ePlex Instrument

K163652 · Genmark Diagnostics, Incorporated · NSU · Jun 9, 2017 · Clinical Chemistry

Device Facts

Record IDK163652
Device NameePlex Instrument
ApplicantGenmark Diagnostics, Incorporated
Product CodeNSU · Clinical Chemistry
Decision DateJun 9, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2570
Device ClassClass 2

Intended Use

The ePlex Instrument is an automated in vitro diagnostic (IVD) device designed to perform multiplexed nucleic acid tests for the simultaneous detection and identification of nucleic acid targets by processing single-use cartridges developed and manufactured by GenMark Diagnostics, Inc.

Device Story

ePlex Instrument is an automated, multiplexed nucleic acid testing system for clinical laboratory use. Input: clinical samples manually loaded into single-use reagent cartridges. Operation: system performs cell lysis, nucleic acid extraction via magnetic beads, PCR/RT-PCR amplification, and target hybridization within the cartridge. Detection: proprietary eSensor technology using electrowetting on a printed circuit board (PCB) and alternating current voltammetry (ACV) to identify targets via ferrocene-labeled signal probes. Output: automated test results displayed on a touchscreen GUI, printable, or exportable to LIS. System is scalable (1-4 towers, 6-24 bays) for random-access testing. Benefits: automates complex molecular workflows, reducing hands-on time for laboratory professionals; provides rapid, multiplexed diagnostic information to support clinical decision-making.

Clinical Evidence

No clinical data provided in this submission. Analytical performance characteristics (accuracy, precision, linearity, carryover, interfering substances) are referenced to K163636.

Technological Characteristics

System utilizes digital microfluidics, electrowetting, and eSensor (ACV) detection. Components: Base unit with Windows 7 PC/touchscreen GUI; scalable towers with 6 bays each. Connectivity: Ethernet communication between bays and base; barcode scanner/LIS integration. Sterilization: Not applicable (single-use cartridges).

Indications for Use

Indicated for use by trained laboratory professionals in a controlled laboratory environment for automated multiplexed nucleic acid testing. Not intended for patient contact.

Regulatory Classification

Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 9, 2017 GENMARK DIAGNOSTICS, INCORPORATED ALAN MADERAZO, Ph.D., RAC VP OF QUALITY, REGULATORY & CLINICAL AFFAIRS 5964 La PLACE COURT CARLSBAD CA 92008 Re: K163652 Trade/Device Name: ePlex Instrument Regulation Number: 21 CFR 862.2570 Regulation Name: Instrumentation for clinical multiplex test systems Regulatory Class: II Product Code: NSU Dated: May 8, 2017 Received: May 9, 2017 Dear Dr. Maderazo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Stephen J. Lovell -S for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163652 Device Name ePlex® Instrument Indications for Use (Describe) The ePlex Instrument is an automated in vitro diagnostic (IVD) device designed to perform multiplexed nucleic acid tests for the simultaneous detection and identification of nucleic acid targets by processing single-use cartridges developed and manufactured by GenMark Diagnostics, Inc. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------|-------------------------------------------------------------| | <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary Summary of Information to Support Substantial Equivalence to Predicate Device # Submitter Information | Submitter: | GenMark Diagnostics, Incorporated<br>5964 La Place Court<br>Carlsbad, CA 92008 | |------------------------------------|---------------------------------------------------------------------------------------| | Manufacturer: | GenMark Diagnostics, Incorporated<br>5964 La Place Court<br>Carlsbad, CA 92008 | | Establishment Registration Number: | 3008632402 | | Contact: | Alan Maderazo, Ph.D., RAC<br>Vice President, Quality, Regulatory and Clinical Affairs | | Phone: | 760-448-4308 | | Fax: | 760-683-6961 | | E-mail: | Al.Maderazo@genmarkdx.com | | Alternate Contact: | Joseph McMullen<br>Consultant, Regulatory Affairs | | Phone: | 760-410-5052 | | Fax: | 760-683-6961 | | E-mail: | Joseph.McMullen@genmarkdx.com | | Date Prepared: | December 21, 2016 | #### Name of Device and Classification | Product Name: | ePlex® Instrument | |------------------------|-------------------------------------------------------------------------| | Device Classification: | 862.2570, Instrumentation for clinical multiplex test systems, Class II | | Product Code: | NSU, Instrumentation for Clinical Multiplex Test Systems. | ## Predicate Device | Predicate: | BioFire Diagnostics FilmArray 2.0 | |------------|-----------------------------------| |------------|-----------------------------------| {4}------------------------------------------------ #### Device Description The ePlex® Instrument is used to run single-use assay cartridges that incorporate digital microfluidics and GenMark's eSensor® detection technology (used by products that are currently FDA-cleared: K073720 and K090901) to automate all aspects of nucleic acid testing. The ePlex Instrument is designed to: provide a nucleic acid amplification testing solution directly from various sample types, provide random access testing capability, and require minimal operator interaction. The ePlex Instrument includes the following components: - . Base: A touchscreen graphical user interface (GUI) powered by a PC with a Windows Operating System 7. The base communicates with the bays to transfer data. The instrument software installed on the ePlex base processes the raw data generated by the individual bays and determines the test result. - Tower: A chassis housing six bays. ePlex is scalable from one to four towers . connected to either side of the base. - . Bay: 6 bays are housed in each tower. Each bay will accept cartridges independent of the testing status of the other bays allowing for random access testing. Each bay has an Ethernet port for communication with the base unit to receive user inputs and deliver test data to the ePlex Instrument software. The touchscreen graphical user interface (GUI) is flanked on either side by a tower with six bays containing a slot for the cartridge and an LED to indicate bay status (in-use or available for use). The instrument is designed to be scalable with configurations to accommodate a single tower with 6 bays or up to four towers with 24 bays. The ePlex system is used to run multiplex microarray-based assays developed by GenMark. This type of assay is based on the principles of competitive nucleic acid hybridization using a sandwich assay format, wherein a single-stranded target binds concurrently to a sequencespecific solution-phase signal probe and a solid-phase electrode-bound capture probe. The test employs nucleic acid extraction, target amplification via polymerase chain reaction (PCR) or reverse transcription PCR (RT-PCR) and hybridization of target DNA. In the process, the double-stranded PCR amplicons are digested with exonuclease to generate single-stranded DNA suitable for hybridization. {5}------------------------------------------------ Nucleic acid extraction from biological samples occurs within the cartridge via cell lysis, nucleic acid capture onto magnetic beads, and release for amplification. The nucleic acid extraction is processed through microfluidic liquid handling. Once the nucleic acid targets are captured and inhibitors are washed away, the magnetic particles are delivered to the electrowetting environment on the printed circuit board (PCB) and the targets are eluted from the particles and amplified. During hybridization, the single-stranded target DNA binds to a complementary, singlestranded capture probe immobilized on the working gold electrode surface. Single-stranded signal probes (labeled with electrochemically active ferrocenes) bind to specific target sequence / region adjacent to the capture probe. Simultaneous hybridization of target to signal probes and capture probe is detected by alternating current voltammetry (ACV). Each working electrode on the array contains specific capture probes, and sequential analysis of each electrode allows detection of multiple analyte targets. #### Intended Use/Indications for Use The ePlex Instrument is an automated in vitro diagnostic (IVD) device designed to perform multiplexed nucleic acid tests for the simultaneous detection and identification of nucleic acid targets by processing single-use cartridges developed and manufactured by GenMark Diagnostics, Inc. ### Summary of Technological Characteristics of the Device Compared to the Predicate Device The GenMark ePlex Instrument ("Subject Device") and the legally marketed device, BioFire FilmArray 2.0 ("Predicate Device") are described below: | Characteristic | Proposed Device | Predicate Device | |--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name | ePlex Instrument | FilmArray 2.0 | | Manufacturer | GenMark Diagnostics | BioFire Diagnostics | | Regulation | 862.2570<br>Instrumentation for clinical multiplex<br>test systems | 862.2570<br>Instrumentation for clinical multiplex<br>test systems | | Product Code | NSU: Instrumentation for Clinical<br>Multiplex Test Systems | NSU: Instrumentation for Clinical<br>Multiplex Test Systems | | Device Class | Class II | Class II | | Characteristic | Proposed Device | Predicate Device | | Intended Use | The ePlex Instrument is an automated<br><i>in vitro</i> diagnostic (IVD) device<br>designed to perform multiplexed<br>nucleic acid tests for the simultaneous<br>detection and identification of nucleic<br>acid targets by processing single-use<br>cartridges developed and<br>manufactured by GenMark<br>Diagnostics, Inc. | The FilmArray 2.0 is an automated in<br><i>in vitro</i> diagnostic (IVD) device designed<br>for use with FilmArray panels. The<br>FilmArray 2.0 is intended for use in<br>combination with assay specific<br>reagent pouches to detect multiple<br>nucleic acid targets contained in<br>clinical specimens. The FilmArray 2.0<br>instrument interacts with the reagent<br>pouch to both purify nucleic acids and<br>amplify targeted nucleic acid<br>sequences using nested multiplex PCR<br>in a closed system. The resulting PCR<br>products are evaluated using DNA<br>melting analysis. The software<br>automatically determines the results<br>and provides a test report.<br><br>The FilmArray 2.0 is composed of one<br>to eight instruments connected to a<br>computer running FilmArray 2.0<br>software, which controls the function<br>of each instrument and collects,<br>analyzes, and stores data generated by<br>each instrument. | | Assays | For use with FDA cleared GenMark<br>developed panels including<br>Respiratory Panel (RP). | For use with FDA cleared FilmArray<br>panels:<br>Respiratory Panel (RP)<br>Blood Culture Identification (BCID)<br>Panel Gastrointestinal (GI) Panel | | Protocol<br>Processing Steps | Cell disruption, nucleic acid<br>extraction, RT-PCR, single stranding<br>and signal detection | Cell disruption, nucleic acid extraction,<br>PCR1 thermocycling, PCR2<br>thermocycling, DNA melt and signal<br>detection. | | Time to result | Approximately 90 minutes (depends<br>on ePlex assay cartridge) | Approximately 1 hour per sample | | Technological<br>Principles | Multiplex nucleic acid amplification<br>(including reverse transcription as<br>appropriate) utilizing GenMark<br>proprietary electrowetting technology,<br>followed by detection of analyte<br>targets utilizing GenMark proprietary<br>eSensor® technology. | Nested multiplex nucleic acid<br>amplification (including reverse<br>transcription as appropriate) followed<br>by high-resolution melting analysis to<br>confirm the identity of the amplified<br>product. | | Required<br>Accessory | GenMark ePlex Cartridge (aka<br>"consumable" or "assay cartridge"),<br>and barcode scanner | FilmArray Reagent Pouch | | Sample<br>Preparation<br>Method | Minimal sample processing and<br>hands-on time. | Minimal sample processing and hands-<br>on time. | | Test<br>Interpretation | Automated test interpretation and<br>report generation. User cannot access | Automated test interpretation and<br>report generation. User cannot access<br>raw data. Report can be printed. | | Characteristic<br>and Results<br>Reporting | Proposed Device | Predicate Device | | | raw data. Report can be printed,<br>exported or sent to an LIS. | | | User Interface | Touch-screen base unit with<br>integrated PC on a Windows<br>operating system. | FilmArray unit(s) plugged into a<br>standalone PC operated on a Windows<br>operating system. | | User Complexity | Moderate | Moderate | | Detection<br>Procedure | Cyclic voltammetry on custom PCB | Complimentary metal-oxide<br>semiconductor (CMOS) camera<br>Hard-coated filters. | | Detection<br>Chemistries | Complimentary nucleic acid binding<br>with ferrocene labeled probes | Melt curve analysis. Fluorescent<br>imaging of PCR reactions. | | Instrument-<br>Software<br>Communication | Communication travels via Ethernet<br>cable/port. | Communication travels via Ethernet<br>cable/port.<br>Communication for multiple<br>instruments mediated by a multi-port<br>switch. | | Device<br>Configuration | A base unit containing an integrated<br>PC; with onscreen key board and<br>touch screen operation, and<br>detachable barcode scanner and<br>magnetic arm rest. Designed to<br>support external keyboard and mouse.<br>System is configurable with 1 to 4<br>towers; each tower contains 6 bays.<br>Towers are attached to the left and<br>right sides of the base, a maximum of<br>2 towers may be attached to each side.<br>Printer provided with System and<br>optional Universal Power Supply<br>(UPS) offered.<br>Each bay has single-sample test<br>capacity.<br>Random-access multi-sample test<br>capacity per system. | Up to eight FilmArray 2.0 instruments<br>to one computer with mouse, barcode<br>scanner and pouch loading station.<br>Single-sample test capacity per<br>instrument with random-access multi-<br>sample test capacity per system.<br>Printer Provided with System.<br>Interlocking two-instrument racks<br>available to stack instruments and<br>reduce system footprint. (Optional)<br>Instrument held at 0° angle when no<br>rack is used. Instrument held at 15°<br>angle on the rack. | {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ Analysis of the similarities and differences indicate that the devices are substantially equivalent in their intended uses/indications for use, and are generally the same regarding user process, ease of use and general operator protocol. Comparison of technological similarities and differences between the proposed device and the predicate do not raise new or different questions of safety and effectiveness, and therefore render the proposed device as substantially equivalent to the predicate device. Comparison of device performance characteristics will be included in each respective GenMark assay 510(k). #### Summary of Performance Data Performance of the ePlex Instrument was evaluated in a reproducibility study using the GenMark ePlex Respiratory Pathogen (RP) Panel. GenMark will submit a traditional 510(k) for the ePlex RP Panel that will be used with the ePlex Instrument. Complete non-clinical and clinical performance data will be included in the traditional 510(k) for the ePlex RP Panel. The reproducibility study utilized samples prepared as a panel at moderate (3x LoD), low (1x LoD) and negative. The study was conducted at three separate testing sites with different ePlex instruments, by multiple operators performing runs over several days. Run validity, agreement, and variability were analyzed. Acceptance criteria were established in advance of conducting the study and all acceptance criteria were met.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...