K982298 · Biotecx Laboratories, Inc. · KLI · Feb 3, 1999 · Clinical Chemistry
Device Facts
Record ID
K982298
Device Name
OPTICOAT T4 EIA KIT
Applicant
Biotecx Laboratories, Inc.
Product Code
KLI · Clinical Chemistry
Decision Date
Feb 3, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1700
Device Class
Class 2
Intended Use
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Device Story
OptiCoat T, EIA Kit is an in vitro diagnostic test system for quantitative measurement of total thyroxine (T4) in human serum. Device utilizes enzyme immunoassay (EIA) technology to detect thyroid hormone levels. Intended for use in clinical laboratory settings by trained laboratory personnel. Healthcare providers use quantitative results to assess thyroid function, aiding in diagnosis and management of thyroid disorders. System provides objective biochemical data to support clinical decision-making regarding patient thyroid status.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Enzyme immunoassay (EIA) test system for serum analysis. Quantitative measurement of total thyroxine (T4).
Indications for Use
Indicated for the measurement of total (free and protein bound) thyroxine (thyroid hormone) in serum to aid in the diagnosis and treatment of thyroid diseases in patients.
Regulatory Classification
Identification
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K991432 — BIOCHECK TOTAL T4 ENZYME IMMUNOASSAY TEST KIT, MODEL BC-1007 · Biocheck, Inc. · Jun 2, 1999
K982954 — FREE T4 MICROWELL EIA, MODEL #7019 · Biomerica, Inc. · Sep 24, 1998
K990993 — BIOCHECK SENSITIVE TSH (S-TSH) ENZYME IMMUNOASSAY TEST KIT; CATALOG NUMBER: BC-1003 · Biocheck, Inc. · Jun 22, 1999
K971709 — TOTAL THROXINE MICROPLATE EIA(225-300) · Monobind · Jun 16, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized symbol. The symbol consists of three abstract human figures, possibly representing individuals or communities, with flowing lines suggesting movement or connection. The overall design is simple and conveys a sense of unity and service.
EEB 3 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mohan Mehra, Ph.D. President Biotecx Laboratories, Inc. 6023 South Loop East Houston, Texas 77033
Re: K982298 Trade Name: OptiCoat T, EIA Kit Regulatory Class: II Product Code: KLI Dated: January 11, 1999 Received: January 12, 1999
Dear Dr. Mehra:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Ref: 1960
Page 2 of 2
510(k) Number (if known): K982298
Device Name: T4 EIA Kit
Indications For Use:
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Dean Hooper
(Division Sign-Off)
Division of Clinical Laboratory Levices
510(k) Number K982298
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.1009)
OR
Over-the-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.