K970420 · Pharmacia, Inc. · JQW · Jun 30, 1997 · Clinical Chemistry
Device Facts
Record ID
K970420
Device Name
MASTERCAP AM 5.0/MASTERCAP RM 5.0
Applicant
Pharmacia, Inc.
Product Code
JQW · Clinical Chemistry
Decision Date
Jun 30, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2750
Device Class
Class 1
Indications for Use
MasterCAP AM 5.0 is a software tool for Assay Management designed to support the set up of an in vitro diagnostic assay run. It is designed to be used with Pharmacia CAP System diagnostic equipment and related devices. It is run either in Microsoft Windows, 3.x, or Windows '95 Environment. MasterCAP AM 5.0 handles sample lists, and creates worklists for the related Pharmacia & Upjohn diagnostic equipment. MasterCAP AM 5.0 also evaluates and calculates the assay results and generates laboratory reports. It can import and export results to main frame computers. MasterCAP AM 5.0 includes a driver function which can send and receive patient data from UniCAP 100 diagnostic equipment. MasterCAP RM 5.0 is a software tool for Request Management designed to be used together with MasterCAP AM 5.0. MasterCAP RM 5.0 includes a database and provides features for request management and data storage. MasterCAP RM handles requests and creates sample lists to be used in MasterCAP AM. When the data are evaluated, MasterCAP RM collects the results and generates a result report for each request. MasterCAP RM can store requests, requester information and test panels, and also provides backup and restore functions.
Device Story
MasterCAP AM/RM 5.0 is a software suite for clinical laboratory management of in vitro diagnostic assays. AM (Assay Management) handles sample lists, creates worklists for Pharmacia CAP System equipment, calculates assay results, and generates reports; RM (Request Management) provides database storage for requests, patient test panels, and requester information. Software runs on Windows 3.x or 95. AM interfaces with diagnostic hardware (e.g., UniCAP 100, RoboCAP) to import/export sample data and results. Laboratory technicians use the system to automate assay workflows, manage data, and facilitate reporting to mainframe computers. System improves laboratory efficiency by centralizing request management and automating result calculation/reporting.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Software-based information management system; runs on Microsoft Windows 3.x or Windows '95; interfaces with Pharmacia CAP System diagnostic equipment via driver functions; supports data import/export to mainframe computers; includes database management for requests and results.
Indications for Use
Indicated for laboratory use to manage in vitro diagnostic assay requests, sample lists, worklists, and result reporting in conjunction with Pharmacia CAP System diagnostic equipment.
Regulatory Classification
Identification
A pipetting and diluting system for clinical use is a device intended to provide an accurately measured volume of liquid at a specified temperature for use in certain test procedures. This generic type of device system includes serial, manual, automated, and semi-automated dilutors, pipettors, dispensers, and pipetting stations.
Predicate Devices
MasterCAP AM Version 1.12
UniCAP 100 RM Software
Related Devices
K965103 — WINMAX SOFTWARE SYSTEMS · Diagnostic Products Corp. · May 9, 1997
K083339 — DIMENSION RXL MAX WITH ADVIA MODULAR AUTOMATION SYSTEM (AMAS) · Siemens Healthcare Diagnostics, Inc. · Mar 9, 2009
K103842 — UNICEL DXC 600 SYNCHRON SYSTEM SOFTWARE VERSION 5.0 AND UNICEL DXC 800 SYNCHRON SYSTEM SOFTWARE VERSION 5.0 · Beckman Coulter, Inc. · Apr 27, 2011
K974762 — STAT TRACKS · Awareness Technology, Inc. · Jan 28, 1998
Submission Summary (Full Text)
{0}
K410420
MasterCAP AM/RM 5.0 510(k) Submission Section 10. Summary of Safety and Effectiveness
JUN 30 1997
# 10. SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safe Medical Devices Act of 1990 (SMDA 1990).
Date of Summary Preparation: January 15, 1997
Distributor: Pharmacia & Upjohn
Manufacturer: Pharmacia & Upjohn AB
S-751 82 Uppsala, Sweden
Company Contact Person: Shelley A. Beadle
Pharmacia & Upjohn
Director, Regulatory Affairs
9200-298-109
7000 Portage Road
Kalamazoo, MI 49001-0199
Device Names: MasterCAP AM 5.0
MasterCAP RM 5.0
Common Name: Software programs for assay and request management designed to support the set up of in vitro diagnostic assay runs with Pharmacia CAP System diagnostic equipment, which includes pipetting and diluting systems for clinical use.
# Classification:
Note: MasterCAP AM and MasterCAP RM are software programs which support the functions of diagnostic equipment classified as "JQW Pipetting and Diluting System for Clinical Use".
| Product Name | Product Code | Class | CFR |
| --- | --- | --- | --- |
| MasterCAP AM 5.0 | JQW | I | 862.2750 |
| MasterCAP RM 5.0 | JQW | I | 862.2750 |
mcsec10.doc
1/29/97
01-00210
{1}
MasterCAP AM/RM 5.0 510(k) Submission
Section 10. Summary of Safety and Effectiveness
## Substantial Equivalence to:
MasterCAP AM Version 1.12
UniCAP 100 RM Software
## Intended Use Statements:
1. MasterCAP AM 5.0 is a software tool for Assay Management designed to support the set up of an in vitro diagnostic assay run. It is designed to be used with Pharmacia CAP System diagnostic equipment and related devices. It is run either in Microsoft Windows, 3.x, or Windows '95 Environment.
MasterCAP AM 5.0 handles sample lists, and creates worklists for the related Pharmacia & Upjohn diagnostic equipment. MasterCAP AM 5.0 also evaluates and calculates the assay results and generates laboratory reports. It can import and export results to main frame computers.
MasterCAP AM 5.0 includes a driver function which can send and receive patient data from UniCAP 100 diagnostic equipment.
2. MasterCAP RM 5.0 is a software tool for Request Management designed to be used together with MasterCAP AM 5.0. MasterCAP RM 5.0 includes a database and provides features for request management and data storage. MasterCAP RM handles requests and creates sample lists to be used in MasterCAP AM. When the data are evaluated, MasterCAP RM collects the results and generates a result report for each request. MasterCAP RM can store requests, requester information and test panels, and also provides backup and restore functions.
mcsec10.doc
1/29/97
01-0021i
{2}
MasterCAP AM/RM 5.0 510(k) Submission
Section 10. Summary of Safety and Effectiveness
## General Description:
1. MasterCAP AM 5.0 is a software tool for assay management designed to support the set up of an in vitro diagnostic immunoassay. It is designed to be used with Pharmacia CAP System diagnostic equipment and related devices. It is run either in Microsoft Windows, 3.x or Windows '95 Environment. The following Pharmacia CAP System diagnostic equipment and related devices may be used with MasterCAP AM: AutoCAP, Positioning Guide 96, RoboCAP Version 2.0, Fluorocount 96, Assay Washer 96, UniCAP 100.
When using the UniCAP 100 Device Driver Function of MasterCAP AM with UniCAP 100, you can transfer sample list data to, and import evaluated assay results from, one or more connected UniCAP 100 devices.
MasterCAP AM 5.0 software directs the diagnostic equipment to perform the following functions:
* import sample lists from a main frame computer or enter the information manually
* create an assay run
* distribute samples and tests
* process an assay
* collect raw data
* evaluate and calculate the results
* export the results to the main frame computer
* define assay methods and method groups
mcsec10.doc
1/29/97
01-00212
{3}
MasterCAP AM/RM 5.0 510(k) Submission
Section 10. Summary of Safety and Effectiveness
2. MasterCAP RM 5.0 is a software tool for Request Management designed to be used with MasterCAP AM 5.0 for laboratories running in vitro diagnostic assays. MasterCAP RM includes a database and provides features for request management and data storage. MasterCAP RM handles requests and creates sample lists to be used in MasterCAP AM. When the data are evaluated, MasterCAP RM collects the results and generates a result report for each request. MasterCAP AM and MasterCAP RM are designed to be used with Pharmacia CAP System and UniCAP 100 diagnostic equipment and related devices.
MasterCAP RM can perform the following functions:
* create sample lists
* store requests, requester information and test panels
* create laboratory, result and patient test reports
* perform database queries
* provide backup and restore functions
## Device Comparison:
Pharmacia CAP System is a complete modular system for in vitro diagnostics of allergy and other clinical areas. The system is based on ImmunoCAP technology and includes reagents, information and assay management software and automated processing equipment.
MasterCAP AM 5.0 and MasterCAP RM 5.0 are updates of previous versions of information management software designed to support the set up and management of in vitro diagnostic assays using Pharmacia CAP System and related diagnostic equipment.
MasterCAP AM 5.0 is substantially equivalent to MasterCAP AM 1.12, and MasterCAP RM 5.0 is substantially equivalent to UniCAP 100 RM software. Several improvements have been made to the updated software resulting in more convenient and versatile software.
mcsec10.doc
1/29/97
01-00213
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
JUN 30 1997
Karen E. Matis
Regulatory Affairs Manager
Pharmacia & Upjohn
5094 St. Andrew Drive
Westerville, Ohio 43082
Re: K970420
MasterCAP AM 5.0/MasterCAP RM 5.0
Regulatory Class: II
Product Code: JQW
Dated: June 6, 1997
Received: June 9, 1997
Dear Ms. Matis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{5}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
MasterCAP AM/RM 5.0 510(k) Submission
Section 1. Intended Use Statements
510(k) Number: 6970420
Device Name: MasterCAP AM 5.0
Indications For Use:
MasterCAP AM 5.0 is a software tool for Assay Management designed to support the set up of an in vitro diagnostic assay run. It is designed to be used with Pharmacia CAP System diagnostic equipment and related devices. It is run either in Microsoft Windows, 3.x, or Windows '95 Environment.
MasterCAP AM 5.0 handles sample lists, and creates worklists for the related Pharmacia & Upjohn diagnostic equipment. MasterCAP AM 5.0 also evaluates and calculates the assay results and generates laboratory reports. It can import and export results to main frame computers.
MasterCAP AM 5.0 includes a driver function which can send and receive patient data from UniCAP 100 diagnostic equipment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patricia A. Bernhardt (for AWM)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 6970420
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
mcsec1.doc
1/21/97
01-00001
{7}
MasterCAP AM/RM 5.0 510(k) Submission
Section 1. Intended Use Statements
510(k) Number: 6970420
Device Name: MasterCAP RM 5.0
Indications For Use:
MasterCAP RM 5.0 is a software tool for Request Management designed to be used together with MasterCAP AM 5.0. MasterCAP RM 5.0 includes a database and provides features for request management and data storage. MasterCAP RM handles requests and creates sample lists to be used in MasterCAP AM. When the data are evaluated, MasterCAP RM collects the results and generates a result report for each request. MasterCAP RM can store requests, requestor information and test panels, and also provides backup and restore functions.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)


OR
Over-The-Counter Use
Optional Format 1-2-96)
mcsec1.doc
1/21/97
01-00002
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.