K962231 · Roche Diagnostic Systems, Inc. · JJE · Aug 30, 1996 · Clinical Chemistry
Device Facts
Record ID
K962231
Device Name
COBAS CORE IMMUNOCHEMISTRY SYSTEM
Applicant
Roche Diagnostic Systems, Inc.
Product Code
JJE · Clinical Chemistry
Decision Date
Aug 30, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2160
Device Class
Class 1
Indications for Use
The COBAS Core II Immunochemistry System is an in vitro diagnostic analyzer intended for enzyme immunoassay procedures. The instrument performs initial sample measurement and pipetting, timed addition of assay reagents, incubation and shaking of assay tubes, aspiration of used reagents and washing of coated polystyrene assay beads and photometric measurement of the chromogen / enzyme reaction.
Device Story
COBAS Core II is an automated in vitro diagnostic analyzer for enzyme immunoassays. It processes clinical samples by performing pipetting, reagent addition, incubation, shaking, bead washing, and photometric measurement. The system utilizes coated polystyrene assay beads. It is operated by laboratory personnel in a clinical setting. The device automates sample handling and data management, providing photometric results that assist clinicians in diagnostic testing. Compared to the predicate, it features hardware upgrades including a laser barcode scanner, increased rack capacity (14 positions), mechanical rack locking, and a PC-486 based processing unit running software on Windows NT.
Clinical Evidence
Bench testing only. Precision and accuracy evaluated using three representative assays (LH, FSH, Ferritin). Precision study: within-run and run-to-run reproducibility tested using reference sera and kit controls; CVs were <10% (except one low-level LH sample). Correlation study: 137 clinical samples compared between Core I and Core II using linear regression; slopes were 1.000 ± 0.014 with correlation coefficients > 0.995.
Technological Characteristics
Automated enzyme immunoassay analyzer. Materials: coated polystyrene assay beads. Sensing: photometric measurement. Hardware: PC-486 DX33 microprocessor, 16 MB RAM, 350 MB hard disk, laser barcode scanner. Connectivity: 14-position rack platform with mechanical locking. Software: embedded in Microsoft Windows NT. Operation: automated pipetting, incubation, and washing.
Indications for Use
Indicated for use in clinical laboratories to perform automated enzyme immunoassay procedures for the measurement of analytes including Ferritin, TSH, Prolactin, T4, FSH, T3, LH, Free T4, Free T3, and IgE.
Regulatory Classification
Identification
A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.
K052794 — ETI-MAX 3000 · DiaSorin, Inc. · Mar 31, 2006
K060373 — COBAS 6000 SERIES SYSTEM · Roche Diagnostics · Mar 13, 2006
Submission Summary (Full Text)
{0}
K962231
Roche
Roche Diagnostic Systems
A Member of the Roche Group
AUG 30 1996
Roche Diagnostic Systems, Inc.
Branchburg Township
1080 US Highway 202
Somerville, New Jersey 08876-3771
Direct Mail
Fax
# 510(k) Summary
Roche COBAS Core II Immunochemistry System
In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary as outlined in 21 CFR 807.92 is provided herewith.
## I. Identification of 510(k) Sponsor:
Roche Diagnostic Systems, Inc.
a subsidiary of Hoffmann-La Roched, Inc.
Branchburg Township
1080 US Highway 202
Somerville, NJ 08876-3771
510(k) Submission dated June 7, 1996
## II. Description of the Device/Statement of Intended Use:
The COBAS Core II Immunochemistry System is an *in vitro* diagnostic analyzer intended for enzyme immunoassay procedures. The instrument performs initial sample measurement and pipetting, timed addition of assay reagents, incubation and shaking of assay tubes, aspiration of used reagents and washing of coated polystyrene assay beads and photometric measurement of the chromogen / enzyme reaction.
## III. Identification of the legally marketed device to which the 510(k) sponsor claims equivalence:
The COBAS Core II Immunochemistry System is substantially equivalent to the currently marketed Roche COBAS Core Immunochemistry System (K921180).
9 282
{1}
3 283
# IV. Summary of the technological characteristics of the new device in comparison to those of the predicate.
## Similarities to predicate
- Both instruments are intended for automation of the currently marketed Roche COBAS Core assays:
- COBAS Core Ferritin (K920829) COBAS Core TSH (K930776)
- COBAS Core Prolatin (K930305) COBAS Core T4 (K932605)
- COBAS Core FSH (K930304) COBAS Core T3 (K932608)
- COBAS Core LH (K930306) COBAS Core Free T4 (K932607)
- COBAS Core IgE (K930890) COBAS Core Free T3 (K942676)
- Both instruments perform the same basic operations, which are, sample handling, pipetting, dilution, incubation, bead washing, measurement and data management.
- Both instruments utilize the coated bead technology and photometric measurement.
## Differences to predicate device
The differences in the above mentioned devices consist hardware and software changes in the following areas:
| Area of Change | COBAS Core I | COBAS Core II |
| --- | --- | --- |
| Barcode Reader | Infra Red Reader | Laser Barcode Scanner |
| Rack Platform | 6 positions for samples and reagents | 14 positions for samples and reagents |
| Test /Sample Racks Identity | Binary Rack Coding | Barcode Label ID |
| Rack Locking Device | Racks could be unlocked during processing | Racks are mechanically locked when in use |
| Pipetting Syringes | Module on left side of instrument | Module moved to the right side of the instrument with no modification to the tube length or syringes |
| Operating Software | Internally developed software | Internally developed software embedded in Microsoft Windows NT |
| Microprocessor | CPU 186 | PC-486 DX33
Real Time CPU 186
Hard Disk - 350 MB
RAM - 16 MB
Diskette - 1.44 MB floppy disk drive 3.5" |
{2}
V. Breif discussion of the clinical and nonclinical tests relied on for a determination of substantial equivalence:
To demonstrate equivalence in performance characteristics, three representative assays were evaluated for precision and accuracy. The three assays are:
- COBAS Core LH
- COBAS Core FSH
- COBAS Core Ferritin
Precision Study
Four internal reference sera and the kit control were tested to evaluate the within-run and run-to-run reproducibility of the COBAS Core II analyzer. For within-run precision the samples were tested seven times in triplicate. For the run-to run precision the same samples were tested in duplicate in 10 independent runs. Except for one serum with a very low LH level, all CV's are far below 10%. The precision range found for the COBAS Core II correspond to the values determined for the COBAS Core I.
Correlation Study
The correlation of the COBAS Core I and COBAS Core II was evaluated using 137 randomly selected clinical samples run in duplicate on both analyzers. The results of Core I and Core II were compared using linear regression analysis. For all three assays, the calculated slopes were $1.000 \pm 0.014$ with a correlation coefficient > 0.995.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.