STRATUS CS ACUTE CARE NT-PROBNP TESTPAK, MODEL CPBNPM

K071834 · Dade Behring, Inc. · NBC · Aug 17, 2007 · Clinical Chemistry

Device Facts

Record IDK071834
Device NameSTRATUS CS ACUTE CARE NT-PROBNP TESTPAK, MODEL CPBNPM
ApplicantDade Behring, Inc.
Product CodeNBC · Clinical Chemistry
Decision DateAug 17, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1117
Device ClassClass 2

Intended Use

The Stratus® CS Acute Care™ NT-proBNP method (pBNP) is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NTproBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. This method is for use by trained health care professionals on the Stratus® CS Stat Fluorometric Analyzer in the clinical laboratory and point of care (POC) settings.

Device Story

The Stratus® CS Acute Care™ NT-proBNP TestPak is an in vitro diagnostic assay used on the Stratus® CS Stat Fluorometric Analyzer. It utilizes a two-site sandwich Radial Partition Immunoassay (RPIA) on glass fiber paper. Input: heparanized plasma sample. Process: sample reacts with immobilized monoclonal antibody; enzyme-labeled monoclonal antibody conjugate is added to form an antibody-antigen-labeled antibody sandwich; unbound conjugate is eluted via substrate wash. Output: enzymatic reaction rate measured via front surface fluorescence, proportional to NT-proBNP concentration. Used in clinical labs and point-of-care settings by trained healthcare professionals. The analyzer's microprocessor performs all data analysis. Results assist clinicians in diagnosing CHF, assessing disease severity, and performing risk stratification for acute coronary syndrome and heart failure, potentially improving patient management.

Clinical Evidence

Bench testing only. Precision study (CLSI EP5-A) showed total CV 3.4-4.6%. Linearity study (17-19,485 pg/mL) showed 90.9-108.9% recovery. Analytical sensitivity (LoB) 1.5 pg/mL; functional sensitivity (LoQ) 15 pg/mL. Interference testing (CLSI EP7-A2) confirmed no significant interference from common substances. Method comparison (n=148) vs predicate yielded slope 1.02, intercept 8.2, r=0.99. Concordance study (n=149) across reference and disease populations showed 96-100% agreement.

Technological Characteristics

Two-site sandwich Radial Partition Immunoassay (RPIA) on glass fiber paper. Uses monoclonal (sheep) antibodies. Energy source: fluorometric optical system. Connectivity: standalone analyzer. Software: microprocessor-based data analysis. Calibration: lot-specific, updated every 30 days.

Indications for Use

Indicated for quantitative measurement of NT-proBNP in human plasma to aid in diagnosis and severity assessment of congestive heart failure (CHF) and for risk stratification of patients with acute coronary syndrome and heart failure.

Regulatory Classification

Identification

The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.

Special Controls

*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K071834 AUG 1 7 2007 #### 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Submitter's Name: | George M. Plummer<br>Dade Behring Inc.<br>P.O. Box 6101<br>Newark, DE 19714-6101 | |---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date of Preparation: | June 30, 2007 | | Name of Product(s): | Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak | | FDA Classification Name(s): | B-type natriuretic peptide test system | | FDA Guidance Documents: | "Class II Special Control Guidance Document for B-Type Natriuretic<br>Peptide Premarket Notifications; Final Guidance for Industry and<br>FDA Reviewers" - 11/30/2000 | | Predicate Device(s):<br>TestPak | Dade Behring Stratus® CS Acute Care™ NT-proBNP (pBNP)<br>(k043476/k060548) | ### Device Description(s): #### Method The Stratus® CS Acute Care™ NT-proBNP method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassav (RPIA) technology. In this procedure, dendrimer linked monoclonal antibody is added to the center portion of a square piece of glass fiber paper in the pBNP TestPak. This antibody recognizes a distinct antigenic site on the NT-proBNP molecule. Sample is then added onto the paper where it reacts with the immobilized antibody. After a short incubation, a conjugate consisting of enzyme-labeled monoclonal antibody directed against a second distinct antigenic site on the NT-proBNP molecule is pipetted onto the reaction zone of the paper. During this second incubation period, enzyme-labeled antibody reacts with the bound NT-proBNP, forming an antibody-antigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS analyzer by applying a substrate wash solution to the center of the reaction zone. By including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of NT-proBNP in the sample. The reaction rate can then be measured by an optical system that monitors the reaction rate via front surface fluorescence: All data analysis functions are performed by the microprocessor within the analyzer. #### Intended Use: #### Method The Stratus® CS Acute Care™ NT-proBNP method (pBNP) is an in vitro diagnostic test for the quantitative measurement of N-terminal pro-brain natriuretic peptide (NT-proBNP) in heparanized plasma. In individuals suspected of having congestive heart failure (CHF), {1}------------------------------------------------ measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. # Comparison to Predicate Device: A summary of the features of the revised Dade Behring Stratus® CS Acute Care™ pBNP TestPak and the current (predicate) Dade Behring Stratus® CS Acute Care™ pBNP immunoassay (k043476/k060548) is provided in the following charts. | Feature | Revised Stratus® CS Acute<br>Care™ pBNP | Current Stratus® CS pBNP | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For the <i>in vitro</i> quantitative<br>determination of N-terminal pro-<br>brain natriuretic peptide in human<br>plasma as an aid in the diagnosis<br>and assessment of severity of<br>individuals suspected of having<br>congestive heart failure. The test is<br>further indicated for the risk<br>stratification of patients with acute<br>coronary syndrome and heart<br>failure. | For the <i>in vitro</i> quantitative<br>determination of N-terminal pro-brain<br>natriuretic peptide in human plasma<br>as an aid in the diagnosis and<br>assessment of severity of individuals<br>suspected of having congestive heart<br>failure. The test is further indicated<br>for the risk stratification of patients<br>with acute coronary syndrome and<br>heart failure. | | Assay Type<br>(detection) | fluorometric immunoassay | fluorometric immunoassay | | Reportable Range | 15- 20,000 pg/mL | 15- 20,000 pg/mL | | Antibody | Monoclonal (sheep) antibody | Polyclonal (sheep) antibody | | Cut-off | 125 pg/mL for patients less than 75<br>years and 450 pg/mL for patients 75<br>years and older | 125 pg/mL for patients less than 75<br>years and 450 pg/mL for patients 75<br>years and older | | Analytical<br>Sensitivity | ≤15 pg/mL | ≤15 pg/mL | | Functional<br>Sensitivity | ≤ 50 pg/mL | ≤ 50 pg/mL | {2}------------------------------------------------ | Analytical<br>Specificity | The pharmaceutical Natrecor®<br>shows no significant cross reactivity<br>at 0 and 125 pg/mL NT-proBNP;<br>sixteen other substances also show<br>no significant cross reactivity | The pharmaceutical Natrecor® shows<br>no significant cross reactivity at 0 and<br>125 pg/mL NT-proBNP; sixteen other<br>substances also show no significant<br>cross reactivity | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Interferences | No significant interference from:<br>bilirubin, conj. up to 60 mg/dL<br>bilirubin, unconjugated up to 60<br>mg/dL, hemoglobin up to 1000<br>mg/dL, lipemia up to 3000 mg/dL,<br>rheumatoid factors up to 500 IU/mL | No significant interference from:<br>bilirubin, conj. up to 60 mg/dL<br>bilirubin, unconjugated up to 60<br>mg/dL, hemoglobin up to 1000<br>mg/dL, triglycerides up to 3000<br>mg/dL, rheumatoid factors up to 750<br>IU/mL | | Hook Effect | No high dose effect<br>( up to 833,585 pg/mL) | No high dose effect<br>( up to 1,400,000 pg/mL) | | Calibration Interval | Calibration curve updated for each<br>lot, using one level and every 30<br>days, thereafter with the same<br>reagent lot. After calibration update<br>at completion of each test,<br>recovered values are calculated<br>from stored calibration coefficients. | Calibration curve updated for each<br>lot, using one level and every 30<br>days, thereafter with the same reagent<br>lot. After calibration update at<br>completion of each test, recovered<br>values are calculated from stored<br>calibration coefficients. | | Sample Volume | 50 uL | 50 uL | # Method performance Summary: ### Analytical Results ### Method Comparison A split sample method comparison demonstrated good agreement between the revised Dade Behring Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak method and the current (predicate) Dade Behring pBNP TestPak method with heparinized plasma patient samples. | Comparative | | Intercept | Correlation | | |-------------------|-------|-----------|-------------|-----| | Method | Slope | (pg/mL) | Coefficient | n | | Dade Behring pBNP | 1.02 | 8.2 | 0.99 | 148 | The model equation for Passing-Bablok linear regression statistics is: [results for revised Stratus CS pBNP] = slope x [comparative method results] + intercept. The range of NT-proBNP values for the monoclonal method in the correlation study was: 15.8 - 18914.2 pg/mL. ## Lithium Heparin versus Sodium Heparin Comparison of lithium heparin versus sodium heparin samples on the Stratus® CS system showed very good agreement. Fifty-one samples were tested in duplicate. The lithium heparin values ranged from {3}------------------------------------------------ 31.2 to 16, 445 pg/mL and sodium heparin values from 32.2 to 17,769 pg/mL. A linear regression gave a slope of 1.06, an intercept of -75 pg /mL and a correlation coefficient of 0.997. ### Comments on Substantial Equivalence: Both the revised Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak and the current Dade Behring pBNP immunoassays are intended for the quantitative determination of NT-proBNP. Comparative data for human plasma samples demonstrate good analytical agreement between the methods. #### Conclusion: The revised Dade Behring ® CS Acute Care™ NT-proBNP (pBNP) TestPak and the current (predicate) Dade Behring pBNP immunoassays (k043476/k060548) are substantially equivalent based on their intended use and performance characteristics as described above. > George M. Plummer Regulatory Affairs and Compliance Manager June 30, 2007 {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Dade Behring Inc. c/o Mr. George M. Plummer Regulatory Affairs & Compliance Manager Bldg. 400, M.S. 514 P.O. Box 6101 Newark, DE 19714-6101 Re: k071834 Trade/Device Name: Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak Regulation Number: 21 CFR §862.1117 Regulation Name: B-type Natriuretic Peptide Test System. Regulatory Class: Class II Product Code: NBC Dated: June 30, 2007 Received: July 03, 2007 Dear Mr. Plummer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. AUG 17 2007 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 1egally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT 510(k) Number (If Known): k 071834 ### Device(s) Name(s): Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak #### Indications for Use: The Stratus® CS Acute Care™ NT-proBNP method (pBNP) is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NTproBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. This method is for use by trained health care professionals on the Stratus® CS Stat Fluorometric Analyzer in the clinical laboratory and point of care (POC) settings. Prescription Use X (Part 21 CFR 801 Subpart D) --------------------------------- and/or Over-the-counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carol Benson Division Sign-Off Office of In Vitro Diagnostic D Evaluation and Safe K071834
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...