This product is intended for in vitro diagnostic use as a control product to assess the analytical performance of immunochemistry systems and clinical chemistry systems using methods for the quantitative measurement of the listed analytes. It is intended to provide the necessary control material required by inspection agencies for the evaluation of system performance. In addition, it will provide assistance when troubleshooting chemistry systems, reagent problems and calibration anomalies.
Device Story
DOCUMENT™ TDM ULTRA Assayed Control is a liquid human serum matrix containing 25 analytes; used as a quality control material for automated and semi-automated immunochemistry and clinical chemistry systems. The device provides a stable, ready-to-use control to assess analytical performance, monitor accuracy/precision, and assist in troubleshooting system, reagent, or calibration issues. It is used by laboratory personnel in clinical settings to verify that diagnostic systems are functioning within expected parameters for the listed therapeutic drugs, thyroid hormones, and ethanol.
Clinical Evidence
Bench testing only. Equivalence established by comparing inter-assay precision of the 25 analytes in the DOCUMENT™ TDM ULTRA Assayed Control against the performance of the three predicate devices. Results indicate the subject device behaves in a similar manner to the predicates and is suitable for use as a control for the listed analytes.
Technological Characteristics
Liquid human serum matrix; contains 25 analytes; provided in three levels (I, II, III); packaged in plastic containers with dropper tips; ready-to-use (no dilution required); 30-day open-vial stability.
Indications for Use
Indicated for use as an in vitro diagnostic control material to assess the analytical performance of immunochemistry and clinical chemistry systems for the quantitative measurement of 25 analytes: acetaminophen, amikacin, carbamazepine, digoxin, disopyramide, ethanol, ethosuximide, gentamicin, lidocaine, methotrexate, N-acetylprocainamide, phenobarbital, phenytoin, primidone, procainamide, quinidine, salicylate, theophylline, thyroid-stimulating hormone, thyroxine, tobramycin, triiodothyronine, T-Uptake, valproic acid, and vancomycin.
Regulatory Classification
Identification
A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.
Predicate Devices
Dade Immunoassay Controls Comprehensive Tri-Level, Levels 1, 2, and 3
Instrumentation Laboratory TheraChem TDC Therapeutic Drug Controls, Low and High Levels
Abbott REA Ethanol Serum Control, Levels L, M, and H
Related Devices
K060429 — TDM CONTROL SET · Roche Diagnostics Corp. · Mar 15, 2006
K973894 — IMMUSURE · Quantimetrix Corp. · Oct 29, 1997
K050536 — QUEST DIAGNOSTICS IMMUNOASSAY/TDM CONTROL · Bio-Rad · Mar 24, 2005
Submission Summary (Full Text)
{0}
K96C344 MAR 22 13:51
Exhibit 15. 510(k) Summary
FEDERAL FOOD, DRUG AND COSMETIC ACT
510(k) SUMMARY
DOCUMENT™ TDM ULTRA Assayed Control, Levels I, II, III
1. Submitted by: CASCO Standards
P.O. Box 970
Yarmouth, ME 04096
Attention: Christine V. Beach
Manager, Regulatory Affairs
(207) 878-7550
(207) 878-7578 FAX
January 5, 1996
2. Product Name:
Proprietary Name: DOCUMENT™ TDM ULTRA Assayed Control
Classification Name: Control, multiple analyte, assayed.
Page • 80 •
{1}
Page • 81 •
# 3. Predicate Device:
Product: Dade Immunoassay Controls Comprehensive Tri-Level, Levels 1, 2, and 3
Catalog: Level 1 (B5700-6), Level 2 (B5700-7), Level 3 (B5700-8)
Manufacturer: BaxterDiagnostics, Inc., Deerfield, IL 60015-4633
Analytes: ACET, AMI, CARB, DIG, DISO, ETHO, GENT, LIDO, NAPA, PHENO, PHENY, PRIM, PROC, QUIN, SAL, THEO, hTSH, T4, TOB, T3, TU, VAL
Product: Instrumentation Laboratory TheraChem TDC Therapeutic Drug Controls, Low and High Levels
Catalog: Low (2841-31), High (2842-31)
Manufacturer: Instrumentation Laboratory Company, Lexington, MA 02173-3190
Analytes: METH, VANCO
Product: Abbott REA Ethanol Serum Control, Levels L, M, and H
Catalog: 9545-11
Manufacturer: Abbott Laboratories, Abbott Park, IL 60064
Analytes: ETOH
# 4. Product Description:
**DOCUMENT® TDM ULTRA Assayed Controls** consist of three levels of a human serum matrix containing 25 commonly monitored analytes. The formulation design provides a liquid matrix intended for use on automated and semi-automated immunochemistry systems and clinical chemistry systems for the determination of blood acetaminophen (ACET), amikacin (AMI), carbamazepine (CARB), digoxin (DIG), disopyramide (DISO), ethanol (ETOH), ethosuximide (ETHO), gentamicin (GENT), lidocaine (LIDO), methotrexate (METH), N-acetylprocainamide (NAPA), phenobarbital (PHENO), phenytoin (PHENY), primidone (PRIM), procainamide (PROC), quinidine (QUIN), salicylate (SAL), theophylline (THEO), thyroid-stimulating hormone (hTSH), thyroxine (T4), tobramycin (TOB), triiodothyronine (T3), T-Uptake (TU), valproic acid (VAL), and vancomycin (VANCO).
# 5. Intended Use:
This product is intended for *in vitro* diagnostic use as a control product to assess the analytical performance of immunochemistry systems and clinical chemistry systems using methods for the quantitative measurement of the listed analytes. It is intended to provide the necessary control material required by inspection agencies for the evaluation of system performance. In addition, it will provide assistance when troubleshooting chemistry systems, reagent problems and calibration anomalies.
{2}
6. Comparison to the Predicate Device:
| Characteristic | DOCUMENT®
TDM ULTRA
Assayed Control | Dade
Immunoassay
Comprehensive
Tri-Level Control | Instrumentation
Laboratories
TheraChem TDC
Therapeutic Drug
Control | Abbott REA
Ethanol Serum
Control |
| --- | --- | --- | --- | --- |
| Intended Use | To assess the analytical performance of analytical systems in the quantitative measurement of therapeutic drugs, thyroid function, and ethanol | Assist in monitoring accuracy and precision in clinical assays | Control material for procedures used to quantitate levels of therapeutic drugs and thyroid hormones in serum specimens | Controls for the validation of calibration of the REA Ethanol assay on the TDxFLx, TDx, or ADx |
| Number of Levels | Three (3):
Level I, Level II
Level III | Three (3):
Level 1, Level 2,
Level 3 | Two (2):
Low, High | Three (3):
L, M, H |
| Type | Assayed | Assayed | Assayed | Assayed |
| Analytes | 25 | 64 | 30 | 1 |
| Volume | 5 mL | 5 mL | 5 mL | 4 mL |
| Matrix | Human serum,
Liquid | Human source,
Lyophilized | Bovine serum,
Lyophilized | Human blood components,
Liquid |
| Dilution | None required | None required | None required | None required |
| Unopened Stability | Until Expiration Date | Until Expiration Date | Until Expiration Date | Until Expiration Date |
| Open Stability | 30 Days | Varies with analyte following reconstitution | Up to four weeks after reconstitution | Until expiration date |
| Container | Plastic,
Dropper Tip | Glass | Glass | Plastic |
Page • 82 •
{3}
Page • 83 •
# 7. Test Results:
The equivalence for this product was carried out by comparing the inter-assay precision of the listed analytes of the **DOCUMENT™** TDM ULTRA Assayed Control to that of Dade Immunoassay Controls Comprehensive Tri-Level, Instrumentation Laboratory TheraChem TDC Therapeutic Drug Controls and Abbott REA Ethanol Serum Control. The results show that the **DOCUMENT™** TDM ULTRA Assayed Control behaves in a similar manner compared to the predicate devices and is suitable for use as a control for the listed analytes.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.