Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer

K191030 · Life Technologies Corporation · PCA · Feb 21, 2020 · Clinical Chemistry

Device Facts

Record IDK191030
Device NameApplied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer
ApplicantLife Technologies Corporation
Product CodePCA · Clinical Chemistry
Decision DateFeb 21, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2570
Device ClassClass 2

Intended Use

The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are in vitro diagnostic devices intended for detection of fluorescently-labeled human genomic deoxyribonucleic acid (DNA) nucleotides by capillary electrophoresis. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are indicated for sequencing and fragment analysis using FDA- cleared or approved assays.

Device Story

Fluorescence-based DNA analysis instrument using capillary electrophoresis (8 or 24 capillaries); processes human genomic DNA samples. System includes capillary array, POP polymer, computer workstation, and integrated software. Operates by migrating labeled DNA fragments from cathode to anode; laser excitation (505 nm) detects fluorescent signals. Software performs instrument control, data collection, quality control, basecalling, and sizecalling. Used in clinical laboratories by trained technologists. Output provides sequencing and fragment analysis data to assist healthcare providers in clinical decision-making based on FDA-cleared/approved assays.

Clinical Evidence

Clinical performance evaluated across 3 US clinical laboratory sites using multiple instruments and a representative fragment analysis assay compared against an appropriate method comparison assay. All pre-established performance criteria were met. Non-clinical reproducibility study performed across multiple sites, instruments, and operators also met acceptance criteria.

Technological Characteristics

Capillary electrophoresis system; 8 or 24 capillary configuration; 50 cm capillary array. Laser: 505 nm solid-state. Electrophoresis voltage up to 20kV. Oven temperature 18°C-70°C. Consumables: POP-6/POP-7 polymers, Hi-Di Formamide, buffers, matrix standards. Software: Data Collection Software 3 IVD v3.2. Connectivity: PC-based workstation (Dell OptiPlex XE2, Windows 7).

Indications for Use

Indicated for sequencing and fragment analysis of fluorescently-labeled human genomic DNA using FDA-cleared or approved assays in clinical laboratory settings.

Regulatory Classification

Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. February 21, 2020 Life Technologies Corporation Darcie Baynes Regulatory Affairs Specialist 5781 Van Allen Way Carlsbad, CA 92008 Re: k191030 Trade/Device Name: Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer Regulation Number: 21 CFR 862.2570 Regulation Name: Instrumentation For Clinical Multiplex Test Systems Regulatory Class: Class II exempt, meets the limitation of exemptions 21 CFR 862.9(a) Product Code: PCA Dated: February 3, 2020 Received: February 4, 2020 Dear Darcie Baynes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. Please note that if you modify your IVD in the future to exceed any of the limitations to the exemption found in 21 CFR 862.9(c), your device will require a new 510(k) prior to marketing this device in the United States and will not be exempt from the premarket notification requirements so long as it exceeds the limitations to the exemption found in 21 CFR 862.9. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K191030 Device Name Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer Indications for Use (Describe) The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are in vitro diagnostic devices intended for detection of fluorescently-labeled human genomic deoxyribonucleic acid (DNA) nucleotides by capillary electrophoresis. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are indicated for sequencing and fragment analysis using FDA- cleared or approved assays. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # Submitter Information - 21 CFR 807.92(a)(1): | Submitter: | Life Technologies Corporation<br>5781 Van Allen Way<br>Carlsbad, CA 92008 | |--------------------------------|---------------------------------------------------------------------------------------------------------------| | Manufacturer: | Life Technologies Holdings Pte Ltd<br>Blk 33, #07-06, Marsiling Industrial Estate,<br>Road 3 Singapore 739256 | | Establishment Registration No: | 3003673482 | | 510(k) Number | K191030 | | Contact: | Darcie Baynes, Sr. Regulatory Affairs<br>Specialist | | Phone: | 760.918.3148 | | Fax: | 760.603.2804 | | E-mail: | darcie.baynes@thermofisher.com | | Alternate Contact: | Jody Schulz, Senior Manager Regulatory<br>Affairs | | Phone: | 262.357.4605 | | Fax: | 414.278.0688 | | E-mail: | jody.schulz@thermofisher.com | | Date Prepared: | February 18, 2020 | # Device Name - 21 CFR 807.92(a)(2): | Device Name: | Applied Biosystems™ 3500 Dx Genetic<br>Analyzer<br>and<br>Applied Biosystems™ 3500xL Dx Genetic<br>Analyzer | |-----------------|---------------------------------------------------------------------------------------------------------------| | Common Name: | DNA Genetic Analyzer | | Classification: | Class II, exempt from the premarket<br>notification requirement subject to the<br>limitations in 21 CFR 862.9 | | Product Code: | PCA | {4}------------------------------------------------ ## Predicate Device 21 CFR 807.92(a)(3) | Predicate | Applied Biosystems 3500 Dx/3500xL Dx<br>Genetic Analyzer CS2 and 3500 Dx Series<br>Software, BK110039 | |-----------|-------------------------------------------------------------------------------------------------------| |-----------|-------------------------------------------------------------------------------------------------------| ### Device Description 21 CFR 807.92(a)(4): The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are fluorescence-based DNA analysis instruments that use capillary electrophoresis technology with 8 and 24 capillaries, respectively. The 8-capillary system and the 24-capillary system include the following components: - 8-capillary or 24-capillary array and POP™ polymer . - . Consumables for system qualification - Computer workstation and monitor - Integrated software for instrument control, data collection, quality control, . basecalling and sizecalling of samples The following consumables (branded with the Applied Biosystems name) are required to operate the Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer. - . 50cm Capillary Array: enable the labeled DNA fragments to migrate from the cathode toward the anode for detection - POP-6TM Polymer: used as a separation matrix to separate DNA fragments by size ● during electrophoresis for sequencing - POP-7TM Polymer: used as a separation matrix for separating DNA fragments by size during electrophoresis for fragment analysis - Hi-Di™ Formamide: sample re-suspension solution used for electrokinetic ● injection and denaturing the DNA - Sequencing Standard v1.1: used for spectral calibration of the instrument and instrument performance check - Cathode Buffer Container: pre-filled with running buffer which maintains a source . of ions and the correct pH for electrophoresis - Anode Buffer Container: pre-filled with running buffer which maintains a source ● of ions and the correct pH for electrophoresis - Conditioning Reagent: pre-filled pouch used for priming the polymer pump, ● washing the pump between polymer type changes, and during instrument shutdown - DS-30 Matrix Standard Dx: used for spectral calibration ● - DS-33 Matrix Standard - Dx: used for spectral calibration - DS-33 GeneScan™ Install Kit Dx: used for instrument operational qualification ● - GeneScan™ 600 LIZ® Size Standard v2.0 Dx: used as a ladder for sizing DNA ● fragments - Other accessories (e.g. sample plate holders, plate retainers, septa) . {5}------------------------------------------------ # Intended Use 21 CFR 807.95(a)(5): The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are in vitro diagnostic devices intended for detection of fluorescently-labeled human genomic deoxyribonucleic acid (DNA) nucleotides by capillary electrophoresis. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are indicated for sequencing and fragment analysis using FDA- cleared or approved assays. # Predicate Device Comparison 21 CFR 807.92(a)(6): A summary of the technological characteristics of the device compared to the predicate device is provided. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer and the legally marketed device the Applied Biosystems 3500 Dx/3500xL Dx Genetic Analyzer CS2 and 3500 Dx Series Software are compared in the table below. | Feature | Applied<br>Biosystems 3500<br>Dx/3500xL Dx CS2<br>and 3500 Dx Series<br>Software<br>(BK110039) | Applied<br>Biosystems™ 3500<br>Dx Genetic<br>Analyzer and the<br>Applied<br>Biosystems™<br>3500xL Dx Genetic<br>Analyzer (with<br>Fragment<br>Analysis) | Comparison | |-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | 510k number | BK110039 | K191030 | N/A | | Regulation | 862.2570 | 862.2570 | Same | | Product Code | PCA | PCA | Same | | Device Class | Class II | Class II | Same | | Intended Use | The Applied<br>Biosystems® 3500<br>Dx / 3500xL Dx<br>Genetic Analyzer<br>CS2 and 3500 Dx<br>Series Software are<br>in vitro diagnostic<br>devices intended for<br>the sequencing<br>(detection and<br>identification) of<br>fluorescently-<br>labeled<br>deoxyribonucleic | The Applied<br>Biosystems™ 3500<br>Dx Genetic<br>Analyzer and the<br>Applied<br>Biosystems™<br>3500xL Dx Genetic<br>Analyzer are in vitro<br>diagnostic devices<br>intended for<br>detection of<br>fluorescently-<br>labeled human<br>genomic | Software upgrade to<br>Data Collection<br>Software (DCS) 3<br>IVD v3.2 and<br>enablement of<br>Fragment Analysis<br>in Diagnostic mode. | | Feature | Applied<br>Biosystems 3500<br>Dx/3500xL Dx CS2<br>and 3500 Dx Series<br>Software<br>(BK110039) | Applied<br>Biosystems™ 3500<br>Dx Genetic<br>Analyzer and the<br>Applied<br>Biosystems™<br>3500xL Dx Genetic<br>Analyzer (with<br>Fragment<br>Analysis) | Comparison | | | acid (DNA) by<br>capillary<br>electrophoresis.<br>The Applied<br>Biosystems® 3500<br>Dx / 3500xL Dx<br>Genetic Analyzer<br>CS2 with 3500 Dx<br>Series Software<br>2011 (v1) are<br>indicated for use<br>with FDA-cleared<br>or approved<br>sequencing assays<br>specifying their use<br>and only by<br>technologists<br>trained in laboratory<br>techniques,<br>procedures, and use<br>of the analyzer. | deoxyribonucleic<br>acid (DNA)<br>nucleotides by<br>capillary<br>electrophoresis.<br>The Applied<br>Biosystems™ 3500<br>Dx Genetic<br>Analyzer and the<br>Applied<br>Biosystems™<br>3500xL Dx Genetic<br>Analyzer are<br>indicated for<br>sequencing and<br>fragment analysis<br>using FDA- cleared<br>or approved assays. | | | Data Collection<br>Software | 3500 Dx Data<br>Collection Software<br>2011 v1.01 | Applied<br>Biosystems™ 3500<br>Dx Series Data<br>Collection Software<br>3 IVD v3.2 | Upgrade to Data<br>Collection Software<br>3 IVD v3.2 | | Firmware | 6228001-05 | 6228001-05 | Same | | Computer | Dell OptiPlex XE | Dell OptiPlex XE2 | Upgrade to new<br>model. | | Windows | Windows® 7 | Windows® 7 | Same | | Laser | Single-line 505 nm,<br>solid-state | Single-line 505 nm,<br>solid-state | Same | | Feature | Applied<br>Biosystems 3500<br>Dx/3500xL Dx CS2<br>and 3500 Dx Series<br>Software<br>(BK110039) | Applied<br>Biosystems™ 3500<br>Dx Genetic<br>Analyzer and the<br>Applied<br>Biosystems™<br>3500xL Dx Genetic<br>Analyzer (with<br>Fragment<br>Analysis) | Comparison | | Electrophoresis<br>Voltage | Up to 20Kv | Up to 20Kv | Same | | Oven Temperature | 18°C to 70°C | 18°C to 70°C | Same | | Operating<br>Environment | Temperature: 15°C-<br>30°C | Temperature: 15°C-<br>30°C | Same | | | Humidity: 20-80%<br>(non-condensing) | Humidity: 20-80%<br>(non-condensing) | Same | | Main Power<br>Voltage | 100-240 V ±10%;<br>50–60 Hz | 100–240 V ±10%;<br>50–60 Hz | Same | | Power<br>Consumption | 320 VA | 320VA | Same | | Dimensions<br>(Width x Depth x<br>Height) | 61cm(122cm) x<br>61cm x 72cm | 61cm(122cm) x<br>61cm x 72cm | Same | | Weight | 82Kg | 82Kg | Same | | Polymer Use | POP-6TM Dx | POP-6TM Dx and<br>POP-7TM Dx | Inclusion of POP-7<br>for Fragment<br>Analysis application<br>in IVD mode of<br>proposed device. | | Capillary Array | 50 cm length<br>available in 8-<br>capillary and 24<br>capillary<br>configuration | 50 cm length<br>available in 8-<br>capillary and 24<br>capillary<br>configuration | Same, however the<br>50 cm array will be<br>used for fragment<br>analysis workflow<br>in IVD mode. | | Applications | Sequencing and<br>Fragment Analysis<br>in RUO mode,<br>Sequencing in IVD<br>mode | Sequencing and<br>Fragment Analysis<br>in RUO mode,<br>Sequencing and<br>Fragment Analysis<br>in IVD mode | Inclusion of<br>Fragment Analysis<br>application in IVD<br>mode of proposed<br>device. | | Feature | Applied<br>Biosystems 3500<br>Dx/3500xL Dx CS2<br>and 3500 Dx Series<br>Software<br>(BK110039) | Applied<br>Biosystems™ 3500<br>Dx Genetic<br>Analyzer and the<br>Applied<br>Biosystems™<br>3500xL Dx Genetic<br>Analyzer (with<br>Fragment<br>Analysis) | Comparison | | Input Sample | DNA | DNA | Same | {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ # Non-Clinical Performance Data 21 CFR 807.92(b)(1) Non-clinical performance of the instrument was evaluated in a reproducibility study using a representative fragment analysis assay. The reproducibility study was performed across multiple sites, using different instruments, multiple operators, and across several days. The pre-established acceptance criteria were met. # Clinical Performance Data 21 CFR 807.92(b)(2) Clinical performance studies were conducted across 3 US clinical laboratory sites, with multiple instruments using a representative fragment analysis assay and an appropriate method comparison assay. All pre-established performance criteria were met. # Conclusion 21 CFR 807.92(b)(3) The proposed Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer with fragment analysis and sequencing, and the legally marketed Applied Biosystems 3500 Dx/3500xL Dx Genetic Analyzer and 3500 Dx Series Data Collection Software with sequencing use the same capillary electrophoresis technology and are the same instrument. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer did not require changes to the instrument hardware or firmware to expand the intended use to include fragment analysis. The change was made in the Applied Biosystems™ 3500 Dx Series Data Collection Software to enable the fragment analysis workflow in the Diagnostic mode. Therefore, the Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer and the legally marketed device, the Applied Biosystems 3500 Dx/3500xL Dx Genetic Analyzer CS2 and 3500 Dx Series Software. are substantially equivalent based on the study data provided in this 510(k). In addition, the performance of the fragment analysis application has been demonstrated by Clinical and Non-Clinical performance studies using a representative assay.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...