THERMO SCIENTIFIC CEDIA CANNABINOIDS OFT ASSAY

K101744 · Microgenics Corp. · LDJ · Apr 8, 2011 · Clinical Toxicology

Device Facts

Record IDK101744
Device NameTHERMO SCIENTIFIC CEDIA CANNABINOIDS OFT ASSAY
ApplicantMicrogenics Corp.
Product CodeLDJ · Clinical Toxicology
Decision DateApr 8, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3870
Device ClassClass 2

Indications for Use

The CEDIA® Cannabinoids OFT Assay is intended for use in the qualitative determination of Cannabinoids in human oral fluid at a cutoff concentration of 3.0 ng/mL in neat oral fluid. The specimen must be collected exclusively with the Oral-Eze™ Saliva Collection System. The assay is calibrated against 1-Δ⁹ THC and performed on the MGC240. This in vitro diagnostic device is intended for clinical laboratory use only. The CEDIA® THC OFT Calibrators are intended for use in the calibration of 1-Δ⁹ THC when used with the CEDIA® Cannabinoids OFT Assay. This in vitro diagnostic device is intended for clinical laboratory use only. The CEDIA® Cannabinoids OFT Assay provides only a preliminary analytical test result. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography-Tandem Mass Spectrometry (LC-MS-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result particularly when preliminary positive results are used.

Device Story

Homogeneous enzyme immunoassay for qualitative detection of cannabinoids in human oral fluid; utilizes recombinant DNA technology with bacterial β-galactosidase enzyme engineered into two inactive fragments. Input: oral fluid collected via Oral-Eze™ Saliva Collection System. Principle: competitive binding; analyte in sample competes with enzyme-conjugated analyte for antibody binding sites; free inactive fragments re-associate to form active enzyme; active enzyme cleaves substrate to produce spectrophotometrically measurable color change. Output: absorbance change proportional to analyte concentration. Used in clinical chemistry laboratories; operated by laboratory technicians. Results provide preliminary screening; clinical judgment required; positive results necessitate confirmatory testing via GC/MS or LC-MS/MS. Benefits: rapid, automated screening for drug abuse in clinical settings.

Clinical Evidence

Bench testing only. Precision studies (n=50 per concentration) showed 100% agreement for negative and near-cutoff samples. Method comparison studies (n=42 and n=81) against GC/MS/LC-MS-MS confirmed 98-100% agreement for positive and negative results. Analytical specificity evaluated against numerous structurally related and unrelated compounds; no significant interference observed.

Technological Characteristics

Homogenous enzyme immunoassay; recombinant β-galactosidase technology. Reagents: lyophilized enzyme donor/acceptor fragments, monoclonal anti-THC antibody, chlorophenol red-β-D-galactopyranoside. Form factor: liquid/lyophilized reagents for MGC240 analyzer. Connectivity: standalone analyzer. Sterilization: N/A. Software: rule-based spectrophotometric analysis.

Indications for Use

Indicated for qualitative determination of cannabinoids in human oral fluid for clinical laboratory use. Requires collection via Oral-Eze™ Saliva Collection System. Not for point-of-care settings. Provides preliminary results requiring confirmation by GC/MS or LC-MS/MS.

Regulatory Classification

Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k101744 B. Purpose for Submission: New device C. Measurand: Cannabinoids in oral fluid D. Type of Test: Qualitative enzyme immunoassay E. Applicant: Microgenics Corporation, Thermo Fisher Scientific Clinical Diagnostic Division F. Proprietary and Established Names: Thermo Scientific CEDIA® Cannabinoids OFT Assay Thermo Scientific CEDIA® THC OFT Calibrators G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LDJ | II | 21 CFR 862.3870– Cannabinoid test system | 91-Toxicology | | DLJ | II | 21 CFR 862.3200– Clinical toxicology calibrator | 91-Toxicology | H. Intended Use: 1. Intended use(s): See indications for use below {1} 2. Indication(s) for use: The CEDIA® Cannabinoids OFT Assay is intended for use in the qualitative determination of Cannabinoids in human oral fluid at a cutoff concentration of 3.0 ng/mL in neat oral fluid. The specimen must be collected exclusively with the Oral-Eze™ Saliva Collection System. The assay is calibrated against 1-Δ⁹ THC and performed on the MGC240. This in vitro diagnostic device is intended for clinical laboratory use only. The CEDIA® THC OFT Calibrators are intended for use in the calibration of 1-Δ⁹ THC when used with the CEDIA® Cannabinoids OFT Assay. This in vitro diagnostic device is intended for clinical laboratory use only. The CEDIA® Cannabinoids OFT Assay provides only a preliminary analytical test result. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography-Tandem Mass Spectrometry (LC-MS-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result particularly when preliminary positive results are used. 3. Special conditions for use statement(s): This is an in vitro diagnostic device. For prescription use For use in the clinical laboratory Not for use in the Point-of-Care settings 4. Special instrument requirements: For use with the MGC240 analyzer I. Device Description: The assay consists of lyophilized reagent and liquid reconstitution Buffer. Reagent 1 (EA Reconstitution Buffer) contains buffer salts, rabbit monoclonal anti-THC antibody, stabilizer and preservative, Reagent 1(a) (EA Reagent) contains Enzyme Acceptor (microbial), buffer salts and preservative. Reagent 2 (ED Reconstitution Buffer) contains buffer salts, stabilizers and preservative, Reagent 2(a) (ED Reagent) contains Enzyme Donor (microbial) conjugated to cannabinoid derivative, chlorophenol red-β-D-galactopyranoside, stabilizers, detergent and preservative. Calibrators: Qualitative assay required calibrators Negative Calibrator (0 ng/mL), Cutoff Calibrator (1.0 ng/mL) and High Calibrator (10.0 ng/mL). {2} The Oral-Eze Saliva Collection System consists of Oral-Eze saliva collector and collection tube with preservative buffer. Oral-Eze saliva collector consists of an absorbent pad attached to a plastic handle. The saliva collector is provided with a volume adequacy indicator. The plastic handle has a round window where blue color will appear when sufficient volume of oral fluid is collected. Samples are collected by placing the collector pad and plastic shield between lower cheek and gum with the plastic shield facing the cheek. Oral fluid collection is done when blue color appears in the window of the handle. The pad is ejected in to the collection tube by placing thumb on the ridges on the handle and pushing the thumb forward. The collection tube is capped and sent to the laboratory for processing and testing. ## J. Substantial Equivalence Information: 1. Predicate device name(s): STC Cannabinoids Intercept Micro-plate EIA, OTI, Orasure Technologies Inc 2. Predicate 510(k) number(s): k002375 3. Comparison with predicate: | Similarities/Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Qualitative detection of cannabinoids | Same | | Assay Type | Enzyme immunoassay | Same | | Matrix | Oral Fluid | Same | | Cutoff | 3.0 ng/mL in Neat Oral Fluid | 1.0 ng/mL when oral fluid collected with the Oral Specimen Collection Device | | Calibrators | 0, 1.0, 10.0 ng/mL | 0, 1.0 ng/mL | ## K. Standard/Guidance Document Referenced (if applicable): CLSI EP05-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guidelines - 2nd edition CLSI EP09-A2 Method Comparison and Bias Estimation Using Patient Samples; Approved Guidelines - 2nd edition {3} L. Test Principle: CEDIA Cannabinoids OFT Assay uses recombinant DNA technology to produce a unique homogenous enzyme immunoassay system. The assay is based on the bacterial enzyme β-galactosidase, which has been genetically engineered into two inactive fragments. Analyte in the sample competes with analyte conjugates to one inactive fragment (enzyme donor) of β-galactosidase for antibody binding site. The amount of active enzyme formed (color change) is measured spectrophotometrically at 570 nm which is directly proportional to the amount of analyte present in the sample. The Oral-Eze Saliva Collection System contains a preservative buffer that dilutes the neat oral fluid sample. The calibrator levels are set at diluted levels so that sample absorbance values can be compared directly to the absorbance values of the calibration curve. The assay result is reported as a positive or negative result relative to the neat oral fluid cutoff of 3.0 ng/mL. M. Performance Characteristics (if/when applicable): 1. Analytical performance: All analytical performance data was collected on neat human oral fluid samples and processed through the Oral-Eze Saliva Collection System and analyzed on the MGC 240 instrument. The Oral-Eze collection device includes a diluent that results in a dilution of approximately 1/3. The assay cannot be used to measure undiluted (neat) samples. Analyte concentrations refer to the neat oral fluid concentration, unless otherwise noted. a. Precision/Reproducibility: Precision studies were performed by spiking an l-isomer-Δ⁹ THC solution into neat oral fluid pools at concentrations of 0, 0.75, 1.50, 2.25, 3.00, 3.75, 4.50, 5.25 and 6.00 ng/mL. Concentrations were confirmed by LC-MS/MS. Each sample was then processed through the Oral-Eze Saliva Collection system to obtain a final concentration at approximately negative, -75%, -50%, -25%, cutoff, +25% and 50%, 75% and 100% of the calibrator cutoff. Testing was performed in replicates of 5, twice a day over 5 non-consecutive days for all concentrations. The results are presented in the table below: 4 {4} | Drug | Concentration of sample | Number of determinations | Results # Neg/ #Pos | | --- | --- | --- | --- | | l-isomer-Δ9THC | Negative | 50 | 50/0 | | | -75% | 50 | 50/0 | | | -50% | 50 | 50/0 | | | -25% | 50 | 50/0 | | | Cutoff | 50 | 2/48 | | | 25% | 50 | 0/50 | | | 50% | 50 | 0/50 | | | 75% | 50 | 0/50 | b. Linearity/assay reportable range: Not applicable, this is a qualitative assay c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrator: Traceability: Calibrators are specified in the labeling but are supplied separately from the reagents. Calibrators consist of a synthetic oral fluid matrix spiked with known concentrations of l-isomer- $\Delta^9$ THC. The concentration of l-isomer- $\Delta^{9}$ THC in the calibrators is verified by LC/MS/MS. Stability: Real-time Stability studies were conducted on two lots of calibrators. The stability protocol was reviewed and found acceptable. The testing supports the stability at $2 - 8^{\circ}\mathrm{C}$ for 18 months. The open vial stability of reconstituted reagents is 50 days $(2 - 8^{\circ}\mathrm{C})$ . Shipment Stability: Conditions simulating ground shipping, air shipping and various climate conditions (desert, tropical) were tested. Oral fluid samples were spiked with l-isomer- $\Delta^{9}$ THC to concentrations of $-50\%$ of the cutoff and $+50\%$ of the cutoff. One set was stored at room temperature and used as the control, while the other set was used for testing. All samples at $-50\%$ of the cutoff recovered as negative and all samples at $+50\%$ of the cutoff recovered as positive. The shipping temperature should not exceed $>40^{\circ}\mathrm{C}$ . {5} Sample Storage and Stability: The stability of oral fluid samples in the preservative buffer was evaluated in real time. The stability protocol was reviewed and found acceptable. Oral fluid samples can be stored at 2-8°C or at room temperature (21-25°C) for 21 days. Reagent Stability: Real time stability testing was conducted. The stability protocol was reviewed and found acceptable. The testing supports the stability at 2-8°C for 24 months. The on board stability of reconstituted reagents is 60 days (2-8°C). d. Detection limit: Analytical performance of the device around the cutoff is described in the precision section 1.a above. e. Analytical specificity: Cross-reactivity was evaluated by spiking various concentrations (which could be found in a neat oral fluid sample) of structurally related compounds into drug-free neat oral fluid pool, than added to the oral fluid collection device. The various concentrations were evaluated against the cutoff calibrator. The table below lists the concentration of each compound that gave a response approximately equal to the cutoff. | Compound | Tested Concentration (ng/mL) | Response equivalent to cutoff | | --- | --- | --- | | 1-11-nor-Δ^{9} THC-COOH | 3.0 | Positive | | 11-OH-Δ^{9} THC | 3.75 | Positive | | Δ^{8} THC | 3.75 | Positive | | Cannabinol | 12 | Positive | | Cannabidiol | 3000 | Positive | Potential interference from structurally unrelated compounds and various common over-the-counter medications were tested by spiking the potentially interfering compound into neat oral fluid, and then processed through the oral fluid collection device. 6 {6} | Compound | Tested Concentration in neat Oral Fluid (ng/mL) | Response Equivalent to the cutoff | | --- | --- | --- | | Acetaminophen | 240,000 | Negative | | Acetylsalicylic acid | 240,000 | Negative | | Alprazolam | 30,000 | Negative | | Amobarbital | 30,000 | Negative | | Amoxicillin | 240,000 | Negative | | Amphetamine | 240,000 | Negative | | Ampicillin | 30,000 | Negative | | Atropine | 30,000 | Negative | | Benzoylecgonine | 120,000 | Negative | | Phenethylamine | 30,000 | Negative | | Butabarbital | 30,000 | Negative | | Butalbital | 30,000 | Negative | | Caffeine | 24,000 | Negative | | Captopril | 120,000 | Negative | | Chlordiazepoxide | 24,000 | Negative | | Chlorpromazine | 30,000 | Negative | | Cimetidine | 120,000 | Negative | | Clonazepam | 30,000 | Negative | | Clorazepate | 30,000 | Negative | | Cocaethylene | 30,000 | Negative | | Cocaine | 1500 | Negative | | Codeine | 240,000 | Negative | | Cyclizine | 30,000 | Negative | | Dextromethorphan | 240,000 | Negative | | Diazepam | 120,000 | Negative | | Digoxin | 24,000 | Negative | | Diphenhydramine | 30,000 | Negative | | Enalapril | 120,000 | Negative | | Fluoxetine | 120,000 | Negative | | Gentisic acid | 30,000 | Negative | | Hydrocodone | 30,000 | Negative | | Hydromorphone | 30,000 | Negative | {7} | Compound | Tested Concentration in neat Oral Fluid (ng/mL) | Response Equivalent to the cutoff | | --- | --- | --- | | Ibuprofen | 120,000 | Negative | | Imipramine | 30,000 | Negative | | l-Ephedrine | 30,000 | Negative | | Levothyroxine | 12,000 | Negative | | Lidocaine | 30,000 | Negative | | Loperamide | 30,000 | Negative | | Medazepam | 30,000 | Negative | | Meperidine | 240,000 | Negative | | Methadone | 240,000 | Negative | | Methamphetamine | 240,000 | Negative | | Metoprolol | 30,000 | Negative | | Morphine | 48,000 | Negative | | Naproxen | 240,000 | Negative | | Niacinamide | 30,000 | Negative | | Nifedipine | 120,000 | Negative | | Norchlordiazepoxide | 30,000 | Negative | | Oxazepam | 120,000 | Negative | | Penicillin | 30,000 | Negative | | Phencyclidine | 240,000 | Negative | | Phenobarbital | 240,000 | Negative | | Phenylephrine | 30,000 | Negative | | Phenylpropanolamine | 30,000 | Negative | | Procainamide | 30,000 | Negative | | Procaine | 30,000 | Negative | | Propoxyphene | 240,000 | Negative | | Pseudoephedrine | 30,000 | Negative | | Quinidine | 30,000 | Negative | | Ranitidine | 120,000 | Negative | | Salbutamol | 30,000 | Negative | | Salicyluric Acid | 120,000 | Negative | | Secobarbital | 240,000 | Negative | | Temazepam | 30,000 | Negative | 8 {8} | Compound | Tested Concentration in neat Oral Fluid (ng/mL) | Response Equivalent to the cutoff | | --- | --- | --- | | Theophylline | 30,000 | Negative | | Tolmetin | 120,000 | Negative | | Verapamil | 120,000 | Negative | | Zomepirac | 30,000 | Negative | Potential interference from endogenous and exogenous substances and pH were spiked into neat oral fluid containing l-isomer- $\Delta^9\mathrm{THC}$ at $+ / - 50\%$ of the cutoff, and then processed through the oral fluid collection device. No interference was observed with the interfering substances and pH 5-9. The results are presented in the table below: | Compounds | Tested Conc. in Neat Oral Fluid (ng/mL) | PCP OFT Assay | | | --- | --- | --- | --- | | | | -50% PCP | +50% PCP | | Cotinine | 0.03 | Negative | Positive | | Nicotine | 0.03 | Negative | Positive | | Hemoglobin | 0.3 | Negative | Positive | | Human serum albumin | 7.5 | Negative | Positive | | Sodium Chloride | 18 | Negative | Positive | | Cholesterol | 0.45 | Negative | Positive | | Acetaminophen | 0.3 | Negative | Positive | | Acetylsalicylic Acid | 0.3 | Negative | Positive | | Caffeine | 0.3 | Negative | Positive | | Ibuprofen | 0.3 | Negative | Positive | | Coffee | 6% v/v | Negative | Positive | | Milk | 1.5% v/v | Negative | Positive | | Orange Juice | 6% v/v | Negative | Positive | | Cranberry Juice | 6% v/v | Negative | Positive | | Soft drink (Coke) | 6% v/v | Negative | Positive | | Toothpaste | 6% v/v | Negative | Positive | | Mouthwash | 6% v/v | Negative | Positive | | Tea | 6% v/v | Negative | Positive | | Denture Adhesive | 6% v/v | Negative | Positive | | Alcohol | 6% v/v | Negative | Positive | | Baking Soda | 6% v/v | Negative | Positive | | Cough Syrup | 6% v/v | Negative | Positive | | Whole Blood | 6% v/v | Negative | Positive | | Hydrogen Peroxide | 3% v/v | Negative | Positive | | pH | 5-9 | Negative | Positive | {9} Potential interference from additional food and dental compounds was tested by collecting neat oral fluid from volunteers after use of the following substances: hard candy, chewing gum, chewing tobacco, cigarettes and tooth whitening strips. | Compounds | Tested Concentration in Neat Oral Fluid | Cannabinoid OFT Assay Results | | | --- | --- | --- | --- | | | | -50% l-isomer-Δ⁹THC | +50% l-isomer-Δ⁹THC | | Water | n/a | Negative | Positive | | Chewing Tobacco | n/a | Negative | Positive | | Cigarettes | n/a | Negative | Positive | | Gum | n/a | Negative | Positive | | Hard Candy | n/a | Negative | Positive | | Tooth Whitening Strips | n/a | Negative | Positive | f. Assay cut-off: Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision above. 2. Comparison studies: a. Method comparison with predicate device: Two method comparison studies were performed. Study 1: 42 unaltered neat oral fluid samples were collected from rehabilitation clinics. The neat oral fluid samples were processed using the Oral-Eze collection device. The diluted samples were tested in the CEDIA® Cannabinoids OFT Assay and compared to the neat and diluted oral fluid samples tested by LC/MS/MS. The results reflect the performance of the entire system including the collection step. 10 {10} 11 | Candidate Device Results | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) | | --- | --- | --- | --- | --- | | Positive | 0 | 0 | 3 | 18 | | Negative | 19 | 2 | 0 | 0 | % Agreement among positive and negative is 100%. LC/MS/MS values used to categorize samples in this table are based on the concentration found in the neat oral fluid sample. Study 2: A total of 81 samples (40 negative and 41 positive) were evaluated by the candidate device and GC/MS. All samples were collected with the Oral-Eze Saliva collection system. Note: this study was performed on samples already collected with the Intercept collection device. When the LC/MS/MS values of the diluted samples were compared to the immunoassay values, the following results were obtained. Therefore the results below do not reflect any inaccuracy inherent in the collection process itself. | Candidate Device Results | Negative | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) | | --- | --- | --- | --- | --- | --- | | Positive | 0 | 0 | 0 | 4 | 36 | | Negative | 34 | 2 | 4 | 1 | 0 | % Agreement among positives is 98% (40/41) % Agreement among negatives is 100% (40/40) | | MGC 240 OFT Assay (POS/NEG) | Drug/Metabolite GC/MS value | | --- | --- | --- | | Sample | Negative | 4.2 (ng/mL) Δ⁹ THC | b. Matrix comparison: Not applicable. The assay is intended for only one sample matrix. {11} 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Not Applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 12
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...