LIQUICHEK URINE TOXICOLOGY CONTROL (SCREEN SERIES)
K033924 · Bio-Rad Laboratories, Inc. · DIF · Feb 3, 2004 · Clinical Toxicology
Device Facts
Record ID
K033924
Device Name
LIQUICHEK URINE TOXICOLOGY CONTROL (SCREEN SERIES)
Applicant
Bio-Rad Laboratories, Inc.
Product Code
DIF · Clinical Toxicology
Decision Date
Feb 3, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3280
Device Class
Class 1
Indications for Use
The product is intended for use as a quality control urine to monitor the performance of urine toxicology screening procedures. The device is for in vitro diagnostic use. The device is for prescription use.
Device Story
Liquichek Urine Toxicology Control is a liquid-form quality control material derived from human urine; supplemented with drugs of abuse, drug metabolites, preservatives, and animal-derived constituents. Used in clinical laboratories to monitor the performance of urine toxicology screening procedures and enzyme immunoassays (EIA). Healthcare providers use the control to verify the accuracy and reliability of their screening assays; ensuring consistent performance of diagnostic tests. The device provides a known reference point for laboratory personnel to validate test results, thereby supporting clinical decision-making regarding patient drug screening.
Clinical Evidence
No clinical data. Bench testing only. Stability studies performed using GC/MS to verify concentrations and monitor drift (acceptance criteria: 10% drift from time zero).
Technological Characteristics
Liquid-form human urine matrix; contains drugs of abuse, metabolites, preservatives (antibiotic cocktail), and animal-derived constituents. Storage at 2-8°C. No software or electronic components.
Indications for Use
Indicated for use as a quality control urine to monitor the performance of urine toxicology screening procedures in clinical laboratory settings.
Regulatory Classification
Identification
A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.
Predicate Devices
Liquichek Urine Toxicology Control Levels S1, S2, and S3 (K021411)
Related Devices
K033366 — LIQUICHEK QUALITATIVE URINE TOXICOLOGY CONTROL, NEGATIVE, MODEL 454; POSITIVE, MODEL 455 · Bio-Rad · Dec 12, 2003
K033404 — LIQUICHEK URINE TOXICOLOGY CONTROL (CONFIRMATORY SERIES) · Bio-Rad · Dec 12, 2003
K050682 — LIQUICHEK URINE TOXICOLOGY CONTROL (LEVEL C1) · Bio-Rad Laboratories, Inc. · Apr 29, 2005
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
K033924
B. Analyte:
Cannabinoids (delta-9-THC), amphetamines (d-amphetamine), barbiturates (secobarbital), benzodiazepines (nordiazepam), cocaine (benzoylecgonine), ethanol, lysergic acid diethlamide (LSD), methadone, methaqualone, opiates (morphine, free), PCP, propoxyphene, TCA (nortryptiline)
C. Type of Test:
Quality Control material
D. Applicant:
Bio-Rad Laboratories
E. Proprietary and Established Names:
Liquichek Urine Toxicology Control (Screen Series), to include:
Liquichek Urine Toxicology S1 Control
Liquichek Urine Toxicology S2 Control
Liquichek Urine Toxicology S3 Control
Liquichek Urine Toxicology S1 Low Opiate Control
Liquichek Urine Toxicology S2 Low Opiate Control
Liquichek Urine Toxicology S1E Control
Liquichek Urine Toxicology S2E Control
Liquichek Urine Toxicology S1E Low Opiate Control
Liquichek Urine Toxicology S2E Low Opiate Control
F. Regulatory Information:
1. Regulation section:
862.3280, Clinical Toxicology Control Material, Drug Mixture Control Materials
2. Classification:
I, reserved
3. Product Code:
DIF
4. Panel:
Toxicology (91)
{1}
Page 2 of 4
G. Intended Use:
1. Intended use(s):
Refer to Indications for use.
2. Indication(s) for use:
The product is intended for use as a quality control urine to monitor the performance of urine toxicology screening procedures.
The device is for in vitro diagnostic use.
The device is for prescription use.
3. Special condition for use statement(s):
None.
4. Special instrument Requirements:
Not applicable.
H. Device Description:
The product is a liquid ready to use quality control material.
I. Substantial Equivalence Information:
1. Predicate device name(s):
Liquichek Urine Toxicology Control Levels S1, S2, and S3
2. Predicate K number(s):
K021411
3. Comparison with predicate:
Both devices have the same matrix and both are quality control materials that measure the same analytes. The devices are different in that the candidate device has added two new levels; the S1E Low Opiate and S2E Low Opiate levels. The preservative of the candidate device is a cocktail of antibiotics not requiring the hazard symbol, whereas the predicate include 0.1% sodium azide which required a hazard symbol.
J. Standard/Guidance Document Referenced (if applicable):
The sponsor did not reference any standards in their submission.
K. Test Principle:
Not applicable.
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
{2}
Page 3 of 4
Not applicable.
c. Traceability (controls, calibrators, or method):
The product consists of nine levels of control material. Controls are human urine spiked with known concentrations of drug. The materials are gravimetrically prepared. The expected concentrations are verified by reference laboratories. The minimum number of laboratories, replicate measurements, or assay systems is not, however, specified.
Although GC/MS analysis is performed during stability studies, (see below), the traceability of control results to GC/MS analysis is not specified or discussed by the sponsor.
Stability studies are summarized for the controls. The sponsor specifies the frequency of testing, the method for testing the materials (GC/MS), environmental conditions of storage, and acceptance criteria for the study (10% drift from time zero). Accelerated studies are being used by the sponsor to estimate the expiration date, however, on-going real time studies are also being performed. All procedures appear to be standard for the industry.
Representative values of the materials are provided and seem appropriate.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
{3}
Page 4 of 4
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
M. Conclusion:
I recommend that this device be found substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.