K072536 · Esa Biosciences, Inc. · MRG · May 7, 2008 · Clinical Chemistry
Device Facts
Record ID
K072536
Device Name
ESA BIOSCIENCES INC. VITAMIN D HPLC TEST
Applicant
Esa Biosciences, Inc.
Product Code
MRG · Clinical Chemistry
Decision Date
May 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1825
Device Class
Class 2
Indications for Use
The ESA Biosciences Inc. Vitamin D HPLC test is for the quantitative determination of 25-hydroxyvitamin D in human serum or EDTA-plasma to be used in the assessment of vitamin D sufficiency. Assay results should be used in conjunction with other clinical or laboratory data to assist the clinician in making individual patient management decisions in an adult population.
Device Story
The Vitamin D HPLC Test is a quantitative assay for 25-hydroxyvitamin D in human serum or EDTA-plasma. The system utilizes high performance liquid chromatography (HPLC) with electrochemical (EC) detection. Input samples undergo pre-analytical processing: mixing with precipitation reagent containing an internal standard, centrifugation, and solid phase extraction (SPE) to isolate 25(OH)D. The prepared sample is analyzed via HPLC using a dual coulometric EC cell; the first cell oxidatively screens interfering components, while the second cell is optimized for selective 25(OH)D detection. The device is used in clinical laboratory settings by trained personnel. Output is a quantitative concentration value used by clinicians alongside other laboratory data to assess vitamin D sufficiency and guide patient management. The internal standard corrects for extraction recovery and volume variability. The device provides a standardized method for vitamin D measurement, aiding in the diagnosis and monitoring of vitamin D status in adults.
Clinical Evidence
Bench testing only. Precision studies (CLSI EP5-A) demonstrated total CVs of 5.43%–12.6%. Linearity (CLSI EP6-A) confirmed a measuring range of 7–200 ng/mL (r² > 0.99). Method comparison (CLSI EP9-A2) against a reference HPLC method (n=95) showed a correlation of r=0.973. Limit of Quantitation (LoQ) was 7.0 ng/mL. Interference testing (CLSI EP7-A2) identified 25(OH)EpiD3 as an interferent in infants, leading to the age-based contraindication.
Technological Characteristics
Quantitative HPLC assay with electrochemical (EC) detection. Components include SPE columns, mobile phase, and reagents. Requires ESA HPLC system with CouloChem III or CoulArray EC detector. Dual coulometric cell configuration for interference screening and analyte detection. Single-point internal standard quantitation. Calibrated via UV absorbance spectrometry against NIST SRM 935a.
Indications for Use
Indicated for quantitative determination of 25-hydroxyvitamin D in human serum or EDTA-plasma to assess vitamin D sufficiency in adults. Contraindicated for patients < 1 year of age due to potential interference from C3 epimers.
Regulatory Classification
Identification
A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.
Special Controls
*Classification.* Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.
DEN170019 — Vitamin D 200M Assay for the Topaz System · Ab Sciex, LLC · May 18, 2017
K021163 — OCTEIA 25-HYDROXY VITAMIN D · Immunodiagnostic Systems , Ltd. · May 10, 2002
K162840 — Elecsys Vitamin D total II, Vitamin D total II CalSet, PreciControl Vitamin D total II, Vitamin D CalCheck 5 · Roche Diagnostics · Feb 8, 2017
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k072536
B. Purpose for Submission:
New device
C. Measurand:
25-Hydroxyvitamin D
D. Type of Test:
Quantitative HPLC assay
E. Applicant:
ESA Biosciences, Inc.
F. Proprietary and Established Names:
Vitamin D HPLC Test
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1825
2. Classification:
II
3. Product code:
MRG
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
The ESA Biosciences Inc. Vitamin D HPLC test is for the quantitative determination of 25-hydroxyvitamin D in human serum or EDTA-plasma to be used in the assessment of vitamin D sufficiency. Assay results should be used in conjunction with other clinical or laboratory data to assist the clinician in making individual patient management decisions in an adult population.
3. Special conditions for use statement(s):
Assay results should be used in conjunction with other clinical laboratory data to assist the clinician in making patient management decisions in an adult population.
This assay should not be used for measurement in samples from patients less than 1 year of age.
This assay is for prescription use.
{1}
4. Special instrument requirements:
For use with the ESA high performance liquid chromatography (HPLC) system with an ESA CouloChem III (k903231) or CoulArray EC Detector (k963938).
I. Device Description:
The Vitamin D HPLC Test system includes:
Vitamin D Calibration Stock (includes alcohol and 1 µg/mL 25(OH)D); Vitamin D Internal Standard Stock (includes acetonitrile and 1 µg/mL 25(OH)D); Vitamin D Mobile Phase (includes acetonitrile, propanol and supporting electrolyte); Vitamin D Reagent A and Vitamin D Reagent B (both include acetonitrile); Vitamin D Reagent C (includes alcohol) and a Vitamin D SPE column.
J. Substantial Equivalence Information:
1. Predicate device name(s):
DiaSorin LIAISON Vitamin D Assay
2. Predicate K number(s):
k032844
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Predicate Device (k032844) | Proposed Device |
| Intended Use | Quantitative determination of 25-Hydroxyvitamin D to be used in the assessment of Vitamin D sufficiency. | Same |
| Matrix | Human serum or plasma | Same |
| Limit of Quantitation | 7.0 ng/mL | Same |
| Differences | | |
| --- | --- | --- |
| Item | Predicate Device (k032844) | Proposed Device |
| Test principle | Chemiluminescent immunoassay | HPLC with electrochemical detection |
| Calibrators | Derived from serum-based standards, extracted identically to controls and patient samples | Working single point calibrator is an aqueous calibration stock solution and internal stock solution |
| Reference range: 2.5th to 97.5th percentile (n) | 9.5 – 52.0 ng/mL (98) | 5.7 – 34.8 ng/mL (150) |
| Total precision (% CV) | 6% - 13% | 5.43% – 12.6% |
{2}
K. Standard/Guidance Document Referenced (if applicable):
- User Evaluation of Precision Performance of Clinical Chemistry Devices: Approved Guideline (EP5-A2)
- Evaluation of the Linearity of Quantitative Measurement Procedures: a Statistical Approach: Approved Guideline (EP6-A2)
- Interference Testing in Clinical Chemistry: Approved Guideline (EP7-A2)
- Method Comparison and Bias Estimation using Patient Samples: Approved Guideline (EP9-A2)
- Protocols for Determination of Limits of Detection and Limits of Quantitation (EP17-A3)
- How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline (C28-A2)
L. Test Principle:
The ESA method is a kit for measurement of total 25(OH) Vitamin D by HPLC with electrochemical (EC) detection. Specific reagents and solid phase extraction columns are included for sample preparation. A 200 uL volume of sample (serum or plasma) is mixed with a precipitation reagent, which contains internal standard. The internal standard is a stable vitamin D analog that is used to correct for variability in extraction recovery and analytical sample volume. After centrifugation, supernatant is poured onto a pre-conditioned SPE column for rapid extraction of 25(OH)D and the IS. SPE columns are washed with 2 different reagents and analytes are eluted with a third reagent. The resulting eluent is diluted before analysis. The prepared sample is analyzed with an ESA HPLC system using an ESA EC detector (Coulochem III or CoulArray) equipped with a dual coulometric EC cell. Calibration is accomplished by direct HPLC analysis of authentic standard solutions (i.e. not taken through the extraction step). 25(OH)D and IS are separated chromatographically followed by EC detection. A dual EC cell is used with the first, upstream, cell maintained at a specific potential to oxidatively screen possible interfering sample components. The second, downstream cell is maintained at a potential that is optimized for selective 25(OH)D detection. Total 24(OH)D sample concentrations are determined by single-point internal standard quantitation.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed using CLSI Document EP5-A as a guideline. Five replicates of three concentration levels of analyte in EDTA plasma pools that were individually prepared and analyzed in a single run each day, repeated over 20 days. Five days were performed at an external site. Three different HPLC-EC systems, four operators, two reagent lots, five analytical columns (3 lots), three lots of SPE columns and four analytical cells were used in these studies. Three samples were prepared by spiking EDTA plasma, pooled from 5 individual blood bank donors, with assayed, commercially procured 25(OH)D.
{3}
| Sample level | Mean ng/mL | Within-run | | Total | | N |
| --- | --- | --- | --- | --- | --- | --- |
| | | SD (ng/mL) | % CV | SD (ng/mL) | % CV | |
| 1 | 43.8 | 4.6 | 10.6 | 5.5 | 12.6 | 100 |
| 2 | 117.2 | 8.0 | 6.8 | 9.9 | 8.4 | 100 |
Two serum and two EDTA plasma samples were acquired from adult human subjects. Four replicates of each sample were individually prepared (taken through all pre-analytical and analytical steps of the procedure) and analyzed in a single run each day, repeated over 20 days. Samples were run on a single instrument, by two operators and using two lots of reagents and extraction columns. Three replicates were lost during pre-analytical sample processing.
| Sample # (matrix) | Mean ng/ml | Within-run | | Within-device | | N |
| --- | --- | --- | --- | --- | --- | --- |
| | | SD (ng/ml) | % CV | SD (ng/ml) | % CV | |
| 1 (plasma) | 22.1 | 0.63 | 2.82 | 1.34 | 6.05 | 79 |
| 2 (plasma) | 21.8 | 0.50 | 2.31 | 1.19 | 5.48 | 80 |
| 3 (serum) | 25.5 | 0.65 | 2.55 | 1.44 | 5.66 | 80 |
| 4 (serum) | 25.6 | 0.55 | 2.16 | 1.39 | 5.43 | 78 |
b. Linearity/assay reportable range:
To establish the linearity of the assay, a study design was used based on CLSI protocol EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: a Statistical Approach: Approved Guideline. The claimed measuring range of this device is 7 - 200 ng/mL.
A volume of low level EDTA plasma was spiked with assayed 25(OH)D concentrate. The low and high level samples were mixed to obtain 13 samples with concentrations spanning the measuring range. Three aliquots of each sample concentration were taken through the pre-analytical step for the assay. Upon completion of this step, extracts were divided into 2 aliquots and one aliquot was analyzed by an HPLC-Coulochem III system and the other was analyzed by an HPLC-CoulArray system.
Linear regression statistics:
HPLC-Coulochem III system
$$
y = 1.08x - 4.46
$$
$$
r^2 = 0.994
$$
HPLC-CoulArray system
$$
y = 1.05x - 2.46
$$
$$
r^2 = 0.995
$$
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
{4}
5
# Calibrator:
Value assignment: An Internal Standard Stock and Master Product Calibrator Stock are created in-house (there is no primary calibrator, recognized reference material or reference measurement procedure for this Measurand at this time). Value assignment of both Stocks is made by UV absorbance spectrometry calibrated with NIST SRM 935a, based on NIST recommended molar extinction coefficients. A working Internal Standard is created by ten-fold volumetric dilution of the Internal Standard Stock. The working Calibrator is a mixture of equivalent volumes of the Master Product Calibration Stock and Internal Standard Stock, followed by dilution to 100 ng/mL of 25(OH)D. Each lot of product calibrator and internal standard are assayed by HPLC to confirm purity and between-lot consistency.
# Stability:
Kit components: The shelf-life, on-board and open stability testing protocols for kit components and the acceptance criteria were described and found to be acceptable. Current shelf life studies support a six-month shelf life, with real-time studies on-going to support twelve months.
Samples: To evaluate the stability of samples stored at different conditions, two sample levels were prepared by spiking freshly drawn EDTA plasma with 25(OH)D. Three aliquots of each sample were processed immediately and analyzed within one hour of processing. Results from these aliquots were used as the basis for comparison to samples that were stored for various time periods and storage conditions before processing. On board stability of processed samples was also assessed.
| Sample | Storage conditions | Stable for at least |
| --- | --- | --- |
| Processed sample | 20 – 23°C (on-board) | 24 hours |
| Sample | -15 to -25°C | 3 freeze-thaw cycles |
| Sample | 2 - 8°C | 7 days |
| Sample | 20 - 23°C | 6 hours |
# d. Detection limit:
The Limit of the Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) for the ESA Vitamin D HPLC test was determined in accordance with the CLSI guideline EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation. One low level sample was prepared by spiking blank matrix (bovine serum albumin in phosphate buffered saline) with 25(OH)D. Two additional low samples were prepared by pooling patient samples. Six replicates of the blank matrix and three replicates each of the low level samples were analyzed in each run. Each replicate was taken through all pre-analytical and analytical steps of the assay. Twelve runs were performed with two analytical systems, 3 EC cells, 3
{5}
analytical columns, 2 lots of SPE columns, 2 reagent lots and two operators. The non-parametric method (CLSI EP17-A) was used to calculate the LoB. The LoB was calculated to be 2.51 ng/ml and the LoD was calculated to be 5.3 ng/ml. The Limit of Quantitation (LoQ) was calculated to be 7.0 ng/ml and was defined as the concentration of the analyte demonstrated imprecision less than or equal to 20% CV.
e. Analytical specificity:
The ESA Vitamin D HPLC test was evaluated for interference consistent with CLSI Guideline EP7-A2.
The following substances were added to two EDTA plasma samples (with 25(OH)D levels of 43 ng/mL and 105 ng/mL) and were found not to interfere with the Vitamin D HPLC test (a bias exceeding 10% is considered interference):
| Substance | Test concentration |
| --- | --- |
| Triglycerides | 1571 mg/dL |
| Albumin | 12.7 g/dL |
| Bilirubin | 60 mg/dL |
| Hemoglobin | 500 ng/dL |
| Paricalcitol (vitamin D analog) | 100 ng/mL |
Cholesterol, tested at a level of 500 mg/dL, was found to produce a bias of more than 10%. (The actual bias seen was 10.1%. ) In addition, 25(OH)EpiD3, a vitamin metabolite, was found to interfere with the Vitamin D HPLC test. This compound is found only in infants 0–1 year of age. As noted in the limitations section of the package insert, this assay should therefore not be used for measurement in samples from patients less than 1 year of age. (The sponsor did not test 25(OH)EpiD2, also found only in infants 0-1 year of age, which is known to interfere with 25(OH)D assays, but referred the user to literature in the product insert for further information.)
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
The ESA Vitamin D HPLC Test was compared to a reference method using CLSI Document EP9-A2 as a guideline for the method comparison study.
The Vitamin D HPLC Test was compared to a reference HPLC method using serum samples from 85 individual patients and fifteen additional samples prepared by augmenting aliquots of a single serum pool (prepared from approximately 50 individual patient samples) with varying amounts of 25(OH)D. Results from 95 specimens with analyte values that spanned the
6
{6}
measuring range of the device (tested over 6 non-consecutive days) were reported. The correlation was found to be:
| Equation: Least Squares | y = 1.027x - 5.3 |
| --- | --- |
| Slope, 95% Confidence Interval | 0.976 to 1.077 |
| Intercept, 95% Confidence Interval (ng/mL) | -8.7 to -1.8 |
| Correlation Coefficient, R | 0.973 |
| N | 95 |
| Standard Error (ng/mL, y-direction) | 9.91 |
| Range (ng/mL) | 9.6 – 195 |
| Equation: Deming | y = 1.056x - 6.9 |
b. Matrix comparison:
19 neat and 28 spiked samples were used to demonstrate equivalent performance between paired serum and EDTA-plasma sample types over the claimed measuring range. Samples were spiked after serum/plasma separation using volumetric technique.
y = 1.01x + 0.24
r = 0.999
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The reference range was determined according to the non-parametric method in CLSI C28-A2.
Samples from 150 apparently healthy adult human subjects were collected from 3 locations to achieve even North-South geographical distribution within the latitudes of the contiguous 48 US States. (Specifically, samples were obtained from 75 individuals living in Massachusetts, 37 in Tennessee and 38 in Texas.) Inclusion and exclusion of individuals was based on a questionnaire. The designation of "apparently healthy adult human subjects" was based on the initial Normal Blood Donor intake questionnaire and by preliminary viral testing. Individuals taking prescription or over-the-counter vitamin D supplements were excluded.
{7}
Sample analysis was performed over a 2-day time frame using one analytical system, one lot of reagents and by one operator. Samples were collected in late winter (i.e., March). The population consisted of 71 (47%) females and 79 (53%) males ranging in age from 20 to 64 years [mean = 40 years, SD = 10 years] and comprised 109 (73%) Caucasian, 36 (24%) African-American and 5 (3%) Hispanic individuals.
Reference range and frequency distribution from the population are as indicated below:
| Population (N) | Median 25(OH)D ng/mL | Observed Range ng/mL 2.5th to 97.5th Percentile |
| --- | --- | --- |
| 150 | 17.4 | 5.7 – 34.8 |

25(OH)D Reference Sample Distribution
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.