in table below). Evaluation of the reproducibility of the analytical (post-extraction) steps of the assay was further assessed with testing of 5 purified genomic DNA samples. This study was designed to distinguish between Site-to-Site, Lot-to-Lot and Operator-to-Operator reproducibility in the CYP2D6 assay. There were 2 operators per site, each performing one run per day across 3 non-consecutive days (3 runs per operator or 6 runs per site). Thus the number of replicates per sample for this study was: (3 sites) x (2 operators/site) x (3 runs/operator) x (2 replicates/run) = 36 replicates. {5} Reproducibility of the xTAG® CYP2D6 Kit v3 was demonstrated with a panel of 13 samples representing the following star (*) alleles (also termed 'star genotypes'): | Sample ID | Star Alleles Present | Sample Type | | --- | --- | --- | | 1 | *1/*5 | genomic DNA | | 2 | *1/*41 | genomic DNA | | 3 | *4/*35 | genomic DNA | | 4 | *2/*4, DUP | genomic DNA | | 5 | *35/*41 | genomic DNA | | 6 | *1/*10 | whole blood | | 7 | *10/*17 | whole blood | | 8 | *1/*2 | whole blood | | 9 | *1/*1 | whole blood | | 10 | *2/*10 | whole blood | | 11 | *2/*17 | whole blood | | 12 | *1/*5 | whole blood | | 13 | *2/*2 | whole blood | | Sample | Genotype | Replicates per site | # Total replicates | # correct calls (Site 1) | # correct calls (Site 2) | # correct calls (Site 3) | Total correct calls^{1} | # No calls^{1} | # Incorrect calls^{1} | 95% LCB | % Agreement | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 1 | *1/*5 | 12 | 36 | 12 | 12 | 12 | 36 | 0 | 0 | 90.25 | 100.00 | | 2 | *1/*41 | 12 | 36 | 12 | 12 | 12 | 36 | 0 | 0 | 90.25 | 100.00 | | 3 | *4/*35 | 12 | 36 | 11 | 12 | 12 | 35 | 1 | 0 | 85.47 | 97.22 | | 4 | *2/*4, DUP | 12 | 36 | 10 | 12 | 12 | 34 | 2 | 0 | 81.34 | 94.44 | | 5 | *35/*41 | 12 | 36 | 12 | 12 | 11 | 35 | 1 | 0 | 85.47 | 97.22 | | 6 | *1/*10 | 12 | 36 | 12 | 12 | 12 | 36 | 0 | 0 | 90.26 | 100.00 | | 7 | *10/*17 | 12 | 36 | 11 | 12 | 7 | 30 | 6 | 0 | 67.19 | 83.33 | | 8 | *1/*2 | 12 | 36 | 12 | 12 | 12 | 36 | 0 | 0 | 90.26 | 100.00 | | 9 | *1/*1 | 12 | 36 | 9 | 12 | 11 | 32 | 4 | 0 | 73.94 | 88.89 | | 10 | *2/*10 | 12 | 36 | 10 | 12 | 10 | 32 | 4 | 0 | 73.94 | 88.89 | | 11 | *2/*17 | 12 | 36 | 10 | 12 | 11 | 33 | 3 | 0 | 77.53 | 91.26 | | 12 | *1/*5 | 12 | 36 | 8 | 12 | 12 | 32 | 4 | 0 | 73.94 | 88.89 | | 13 | *2/*2 | 12 | 36 | 11 | 12 | 12 | 35 | 1 | 0 | 85.47 | 97.22 | 1 across all three sites Overall percent agreement after the first test was 94.44% with a 95% lower bound confidence interval of 91.97%. At Site 1, one replicate of one sample required two re-tests to resolve the No Call. The root cause was identified to be extraction failure. All other No Calls were resolved after one re-test. All Site 3 No Calls were resolved after one re-test. {6} Reproducibility results presented (# of correct calls per 12 replicates) by site, operator and lot: | Sample | Genotype | Site 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Operator 1 | | | Operator 2 | | | | | | Day A Lot 1 | Day B Lot 2 | Day C Lot 3 | Day A Lot 1 | Day B Lot 2 | Day C Lot 3 | | 1 | *1/*5 | 12 | 12 | 12 | 12 | 12 | 12 | | 2 | *1/*41 | 12 | 12 | 12 | 12 | 12 | 12 | | 3 | *4/*35 | 12 | 11 | 12 | 12 | 12 | 12 | | 4 | *2/*4, DUP | 11 | 12 | 12 | 12 | 11 | 12 | | 5 | *35/*41 | 12 | 12 | 12 | 12 | 12 | 12 | | 6 | *1/*10 | 12 | 12 | 12 | 12 | 12 | 12 | | 7 | *10/*17 | 12 | 12 | 12 | 12 | 12 | 11 | | 8 | *1/*2 | 12 | 12 | 12 | 12 | 12 | 12 | | 9 | *1/*1 | 11 | 10 | 12 | 12 | 12 | 12 | | 10 | *2/*10 | 12 | 12 | 10 | 12 | 12 | 12 | | 11 | *2/*17 | 11 | 11 | 12 | 12 | 12 | 12 | | 12 | *1/*5 | 12 | 12 | 12 | 11 | 11 | 10 | | 13 | *2/*2 | 11 | 12 | 12 | 12 | 12 | 12 | | Sample | Genotype | Site 2 | | | | | | | 1 | *1/*5 | 12 | 12 | 12 | 12 | 12 | 12 | | 2 | *1/*41 | 12 | 12 | 12 | 12 | 12 | 12 | | 3 | *4/*35 | 12 | 12 | 12 | 12 | 12 | 12 | | 4 | *2/*4, DUP | 12 | 12 | 12 | 12 | 12 | 12 | | 5 | *35/*41 | 12 | 12 | 12 | 12 | 12 | 12 | | 6 | *1/*10 | 12 | 12 | 12 | 12 | 12 | 12 | | 7 | *10/*17 | 12 | 12 | 12 | 12 | 12 | 12 | | 8 | *1/*2 | 12 | 12 | 12 | 12 | 12 | 12 | | 9 | *1/*1 | 12 | 12 | 12 | 12 | 12 | 12 | | 10 | *2/*10 | 12 | 12 | 12 | 12 | 12 | 12 | | 11 | *2/*17 | 12 | 12 | 12 | 12 | 12 | 12 | | 12 | *1/*5 | 12 | 12 | 12 | 12 | 12 | 12 | | 13 | *2/*2 | 12 | 12 | 12 | 12 | 12 | 12 | {7} | Sample | Genotype | Site 3 | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Operator 1 | | | Operator 2 | | | | | | Day A Lot 1 | Day B Lot 2 | Day C Lot 3 | Day A Lot 1 | Day B Lot 2 | Day C Lot 3 | | 1 | *1/*5 | 12 | 12 | 12 | 12 | 12 | 12 | | 2 | *1/*41 | 12 | 12 | 12 | 12 | 12 | 12 | | 3 | *4/*35 | 12 | 12 | 12 | 12 | 12 | 12 | | 4 | *2/*4, DUP | 12 | 12 | 12 | 12 | 12 | 12 | | 5 | *35/*41 | 12 | 12 | 12 | 11 | 12 | 12 | | 6 | *1/*10 | 12 | 12 | 12 | 12 | 12 | 12 | | 7 | *10/*17 | 12 | 10 | 10 | 12 | 12 | 11 | | 8 | *1/*2 | 12 | 12 | 12 | 12 | 12 | 12 | | 9 | *1/*1 | 12 | 12 | 12 | 12 | 11 | 12 | | 10 | *2/*10 | 12 | 12 | 11 | 11 | 12 | 12 | | 11 | *2/*17 | 12 | 12 | 12 | 12 | 12 | 11 | | 12 | *1/*5 | 12 | 12 | 12 | 12 | 12 | 12 | | 13 | *2/*2 | 12 | 12 | 12 | 12 | 12 | 12 | b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Quality control materials are not provided with the kit. The package insert states that previously characterized clinical samples or commercially available 2D6 controls are recommended for quality control testing. All quality control requirements and testing should be performed in conformance with local, state and/or federal regulations or requirements. Stability: The stability performance of the xTAG® CYP2D6 Kit v3 was evaluated using three kit lots. The stability protocol includes testing under real-time and accelerated conditions. Real-time data is based on kit components stored at the recommended temperature of $-20^{\circ}\mathrm{C}$ . Accelerated data is based on stressed conditions where components are stored at $25^{\circ}\mathrm{C}$ . The current claimed shelf-life for the xTAG® CYP2D6 Kit v3 is 12 months. Real-time stability testing is on-going. In addition, repeated freeze-thaw cycles (up to 6) of the xTAG CYP2D6 Kit v3 are acceptable; stability testing protocols, acceptance criteria and summary of the data was reviewed and found to be acceptable. d. Detection limit: The limit of detection (LoD) and range of the xTAG® CYP2D6 Kit v3 was assessed by analyzing serial dilutions of genomic DNA samples from cell lines and whole blood samples containing one or more mutations or polymorphisms analyzed by the assay. DNA concentrations tested ranged {8} from 0.1 – 300 ng/uL. Determination of the DNA range of the assay was performed at 1 site, with 3 operators running a total of five samples across three consecutive days, 2 lots of reagents, and 3 different sets of equipment. Genotypes for all samples were established by bi-directional DNA sequencing. Five samples used in the study had the following genotypes: *1/*41, *1/*5, *35/*41, *17/*29, and *2/*4 DUP. Results for the Limit of Detection study: | Input DNA (ng/μL) | Number of Samples | # Correct Calls | # Incorrect Calls | # No calls | % Agreement | Lower Bound of 95% Confidence Interval | Upper Bound of 95% Confidence Interval | | --- | --- | --- | --- | --- | --- | --- | --- | | 100 | 160 | 160 | 0 | 0 | 100.00 | 97.72 | 100.00 | | 82.5^{1} | 40 | 40 | 0 | 0 | 100.00 | 91.19 | 100.00 | | 50 | 200 | 199 | 0 | 1 | 99.50 | 97.25 | 99.99 | | 10 | 200 | 200 | 0 | 0 | 100.00 | 98.17 | 100.00 | | 5 | 200 | 196 | 0 | 4 | 98.00 | 94.96 | 99.45 | | 2 | 200 | 200 | 0 | 0 | 100.00 | 98.17 | 100.00 | | 1 | 200 | 200 | 0 | 0 | 100.00 | 98.17 | 100.00 | | 0.1 | 200 | 105 | 1^{2} | 94 | 52.50 | 45.34 | 59.59 | 1 The highest DNA concentration extracted from the whole blood sample was 82.5 ng/μL. 2 For one replicate, the *1/*5 sample was incorrectly called as *1/*1. The recommended DNA concentration for this assay is 10 ng/μL and the lowest level of DNA that should be used is 2 ng/μL. e. Analytical specificity: Six patient blood samples were tested in this study. The genotypes of the samples were: *1/*1, *1/*5, *2/*2, *1/*10, *2/*17 and *10/*17. A given blood sample was split into 6 equivalent parts. One part was bi-directionally sequenced. One part was left as a control. The other parts of the blood sample were mixed, respectively, with hemoglobin, albumin, bilirubin, or triglycerides. Each was then split into three, and underwent DNA extraction by one of three different DNA extraction methods. The following potential interferents and the levels tested were: 500 mg/dL hemoglobin, 20 mg/dL bilirubin, 3000 mg/dL albumin and 2000 mg/dL triglycerides. Genotypes for all samples were established by bi-directional DNA sequencing. None of the substances tested interfered with the assay, since test results of all samples that had added compounds were 100% in agreement with the expected genotype determined by sequencing. {9} f. Assay cut-off: Not applicable. ## 2. Comparison studies: a. Method comparison with predicate device: Four hundred and fifty-nine clinical samples were analyzed in the method comparison study. Samples were either EDTA or citrate whole blood samples. Results using the xTAG® CYP2D6 Kit v3 was compared to bidirectional dideoxy sequencing. The percent agreement for genotype detection of the xTAG® CYP2D6 Kit v3 was calculated by determining the percentage of tested samples with the correct genotype assigned, compared to the total number of samples tested of that genotype. | CYP2D6 Allele Star (*) Genotype | Number of Alleles Sequenced | xTAG® CYP2D6 Kit v3 First Run Results | | | | | --- | --- | --- | --- | --- | --- | | | | # Correct Calls | # Incorrect Calls | # No Calls | % Agreement | | *1 | 199 | 198 | 0 | 1 | 99.45 | | *2 | 143 | 139 | 1¹ | 3 | 97.20 | | *3 | 15 | 15 | 0 | 0 | 100.00 | | *4 | 124 | 123 | 0 | 1 | 99.14 | | *5 | 59 | 57 | 0 | 2 | 96.61 | | *6 | 16 | 16 | 0 | 0 | 100.00 | | *7 | 2 | 2 | 0 | 0 | 100.00 | | *9 | 28 | 28 | 0 | 0 | 100.00 | | *10 | 82 | 79 | 0 | 3 | 95.12 | | *11 | 2 | 2 | 0 | 0 | 100.00 | | *15 | 1 | 1 | 0 | 0 | 100.00 | | *17 | 73 | 72 | 0 | 1 | 98.61 | | *29 | 23 | 22 | 0 | 1 | 95.64 | | *35 | 45 | 45 | 0 | 0 | 100.00 | | *41 | 106 | 106 | 0 | 0 | 100.00 | | Total | 918 | 905 | 1 | 12 | 98.4 | ¹ Incorrect call for *2/*10; result given by xTAG® CYP2D6 Kit v3 was *10/*35 because of a HET call at the 31 G>A loci {10} Results by sample genotype: | CYP2D6 Genotype | Total Unique Samples | xTAG® CYP2D6 Kit v3 First Run Results | | | | | --- | --- | --- | --- | --- | --- | | | | # Correct Calls | # No Calls | # Incorrect Calls | % Agreement | | *1 / *1 | 17 | 17 | 0 | 0 | 100 | | *1 / *1, DUP | 1 | 1 | 0 | 0 | 100 | | *1 / *10 | 16 | 16 | 0 | 0 | 100 | | *1 / *17 | 21 | 21 | 0 | 0 | 100 | | *1 / *2 | 29 | 28 | 1 | 0 | 96.55 | | *1 / *2, DUP | 5 | 5 | 0 | 0 | 100 | | *1 / *29 | 6 | 6 | 0 | 0 | 100 | | *1 / *29, DUP | 1 | 1 | 0 | 0 | 100 | | *1 / *3 | 6 | 6 | 0 | 0 | 100 | | *1 / *3, DUP | 1 | 1 | 0 | 0 | 100 | | *1 / *35 | 12 | 12 | 0 | 0 | 100 | | *1 / *4 | 25 | 25 | 0 | 0 | 100 | | *1 / *41 | 17 | 17 | 0 | 0 | 100 | | *1 / *5 | 13 | 13 | 0 | 0 | 100 | | *1 / *5, DUP | 1 | 1 | 0 | 0 | 100 | | *1 / *6 | 5 | 5 | 0 | 0 | 100 | | *1 / *9 | 5 | 5 | 0 | 0 | 100 | | *10 / *10 | 18 | 18 | 0 | 0 | 100 | | *10 / *17 | 8 | 7 | 1 | 0 | 87.50 | | *10 / *35 | 1 | 1 | 0 | 0 | 100 | | *10 / *41 | 5 | 5 | 0 | 0 | 100 | | *15 / *41 | 1 | 1 | 0 | 0 | 100 | | *17 / *17 | 4 | 4 | 0 | 0 | 100 | | *17 / *29 | 4 | 4 | 0 | 0 | 100 | | *17 / *29, DUP | 1 | 1 | 0 | 0 | 100 | | *17 / *41 | 4 | 4 | 0 | 0 | 100 | | *2 / *10 | 6 | 4 | 1 | 1¹ | 66.67 | | *2 / *17 | 10 | 10 | 0 | 0 | 100 | | *2 / *17, DUP | 1 | 1 | 0 | 0 | 100 | | *2 / *2 | 7 | 7 | 0 | 0 | 100 | | *2 / *2, DUP | 3 | 3 | 0 | 0 | 100 | | *2 / *29 | 3 | 3 | 0 | 0 | 100 | | *2 / *3 | 1 | 1 | 0 | 0 | 100 | | *2 / *35 | 6 | 6 | 0 | 0 | 100 | | *2 / *11 | 2 | 2 | 0 | 0 | 100 | | *2 / *4 | 21 | 21 | 0 | 0 | 100 | | *2 / *41 | 14 | 14 | 0 | 0 | 100 | ¹ The number of samples was calculated from the mean ± SD of the mean ± SD of the 2000 samples. {11} | *2 / *41, DUP | 2 | 2 | 0 | 0 | 100 | | --- | --- | --- | --- | --- | --- | | *2 / *5 | 9 | 8 | 1 | 0 | 88.89 | | *2 / *6 | 3 | 3 | 0 | 0 | 100 | | *2 / *7 | 2 | 2 | 0 | 0 | 100 | | *2 / *9 | 9 | 9 | 0 | 0 | 100 | | *29 / *41 | 3 | 3 | 0 | 0 | 100 | | *3 / *3 | 1 | 1 | 0 | 0 | 100 | | *3 / *35 | 2 | 2 | 0 | 0 | 100 | | *3 / *4 | 1 | 1 | 0 | 0 | 100 | | *3 / *5 | 1 | 1 | 0 | 0 | 100 | | *3 / *9 | 1 | 1 | 0 | 0 | 100 | | *35 / *35 | 2 | 2 | 0 | 0 | 100 | | *35 / *41 | 6 | 6 | 0 | 0 | 100 | | *4 / *10 | 6 | 6 | 0 | 0 | 100 | | *4 / *17 | 4 | 4 | 0 | 0 | 100 | | *4 / *17, DUP | 2 | 2 | 0 | 0 | 100 | | *4 / *29 | 1 | 1 | 0 | 0 | 100 | | *4 / *29 DUP | 2 | 1 | 1 | 0 | 50.00 | | *4 / *35 | 7 | 7 | 0 | 0 | 100 | | *4 / *35, DUP | 1 | 1 | 0 | 0 | 100 | | *4 / *4 | 12 | 12 | 0 | 0 | 100 | | *4 / *4, DUP | 1 | 1 | 0 | 0 | 100 | | *4 / *41 | 20 | 20 | 0 | 0 | 100 | | *4 / *5 | 3 | 3 | 0 | 0 | 100 | | *4 / *5, DUP | 2 | 2 | 0 | 0 | 100 | | *4 / *6 | 1 | 1 | 0 | 0 | 100 | | *4 / *9 | 2 | 2 | 0 | 0 | 100 | | *41 / *41 | 13 | 13 | 0 | 0 | 100 | | *5 / *10 | 4 | 3 | 1 | 0 | 75.00 | | *5 / *17 | 10 | 10 | 0 | 0 | 100 | | *5 / *29 | 2 | 2 | 0 | 0 | 100 | | *5 / *35 | 3 | 3 | 0 | 0 | 100 | | *5 / *41 | 2 | 2 | 0 | 0 | 100 | | *5 / *5 | 3 | 3 | 0 | 0 | 100 | | *5 / *6 | 3 | 3 | 0 | 0 | 100 | | *6 / *35 | 2 | 2 | 0 | 0 | 100 | | *6 / *6 | 1 | 1 | 0 | 0 | 100 | | *9 / *35 | 1 | 1 | 0 | 0 | 100 | | *9 / *41 | 6 | 6 | 0 | 0 | 100 | | *9 / *9 | 2 | 2 | 0 | 0 | 100 | | Total | 459 | 446 | 6 | 1 | 98.4² | 12 {12} 13 1 Incorrect call for *2/*10; result given by xTAG® CYP2D6 Kit v3 was *10/*35 because of a HET call at the 31 G>A loci 2 95% Lower Bound Confidence Interval = 96.89% For rare alleles, the performance of the xTAG® CYP2D6 Kit v3 was evaluated by blending plasmids with mutations for rare alleles with homozygous genomic DNA. Plasmid clones for the *8 allele were created by blending plasmids harboring mutations for rare alleles with genomic DNA. A total of 5 blends of genomic DNA and plasmid were tested. Four replicates of each blend were amplified (total of 20 samples) and assayed by xTAG® CYP2D6 Kit v3. Genotype results using plasmid clone-genomic DNA blends are shown in the table below. The CYP2D6 genotype call rate was 100% for plasmid clone genomic DNA blends. | Plasmid Genotype | Genomic DNA Genotype | Blended sample Expected genotype | Blended samples (n) | # Correct Calls | # Incorrect calls | # No calls | % Agreement | | --- | --- | --- | --- | --- | --- | --- | --- | | *8 | *1/*1 | *1/*8 | 4 | 4 | 0 | 0 | 100 | | *8 | *35/*2 | *35/*8 | 4 | 4 | 0 | 0 | 100 | | *8 | *17/*17 | *17/*8 | 4 | 4 | 0 | 0 | 100 | | *8 | *2/*2 | *2/*8 | 4 | 4 | 0 | 0 | 100 | | *8 | *4/*4 | *4/*8 | 4 | 4 | 0 | 0 | 100 | | Total | | | 20 | 20 | 0 | 0 | 100 | b. Matrix comparison: Twenty five specimens were tested that had been collected both in EDTA and citrate, extracted and run on the Luminex 2D6 assay. There were no No Calls or Incorrect calls. In addition, citrate anticoagulated whole blood samples were tested in the method comparison study in 2a above. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. {13} # 5. Expected values/Reference range: | Final TDAS Output (*genotype or * allele) | SNPs that xTAG® CYP2D6 v3 kit detects | Frequency in the U.S. Caucasian population1 | Frequency in the African American population2 | Predicted Enzyme Activity | Reference where the effect of the genotype on drug metabolism is described | | --- | --- | --- | --- | --- | --- | | *1 | None | 37 – 40% | 29 – 35% | Normal | (Kimura, Umeno et al. 1989; Marez, Legrand et al. 1997; Sachse, Brockmoller et al. 1997) | | *2 | -1584C>G, 1661G>C, 2850C>T, 4180G>C | 26 - 33% | 18 - 27% | Normal | (Johansson, Lundqvist et al. 1993; Panserat, Mura et al. 1994; Marez, Legrand et al. 1997; Raimundo, Fischer et al. 2000; Sakuyama, Sasaki et al. 2008) | | *3 | 2549A>del | 1% | 0.2%-0.6% | None | (Kagimoto, Heim et al. 1990; Marez, Legrand et al. 1997) | | *4 | 100C>T, 1661G>C, 1846G>A, 4180G>C, 2850C>T | 18 - 20% | 6 - 9% | None | (Gough, Miles et al. 1990; Hanioka, Kimura et al. 1990; Kagimoto, Heim et al. 1990; Sachse et al, 1997; Marez et al, 1997) | | *5 | deletion | 2 - 4% | 6 - 7% | None | (Gaedigk, Blum et al. 1991; Steen, Molven et al. 1995) | | *6 | 1707T>del, 4180G>C | 1% | 0.5% | None | (Evert, Griese et al. 1994; Saxena, Shaw et al. 1994; Daly, Leathart et al. 1995; Marez, Legrand et al. 1997) | | *7 | 2935A>C 1661G>C, 1758G>T, 2850C>T, 4180G>C | Not known | Not known | None | (Evert, Griese et al. 1994) | | *8 | 1661G>C, 1758G>T, 2850C>T, 4180G>C | Not known | Not known | None | (Broly, Marez et al. 1995) | {14} | *9 | 2613delAGA | 2 - 3% | 0.3% | Reduced | (Tyndale, Aoyama et al. 1991; Broly and Meyer 1993) | | --- | --- | --- | --- | --- | --- | | *10 | 100C>T, 1661G>C, 4180G>C | 2 - 8% | 0.3%-0.4% | Reduced | (Yokota, Tamura et al. 1993; Johansson, Oscarson et al. 1994; Ishiguro, Kubota et al. 2004; Sakuyama, Sasaki et al. 2008) | | *11 | 883G>C, 1661G>C, 2850C>T, 4180G>C | Not known | Not known | None | (Marez, Sabbagh et al. 1995) | | *15 | 138insT | Not known | Not known | None | (Sachse, Brockmoller et al. 1996) | | *17 | 1023C>T, 1661G>C, 2850C>T, 4180G>C | 0.2-0.3% | 15 - 26% | Reduced | (Masimirembwa, Persson et al. 1996; Oscarson, Hidestrand et al. 1997) | | *29 | 1659G>A, 1661G>C, 2850C>T, 3183G>A, 4180G>C | Not known3 | Not known3 | Reduced | (Marez, Legrand et al. 1997; Wennerholm, Johansson et al. 2001; Wennerholm, Dandara et al. 2002) | | *35 | -1584C>G, 31G>A, 1661G>C, 2850C>T, 4180G>C | 7.4%4 | 1%4 | Normal | (Marez, Legrand et al. 1997; Gaedigk, Ryder et al. 2003) | | *41 | 1661G>C, 2850C>T, 2988G>A, 4180G>C | 9%4 | 11%4 | Reduced | (Raimundo, Fischer et al. 2000; Raimundo, Toscano et al. 2004) | $^{1,2}$ Bradford, L. D. (2002). "CYP2D6 allele frequency in European Caucasians, Asians, Africans and their descendants." Pharmacogenomics 3(2): 229-43. 3 The frequency of the *29 allele in the U.S. population is not known; however, it is very common in Tanzanian Africans, with an allele frequency of 20 percent (Wennerholm, A., I. Johansson, et al. (2001). "Characterization of the CYP2D6*29 allele commonly present in a black Tanzanian population causing reduced catalytic activity." Pharmacogenetics 11 (5): 417-27). 4 Gaedigk, A., D. L. Ryder, et al. (2003). "CYP2D6 poor metabolizer status can be ruled out by a single genotyping assay for the -1584G promoter polymorphism." Clinical Chemistry 49(6 Pt 1): 1008-11. {15} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 16
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...