GENZYME CYSTATIN C REAGENT ANF GENZYME CYSTATIN C CALIBRATOR
Applicant
Genzyme Corporation
Product Code
NDY · Clinical Chemistry
Decision Date
May 15, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Intended Use
The Genzyme Cystatin C is an in vitro diagnostic test intended for the quantitative measurement of Cystatin C concentration in human serum, heparinized plasma and EDTA plasma. Cystatin C measurements are used as an aid to the diagnosis and treatment of renal diseases. Calibrator: For the calibration of Genzyme Cystatin assay. For In Vitro Diagnostic Use.
Device Story
In vitro diagnostic assay; measures Cystatin C concentration in human serum, heparinized plasma, or EDTA plasma; utilizes particle-enhanced turbidimetric immunoassay (PETIA) principle; colloidal gold particles coated with rabbit polyclonal anti-Cystatin C antibodies; reaction between particles and sample Cystatin C causes agglutination; change in absorbance signal proportional to Cystatin C concentration; performed on automated clinical chemistry analyzers (e.g., Abbott Aeroset); results compared against standard curve; used by laboratory professionals to aid diagnosis and treatment of renal disease.
Clinical Evidence
Bench testing only. Performance validated on Abbott Aeroset analyzer. Precision: CV% 0.7-6.4%. Linearity: 0.2-8.0 mg/L (R²=0.9917). LoB 0.02 mg/L, LoD 0.05 mg/L, LoQ 0.20 mg/L. Interference testing showed no significant interference from bilirubin, triglycerides, hemoglobin, or rheumatoid factor. Method comparison (n=148) against predicate showed strong correlation (r²=0.997). Reference range established using 196 healthy individuals.
Technological Characteristics
Sol particle turbidimetric immunoassay. Reagents contain colloidal gold particles coated with anti-cystatin C polyclonal antibodies. Detection via light absorption on commercially available clinical analyzers. Calibrators are bovine serum albumin liquid matrix with sodium azide preservative. No specific ASTM standards cited.
Indications for Use
Indicated for quantitative measurement of cystatin C in human serum, heparinized plasma, and EDTA plasma to aid in diagnosis and treatment of renal diseases.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Dade Behring N Latex Cystatin C (k041878)
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 15 2008
## 510(k) Summary of Safety and Effectiveness for Genzyme Cystatin C
## This summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is: K080662
| 1. Manufacturer's Name, Address, Telephone, Contact, Date of Preparation | |
|--------------------------------------------------------------------------|---------------------|
| Manufacturer: | Genzyme Corporation |
Genzyme Diagnostics 31 New York Avenue Framingham, MA 01701 Contact Information: Genzyme Corporation 31 New York Avenue Framingham, MA 01701 Attn: Eric Lawson Tel: 800.332.1042 March 6, 2008 Preparation date: 2. Device Name / Classification: Trade or Proprietary Name: Genzyme Cystatin C Reagent Genzyme Cystatin C Calibrator Test. Cystatin C Common or Usual Name: Calibrator for Cystatin C test
Classification Name:
Creatinine Test System Calibrator, Secondary
#### 3. Identification of the Legally Marketed Device:
Dade Behring N Latex Cystatin C - k041878 Dade Behring N Protein Standard UY, calibrator - k003501
#### 4. Device Description:
Genzymc Cystatin C assay reagent is based on the sol particle turbidimetric immunoassay principle. It contains colloidal gold particles coated with anticystatin C specific polyclonal antibodies. The reaction between the particles and any cystatin C in samples results in the formation of agglutinates and an associated change in absorbance signal. The change in absorbance signal is proportional to the amount of cystatin C in the sample. Cystatin C concentration in the sample is determined by comparison with a standard curve.
{1}------------------------------------------------
Genzyme Cystatin C calibrators consist of a bovine serum albumin liquid matrix with assigned concentrations of cystatin C. The calibrators are preserved with sodium azide and are ready to use.
## 5. Device Intended Use:
### Reagents:
The Genzyme Cystatin C assay reagent is an in vitro diagnostic test intended for the quantitative measurement of cystatin C concentration in human serum, heparinized plasma and EDTA plasma. Cystatin C measurements are used as an aid to the diagnosis and treatment of renal diseases. For In Vitro Diagnostic Use.
### Calibrator:
Genzyme Cystatin C calibrator is an in vitro diagnostic product for the calibration of Genzyme Cystatin C assay. For In Vitro Diagnostic Usc.
## 6. Medical device to which equivalence is claimed and comparison information:
The Genzyme Cystatin C diagnostic test is substantially equivalent to the Dade Behring, Inc. N Latex Cystatin C method (k041878) with respect to the indications for use, the analyte tested, device design, the type of antibody materials, and the gencral sample matrix. The basic differences between the new device and the predicate are the assay technology and the instruments used for testing. The Genzyme device is a turbidimetric immunoassay that can be used on commercially available clinical analyzers using light absorption detection systems; while, the Dade Behring test is an immunonephelometric assay applicable only on Dade Behring, Inc. Nephelometer Systems.
The Genzyme Cystatin C calibrator is substantially equivalent to the Dade Behring N Protein Standard UY calibrator (k003501) with regard to the technological characteristics and in terms of the intended use. The difference is that the Genzymc calibrator is provided as a set of six liquid solutions of assigned cystatin C concentration levels and is ready to use for calibration; while, the Dade Behring N Protein Standard UY is a single lyophilized concentration of cystatin C calibrator that is prepared to various levels in order to establish values for calibration.
## 7. Device Performance Characteristics:
Performance characteristics and data from nonclinical tests submitted in the premarket notification submission support a determination of substantial equivalence of the Genzyme Cystatin C devices with the Dade Behring predicate devices. This testing includes: analytical limits and sensitivity, within run and total precision testing over 20 days, linearity, reportable range, stability, analytical specificity, interfering substances, matrix comparison, and reference range. Method comparison studies performed with clinical
{2}------------------------------------------------
specimens demonstrate comparable performance and accuracy, good correlation, and substantial equivalence.
#### 8. Conclusion:
In conclusion, the new devices Genzyme Cystatin C Reagent and Genzyme Cystatin C Calibrator, do not raise any new issues of safety or effectiveness, and are substantially equivalent to the currently marketed Dade Behring devices N Latex Cystatin C test kit (k041878), and N Protein Standard UY calibrator (k003501) respectively, as demonstrated by the descriptions and performance data acquired by Genzyme and summarized in this submission.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings outstretched and its head turned to the right.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Genzyme Corporation c/o Mr. Eric Lawson Manger, Biomedical Regulatory Affairs 31 New York Avenue Framingham, MA 01701
MAY 15 2008
Rc: K080662
> Trade/Device Name: Genzyme Cystatin C Reagent and Calibrator Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Codc: NDY and JIT : Dated: March 06, 2008 Received: March 10, 2008
Dear Mr. Lawson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number (if known):
Device Name:
Genzyme Cystatin C Reagent Genzyme Cystatin C Calibrator
Indication For Use:
Reagents:
For the quantitative measurement of cystatin C concentration in human serum, heparinized plasma and EDTA plasma. Cystatin C measurements are used as an aid to the diagnosis and treatment of renal diseases. For In Vitro Diagnostic Use.
Calibrator: For the calibration of Genzyme Cystatin C assay. For In Vitro Diagnostic Use.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use __ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Concurrence in 12.11.y
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K080662
Page 1 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.