← Product Code [DKB](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DKB) · K062121

# DIMENSION VISTA SYSTEM DRUG 2 CALIBRATOR (DRUG 2 CAL - KC420) (K062121)

_Dade Behring, Inc. · DKB · Aug 25, 2006 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-d%E2%80%94clinical-toxicology-test-systems/DKB/K062121

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [DKB](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DKB.md)
- **Decision Date:** Aug 25, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3200
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Intended Use

The DRUG 2 CAL is an in vitro diagnostic product for the calibration of acetaminophen (ACTM), carbamazepine (CRBM), digitoxin (DGTX), gentamicin (GENT), lidocaine (LIDO), N-acetylprocainamide (NAPA), procainamide (PROC), tobramycin (TOBR), valproic acid (VALP), and vancomycin (VANC) methods on the Dimension Vista™ System.

## Device Story

DRUG 2 CAL is a multi-analyte, liquid, bovine serum-based calibrator kit for the Dimension Vista™ System. It contains ten specific drug analytes. The kit includes two levels of calibrator (A and B). The system uses these to automatically prepare intermediate levels and calculate calibration values. It is used in clinical laboratory settings by trained personnel to calibrate diagnostic assays. The output is a calibrated instrument state, enabling accurate patient sample testing for therapeutic drug monitoring.

## Clinical Evidence

No clinical data. Bench testing only. Stability testing performed per CEN 13640. Assigned values verified via N=20 replicates across multiple reagent lots and instruments.

## Technological Characteristics

Liquid, bovine serum-based multi-analyte calibrator. Traceable to USP reference materials (except NAPA, which uses Alltech-Applied Sciences standards). Two-level configuration. Designed for automated dilution and calculation on the Dimension Vista™ System. Stability testing per CEN 13640; risk management per ISO 14971:2000.

## Regulatory Identification

A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Dimension® Drug Calibrator II ([K033809](/device/K033809.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kouzi21

AUG 25 2006

### 510(k) Summary for the Dimension Vista™ System Drug 2 Calibrator (DRUG 2 CAL - KC420)

### A. 510(k) Number:

| B. Analytes:     | Acetaminophen (ACTM), carbamazepine (CRBM), digitoxin<br>(DGTX), gentamicin (GENT), lidocaine (LIDO), N-<br>acetylprocainamide (NAPA), procainamide (PROC), tobramycin<br>(TOBR), valproic acid (VALP), and vancomycin (VANC). |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| C. Type of Test: | Calibrator Material                                                                                                                                                                                                            |
| D. Applicant:    | Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101<br>Victor M. Carrio, Regulatory Affairs and Compliance Manager<br>Office: (302) 631-0376 Fax: (302) 631-6299                                                           |

#### E. Proprietary and Established Names:

Dimension Vista™ System Drug 2 Calibrator (DRUG 2 CAL - KC420)

#### F. Regulatory Information:

- 1. Regulation section: 21 CFR § 862.3200 -- Clinical Toxicology Calibrator
- 2. Classification: Class II
- 3. Product Code: DKB Calibrator, Drug Mixture
- 4. Panel: Toxicology
- G. Intended Use: The DRUG 2 CAL is an in vitro diagnostic product for the calibration of acetaminophen (ACTM), carbamazepine (CRBM), digitoxin (DGTX), gentamicin (GENT), lidocaine (LIDO), Nacetylprocainamide (NAPA), procainamide (PROC), tobramycin (TOBR), valproic acid (VALP), and vancomycin (VANC) methods on the Dimension Vista TM System.

### H. Device Description:

DRUG 2 CAL is a multi-analyte, liquid, bovine serum based product containing acetaminophen, carbamazepine, digitoxin, gentamicin, lidocaine, N-acetylprocainamide, procainamide, tobramycin, valproic acid, and vancomycin.

{1}------------------------------------------------

The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 2.5 mL. Intermediate levels are automatically prepared and corresponding values calculated by the Dimension Vista™ System.

### I. Substantial Equivalence Information:

|                     | New Device                                                                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                                                                                                                             |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                | Dimension Vista™ System<br>Drug 2 Calibrator                                                                                                                                                                                                                                                                                                                 | Dimension® Drug Calibrator II<br>K033809                                                                                                                                                                                                                                                                                                                                                     |
| Intended<br>Use     | The DRUG 2 CAL is an in vitro diagnostic<br>product for the calibration of<br>acetaminophen (ACTM), carbamazepine<br>(CRBM), digitoxin (DGTX), gentamicin<br>(GENT), lidocaine (LIDO), N-<br>acetylprocainamide (NAPA), procainamide<br>(PROC), tobramycin (TOBR), valproic acid<br>(VALP), and vancomycin (VANC) methods<br>on the Dimension Vista™ System. | Drug Calibrator II is an in vitro diagnostic<br>product intended for the calibration of the<br>following methods packaged in Flex®<br>reagent cartridges: acetaminophen (ACTM),<br>carbamazepine (CRBM), digitoxin<br>(DGTX), gentamicin (GENT), lidocaine<br>(LIDO), N-acetylprocainamide (NAPA),<br>procainamide (PROC), tobramycin<br>(TOBR), valproic acid (VALP),<br>vancomycin (VANC). |
| Analytes            | Acetaminophen (ACTM), carbamazepine<br>(CRBM), digitoxin (DGTX), gentamicin<br>(GENT), lidocaine (LIDO), N-<br>acetylprocainamide (NAPA), procainamide<br>(PROC), tobramycin (TOBR), valproic acid<br>(VALP), and vancomycin (VANC).                                                                                                                         | Acetaminophen (ACTM), carbamazepine<br>(CRBM), digitoxin (DGTX), gentamicin<br>(GENT), lidocaine (LIDO), N-<br>acetylprocainamide (NAPA), procainamide<br>(PROC), tobramycin (TOBR), valproic acid<br>(VALP), and vancomycin (VANC).                                                                                                                                                         |
| Form                | Liquid.                                                                                                                                                                                                                                                                                                                                                      | Liquid.                                                                                                                                                                                                                                                                                                                                                                                      |
| Traceability        | USP¹ for all analytes except NAPA.<br>NAPA - Alltech-Applied Sciences<br>Reference Standards.                                                                                                                                                                                                                                                                | USP¹ for all analytes except NAPA.<br>NAPA - Alltech-Applied Sciences<br>Reference Standards.                                                                                                                                                                                                                                                                                                |
| Matrix              | Bovine serum based.                                                                                                                                                                                                                                                                                                                                          | Bovine serum based.                                                                                                                                                                                                                                                                                                                                                                          |
| Number of<br>Levels | Two levels².                                                                                                                                                                                                                                                                                                                                                 | Five levels.                                                                                                                                                                                                                                                                                                                                                                                 |

' United States Pharmacopeia.

2 Intermediate levels are automatically prepared and corresponding values calculated by the Dimension Vista™ System.

### J. Standard/Guidance Document Referenced:

- 1. Guidance: Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final, 02/22/1999 Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, 11/30/2004
- 2. Standards: CEN 13640 Stability testing of In-Vitro Diagnostic Devices ISO 14971:2000 Medical devices -Application of risk management to medical devices

{2}------------------------------------------------

### K. Performance Characteristics:

- Target shelf life for the Dimension Vista™ System Drug 2 1. Stability: Calibrator is 12 months. Calibrator shelf life is determined by comparing results of the product stored at 4℃ with control stored at -20°C. The method is calibrated from this stored material. The 4℃ material values are recovered versus the calibration. Recovery versus time is monitored and percent change over time is determined. Percent change should be less than or equal to 5%. Shelf-life stability (expiration) dating assignment at commercialization reflects the real-time data on file at Dade Behring, Inc. A vial punctured by the instrument and stored on board is stable for 24 hours. An open vial not on instrument, but recapped and stored in a refrigerator is stable for 31 days. For testing, vials are opened /punctured on day zero. A quantity
sufficient for multiple calibrations is removed and the vials are recapped and stored at 2 - 8 °C. Opened/punctured vials are tested on days 0, 1, 2, 8, and 32 versus freshly opened vials.

- 2. Traceability: The assigned values of the DRUG 2 CAL are traceable to United States Pharmacopeia Reference Materials except for NAPA, which is traceable to Alltech-Applied Sciences Reference Standards.
- 3. Bottle Value Assignment:

Acetaminophen, Carbamazepine, Digitoxin, Gentamicin, Lidocaine, N-acetylprocainamide, Procainamide, Tobramycin, Valproic Acid and Vancomycin Reference Material is weighed into bovine serum albumin at six levels and stored frozen. The verification of the Master Pool values are compared against previously approved Master Pool values. The approved Master Pool values were assigned with an instrument calibrated with the corresponding standard reference material. The stock solution is made by adding Acetaminophen. Carbamazepine, Digitoxin, Gentamicin, Lidocaine, Nacetylprocainamide, Procainamide, Tobramycin, Valproic Acid and Vancomycin Reference Materials gravimetrically to stock solution at target concentrations and verified on an instrument calibrated with a previously approved Master Pool. The commercial lot is made by adding calculated quantities of stock solution to bovine serum albumin in appropriate concentrations for each of the calibrator levels. The concentration 

{3}------------------------------------------------

of each level is verified by using an instrument calibrated with Master Pools.

The final bottle values for each level of the commercial lot is assigned and verified using multiple instruments by testing N = 20 replicates per level. A previously released commercial lot is used as a control.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

## AUG 2 5 2006

Mr. Victor M. Carrio RA/OS Compliance Manager Dade Behring, Inc. P.O. Box 6101 Mailstop 514 Newark, DE, 19714-6101

k062121 Re:

Trade/Device Name: Dimension Vista™ Drug 2 Calibrator (KC420) Regulation Number: 21 CFR 862.3200 Regulation Name: Clinical Toxicology Calibrator Regulatory Class: Class II Product Code: DKB Dated: July 24, 2006 Received: July 25, 2006

Dear Mr. Carrio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

### Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market,

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Guti

Alberto Guticarez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known): K062121

Device Name:

Dimension Vista™ Drug 2 Calibrator (KC420)

Indications For Use:

The DRUG 2 CAL is an in vitro diagnostic product for the calibration of acetaminophen (ACTM), carbamazepine (CRBM), digitoxin (DGTX), gentamicin (GENT), lidocaine (LIDO), N-acetylprocainamide (NAPA), procainamide (PROC), tobramycin (TOBR), valproic acid (VALP), and vancomycin (VANC) methods on the Dimension Vista™ System.

Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of 1

K062121

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-d%E2%80%94clinical-toxicology-test-systems/DKB/K062121](https://fda.innolitics.com/submissions/CH/subpart-d%E2%80%94clinical-toxicology-test-systems/DKB/K062121)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
