RANGE + QUALICHECK LEVEL 1, RANGE + QUALICHECK LEVEL 2, RANGE + QUALICHECK LEVEL 3

K130236 · Radiometer Medical Aps · JJS · Mar 15, 2013 · Clinical Chemistry

Device Facts

Record IDK130236
Device NameRANGE + QUALICHECK LEVEL 1, RANGE + QUALICHECK LEVEL 2, RANGE + QUALICHECK LEVEL 3
ApplicantRadiometer Medical Aps
Product CodeJJS · Clinical Chemistry
Decision DateMar 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil

Device Story

Range+ QUALICHECK is an assayed quality control system consisting of three levels of aqueous solutions containing biological buffers, salts, glucose, lactate, dyes, and preservatives. It is used to verify the accuracy and precision of Radiometer ABL700, ABL800/8x7, ABL77, ABL80, and ABL90 series analyzers. The solution is equilibrated with carbon dioxide and oxygen. The healthcare provider uses the control solution by conditioning ampoules at 18-32°C, then running the sample on the analyzer. The analyzer measures the parameters and compares results against the provided control ranges. This process ensures the analyzer is functioning correctly, allowing the clinician to rely on patient test results for clinical decision-making.

Clinical Evidence

No clinical data. Bench testing only. Stability and shelf-life were validated using three lots of each level, measured on ABL735 analyzers over 36 months. Value assignment was performed using 144 measurements across multiple validated ABL7xx series devices to establish mean ±2SD target ranges. Traceability to NIST and IUPAC standards was documented.

Technological Characteristics

Aqueous solution containing biological buffer, salts, glucose, lactate, dyes, and preservative. Equilibrated with CO2 and O2. Provided in 2mL ampoules. Three-level system. Traceable to NIST SRM and IUPAC pH scale. Storage: 2-8°C. In-use: 18-32°C.

Indications for Use

Indicated for use as an assayed quality control system for evaluating the accuracy and precision of blood gas, oximetry, electrolyte, metabolite, and hemoglobin parameters on specified Radiometer ABL analyzers.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) Summary #### 1 Administrative ## k130236 # MAR 1 5 2013 Device Information Device | Device Name: | Range+ QUALICHECK, LEVEL 1 | |-----------------------|----------------------------| | | Range+ QUALICHECK, LEVEL 2 | | | Range+ QUALICHECK, LEVEL 3 | | Common Name: | Quality Control | | Product Code: | JJS | | Registration Number: | 21 CFR 862.1660 | | Classification: | Class I, Reserved | | Classification Panel: | Clinical Chemistry | | Submitter | | |---------------|---------------------------------------------| | Company Name: | Radiometer Medical ApS | | ER Number: | 3002807968 | | Address: | Aakandevej 21<br>2700 Broenshoej<br>Denmark | | Phone: | +45 3827 3827 | | Fax: | +45 3827 2727 | ## 2 Description of Device Range+ QUALICHECK is a three level quality control system consisting of: - Range+ QUALICHECK, LEVEL 1 (S7930) . - Range+ QUALICHECK, LEVEL 2 (S7940) . - Range+ QUALICHECK, LEVEL 3 (S7950) . Each level consists of 30 ampoules per box. One ampoule contains 2 mL of solution. The quality control solution is an aqueous solution containing a biological buffer, salts, glucose, lactate, dyes and a preservative, and it is equilibrated with carbon dioxide and oxygen. #### 3 Intended Use For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: {1}------------------------------------------------ pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF , cK+, cNa+, cCa2+, gCl+, cGlu, cLac, ctBil ## 4 Substantial Equivalence The Range+ QUALICHECK Level 1-3 is substantially equivalent in Intended Use, fundamental scientific technology, features, and characteristics to the predicate: 510(k) Number/Device Manufacture: . ← ・・ :-- k980135 QUALICHECK5+, Radiometer Medical ApS | Issue | SE Device | Predicate: QUALICHECK5+ (K980135) | |----------------------------|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Form | Same | Liquid | | Base matrix | Same | Aqueous solution | | Preservatives | Same | ProClin 950 | | Open Vial Claim | Same | Ampoules should be conditioned for at least five hours<br>at a constant temperature between 18 °C and 32 °<br>before use. To maintain the reliability of the blood gas<br>parameters, you must use the contents of the ampoule<br>immediately after opening and according to the<br>instructions in the operator's manual for the relevant<br>analyser. | | Blood Gas<br>Measurement | Same | pH, pO₂, pCO₂ | | Oximetry<br>Measurement | Same | ctHb, sO₂ FO₂Hb, FCOHb, FMetHb | | Electrolyte<br>Measurement | Same | cK⁺, cNa⁺, cCa²⁺, cCl⁻ | | Metabolite<br>Measurement | Same | cGlucose, cLactate | | Hemoglobin<br>Measurement | Same | ctBil | | Compatible<br>analyzers | Same | ABL700/705/710/715/720/725/730/735 ABL805/810/815/820/825/830/835 ABL80 - OSM SW ABL90 FLEX | {2}------------------------------------------------ . | Differences | | | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Issue | SE Device | Predicate: QUALICHECK5+<br>(K980135) | | Intended Use | For In Vitro Diagnostic Use.<br>This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges.<br><br>Analytes are:<br>pH, pCO2, pO2, ctHb, sO2,<br>FO₂Hb, FCOHb, FMetHb, FHbF ,<br>cK+, cNa+, cCa2+, cCl-, cGlu, | For In Vitro Diagnostic Use.<br>This QUALICHECK5+ solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges.<br>For professional use. | | Storage | 2 °C to 8 °C until expiration date<br>5 hours between 18 – 32°C | 2 °C to 25 °C until expiration date, including up to a total of 15 days at up to 32 °C | | Oximetry Measurement | FHbf in Level 1 | FHbf in all four levels | | Levels | Three levels | Four levels | | Target ranges | Target ranges, see Section<br>12.02 Target values | Target ranges, see Section<br>12.02 Target values | #### 5 Stability Stability studies and claims are based on real-time study. Control solutions are stable for 2 years at 2-8°C. In-use Stability: Ampoules should be conditioned for at least 5 hours at a constant temperature between 18-32℃ before use. The contents should be used immediately after opening. #### 6 Value Assignment To determine the assigned values and control ranges for Range+ QUALICHECK, six (6) trays of 1000 ampoules are sampled randomly from the Range+ QUALICHECK batch. Twelve ampoules are sampled from each of the trays and 72 ampoules are sampled from the reference batch. The samples are conditioned and shaken at 25°C in a water bath for 6 hours. Measurements are performed on a minimum of 3 validated ABL7xx series devices with data collection. Measurement of each parameter is performed alternately on the reference ampoule and the sample ampoule and repeated 12 times on each ABL7xx for a total of 144 measurements. Target ranges are calculated based on the mean ±2SD. . {3}------------------------------------------------ ## 7 Conclusion . : The Range+ QUALICHECK Level 1, Level 2 and Level 3 are substantially equivalent in Intended Use, fundamental scientific technology, features, and characteristics to the predicate device: QUALICHECK5+(K980135). .. · {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 15, 2013 Radiometer Medical Aps c/o Martin Gabler Regulatory Affairs Akandevej 21 2700 Bronshoj, Denmark Re: k130236 Trade/Device Name: Range+ QUALICHECK, LEVEL 1 Range+ QUALICHECK, LEVEL 2 Range+ QUALICHECK, LEVEL 3 Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: I, reserved Product Code: JJS Dated: January 24, 2013 Received: February 8, 2013 Dear Mr. Gabler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register: Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2-Mr. Gabler forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): k130236 Device Name: Range+ QUALICHECK, LEVEL 1 Range+ QUALICHECK, LEVEL 2 Range+ QUALICHECK, LEVEL 3 Indications for Use: For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF , cK , cNa , cCa2+, cCI7, cGlu, cLac, ctBil Prescription Use_ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Yung W.O.Chan-S Division Sign-Off Office of In Vitro Diagnostics and Radiological Health k130236 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...