NOVA STATSTRIP LACTATE HOSPITAL METER, NOVA STATSTRIP LACTATE TEST STRIPS
K100602 · Nova Biomedical Corp. · KHP · Aug 17, 2011 · Clinical Chemistry
Device Facts
Record ID
K100602
Device Name
NOVA STATSTRIP LACTATE HOSPITAL METER, NOVA STATSTRIP LACTATE TEST STRIPS
Applicant
Nova Biomedical Corp.
Product Code
KHP · Clinical Chemistry
Decision Date
Aug 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1450
Device Class
Class 1
Intended Use
The Nova StatStrip Lactate Hospital Meter System is intended for in vitro diagnostic use by healthcare professionals for multiple patient use in a professional healthcare setting for clinical and for point-of-care usage for the quantitative determination of Lactate (Lac) in fresh venous and arterial whole blood specimens as an aid to evaluate the acid-base status of patients suspected of having lactic acidosis. It is not for use on capliary blood specimens. It is intended to provide plasma equivalent results to laboratory methods. Nova StatStrip Lactate Test Strips are intended for use only with the Nova StatStrip Lactate Hospital Meter for quantitative determination of lactate in fresh venous and arterial whole blood specimens. It is not for use on capillary blood specimens. The performance characteristics of the device for lactate measurements on capillary specimens have not been established. Nova StatStrip Lactate Test Strips are for testing outside the body (in vitro diagnostic use only). Nova StatStrip Lactate Control Sclutions are intended for use with the Nova StatStrip Lactate Hospital Meter and Nova StatStrip Lactate Test Strips as a quality control check to verify the accuracy of blood lactate test results. There are 2 levels of controls, (Levels 1 and Level 2). Nova StatStrip Lactate Linearity Kit solutions are used to check the linearity of the Nova StatStrip Lactate Hospital Meter. There are 4 levels of lactate linearity solutions: Level 1, Level 2, Level 3, and Level 4.
Device Story
System comprises handheld meter, test strips, control/linearity solutions, and docking station. Input: fresh venous or arterial whole blood (0.6 μL) applied to test strip. Principle: electromechanical biosensor using lactate oxidase; lactate in sample reacts with reagent to generate electric current proportional to lactate concentration. Output: quantitative lactate measurement (mmol/L) displayed on meter. Used in clinical/point-of-care settings by healthcare professionals. Meter includes barcode scanner for patient/QC ID. Data storage for 1000 patient tests, 200 QC tests, 4000 operators. Provides rapid (13s) plasma-equivalent results to aid clinical decision-making regarding patient acid-base status. Benefits include portability and rapid turnaround compared to central laboratory analysis.
Clinical Evidence
Laboratory and clinical testing performed. Studies demonstrated blood lactate results are substantially equivalent to current methods for blood lactate measurements. No specific sensitivity/specificity metrics provided in summary.
Technological Characteristics
Handheld meter system; electrochemical sensing principle using test strips. Results provided within 13 seconds. Includes meter, test strips, control solutions, and docking station. Intended for venous and arterial whole blood.
Indications for Use
Indicated for healthcare professionals in professional healthcare settings for quantitative determination of lactate in fresh venous and arterial whole blood to evaluate acid-base status in patients suspected of lactic acidosis. Not for use on capillary blood, neonates, or patients suspected of sepsis.
Regulatory Classification
Identification
A lactic acid test system is a device intended to measure lactic acid in whole blood and plasma. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Nova Biomedical StatProfile pHOx Plus L (K023567, K012058)
Related Devices
K112955 — NOVA STATSTRIP LACTATE HOSPITAL METER SYSTEM · Nova Biomedical Corporation · Jan 5, 2012
K980908 — LACTATE PRO SYSTEM · Kdk Corp. · Sep 11, 1998
K112430 — I-STAT LACTATE TEST/I-SAT CG4+CARTRIDGE · Abbott Pont of Care, Inc. · Dec 20, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
5. 510(k) Summary
K/00602
# 510(k) Summary
(as required by 21 CFR 807.92)
AUG 1 7 2011
#### Submitter: Nova Biomedical Corporation 200 Prospect Street Waltham, MA 02454 U.S.A.
Correspondent: Paul W. MacDonald Chief Quality Assurance and Regulatory Affairs Officer
> Phone: 781-894-0800 ex 211; Fax: 781-891-4806 Email: pmacdonald@novabio.com
# Device Name:
Nova StatStrip Lactate Hospital Meter System
## Common Name:
Blood Lactate Test System
### Classification:
Division of Clinical Laboratory Devices Clinical Chemistry Lactate dehydrogenase test system. Class II per 21 CFR 862.1440
# Product Codes:
CFJ, KHP
### Predicate Devices:
KDK Corporation Lactate Pro System, K980908 (handheld meter) Nova Biomedical StatProfile pHOx Plus L K023567, K012058 (analyzer)
### Description of the Device:
The Nova StatStrip Lactate Hospital Meter System consists of:
- 1. Nova StatStrip Lactate Hospital Meter
- 2. Nova StatStrip Lactate Test Strips
- 3. Nova StatStrip Lactate Control Solutions (Levels 1 and 2)
- 4. Nova StatStrip Lactate Linearity Solutions (Levels 1,2,3 and 4)
- 5. Meter Docking Station
{1}------------------------------------------------
#### Intended Use/Indications for Use:
#### Meter:
The Nova StatStrip Lactate Hospital Meter System is intended for in vitro diagnostic use by health care professionals for clinical and for point-of-care usage for the quantitative determination of Lactate (Lac) in fresh venous and arterial whole blood specimens. It is not for use on capillary blood specimens. It is intended to provide plasma equivalent results to laboratory methods. The Nova StatStrip Lactate Hospital Meter System is indicated for use in a clinical setting by healthcare professionals as an aid to evaluate the acid-base status of patients suspected of having lactic acidosis.
#### Test Strips:
Nova StatStrip Lactate Test Strips are intended for use only with Nova StatStrip Lactate Hospital Meter for quantitative determination of lactate in fresh venous and arterial whole blood specimens. It is not for use on capillary blood specimens. The performance characteristics of the device for lactate measurements on capillary specimens have not been established. Nova StatStrip Lactate Test Strips are for testing outside the body (in vitro diagnostic use only).
#### Control Solutions:
Nova StatStrip Lactate Control Solutions are intended for use with the Nova StatStrip Family of Meters and Nova StatStrip Lactate Test Strips as a quality control check to verify the accuracy of blood lactate test results. There are 2 levels of controls. (Level 1 and Level 2).
Nova StatStrip Lactate Linearity Kit solutions are used to check the linearity of the Nova StatStrip Family of Meters. There are 4 levels of lactate linearity solutions: Level 1, Level 2, Level 3, and Level 4.
### Summary of Technological Characteristics:
The Nova StatStrip Lactate Hospital Meter System has similar fundamental scientific technology and a similar intended use as the currently marketed Lactate Pro System (K980908).
Both the Lactate Pro and the proposed Nova StatStrip Lactate Hospital Meter System are hand held devices with similar intended use to quantitatively measure the lactate levels in whole blood. The principle of operation is the same for the proposed and predicate device. Each utilizes a test strip that is inserted into a meter for results within 13 seconds.
### Comparison to Predicate Devices:
The proposed Nova StatStrip Lactate Hospital Meter System uses the same fundamental scientific technology, similar specifications, and has a similar intended use as the predicate Lactate Pro System (K980908). In addition, the proposed Nova StatStrip Lactate Hospital Meter System can produce equivalent results to the previously cleared Nova Biomedical StatProfile pHOx Plus L (K023567, K012058), which is a tabletop multianalyte chemistry analyzer,
{2}------------------------------------------------
# Performance Studies:
Laboratory and clinical testing was performed on the proposed Nova StatStrip Lactate Hospital Meter System. The studies demonstrated that the blood lactate results were substantially equivalent to the current methods for blood lactate measurements.
### Conclusion:
Results of laboratory and clinical testing demonstrate that the performance of the Nova StatStrip Lactate Hospital Meter System has the same intended uses, with similar technological characteristics and can produce results that are substantially equivalent to results obtained on the predicate devices. The system performs as intended and raises no new safety or effectiveness issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping shapes representing the head, body, and legs.
Re:
Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
Nova Biomedical Corporation c/o Mr. Paul W. MacDonald Chief Quality Assurance/Regulatory Affairs Officer 200 Prospect Street Waltham, MA 02454-9141
AUG 1 7 2011
k100602 Trade/Device Name: Nova StatStrip Lactate Hospital Meter System. Regulation Number: 21 CFR 862.1450 Regulation Name: Lactic acid test system. Regulatory Class: Class 1, meets limitations of exemptions per 21 CFR 862.9(c)(9) Product Code: KHP, JJX Dated: August 15, 2011 Received: August 16, 2011
Dear Mr. MacDonald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'' (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Signature
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number:
K100602
Device Name:
Nova StatStrip Lactate Hospital Meter System
#### Indications for Use:
#### Meter:
The Nova StatStrip Lactate Hospital Meter System is intended for in vitro diagnostic use by healthcare professionals for multiple patient use in a professional healthcare setting for clinical and for point-of-care usage for the quantitative determination of Lactate (Lac) in fresh venous and arterial whole blood specimens as an aid to evaluate the acid-base status of patients suspected of having lactic acidosis. It is not for use on capliary blood specimens. It is intended to provide plasma equivalent results to laboratory methods.
#### Test Strips:
Nova StatStrip Lactate Test Strips are intended for use only with the Nova StatStrip Lactate Hospital Meter for quantitative determination of lactate in fresh venous and arterial whole blood specimens. It is not for use on capillary blood specimens. The performance characteristics of the device for lactate measurements on capillary specimens have not been established. Nova StatStrip Lactate Test Strips are for testing outside the body (in vitro diagnostic use only).
#### Control Solutions:
Nova StatStrip Lactate Control Sclutions are intended for use with the Nova StatStrip Lactate Hospital Meter and Nova StatStrip Lactate Test Strips as a quality control check to verify the accuracy of blood lactate test results. There are 2 levels of controls, (Levels 1 and Level 2).
Nova StatStrip Lactate Linearity Kit solutions are used to check the linearity of the Nova StatStrip Lactate Hospital Meter. There are 4 levels of lactate linearity solutions: Level 1, Level 3, and Level 4.
Prescription Use *X*
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Initial Sign Off
Division Sign-Off
Office of In Vitro Diagnostic Onice Oraluation and Safety
510(k) L4100602
(Updated February 3, 2005)
Page ( of .
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.