STATUS DS NICOTINE, ACCUSIGN NICOTINE, BIOSIGN NICOTINE

K062565 · Princeton BioMeditech Corp. · MKU · Feb 20, 2007 · Clinical Toxicology

Device Facts

Record IDK062565
Device NameSTATUS DS NICOTINE, ACCUSIGN NICOTINE, BIOSIGN NICOTINE
ApplicantPrinceton BioMeditech Corp.
Product CodeMKU · Clinical Toxicology
Decision DateFeb 20, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3220
Device ClassClass 1

Indications for Use

Status DS™ Nicotine is designed for the qualitative detection of cotinine at the cutoff of 500 ng/mL in urine to assist in screening for exposure to nicotine. For In vitro Diagnostic, Professional Use. The Status DS™ Nicotine test provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography, mass spectrometry (GC/MS) is the preferred confirmatory method.

Device Story

Status DS Nicotine and AccuSign Nicotine are lateral flow immunochromatographic assays for qualitative cotinine detection in urine. Device consists of membrane strip coated with monoclonal anti-cotinine antibody and dye pad with cotinine-protein conjugate. Urine sample migrates up membrane; cotinine in sample competes with conjugate for limited antibody binding sites. If cotinine concentration is ≥500 ng/mL, antibodies are saturated, preventing dye-conjugate binding and resulting in absence of test line (T). Control line confirms proper sample flow. Used in clinical settings; operated by healthcare professionals. Provides preliminary results; requires GC/MS confirmation. Benefits include rapid, point-of-care screening for nicotine exposure.

Clinical Evidence

Bench testing only. Precision studies (n=120 per concentration) showed 100% agreement at 0 and 1000 ng/mL. Method comparison against GC/MS (n=86) demonstrated performance around the 500 ng/mL cutoff. No clinical studies performed.

Technological Characteristics

One-step immunochromatographic lateral flow assay. Qualitative detection of cotinine in urine. Cutoff concentration: 500 ng/mL.

Indications for Use

Indicated for the qualitative detection of cotinine, a major nicotine metabolite, in human urine at a 500 ng/mL cutoff to aid in detecting tobacco or nicotine product use. For in vitro diagnostic use.

Regulatory Classification

Identification

A carbon monoxide test system is a device intended to measure carbon monoxide or carboxyhemoglobin (carbon monoxide bound to the hemoglobin in the blood) in blood. Measurements obtained by this device are used in the diagnosis and treatment of or confirmation of carbon monoxide poisoning.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k062565 B. Purpose for Submission: New device C. Measurand: Cotinine in urine D. Type of Test: Qualitative, visually read immunochromatographic assay E. Applicant: Princeton Biomeditech Corp. F. Proprietary and Established Names: Status DS Nicotine AccuSign Nicotine G. Regulatory Information: 1. Regulation section: 21 CFR 862.3220, Carbon Monoxide test system 2. Classification: Class I, reserved 3. Product code: MKU, Enzyme Immunoassay, Nicotine and Nicotine Metabolites 4. Panel: {1} Toxicology (91) ## H. Intended Use: 1. Intended use(s): See Indications for use. 2. Indication(s) for use: ### Status DS Nicotine Immunoassay for the qualitative detection of cotinine, a metabolite of nicotine, at the cut-off of 500 ng/mL in human urine. Status DS Nicotine is used as an aid in the detection of cotinine after use of tobacco products or other products containing nicotine. For In Vitro Diagnostic Use. The Status DS Nicotine test provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography, mass spectrometry (GC/MS) is the preferred confirmatory method. ### AccuSign Nicotine Immunoassay for the qualitative detection of cotinine, a metabolite of nicotine, at the cut-off of 500 ng/mL in human urine. AccuSign DS Nicotine is used as an aid in the detection of cotinine after use of tobacco products or other products containing nicotine. For In Vitro Diagnostic Use. The AccuSign Nicotine test provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography, mass spectrometry (GC/MS) is the preferred confirmatory method. 3. Special conditions for use statement(s): For prescription use. 4. Special instrument requirements: Not applicable ## I. Device Description: The device contains a membrane strip and a dye pad. The membrane strip is coated with monoclonal anti-cotinine antibody and the dye pad contains dye coated with 2 {2} cotinine protein conjugate. A disposable urine dropper and instructions for use are included in the kit. ## J. Substantial Equivalence Information: 1. Predicate device name(s): OraSure Technologies Auto-Lyte Cotinine EIA 2. Predicate 510(k) number(s): k972481 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Test Type | Qualitative | Qualitative/Semi-quantitative | | Cutoff | 500 ng/mL | Same | | Matrix | Urine | Same | | Analyte | Cotinine | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Results | Visual | Instrument | | Principle | Chromatographic Immunoassay | Enzyme Immunoassay | ## K. Standard/Guidance Document Referenced (if applicable): None Referenced. ## L. Test Principle: The device uses a solid phase chromatographic membrane immunoassay technology and is based on the principle of specific immunochemical reactions between antigens and antibodies. The test relies on antibody binding competition between cotinine conjugate and the cotinine that may be present in the urine sample. In the test procedure, a sample of urine is placed in the sample well of the device and allowed to migrate up the membrane. If cotinine is present in the urine sample, it competes with the cotinine conjugate, which is bound to the dye, for the limited antibodies immobilized on the membrane. If the cotinine is present at or above the 500 ng/mL cutoff, cotinine will saturate the antibodies, thus inhibiting the binding of the dye coated with cotinine conjugate to the antibodies on the membrane. This prevents the formation of a line the membrane. A cotinine-positive urine sample will not generate {3} a line at the test position (T). ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Qualitative imprecision was assessed by conducting a lot-to-lot precision study. The samples were prepared by adding cotinine to a known amount of drug-free specimen at concentrations 50% below the cutoff, 25% below the cutoff, 25% above the cutoff, 50% above the cutoff and 100% above the cutoff. The samples were tested with two lots of devices for two consecutive days. 10 samples for each of the 6 concentrations were tested daily for each lot. The results are summarized in the table below: Number of operators: 3 Number of lots: 2 Number of days: 2 | Cotinine ng/mL | % Cutoff | Number Tested | Positive | Negative | % of Correct Results | | --- | --- | --- | --- | --- | --- | | 0 | 0 | 120 | 0 | 120 | 100 | | 250 | 50% | 120 | 1 | 119 | 99 | | 375 | 75% | 120 | 19 | 101 | 84 | | 625 | 125% | 120 | 99 | 21 | 83 | | 750 | 150% | 120 | 117 | 3 | 98 | | 1000 | 200% | 120 | 120 | 0 | 100 | A second qualitative imprecision study was conducted at three physician office laboratories using 15 blinded, randomized control samples: 5 negative samples, 5 samples at 50% below the cutoff and 5 samples at 100% above the cutoff. One operator performed the test at each site. The operators consisted of 2 nurses and one medical technician. There was 100% agreement at all concentration levels at all sites. #### b. Linearity/assay reportable range: Not applicable. The assay is intended for qualitative use. #### c. Traceability, Stability, Expected values (controls, calibrators, or methods): This device has internal process controls. A visible line appearing in the control region confirms that sufficient sample volume has been applied. In the labeling, users are instructed not to interpret the test if a visible line failed 4 {4} to appear in the control region. External control materials are not supplied with this device. External controls are available from commercial sources to ensure proper kit performance. In the labeling the sponsor recommends that the user run controls at regular intervals following federal, state and local guidelines. Protocols and acceptance criteria were described for stability testing and found to be acceptable. d. Detection limit: The claimed cutoff for this assay is 500 ng/mL. The precision studies (above) demonstrate performance of the device around the cutoff. e. Analytical specificity: The analytical specificity studies were performed on compounds structurally related to nicotine. All compounds were prepared in drug-free normal human urine. The following compounds produce positive results when tested at levels greater than the concentrations listed below. | Compound | Concentration (ng/ml) | | --- | --- | | Cotinine | 500 | | Niacinamide | >100,000 | | (-) Nicotine | >100,000 | | Nicotinic Aid | >100,000 | | Nicotinic Acid N-oxide | >100,000 | | (±)-Anabasin | >100,000 | | (±)-Nornicotine | >100,000 | Interference: To evaluate the potential for interference by certain exogenous compounds, urine samples with concentrations at ± 25% of the cutoff were spiked with potential interferants and tested. No interferences were observed at the concentrations tested. Results are summarized below: | Substance | Concentration (mg/dL) | | --- | --- | | Ascorbic Acid | 50 | | Protein | 2000 | | Hemoglobin | 100 | | Bilirubin | 10 | | Glucose | 1500 | | Creatinine | 20 | The table below lists compounds that showed no cross-reactivity when tested with the device. The interferents tested were at concentrations up to 100 5 {5} $\mu \mathrm{g} / \mathrm{mL}$ . A complete list is presented in the package insert. | Acetaminophen | Chlorpromazine | Doxylamine | | --- | --- | --- | | Acetylsalicylate | Chlorquine | Ecgonine hydrochloride | | Aminopyrine | Cholesterol | B-Estradiol | | Amitryptiline | Chlomipramine | Ethyl-p-aminobenzoate | | Amobarbital | Clonidine | Fenoprofen | | Amoxicillin | Cocaine hydrochloride | Furoximide | | Apomorphine | Codeine | Gentisic acid | | Aspartame | Cortisone | Glutethimide | | Atropine | Creatinine | Guaifenesin | | Benzilic acid | Deoxycorticosterone | Hippuric Acid | | Benzoic acid | Dextromethorphan | Ibuprofen | | Butabarbital | Diazepam | (+/-)-Isoproterenol | | Cannabidiol | Diclofenac | Ketamine | | Chloralhydrate | Diethylpropion | Levorphanol | | Chloramphenicol | Diflunsial | Lidocaine | | Chlordiazepoxide | Digoxin | Methylphenidate | | Chlorothiazide | Diphenhydrine | Naproxen | | Dextromethorphan | Erythromycin | Oxalic acid | | 4-Dimethylaminoantipyrine | Ethanol | Penicillin-G | | Dopamine | Glucose | Quinidine | ## Urinary pH studies: Sample solutions containing drug concentrations that were 50% above and 50% below the cutoff were adjusted for pH between the ranges of 4.7 to 8. The pH adjusted sample solutions were tested ten times with the device. The results are summarized in the table below: | Cotinine Concentration (ng/mL) | pH | Number Tested | # Positive | # Negative | % Agreement | | --- | --- | --- | --- | --- | --- | | 250 | 4.7 | 10 | 0 | 10 | 100 | | 250 | 6.5 | 10 | 0 | 10 | 100 | | 250 | 8.0 | 10 | 0 | 10 | 100 | | 1000 | 4.7 | 10 | 10 | 0 | 100 | | 1000 | 6.5 | 10 | 10 | 0 | 100 | | 1000 | 8.0 | 10 | 10 | 0 | 100 | No interferences due to pH were observed. ## Urinary Specific Gravity studies: Sample solutions containing drug concentrations that were 50% above and 50% below the cutoff were adjusted to specific gravities of 1.002 and 1.40 and tested ten times with the device. The results are summarized in the table below: {6} 7 | Cotinine Concentration (ng/mL) | Specific Gravity | Number Tested | # Positive | # Negative | % Agreement | | --- | --- | --- | --- | --- | --- | | 250 | 1.002 | 10 | 0 | 10 | 100 | | 250 | 1.040 | 10 | 0 | 10 | 100 | | 1000 | 1.002 | 10 | 10 | 0 | 100 | | 1000 | 1.040 | 10 | 10 | 0 | 100 | No interferences due to different specific gravities were observed. f. Assay cut-off: Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section above. 2. Comparison studies: a. Method comparison with predicate and GC/MS: The device was compared to the commercially available Auto-Lyte Cotinine EIA assay on seventy-nine unaltered clinical samples purchased from a laboratory (43 negatives and 36 positives). The results are summarized in the tables below: Candidate device vs. Predicate device: | New device | Predicate | | Total | % Agreement | | --- | --- | --- | --- | --- | | Positive | 36 | 0 | 36 | 100 | | Negative | 0 | 43 | 43 | 100 | | Total | 36 | 43 | 76 | 100 | The device was compared to the reference method, Gas Chromatography/Mass Spectroscopy on eighty-six unaltered clinical samples purchased from a laboratory (39 negatives and 47 positives). The cotinine values ranged from 0 to $2155\mathrm{ng / mL}$ as measured by GC/MS. The results are summarized in the tables below: {7} Candidate device vs. GC/MS | | GC/MS | | | | Total | | --- | --- | --- | --- | --- | --- | | New Device | Negative 0.75% Cutoff | Near Cutoff Negative (0.75% and cutoff) | Near Cutoff Positive (cutoff to 1.25%) | Positive (>1.25%) | | | Positive | 2* | 5* | 7 | 33 | 47 | | Negative | 39 | 0 | 0 | 0 | 39 | *The false positive samples contained cotinine values of 247, 360, 417, 427, 434, 437 and 470 ng/mL. b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...