Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase (CK-MB) in human serum and plasma. A creatine phosphokinase / creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Device Story
Elecsys CK-MB STAT is an automated immunoassay for quantitative measurement of CK-MB in human serum/plasma. Principle: two-step sandwich immunoassay using biotinylated monoclonal CK-MB-specific antibody, ruthenium-labeled monoclonal CK-MB-specific antibody, and streptavidin-coated microparticles. Process: 15 μl sample incubation; magnetic capture of microparticles on electrode surface; removal of unbound substances via ProCell; electrochemiluminescence induction via voltage application; light emission measured by photomultiplier. Results determined via 2-point calibration curve and master curve from reagent bar code. Used in clinical laboratory settings by trained personnel. Provides rapid (9-minute) CK-MB activity levels to assist clinicians in diagnosing myocardial infarction and muscle diseases.
Clinical Evidence
No clinical data provided; substantial equivalence based on analytical performance and technological similarity to the predicate device.
Technological Characteristics
Two-step sandwich immunoassay; electrochemiluminescence detection; streptavidin-coated microparticles; ruthenium-labeled antibodies; biotinylated antibodies. 9-minute assay time. Instrument-based measurement using photomultiplier. Reagent bar code system for master curve data.
Indications for Use
Indicated for in vitro quantitative determination of CK-MB isoenzyme in human serum and plasma to aid in diagnosis and treatment of myocardial infarction and muscle diseases (e.g., Duchenne-type muscular dystrophy).
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
{0}
Boehringer Mannheim Corporation
Elecsys® CK-MB STAT 510(k)
December, 1997
DEC 17 1997
510(k) Summary
K974421
## Introduction
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
1. Submitter name, address, contact
Boehringer Mannheim Corporation
4300 Hacienda Drive
Pleasanton, CA 94588-2722
(510) 730 - 8415
Contact Person: Jody J. Savage
Date Prepared: November 21, 1997
2. Device name
Proprietary name: Elecsys CK-MB STAT Assay
Common name: Creatine Kinase Test
Classification name: Colorimetric method, CPK or Isoenzymes
3. Predicate device
The Boehringer Mannheim Elecsys CK-MB STAT is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed first generation Boehringer Mannheim Elecsys CK-MB STAT (K961501).
Continued on next page
U:\QA_QC\QA\RA\Elecsys\510(k)\CK-MB\CK-MB 510k FDA.doc
{1}
Boehringer Mannheim Corporation
Elecsys® CK-MB STAT 510(k)
November, 1997
# 510(k) Summary, Continued
## 4. Device Description
The Elecsys® CK-MB test principle is based on a two-step sandwich with Streptavidin microparticles and electrochemiluminescence detection.
Total duration of the assay: 9 minutes
- 1st incubation: 15 μl of sample, a biotinylated monoclonal CK-MB-specific antibody and a monoclonal CK-MB-specific antibody labeled with a ruthenium complex
- 2nd incubation: after the addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
- Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.
Continued on next page
U:\QA_QC\QA\RA\Elecsys\510(k)'s\CK-MB\CK-MB 510k FDA.doc
{2}
Boehringer Mannheim Corporation
Elecsys® CK-MB STAT 510(k)
November, 1997
# 510(k) Summary, Continued
## 5. Intended use
Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase (CK-MB) in human serum and plasma.
A creatine phosphokinase / creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
## 6. Comparison to predicate device
The Boehringer Mannheim Elecsys CK-MB STAT as described in this submission is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed first generation Boehringer Mannheim Elecsys CK-MB STAT assay (K961501).
The first generation Boehringer Mannheim Elecsys CK-MB assay (K961501) was cleared for use on June 11, 1996, and is in commercial distribution. Subsequent to clearance, the Elecsys CK-MB was renamed the Elecsys CK-MB STAT, to emphasize the rapid (9 minute) assay time. This current submission reflects the addition of an anti-BB antibody to Reagent 1 of the Elecsys CK-MB STAT assay. This modification does not affect the intended use of the device or the basic fundamental scientific technology of the device.
U:\QA_QC\QA\RA\Elecsys\510(k)\CK-MB\CK-MB 510k FDA.doc
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
DEC 17 1997
Jody J. Savage
Consultant, Regulatory Affairs
Boehringer Mannheim Corporation
4300 Hacienda Drive
P.O. Box 9002
Pleasanton, California 94588-2722
Re: K974421
Elecsys® CK-MB STAT Assay
Regulatory Class: II
Product Code: JHY
Dated: November 21, 1997
Received: November 24, 1997
Dear Ms. Savage:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Boehringer Mannheim Corporation
Elecsys® CK-MB STAT 510(k)
November, 1997
510(k) Number (if known): N/A
Device Name: Elecsys CK-MB STAT Assay
Indications For Use:
Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase (CK-MB) in human serum and plasma.
A creatine phosphokinase / creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 10979421
U:\QA_QC\QA\RA\Elecsys\510(k)'s\CK-MB\CK-MB 510k FDA.doc
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.