PHAMATECH QUICKSCREEN BENZODIAZEPINES TEST, MODELS 9025, 9026, 9027T, 9153T AND 9195X

K043167 · Phamatech, Inc. · JXM · Dec 29, 2004 · Clinical Toxicology

Device Facts

Record IDK043167
Device NamePHAMATECH QUICKSCREEN BENZODIAZEPINES TEST, MODELS 9025, 9026, 9027T, 9153T AND 9195X
ApplicantPhamatech, Inc.
Product CodeJXM · Clinical Toxicology
Decision DateDec 29, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3170
Device ClassClass 2

Intended Use

The QuickScreen™ Benzodiazepines Test is a rapid, qualitative immunoassay for the detection of benzodiazepines in urine. The cutoff concentration for this test is 200 ng/ml. This assay is intended for professional use.

Device Story

QuickScreen™ Benzodiazepines Test is a rapid, qualitative, one-step immunoassay for urine drug screening. Device utilizes visual color sandwich immunochemical technology to detect benzodiazepines at a 200 ng/ml cutoff. Professional users perform the test in clinical settings. The assay relies on specific antibody-target analyte-antibody complexes to produce a visual result. Healthcare providers use the qualitative output to aid in the diagnosis and treatment of drug abuse. The device provides a rapid screening tool for clinical decision-making regarding patient drug use status.

Clinical Evidence

Clinical performance evaluated via clinical sample correlation study and blind labeled spiked study. Results showed 97.9% correlation with predicate devices for the qualitative detection of benzodiazepines in urine.

Technological Characteristics

Rapid, qualitative, one-step visual color sandwich immunoassay. Designed for professional use. Detects benzodiazepines at 200 ng/ml cutoff. No electronic components, software, or external energy sources required.

Indications for Use

Indicated for professional use in the qualitative detection of benzodiazepines (and other drugs of abuse including amphetamine, cocaine, methamphetamine, opiates, PCP, barbiturates, methadone, and THC depending on model) in human urine to assist in the diagnosis and treatment of drug abuse.

Regulatory Classification

Identification

A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # DEC 2 9 2004 K043/67 #### 510 (k) SUMMARY AS REQUIRED BY SECTION 807.92(C) QuickScreen™ Benzodiazepines Test (Models 9025, 9026, 9027T, 9153 and ldentification: 9195X) Description: Immunoassay for the qualitative detection of Benzodiazepines in urine | Name Of Manufacturer: | Phamatech<br>10151 Barnes Canyon Road<br>San Diego, California 92121, USA | |-----------------------|---------------------------------------------------------------------------| |-----------------------|---------------------------------------------------------------------------| Intended Use: The QuickScreen™ Benzodiazepines Test is a rapid, qualitative immunoassay for the detection of benzodiazepines in urine. The cutoff concentration for this test is 200 ng/ml. This assay is intended for professional use. The QuickScreen™ Benzodiazepines Test, like many commercially available Technology: oxycodone screening test kits, qualitatively measures the presence of benzodiazepines by visual color sandwich one step immunoassay technology. Examples of such predicate devices include the ABMC RapidOne BZD test (Kinderhook, NY). These devices rely on the basic immunochemical sandwich assay principle of recognition of specific antibody / target analyte / antibody / complexes. The product performance characteristics of the QuickScreen™ Performance: Benzodiazepines Test were evaluated in a clinical sample correlation study and a blind labeled spiked study. The results of these studies demonstrate the QuickScreen™ Benzodiazepines Test to be substantially equivalent to the reported performance characteristics of other commercially available tests for the qualitative detection of benzodiazepines in urine. Laboratory studies, using clinical specimens, produced a 97.9% correlation when compared to the predicate devices. For the reasons mentioned above, it may be concluded that the Conclusion: Phamatech QuickScreen™ Beenzodiazepines Test is substantially equivalent to a variety of detection tests currently in commercial distribution and is safe in the hands of the professional user. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized image of what appears to be a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 DEC 2 9 2004 Mr. Carl A. Mongiovi Vice President Phamatech, Inc. 10151 Barnes Canyon Road San Diego, CA 92121 Re: k043167 Trade/Device Name: Phamatech QuickScreen™ Benzodiazepines Test Models 9025, 9026 and 9027T Phamatech QuickScreen™ Pro Multi Drug Screening Test Models 9153T Phamatech QuickScreen™ Pro Drug Cup Model 9195X Regulation Number: 21 CFR 862.3170 Regulation Name: Benzodiazepine test system Regulatory Class: Class II Product Code: JXM Dated: December 21, 2004 Received: December 22, 2004 Dear Mr. Mongiovi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours Corneria B. Rooks, MA Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### INDICATIONS FOR USE Applicant: Phamatech 510 (k) Number (if known): __________________________________________________________________ Device Name: Phamatech QuickScreen™ Benzodiazepines Test Models 9026 and 9027T Indications for Use: The QuickScreen Benzodiazepines Test is an in-vitro diagnostic screen that provides a rne QuiloRooroon Bonzoulatespines of benzodiazepines in urine. It is intended for professional use only. ____ AND/OR Prescription Use: ____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D 510/K Over the Counter Use: (21 CFR 807 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of the CDRH Office off In Vitro Diagnostic Devices (OIVD) *Acting* *Division/Sign-Off* Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1 {4}------------------------------------------------ ### INDICATIONS FOR USE Applicant: Phamatech 510 (k) Number (if known): Kou 3167- Device Name: Phamatech QuickScreen™ Pro Multi Drug Screening Test Model 9153T Indications for Use: An invitro diagnostic test for the qualitative detection of amphetamine, cocaine, methamphetamine, opiates, PCP, benzodiazepines, barbiturates, methadone and THC in urine. Measurements obtained by this device are used in the diagnosis and treatment of drug abuse. It is intended for professional use only. AND/OR X Prescription Use: (Part 21 CFR 801 Subpart D Over the Counter Use: (21 CFR 807 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of the CDRH Office of In Vitro Diagnostic Devices (OIVD) Division/Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K043167 Page 1 of 1 {5}------------------------------------------------ #### INDICATIONS FOR USE Applicant: Phamatech 510 (k) Number (if known): _K 043 (67- - Device Name: Phamatech QuickScreen™ Pro Drug Cup Model 9195X Indications for Use: An invitro diagnostic test for the qualitative detection of amphetamine, cocaine, An Inntro ultagines, PCP, benzodiazepines, barbiturates, methadone and THC in urine. Measurements obtained by this device are used in the diagnosis and treatment of drug abuse. It is intended for professional use only. _X____________________________________________________________________________________________________________________________________________________________________________ AND/OR Prescription Use: (Part 21 CFR 801 Subpart D Over the Counter Use: (21 CFR 807 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of the CDRH Office of In Vitro Diagnostic Devices (OIVD) Division/Sign-6 Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K043167 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...