Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card
Device Facts
| Record ID | K150356 |
|---|---|
| Device Name | Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card |
| Applicant | Healgen Scientific,, LLC |
| Product Code | JXM · Clinical Toxicology |
| Decision Date | May 13, 2015 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3170 |
| Device Class | Class 2 |
Indications for Use
The Healgen multi-test devices are intended for the qualitative detection of drugs of abuse in human urine. The devices are intended for professional use only.
Device Story
Multi-test device for qualitative detection of up to 7 drugs of abuse in human urine; combines previously cleared single-drug test strips, cassettes, dip cards, and cups into a single platform; intended for professional use; provides visual qualitative results; assists clinicians in identifying drug presence; utilizes lateral flow immunoassay technology; no change to fundamental scientific technology from predicate devices.
Clinical Evidence
Bench testing only. Verification studies included interference studies and a lay-user study to support the multi-drug format modification. No clinical trial data provided.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding format. Multi-drug cup or dip card configuration containing multiple test strips. Qualitative visual readout. No electronic components, software, or energy source required.
Indications for Use
Indicated for qualitative, simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, and Oxycodone in human urine. Intended for both prescription and over-the-counter (OTC) use. Provides preliminary results; requires confirmatory testing (GC/MS or LC/MS).
Regulatory Classification
Identification
A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Healgen Oxazepam and Morphine Tests (K142280)
- Healgen Amphetamine and Oxycodone Tests (K143187)
- Healgen Cocaine and Morphine Tests (K141647)
- Healgen Marijuana and Methamphetamine Tests (K140546)
Related Devices
- K121166 — WONDFO MULTI-DRUG URINE TEST CUP WONDFO MULTI-DRUG URINE TEST PANEL · Guangzhou Wondfo Biotech Co., Ltd. · May 8, 2012
- K151642 — Heagen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card · Healgen Scientific,, LLC · Aug 13, 2015
- K180255 — CLUNGENE Multi-Drug Test Dip Card, CLUNGENE Multi-Drug Test Easy Cup · Hangzhou Clongene Biotech Co., Ltd. · Feb 28, 2018